K190305 · International Medical Industries, Inc. · KPE · Apr 30, 2019 · General Hospital
Device Facts
Record ID
K190305
Device Name
Additive Cap
Applicant
International Medical Industries, Inc.
Product Code
KPE · General Hospital
Decision Date
Apr 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5025
Device Class
Class 2
Indications for Use
The Additive Port Cap is indicated for use on the medication port of Baxter VIAFLEX, Baxter INTRAVIA, Baxter AVIVA, and Baxter ALL-IN-ONE EVA containers to provide both visual evidence that medication has been added and tamper evidence once the device is closed.
Device Story
Additive Port Cap (APC) is a single-use, non-sterile, polypropylene cap designed to snap over medication ports of specific Baxter IV containers. Device provides visual indication of medication addition via bright red coloration; functions as tamper-evident seal once closed. If forced open or removed, device causes visible damage to IV container or cap, rendering it unusable. Used in clinical settings by healthcare personnel to secure medication ports after additive injection. Benefits include enhanced medication safety through visual confirmation and tamper deterrence.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included visual inspection, axial detachment force, opening/closing force, dimensional verification, leakage testing, packaging/labeling integrity, simulated transportation, shelf-life testing, and human factors/usability validation. All pre-determined acceptance criteria were met.
Technological Characteristics
Material: Polypropylene. Form factor: Snap-on cap. Energy source: None (mechanical). Connectivity: None. Sterilization: Non-sterile. Software: None.
Indications for Use
Indicated for use on medication ports of Baxter VIAFLEX, INTRAVIA, AVIVA, and ALL-IN-ONE EVA IV containers to provide visual confirmation of medication addition and tamper evidence.
Regulatory Classification
Identification
An I.V. container is a container made of plastic or glass used to hold a fluid mixture to be administered to a patient through an intravascular administration set.
Predicate Devices
Baxter Additive Cap for Viaflex Containers (K111217)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 30, 2019
International Medical Industries, Inc. Peter Lehel Senior Manager Engineering 2981 Gateway Drive Pompano Beach, Florida 33069
Re: K190305
Trade/Device Name: Additive Cap Regulation Number: 21 CFR 880.5025 Regulation Name: IV Container Regulatory Class: Class II Product Code: KPE Dated: February 7, 2019 Received: February 12, 2019
Dear Peter Lehel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190305
Device Name Additive Port Cap
Indications for Use (Describe)
The Additive Port Cap is indicated for use on the medication port of Baxter VIAFLEX, Baxter AVIVA, and Baxter ALL-IN-ONE EVA containers to provide both visual evidence that medication has been added and tamper evidence once the device is closed.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------|
| □ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for IMI. The logo consists of a blue square with three curved white lines inside, followed by the letters "IMI" in black. There is a registered trademark symbol to the upper right of the last "I" in the logo.
# K190305 510(k) Summary
| Date Summary Prepared: | February 7, 2019 | |
|-------------------------|--------------------------------------------------|--|
| Manufacturing Company: | International Medical Industries, Inc. | |
| Manufacturing Address: | 2981 Gateway Drive, Pompano Beach, Florida 33069 | |
| Corresponding Official: | Peter Lehel | |
| | Senior Manager Engineering | |
| Telephone Number: | 954.917.9570 x 227 | |
| Fax Number: | 954.917.9244 | |
| Email Address: | plehel@imiweb.com | |
| Trade Name: | Additive Port Cap | |
| Device Common Name: | Additive Cap | |
| Regulation Number: | 21 CFR 880.5025 | |
| Regulation Name: | I.V. Container | |
| Product Class: | II | |
| Product Panel: | General Hospital | |
| Product Code: | KPE | |
| Predicate Device: | Baxter Additive Cap for Viaflex Containers | |
| | K111217 cleared June 14, 2011 | |
### 5.1 Device Description
The Additive Port Cap (APC) is a polypropylene, single use device designed to snap over the outside of the medication port of compatible Baxter IV container (Baxter VIAFLEX, Baxter INTRAVIA, Baxter AVIVA, and Baxter ALL-IN-ONE EVA) after the addition of medication. Once closed, the device prevents the port from being accessed without causing visible damage to the IV container. The bright red coloration serves as a clear indicator that medication has been added. The APC device is non-fluid path and non-sterile. The APC device is marketed as a stand-alone device and packaged in bulk.
## 5.2 Indications for Use
| Device:<br>Additive Port Cap (APC) | Predicate:<br>Baxter Additive Cap for Viaflex<br>Containers K111217 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The Additive Port Cap is indicated for use<br>on the medication port of Baxter<br>VIAFLEX, Baxter INTRAVIA, Baxter<br>AVIVA, and Baxter ALL-IN-ONE EVA<br>containers to provide both visual evidence<br>that medication has been added and<br>tamper evidence once the device is<br>closed. | Additive Cap is indicated for use on the<br>medication port of VIAFLEX and<br>AVIVA containers to provide visual<br>evidence that medication has been<br>added. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the IMI logo. The logo consists of a blue square with three curved lines inside, followed by the letters "IMI" in black. The "I" on the right has a registered trademark symbol next to it. The logo is simple and modern.
#### 5.3 Technological Characteristics and Basis for Substantial Equivalence
The APC is substantially equivalent to the predicate device, Baxter Additive Cap for VIAFLEX and AVIVA Containers (K111217). The devices are both indicated as non-fluid path caps which provide visual evidence that medication has been added to the specified IV containers. Both devices are injection molded polypropylene and are provided non-sterile.
| Parameter | Proposed Device | Predicate Device |
|---------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|
| Image | Image: Proposed Device Images: Injection port fit. and Installed. | Image: Predicate Device Images: Injection port fit. and Installed. |
| Proprietary Device<br>Name | Additive Port Cap (APC) | Additive Cap for Viaflex Containers |
| Company Name | International Medical Industries, Inc. | Baxter Healthcare Corporation |
| 510(k) # | K190305 | K111217 |
| Product Code | KPE | KPE |
| Regulation No. | 21 CFR 880.5025 | 21 CFR 880.5025 |
| Classification | Class II | Class II |
| Sterility | Non-Sterile | Non-Sterile |
| Number of Uses | Single Use | Single Use |
| Cap Material | Polypropylene | Polypropylene |
| Biocompatibility<br>Contact and<br>Duration | Device does not contact the patient's body<br>directly or indirectly. | Device does not contact the patient's body<br>directly or indirectly. |
#### Table 5.3.1 - Comparison Between APC and Predicate Devices
The label modification is to expand the indications for use statement by adding compatibility with the Baxter INTRAVIVA and Baxter ALL-IN-ONE EVA containers which are functionally equivalent to the VIAFLEX and AVIVA. The second label modification is to add tamper evidence to the indications for use. Performance testing to demonstrate tamper evidence was conducted. Once the Additive Port Cap is attached and closed, it cannot be opened manually. If a closed cap is opened (e.g. with a tool), it can cause damage to the bag or the cap making it unusable. Additionally, performance testing demonstrated that pulling on the cap in order to remove from the bag tears the bag.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for IMI. The logo consists of a blue square with white curved lines inside, followed by the letters "IMI" in black. There is a registered trademark symbol to the right of the last "I".
| Feature of the Device | Proposed Device | Predicate Device | Discussion / Comment |
|------------------------------------------------|-------------------------------------------------------------------------------------------------|--------------------------------------|------------------------------------------------------------------------------|
| Compatibility with Baxter<br>fluid containers. | Baxter VIAFLEX,<br>Baxter INTRAVIA,<br>Baxter AVIVA, and<br>Baxter ALL-IN-ONE<br>EVA containers | VIAFLEX and<br>AVIVA containers | The medication port are<br>equivalent for all the<br>containers. |
| Tamper Evidence | Includes tamper<br>evidence | Does not include<br>tamper evidence. | Performance testing was<br>conducted to support claim<br>of tamper evidence. |
#### Table 5.3.2 - Differences Between APC and Predicate Devices
The safety and effectiveness of the APC are adequately supported by the material information, comparison of design characteristics with the predicate device, testing rationale and data provided within this premarket notification. This testing supports the substantial equivalence of the APC to the predicate and shows that no different questions of safety and effectiveness have been introduced with this device.
### 5.4 Performance Data
Design verification and validation were performed to ensure that the APC device meets its performance specifications and demonstrates substantial equivalence to the predicate device. There are no known performance standards for this device. A summary of the performance testing conducted is presented below. All pre-determined acceptance criteria were met. The data demonstrates that the proposed cap device is substantially equivalent to the predicate device.
| Test Name | Test Description |
|--------------------------|---------------------------------------------------------------------------------|
| Visual Inspection | Visual Inspection to ensure no gross damage, coverage<br>of drug port etc. |
| Axial Detachment | Force required to remove the additive port cap from IV<br>container |
| Opening Force | Force required to open a properly installed cap |
| Closing Force | Force required to properly install product on injection<br>port of IV container |
| Dimensional Verification | Dimensional measurements of cap features |
| Leakage | Drug port freedom from leakage |
Table 5.4.1 - Summary of Performance Testing Conducted on APC Device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for IMI. The logo consists of a blue square with white curved lines inside, followed by the letters "IMI" in black. There is a registered trademark symbol next to the last "I".
| Packaging and Labeling | Visual inspections to verify integrity of the packaging to<br>protect the device and of the label to be legible. |
|----------------------------|------------------------------------------------------------------------------------------------------------------|
| Transportation, Shelf Life | Simulated transportation conditions and shelf life<br>followed by design verification testing. |
| Product Validation | Human factors and Usability |
### 5.5 Biocompatibility Testing
The device does not contact the patient's body either directly. In accordance to FDA Guidance Document "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process; Guidance for Industry and Food and Drug Administration Staff" issued June 16, 2016, no biocompatibility testing is required for the APC device.
### Conclusion
Review of the performance test data as well as comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility demonstrate that the subject device, Additive Port Cap (APC) is substantially equivalent to the predicate cap Baxter Additive Cap for VIAFLEX and AVIVA Baxter containers K111217. Any differences between the subject and the predicate devices do not raise any different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.