← Product Code [LHI](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LHI) · K070584

# SMART-ROD (K070584)

_Duoject Medical Systems, Inc. · LHI · Apr 2, 2008 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LHI/K070584

## Device Facts

- **Applicant:** Duoject Medical Systems, Inc.
- **Product Code:** [LHI](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LHI.md)
- **Decision Date:** Apr 2, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Smart-Rod™ is a single use reconstitution device intended for the transfer of a diluent contained in a prefilled syringe to a vial containing a drug product, and the return transfer of the admixture to the syringe. The Smart-Rod™ is intended for use by physicians, nurses, and other practitioners who routinely administer injections, or patients and other individuals authorized by a physician to administer injections of prescribed medication.

## Device Story

Smart-Rod™ is a single-use reconstitution device comprising a vial socket and plunger rod; replaces standard plunger rod of pre-filled syringe. Device facilitates fluid transfer between pre-filled syringe and drug vial. Integral needle perforates syringe plunger to create fluid path; vial socket attaches to drug vial. After reconstitution, admixture is transferred back to syringe; needle retracted into plunger rod; vial socket and vial discarded. Used in clinical or home settings by physicians, nurses, or authorized patients. Simplifies reconstitution process; maintains sterility; enables injection using syringe's own needle.

## Clinical Evidence

Bench testing only. 16 qualification tests performed to verify operational performance, including attachment to vials and syringes, interaction with vial stoppers and syringe plungers, and overall device integrity during use.

## Technological Characteristics

Materials: polycarbonate, polyethylene, ABS, and steel. Single-use, sterile, manual fluid transfer device. Consists of vial socket and plunger rod with integral needle. Mechanical operation; no energy source or software.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Clip'n'Ject® Reconstitution System ([K041691](/device/K041691.md))
- Bioject® Needle-Free Vial Adapter ([K963012](/device/K963012.md))
- Bioject® Needle-Free Vial Adapter ([K043304](/device/K043304.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) SUMMARY 5.

K070.584

APR - 2 2008

| Submitter's Name:    | Duoject Medical Systems Inc.                                                                             |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Address:             | 50 chemin de Gaspe, Complex B-5                                                                          |
|                      | Bromont, Quebec                                                                                          |
|                      | Canada J2L 2N8                                                                                           |
| Telephone:           | (450) 534-3666 ext.24                                                                                    |
| Fax:                 | (450) 534-3700                                                                                           |
| Contact Person:      | Maud Richard                                                                                             |
| Date of Summary:     | April 1, 2008                                                                                            |
| Trade Name:          | Smart-Rod™ Reconstitution System                                                                         |
| Common Name:         | Reconstitution system                                                                                    |
| Classification Name: | Fluid Transfer Set (includes vial adapter) per 21 CFR                                                    |
|                      | 880.5440                                                                                                 |
| Predicate Device:    | Clip'n'Ject® Reconstitution System (K041691)<br>Bioject® Needle-Free Vial Adapter (K963012 &<br>K043304) |

#### Device Description:

The Smart-Rod™ is a single-use reconstitution device consisting of a preassembled vial socket and plunger rod. The Smart-Rod™ is designed to be used together with a pre-filled syringe. The pre-filled syringe would be legally authorized through its own clearance or customer New Drug Application and is not intended to be cleared as part of this premarket notification. The Smart-Rod™ plunger rod replaces the usual plunger rod of the pre-filled syringe and has an integral needle designed to perforate the plunger of the pre-filled syringe for fluid transfer from the syringe to the drug vial (connected using the vial socket) through the internal fluid path in the rod. Once reconstitution is complete, the device permits the transfer of the reconstituted drug product back to the The integral needle is retracted into the plunger rod by unscrewing the syringe. The vial socket and vial are discarded, and the filled syringe vial socket. containing reconstituted drug is ready for use and the drug product may be injected by means of the yet unexposed needle equipping the syringe.

pg 1 + 2

{1}------------------------------------------------

## Intended Use:

The Smart-Rod™ is a single use reconstitution device intended for the transfer of a diluent contained in a prefilled syringe to a vial containing a drug product, and the return transfer of the admixture to the syringe.

The Smart-Rod™ is intended for use by physicians, nurses, and other practitioners who routinely administer injections, or patients and other individuals authorized by a physician to administer injections of prescribed medication.

## Substantial Equivalence:

The Smart-Rod™ has the same intended use as the predicate devices, with the primary purpose of permitting the transfer of a diluent contained in a syringe to a vial containing a drug product, in order to allow drug reconstitution prior to injection. All devices are single-use and intended to be marketed sterile in individual packaging.

It combines design concepts found in the predicate devices for vial attachment, vial closure piercing and use of the syringe plunger to open a fluid path.

The Smart-Rod™ device is similar in design to Bioject®'s Needle-Free Vial Adapter and West's Clip'n'iect® with respect to the mechanism of attachment to a vial. Its vial closure piercing mechanism is identical to that of Bioject®'s Needle-free Vial Adapter, however unlike the Bioject® Needle-Free vial adapter, the Smart-Rod™ is intended for use with a syringe equipped with a needle, like the West Clip'n'ject® device.

Construction materials include polycarbonate, polyethylene, ABS and steel. Predicate devices use polycarbonate, polyethylene and steel in the operation of the device. ABS is a well established material for other devices of similar intended use.

## Performance testing:

16 qualification tests were conducted to demonstrate the operational performance of the device, including attachment to vial and syringe, interaction with the vial stopper and the syringe plunger, operational performance and integrity during use.

070584 pg 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in a simple, bold design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 2 2008

Ms. Maud Richard Compliance Engineer Duoject Medical Systems, Incorporated 50 rue de Gaspe, Complex B-5 Bromont, (Quebec) CANADA J2L 2N8

Re: K070584

Trade/Device Name: Smart-Rod Reconstitution System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: March 12, 2008 Received: March 17, 2008

Dear Ms. Richard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Richard

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 4. INDICATIONS FOR USE STATEMENT - -------------------------------------------------------------------------------------------------------------------------------------------

510(k) Number (if known): K070584 Device Name: Smart-Rod™

Indications for Use:

The Smart-Rod™ is a single use reconstitution device intended for the transfer of a diluent contained in a prefilled syringe to a vial containing a drug product, and the return transfer of the admixture to the syringe.

The Smart-Rod™ is intended for use by physicians, nurses, and other practitioners who routinely administer injections, or patients and other individuals authorized by a physician to administer injections of prescribed medication.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ Prescription Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Ams 20

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control. Dental Devices

510(k) Number: 1 1 < 4 20194

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LHI/K070584](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LHI/K070584)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
