THERMOSTAT 900 ADMINISTRATION SET, MODEL ASH 100, 200 AND 300

K073363 · Paladin Biomedical Corporation · LGZ · Mar 20, 2008 · General Hospital

Device Facts

Record IDK073363
Device NameTHERMOSTAT 900 ADMINISTRATION SET, MODEL ASH 100, 200 AND 300
ApplicantPaladin Biomedical Corporation
Product CodeLGZ · General Hospital
Decision DateMar 20, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Paladin Biomedical Corporation ThermoStat 900 Administration Set is intended to be used with the MMS Thermostat 900 Blood and Fluid Warmer. The disposable cartridge and administration set is intended to deliver high volume, rapid administration of blood and physiological fluids when connected to a large bore catheter. The set is intended to help prevent hypothermia in patients receiving large volumes of fluids.

Device Story

Disposable administration set (models ASH200, ASH300) used with MMS Thermostat 900 Blood and Fluid Warmer; delivers high-volume, rapid blood/fluid infusion via large-bore catheter. Components include tubing, spikes, clamps, luers, drip chambers, filter/vent assembly, and cartridge. Used in clinical settings to prevent hypothermia during rapid fluid resuscitation. Healthcare providers connect set to warmer and patient access site; device facilitates thermal regulation of fluids before delivery to patient.

Clinical Evidence

Bench testing only.

Technological Characteristics

Disposable administration set; materials are biocompatible. Components: tubing, spikes, clamps, luers, drip chambers, filter/vent assembly, cartridge. Configurations: ASH200 (2-spike) and ASH300 (3-spike).

Indications for Use

Indicated for patients requiring high volume, rapid administration of blood and physiological fluids to prevent hypothermia.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the shape of the bird's body and wings. The text is in all capital letters and is evenly spaced around the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Mr. Robert Bickford COO Paladin Biomedical Corporation 506 Boston Post Road Weston, Massachusetts 02493 # JAN 1 0 2017 Re: K073363 Trade/Device Name: ThermoStat 900 Administration Set, Models ASH200 and ASH300 Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LGZ Dated: March 10, 2008 Received: March 13, 2008 Dear Mr. Bickford: This letter corrects our substantially equivalent letter of March 20, 2008. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting of {1}------------------------------------------------ Page 2 - Mr. Robert Bickford medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Image /page/1/Picture/8 description: The image shows the name "Erin I. Keith -S" in a bold, sans-serif font. The letters are black against a white background. The name appears to be a signature or label. Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: Paladin Biomedical Corporation ThermoStat 900 Administration Set Indications For Use: Paladin Biomedical Corporation ThermoStat 900 Administration Set is intended to be used with the MMS Thermostat 900 Blood and Fluid Warmer. The disposable cartridge and administration set is intended to deliver high volume, rapid administration of blood and physiological fluids when connected to a large bore catheter. The set is intended to help prevent hypothermia in patients receiving large volumes of fluids. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cintan (Division Sign-Off) Division of Anesthesiology. General Hospital Infection Control, Dental Devices Page 1 of 510(k) Number: _______________________________________________________________________________________________________________________________________________________________ 000011 {3}------------------------------------------------ ### MAR 2 0 2008 K073363 (A. lotz) 510K Summary of Safety and Effectiveness Paladin Biomedical Corporation ThermoStat 900 Administration Set Models ASH 200 and ASH 300 506 Boston Post Road Weston, MA 02493 March 18, 2008 - 1. Sponsor Name Paladin Biomedical Corporation 506 Boston Post Road Weston, MA 02493 - 2. Device Name Proprietary Name: ThermoStat 900 Administration Set Common/Usual Name: warmer, thermal, infusion fluid - 3. Identification of Predicate or Legally Marketed Device BK970045 Microwave Medical System, 310-312 School St., Acton, MA 01720 Trade Name: ThermoStat 900 Blood & IV Fluid Adm. Set Cleared Date: 06-MAR-1998 #### 4. Device Description There are two configurations of the administration set the ASH200 (High flow - 2 spike) and ASH300 (high flow - 3 spike). Materials of construction are all known biocompatible materials. The set consists of three major sub assemblies: - tubing and administration set consisting of tubing, spikes, clamps, luers, . drip chambers - filter/vent assembly . - cartridge assembly. . - న. Intended Use Paladin Biomedical Corporation ThermoStat 900 Administration Set is intended to be used with the MMS Thermostat 900 Blood and Fluid Warmer. The disposable cartridge and administration set is intended to deliver high volume, rapid administration of blood and physiological fluids when connected to a large bore catheter. The set is intended to help prevent hypothermia in patients receiving large volumes of fluids. #### Comparison of Technological Characteristics ર્ભ. The Paladin Biomedical Administration Set and the MMS Administration Sets were compared with respect to various attributes (Dimensions, warming method, filter size, flow rate, and maximum pressure) and determined to be the same with respect to materials, design, specifications, construction, and performance. {4}------------------------------------------------ K 473363 (f. 20f2) - 7 Performance Testing Bench testing was performed to determine equivalency and performance aspects of the device. - 8. Statement of Equivalency The Paladin System is substantially equivalent in design, materials, construction and intended use as that of the predicate. Since the Paladin device is the same in intended use and technological characteristics as the predicate device, the Paladin device does not raise any new safety and efficacy concerns when compared to these similar legally marketed devices. The descriptive characteristics demonstrate that the Paladin device is substantially equivalent to the predicate device and is capable of safely and accurately performing the stated intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...