The Amsco Warming Cabinet is designed to raise the temperature of blankets, linens and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications.
Device Story
Amsco Warming Cabinet is a wall-mounted, pass-thru thermal storage unit for hospital/surgical outpatient use. Device features upper and lower compartments with front and rear doors, allowing loading from outside the operating room and access from inside. Heating system utilizes electric heater and fan blower (convection). Controls include digital push-button keypad, LCD temperature display, and over-temperature alarm. Device maintains set temperatures (90°F-160°F) for IV/irrigation solutions and linens. Healthcare providers use the cabinet to ensure supplies are at an acceptable temperature for patient care, potentially improving patient comfort and clinical outcomes. Safety features include temperature lock-out, door ajar indicators, and automatic heater shut-off during over-temperature conditions.
Clinical Evidence
Bench testing only. Verification and validation activities included electrical safety testing (dielectric withstand, leakage current, ground bond), system function testing, and thermal performance testing (empty and full load conditions). Results confirmed compliance with UL 61010-1 and CAN/CSA C22.2 61010-1 standards. Heating performance met set-point accuracy requirements (±3°F upper, ±5°F lower) and reached set-points within specified timeframes.
Technological Characteristics
Materials: Stainless steel and laminated galvanized steel. Heating: Electric heater and fan blower (convection). Dimensions: 24" wide, recessed wall-mount. Connectivity: Data port for retrieval of stored temperatures. Power: 110/120 Vac, 50/60 Hz. Controls: Digital push-button keypad, LCD display. Safety: Over-temperature alarm, temperature lock-out, manual mechanical door locks. Standards: UL 61010-1, CAN/CSA C22.2 61010-1.
Indications for Use
Indicated for warming blankets, linens, sterile surgical irrigation solutions, and IV solutions for use in surgical, obstetrical, emergency, critical care, and other healthcare settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
JUL 2 9 2011
KI11870
STERIS®
## 510(k) Summary For Amsco Warming Cabinet 24" Wide Pass-thru Model
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact:
Robert F. Sullivan -Senior Director, Regulatory Affairs
| Telephone: | 440 392 7695 |
|------------|--------------|
| Fax No: | 440 357 9198 |
June 30, 2011 Summary Date:
STERIS Corporation ■ 5960 Heisley Road ■ Mentor. OH 44060-1834 USA ■ 440-354-2600
{1}------------------------------------------------
#### Device Name 1.
| Trade Name: | Amsco Warming Cabinet, 24" Wide Pass-thru-<br>Model |
|----------------------|--------------------------------------------------------------------|
| Common/usual Name: | Warming Cabinet |
| Classification: | Unclassified |
| Classification Name: | Warmer, Thermal, Infusion Fluid; Unclassified,<br>Product Code LGZ |
#### Predicate Device 2.
Amsco Warming Cabinet (K092823)
#### Description of Device 3.
The Amsco Warming Cabinet is designed to store and warm sterile IV solutions, surgical irrigation solutions, linens and/or blankets to an acceptable level for hospital and surgical outpatient center applications.
The 24" Wide Pass-thru model will be mounted in a wall opening. Doors (upper and lower compartment) added to the rear of the cabinet will allow users to load supplies in compartments from outside the operating room, while doors on the front side will allow access to compartments when inside the operating room.
The upper compartment of this model holds up to 20( 1-liter) liquid bottles, or 6 ( 1liter) liquid (IV or Irrigation solution) bags; the lower compartment holds up to 40 (1-liter) liquid bottles or 12 (1 liter) liquid bags.
#### 4. Intended Use
The Amsco Warming Cabinet is designed to raise the temperature of blankets, linens and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications.
{2}------------------------------------------------
#### Description of Safety and Substantial Equivalence રાં
Substantial Equivalence
The 24" Wide Pass-thru Amsco Warming Cabinet is identical in technology and intended use to the predicate Amsco Warming Cabinet models. A table comparing the technological characteristics of the proposed Amsco Warming Cabinet to the predicate is provided in Table 8-1.
路站,同: 上海: 下:: 上一篇: 上一篇: 上一篇: 上一篇:
| Features | PREDICATE.<br>Amsco Warming Cabinet.<br>(K092823) | PROPOSED<br>Amsco Warming Cabinet.<br>24" Wide Pass-thru Model |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Amsco Warming Cabinet is<br>designed to raise the temperature<br>of blankets, linens and sterile<br>surgical irrigation solutions and<br>IV solutions to an acceptable<br>level for various surgical,<br>obstetrical, emergency, critical<br>care and other healthcare<br>applications. | Identical |
| Heating System | Electric heater and fan blower<br>(Convection heating) | Identical |
| Unit Configuration | Single/Double chamber | Double chamber |
| Unit Depth | 18" or 24" | 18" |
| Model | Freestanding, mobile base (24"<br>dual chamber only), or Counter<br>(single chamber only) | Freestanding |
| Interior and<br>Exterior Surfaces | Stainless Steel, ABS Plastic and<br>laminated galvanized steel | Stainless Steel and laminated<br>galvanized steel |
| Installation | Free-Standing, Recessed, Mobile<br>(24" dual chamber only), or<br>Counter (single chamber only) | Recessed only |
| Door | Laminated steel exterior and<br>Stainless Steel interior (Solid and<br>Glass) | Stainless steel exterior and<br>interior (Solid only) |
| Features | PREDICATE:<br>Amsco Warming Cabinet<br>(K092823) | PROPOSED:<br>Amsco Warming Cabinet<br>24"-Wide-Pass-thru-Model |
| Cabinet Storage<br>Capacity and<br>Volume | 30" wide by 18" deep upper / single chamber - 3.2 cu ft - up to 24 (1-liter) bottles lower chamber - 8.5 cu ft - up to 72 (1-liter) bottles | Upper chamber - 2.1 cu ft - up to 20 (1-liter) bottles<br>Lower chamber - 4.1 cu ft - up to 40 (1-liter) bottles |
| | 30" wide by 24" deep upper / single chamber - 4.3 cu ft - up to 30 (1-liter) bottles lower chamber - 11.6 cu ft - up to 90 (1-liter) bottles | |
| | 30" wide by 18" deep OR Console upper chamber - 3.3 cu ft - up to 24 (1-liter) bottles lower chamber - 6.2 cu ft - up to 48 (1-liter) bottles | |
| | 24" wide by 18" deep OR Console upper chamber - 2.4 cu ft - up to 20 (1-liter) bottles lower chamber - 4.7 cu ft - up to 40 (1-liter) bottle | |
| Controls | Digital Push Button keypad / power switch / Digital LCD temperature display / mode selection buttons / door ajar indicator / Over-temperature light for each compartment / Data port for retrieval of stored temperatures. | Identical |
| Software | Unit contains software | Identical - no changes were made to the software |
| Temperature<br>Selection Range | 90°F (32°C) to 160°F (71°C) | Identical |
| Temperature Lock | Temperature lock-out function to prevent unauthorized temperature changes. | Identical |
| Features | PREDICATE<br>Amsco Warming Cabinet<br>(K092823) | PROPOSED<br>Amsco Warming Cabinet<br>24"-Wide-Pass-thru-Model |
| Door Lock | All configurations will be<br>equipped with either a manual<br>mechanical door lock or optional<br>electronic door lock system for<br>each compartment | All configurations will be<br>equipped with a manual<br>mechanical door lock for each<br>door on each compartment |
| Over Temperature<br>Alarm Point | Visual and audible alarm if unit<br>has a chamber temperature<br>greater than 10°F (5.5°C) above<br>set temperature. In the event of an<br>over temp condition, the controls<br>automatically turn off the<br>heater(s). | Identical |
| Voltage<br>Requirements | 110/120 Vac, 220/240 Vac<br>nominal,<br>50/60 HZ | 110/120 Vac, 50/60 HZ |
Table 8-1: Summary of the Proposed Device and Predicate Devices Technological Characteristics
{3}------------------------------------------------
## STERIS SPECIAL 510(k) – DEVICE MODIFICATION TO K092823
AMSCO WARMING CABINET – 24" WIDE PASS-THRU MODEL
1
1
{4}------------------------------------------------
#### STERIS SPECIAL 510(k) - DEVICE MODIFICATION TO K092823 AMSCO WARMING CABINET - 24" WIDE PASS-THRU MODEL
### Safety and Effectiveness
Mechanical, electrical and heating performance testing demonstrate that the proposed Amsco Warming Cabinet, 24" Wide Pass-thru model, operates as intended and is as safe and effective as the predicate. Table 8-2 summarizes the verification and validation activities that were performed to ensure that rorifications do not affect the safety or effectiveness of the Amsco Warming Cabinet.
| Table 8-2: Summary of Verification and Validation Activities | |
|---------------------------------------------------------------|--|
| for 24" Wide Pass-thru Warming Cabinet | |
| Test Description | Acceptance Criteria | Results |
|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Electrical Testing | Dielectric Withstand<br>• Achieve passing results thru automated dielectric<br>strength tester.<br>Earth and Enclosure Leakage Current<br>• Must not exceed requirements listed in UL 61010-1<br>clause 6.3.1b (must not exceed .5mA normal<br>condition and 3.5mA single fault condition)<br>Ground bond<br>• Must not exceed 0.1 ohm and maintain ground<br>continuity.<br>Power Input & Line Voltage<br>• The power input of equipment at rated voltage and<br>steady state current shall not exceed the marked<br>rating by 10%. | PASS |
| Test Description | Acceptance Criteria | Results |
| System Functions | • Main Power ON indicator at rear upper door is - illuminated when Main Power Switch is set to the ON position.<br>• Exhaust fans and blowers must operate continuously when the main power is set to the ON position<br>• Opening or closing the doors shall not cause fans or blowers to turn off<br>• Input current must be within 11.4 - 14.25 Amps.<br>• Over-temp alarm must activate and displayed temperature flashes alternately with the error "Hi" when compartment temperature is more than 10°F above set point temperature<br>• Door must remain locked when 40 pounds of force is applied | PASS |
| Thermal Test | Temperature Limits (recessed mounted, normal condition)<br>• Temperatures of easily touched surfaces and identified components must comply with UL 61010- 1 clauses 10.1 and 10.2, UL 61010-2-010 clause 10.1, temperature limits under normal conditions.<br>• Maintain normal operation without faults occurring. | PASS |
| Single Fault Conditions | Cooling (single fault condition)<br>• Pass dielectric strength test per UL 61010-1, section 6, clause 6.8<br>• Cabinet outer surface must not exceed 105°C at 40°C ambient<br>• Transformer, blower and exhaust fan windings must not exceed 150°C under single fault condition at 40°C ambient<br>• Voltage measurement of cabinet outer surfaces must not exceed test voltages per UL 61010-1, section 6, clause 6.3.2a<br>• No signs of molten metal, burning insulation, flaming particles, etc.<br>• No signs of charring, glowing, or flaming of tissue paper or cheesecloth.<br>Heating Devices<br>• Pass dielectric strength test per UL 61010-1, section 6, clause 6.8<br>• Cabinet outer surface must not exceed 105°C at 40°C ambient<br>• Transformer, blower and exhaust fan windings must not exceed 150°C under single fault condition at 40°C ambient<br>• Voltage measurement of cabinet outer surfaces must not exceed test voltages per UL 61010-1, section 6, clause 6.3.2a<br>• No signs of molten metal, burning insulation, flaming particles, etc.<br>• No signs of charring, glowing, or flaming of tissue paper or cheesecloth. | PASS |
| Test.Description | Acceptance Criteria | Results |
| Protection against<br>Hazards from Fluids | - Pass dielectric strength test per-UL.61010-1, section.<br>6, clause 6.8<br>- Voltage measurement of cabinet outer surfaces must<br>not exceed test voltages per UL 61010-1, section 6,<br>clause 6.3.2a. | PASS |
| Heating Performance<br>(empty compartment) | - Temperature reading of each thermocouple must not<br>vary from set-point temperature by more than $\pm$ 3°F in<br>upper compartment and $\pm$ 5°F in lower compartment.<br>- At a 160°F set-point temperature, the upper<br>compartment display must reach set-point within 35<br>minutes and lower compartment display must be<br>within 65 minutes.<br>- Normal system operation is maintained without over-<br>temperature condition or faults occurring. | PASS |
| Heating Performance<br>(Full IV solution load in<br>upper compartment and<br>full blank load in lower<br>compartment) | - Temperature reading of each thermocouple must not<br>vary from set-point temperature by more than $\pm$ 3°F in<br>upper compartment and $\pm$ 5°F in lower compartment.<br>- Upper and lower compartment controls temperature<br>display must reach set-point within 12 hours.<br>- Normal system operation is maintained without over-<br>temperature condition or faults occurring. | PASS |
| Heating Performance<br>(Full irrigation solution<br>load in upper and lower<br>compartments | - With a fully loaded compartment of Irrigation<br>solution, bottles of solution shall not overbalance and<br>fall off shelf when doors are opened.<br>- Temperature reading of each thermocouple must not<br>vary from set-point temperature by more than $\pm$ 3°F in<br>upper compartment and $\pm$ 5°F in lower compartment<br>- Upper and lower compartment controls temperature<br>display must reach set-point within 12 hours.<br>- Normal system operation is maintained without over-<br>temperature condition or faults occurring. | PASS |
| ETL/cETL Code<br>Compliance | - Meet UL 61010-1 Standard for Safety Electrical<br>Equipment for Measurement, Control, and<br>Laboratory Use.<br>- Meet CAN/CSA C22.2 61010-1 Standard for Safety<br>Electrical Equipment for Measurement, Control and<br>Laboratory Use. | PASS |
{5}------------------------------------------------
## STERIS SPECIAL 510(k) – DEVICE MODIFICATION TO K092823
AMSCO WARMING CABINET – 24" WIDE PASS-THRU MODEL
{6}------------------------------------------------
#### STERIS SPECIAL 510(k) - DEVICE MODIFICATION TO K092823 AMSCO WARMING CABINET - 24" WIDE PASS-THRU MODEL
### Conclusion
Verification and validation testing demonstrate that the proposed Amsco Warming Cabinet, 24" Wide Pass-thru model, operates as intended and is as safe and effective as the predicate. The differences between the proposed and predicate device are limited to the described modifications of the device and these proposed changes raise no new concerns of safety and effectiveness when compared to the predicate device. The proposed Amsco Warming Cabinet, 24" Wide Pass-thru model, is substantially equivalent to the predicate.
{7}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the top half of the circle. Inside the circle is a stylized image of an eagle or bird-like figure with three tail feathers.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Robert F. Sullivan Senior Director, Regulatory Affairs Steris Corporation 5960 Heisley Road Mentor, Ohio 44060
## JAN 1 0 2017
Re: K111870
Trade/Device Name: Amsco Warming Cabinet Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LGZ Dated: June 30, 2011 Received: July 1, 2011
Dear Mr. Sullivan:
This letter corrects our substantially equivalent letter of July 29, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting (reporting of
{8}------------------------------------------------
Page 2 - Mr. Robert F. Sullivan
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name:
Amsco Warming Cabinet
Indications For Use:
The Amsco Warming Cabinet is designed to raise the temperature of blankets, linens and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ian Sti for Richard Chapman
(Division Sign-Off)
Division of Sign Only
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111876
Page 1 of 1
June 30, 2011
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.