SIDEKICK INFUSION KIT

K990425 · I-Flow Corp. · FRN · Apr 23, 1999 · General Hospital

Device Facts

Record IDK990425
Device NameSIDEKICK INFUSION KIT
ApplicantI-Flow Corp.
Product CodeFRN · General Hospital
Decision DateApr 23, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SideKick Infusion Kit is intended to provide continuous infusion of a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management. Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion. The SideKick pump is re-usable. The disposable SideKick administration set is single patient use only. No testing has been conducted to determine the efficacy of the SideKick for the delivery of blood, blood products, lipids or fat emulsions. The SideKick is not intended for the delivery of blood, blood products, lipids or fat emulsions. The SideKick is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.

Device Story

SideKick Infusion Kit is a mechanical, spring-powered ambulatory infusion pump system. It delivers local anesthetics at a continuous, fixed flow rate (1 or 2 ml/hr) to intraoperative sites or via percutaneous, subcutaneous, intramuscular, or epidural routes for postoperative pain management. The system consists of a reusable mechanical pump and a disposable administration set, which includes a catheter, needle, syringe, and accessories. Used in hospitals, homes, or alternative care sites, the device is operated by clinicians or patients. The pump provides a fixed flow rate without electronic components, eliminating fluid runaway risks associated with electronic pumps. The healthcare provider sets up the device to provide consistent pain relief, benefiting the patient through ambulatory postoperative pain management.

Clinical Evidence

Bench testing only. Performance data confirmed flow rate accuracy within ±15% at a 95% confidence interval under standard operating conditions (31°C, 0.9% NaCl test solution). Biological testing of fluid path components conforms to ISO 10993 Part 1.

Technological Characteristics

Mechanical spring-powered infusion pump. Disposable administration set; reusable pump. Flow rates: 1 or 2 ml/hr. Accuracy: ±15%. Sterilization: ETO gas. Fluid path materials conform to ISO 10993. No external power source required.

Indications for Use

Indicated for continuous infusion of local anesthetics for postoperative pain management in patients undergoing general surgery. Routes of administration include intraoperative (soft tissue/body cavity), percutaneous, subcutaneous, intramuscular, and epidural. Not indicated for delivery of blood, blood products, lipids, or fat emulsions.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the date 4/23/99 at the top left. Below the date is the I-FLOW logo, which is a stylized oval shape with the words "I-FLOW" inside. Underneath the logo, the words "I-FLOW CORPORATION" are printed in a simple, sans-serif font. The logo and text are centered. 20202 Windrow Drive Lake Forest. CA 92630 (800) 448-3569 (949) 206-2700 Fax (949) 206-2600 1990425 # SUMMARY OF SAFETY AND EFFECTIVENE February 9, 1999 Trade Name: SideKick Infusion Kit Common Name: Infusion Pump Kit Classification Name: Pump, Infusion All questions and/or comments concerning this document should be made to: Robert J. Bard, Esq., R.A.C. Vice President of Regulatory and Legal Affairs I-Flow Corporation 20202 Windrow Drive Lake Forest, CA 92630 Telephone: 949.206.2700 Fax: 949.206.2600 {1}------------------------------------------------ #### 1.0 GENERAL INFORMATION #### 1.1 Purpose of Submission - This submission is intended to notify the Federal Food and Drug 1.1.1 Administration that I-Flow Corporation intends to market a new kit, the SideKick Infusion Kit. - Trade Name: SideKick Infusion Kit 1.1.2 - 1.1.3 Common Name: Infusion Pump Kit - 1.1.4 Classification Name: Pump, Infusion - Classification Panel: General Hospital and Personal Use Device 1.1.5 #### 1.2 Statement of Equivalence - 1.2.1 The SideKick Infusion Kit includes components that are legally marketed (either pre-amendment devices or devices that have been granted permission to market via premarket notification regulation). - 1.2.2 The SideKick Kit is substantially equivalent to the I-Flow Paragon Infusion Kit (K984146), the I-Flow Paragon Infusion System (K923875), the I-Flow PainBuster Infusion Kit (K980558, K982946), the Sgarlato Pain Control Infusion Pump (PCIP) (K896422), the I-Flow Homepump C-Series (K944692) and the McKinley Outbound Disposable Syringe Infuser (K982256). #### 2.0 PHYSICAL SPECIFICATIONS AND DESCRIPTIONS #### 2.1 Description of the SideKick Infusion Kit - 2.1.1 The SideKick Infusion Kit is identical to the I-Flow Paragon Infusion Kit with the exception of the SideKick pump and administration set replacing the Paragon pump and administration set. - 2.1.2 The kit is comprised of a SideKick pump and administration set and various kit components such as catheter, needle, syringe, Y adapter, dressing, tape, gauze and carry case. - 2.1.2.1 The Paragon Infusion Kit contains all the above components except for a Paragon pump and administration set instead of the SideKick pump and administration set. - 2.1.3 The SideKick administration set is intended to attach to the kit catheter at the distal end of the set to provide continuous infusion of a local anesthetic directly into the intraoperative site for general surgery for postoperative pain management. - 2.1.4 The SideKick administration set is a disposable device intended for single patient use. The SideKick pump is reusable. - 2.1.5 The SideKick is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites. #### 2.2 Product Configuration - 2.2.1 The SideKick Infusion Kit models are available in 100 ml fill volumes with 1 or 2 ml/hr flow rates. - 2.2.2 Each model consists of a SideKick administration set with the following optional components/accessories: {2}------------------------------------------------ - SideKick pump, catheter, needle, syringe, dressing, carry case, 2.2.2.1 antiseptic skin swabs, tape, gauze and Y adapter. #### 2.3 Components and Materials All fluid path components of the SideKick administration set are identical to the fluid path components of the Paragon administration set. #### 2.4 Power Requirements - The SideKick pump is a mechanical pump that utilizes spring energy for 2.4.1 power. No additional external power source is required. #### 3.0 OPERATIONAL SPECIFICATIONS AND DESCRIPTIONS #### 3.1 Standard Operating Conditions: | Residual Volume: | < 5 ml | |------------------------|---------------------------------| | Operating Temperature: | 31°C skin temperature (90°F) | | Test Solution: | 0.9% NaCl | | Operating Pressure: | 9 to 1 psi pressure source | | Head Height: | 0" | | Accuracy: | ±15% at 95% confidence interval | - 3.2 Flow Rate Performance Data: Testing occurred at standard operating conditions. All models produced an average flow rate within the ±15% accuracy claim. #### 3.3 Safety / Alarm Functions - 3.3.1 The SideKick pump and administration set provide a continuous fixed flow and as such is not subject to fluid runaway conditions similar to that of some electronic pumps. ## 4.0 BIOLOGICAL SPECIFICATIONS - 4.1 Biological testing is in conformance with ISO 10993 Part 1 for all fluid path components of the SideKick administration set. ## 5.0 CHEMICAL AND DRUG SPECIFICATIONS - 5.1 Compatibility - 5.1.1 There are no specific drugs referenced in the labeling for the SideKick Infusion Kit. - 5.1.2 The SideKick Infusion Kit is intended for use with general local anesthetics and epidural medications. ## 6.0 INTENDED USE - 6.1 The SideKick Infusion Kit is intended to provide continuous infusion of a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management. - 6.2 Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion. - 6.3 The SideKick pump is re-usable. The disposable SideKick administration set is single patient use only. {3}------------------------------------------------ - 6.4 No testing has been conducted to determine the efficacy of the SideKick for the delivery of blood, blood products, lipids or fat emulsions. The SideKick is not intended for the delivery of blood, blood products, lipids or fat emulsions. - The SideKick is suitable for use as an ambulatory device and is intended for use in the 6.5 hospital, home environment or alternative care sites. #### 7.0 PACKAGING - Packaging is suitable for either radiation or ETO sterilization. 7.1 #### 8.0 STERILIZATION INFORMATION - 8.1 The method of sterilization is ETO gas. #### 9.0 COMPARISON TO LEGALLY MARKETED DEVICES - 9.1 The SideKick Infusion Kit has similar routes of administration and components as the following predicate devices: the Paragon Infusion Kit, the Paragon Infusion System, PainBuster Infusion Kit, Sgarlato Pain Control Infusion Pump (PCIP), Homepump C-Series and McKinley Outbound Syringe Infuser. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circle's perimeter. The text is in all capital letters and is arranged to follow the curve of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 3 1999 Robert J. Bard, Esq., R.A.C. Vice President Regulatory and Legal Affairs I-Flow Corporation 20202 Window Drive Lake Forest, California 92630 Re: K990425 SideKick Infusion Kit Trade Name: Regulatory Class: II Product Code: FRN Dated: February 9, 1999 February 11, 1999 Received: Dear Mr. Bard: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {5}------------------------------------------------ Page 2 - Mr. Bard This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". sincerely yours L. G. Whatrust Timot Wy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ K990425 Image /page/6/Picture/1 description: The image shows the logo for I-Flow Corporation. The logo consists of the words "I-Flow" inside of a rounded rectangle. The word "Corporation" is printed below the logo. 20202 Windrow Drive Lake Forest, CA 92630 (800) 448-3569 (949) 206-2700 Fax (949) 206-2600 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: __ SideKick Infusion Kit # Indications for Use: 1. The SideKick Infusion Kit is intended to provide continuous infusion of a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management. Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUED ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number K990425 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use __ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...