K101309 · Exelint International, Co. · LHI · Oct 20, 2010 · General Hospital
Device Facts
Record ID
K101309
Device Name
EXELINT ALUMINUM HUB BLUNT NEEDLES
Applicant
Exelint International, Co.
Product Code
LHI · General Hospital
Decision Date
Oct 20, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The EXELINT Aluminum Hub Blunt Needles are intended to be used in a variety of medical, pharmaceutical and laboratory procedures to extract fluid or medication from vials or ampoule for transfer or for irrigation.
Device Story
EXELINT Aluminum Hub Blunt Needles are sterile, single-use fluid transfer devices. Construction consists of Type 304 stainless steel blunt cannulas staked into aluminum luer lock hubs. Packaged in polypropylene tubes with caps. Used by clinicians in medical, pharmaceutical, or laboratory settings to extract fluids or medications from vials/ampoules for transfer or irrigation. Device facilitates safe fluid handling; blunt tip design reduces risk of accidental needle sticks compared to sharp needles.
Clinical Evidence
Bench testing only. Compliance with ISO 7864 (Needles) and ISO 9626 (Stainless Steel Tubing) demonstrated. Packaging seal strength testing performed. No clinical data required.
Technological Characteristics
Materials: Type 304 stainless steel cannula, aluminum luer lock hub. Design: Blunt tip, single lumen. Sterilization: Ethylene Oxide (EtO). Standards: ISO 7864, ISO 9626.
Indications for Use
Indicated for use in medical, pharmaceutical, and laboratory procedures for fluid/medication extraction from vials or ampoules for transfer or irrigation.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K120420 — STERILE BLUNT FOR SINGLE USE · Shanghai Double Dove Industrial Co., Ltd. · Oct 18, 2012
K102328 — BLUNT FILL NEEDLE AND BLUNT PLASTIC CANNULA · Shandong Weigao Group Medical Polymer Co., Ltd. · May 13, 2011
K223334 — Sterile Hypodermic Needles for Single Use · Zhejiang Kindly Medical Devices Co., Ltd. · Dec 16, 2022
K083851 — INVIROBLUNT CANNULA WITH AND WITHOUT EZ WINGS, MODELS 120522 AND 120022 · Inviro Medical Devices, Inc. · May 15, 2009
K231720 — Sterile Hypodermic Needles for Single Use · Wepon Medical Technology Co., Ltd. · Sep 8, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(K) SUMMARY
Date Revised: July 22, 2010
1.1 SUMMARY OF SAFETY AND EFFECTIVENESS Submitted By: Armand Hamid
.
Director EXELINT International, Co. 5840 West Centinela Avenue, Los Angeles, CA 90045 Telephone: (310)-649-0707 Fax: (310) 649-1168
OCT 2 0 2010
- 1.1 Trade/Proprietary Name
- 1.2 Common/Usual Name
- 1.3 Classification Name
1.4 Classification
EXELINT Aluminum Hub Blunt Needles Blunt Fluid Transfer Needle Needle, hypodermic, single lumen
- Class: Il Panel: 80 Product Code: FMI Cite: 21 CFR 880.5470
- 1.5 Description
The EXELINT Aluminum Hub Blunt Needles are sterile, non-pyrogenic, single use fluid transfer needles consisting of blunt Type 304 stainless steel cannulas staked into aluminum luer lock hubs. The gauge and length of the cannula determines the final gauge and length of the product. The hubs are identical for all of the needles. Each needle is packaged into a Polypropylene tube and sealed with a polypropylene cap. The needles are final packaged into a multineedle box and ETO sterilized.
#### 1.6 Indication for Use
The product is intended to be used in a variety of medical, pharmaceutical and laboratory procedures to extract fluid or medication from vials or ampoule for transfer or for irrigation.
- 1.7 Substantial Equivalence
The devices are substantially equivalent to the MONOJECT® Rigid Pack Blunt Cannula, cleared as the Monoject Sterile M200 Aluminum Hub Blunt Cannula Device under 510(k) file number K854547.
#### 1.8 Technological Characteristics
The EXELINT Aluminum Hub Blunt Needles are constructed of the same materials (aluminum and Stainless steel), have the same dimension (lengths, gauges), are sterilized with EtO in packaging of the same materials, design and colors, and comply with the same standards as the currently marketed predicate products.
#### 1.9 Performance Data
The devices were tested and found in compliance with the ISO standards for Needles (ISO 7864) and Stainless Steel Tubing (ISO 9626). The packaging was tested for seal strength. These tests demonstrated that the EXELINT Aluminum Hub Blunt Needles are safe and effective and that their performance meets the requirements of their pre-defined acceptance criteria and intended use.
- 1.10 Conclusion
EXELINT International concludes based on the information presented that the new products lines are substantially equivalent to products currently legally marketed in the USA.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Armand Hamid Director Exelint International Company 5840 W. Centinela Avenue Los Angeles, California 90045
UCI 2 0 200
Re: K101309
Trade/Device Name: EXELINT Aluminum Hub Blunt Needles : Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: September 27, 2010 Received: September 30, 2010
Dear Mr. Hamid:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however .that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. : ·
{2}------------------------------------------------
## Page 2- Mr. Hamid
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
OCT 2 0 2010
| 510(k) Number: | K101309 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | EXELINT Aluminum Hub Blunt Needles |
| Indications for Use: | The EXELINT Aluminum Hub Blunt Needles are intended to<br>be used in a variety of medical, pharmaceutical and<br>laboratory procedures to extract fluid or medication from<br>vials or ampoule for transfer or for irrigation. |
Prescription Use (Per 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ri C. A 10/20/10
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.