CADD-LEGACY 1 AMBULATORY INFUSION SYSTEM MODEL NUMBER 6400
K982838 · Sims Deltec, Inc. · FRN · Nov 4, 1998 · General Hospital
Device Facts
Record ID
K982838
Device Name
CADD-LEGACY 1 AMBULATORY INFUSION SYSTEM MODEL NUMBER 6400
Applicant
Sims Deltec, Inc.
Product Code
FRN · General Hospital
Decision Date
Nov 4, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CADD-Legacy™ pump is suitable for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion.
Device Story
CADD-Legacy™ 1 Model 6400 is an ambulatory infusion pump for continuous drug delivery. Device provides measured therapy via intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, or subarachnoid routes. Used in hospital or outpatient settings; operated by healthcare professionals or patients. Pump stores one continuous delivery application. Healthcare providers program the device to manage infusion rates; output is the controlled delivery of medication. Benefits include portable, measured drug administration. Potential risks include over-infusion, under-infusion, or no infusion.
Clinical Evidence
No clinical data. Clinical studies were not deemed necessary for this device. Evidence consists of functional testing, including software validation and verification of software-controlled programming functions and pump operation.
Technological Characteristics
Ambulatory infusion pump for continuous delivery. Software-controlled programming. Designed for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, or subarachnoid administration. Standalone device.
Indications for Use
Indicated for patients requiring intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, or subarachnoid space infusion therapy. Suitable for use in hospital or outpatient settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
CADD-1® Ambulatory Infusion Pump
CADD-Prizm® VIP Model 6100 Ambulatory Infusion System
Submission Summary (Full Text)
{1}------------------------------------------------
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS 4 1998 NOY for the CADD-Legacy™ 1 MODEL 6400 AMBULATORY INFUSION SYSTEM
#### GENERAL INFORMATION I.
| Common/Usual Name: | Ambulatory Infusion Pump |
|--------------------------------|--------------------------------------------------------------------------------------------------|
| Proprietary Name: | CADD-Legacy™ 1 Model 6400<br>Ambulatory Infusion System |
| Applicant's name and address: | SIMS Deltec, Inc.<br>1265 Grey Fox Road<br>St. Paul, MN 55112 |
| Equivalence device comparison: | CADD-1® Ambulatory Infusion Pump and<br>CADD-Prizm® VIP Model 6100 Ambulatory<br>Infusion System |
#### DEVICE DESCRIPTION II.
The CADD-Legacy™ 1 Model 6400 ambulatory infusion pump is similar in design, function, and intended use to Deltec's CADD-1 " Ambulatory Infusion Pump and the CADD-Prizm® VIP Model 6100 Ambulatory Infusion System . The Model 6400 pump provides measured drug therapy for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space delivery to patients in hospital or outpatient settings. The pump is capable of storing one delivery application. The delivery application resident in the Model 6400 pump and the subject of this 510(k) Notification is the continuous delivery application.
#### Ш. ALTERNATIVES
Alternatives to the CADD-Legacy™ 1 Model 6400 pump include the use of other commercially available ambulatory infusion pumps, such as Deltec's CADD-1® and CADD-Prizm® VIP Ambulatory Infusion Pumps.
#### POTENTIAL ADVERSE EFFECTS IV.
The potential direct adverse effects that may occur when using the CADD-Legacy™ 1 Model 6400 pump, as well as other commercially available non-implantable ambulatory infusion pumps, include the possibility of overinfusion, under-infusion, or no infusion.
{2}------------------------------------------------
K982838
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS for the CADD-Legacy™ 1 MODEL 6400 AMBULATORY INFUSION SYSTEM
#### SUMMARY OF STUDIES V.
#### Functional Testing A.
Test plans associated with software validation, verification of software controlled programming functions, and software related to proper pump operation were certified for the CADD-Legacy™ 1 Model 6400 Ambulatory Infusion System.
#### B. Clinical Studies
Clinical studies were not deemed necessary regarding the use of the CADD-Legacy™ 1 Model 6400 Ambulatory Infusion System.
#### CONCLUSIONS DRAWN FROM THESE STUDIES VI.
Based upon the information provided above, the CADD-Legacy™ 1 Model 6400 Ambulatory Infusion System is substantially equivalent to other commercially available ambulatory infusion systems.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird or eagle with three lines forming the wings and head. The image is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 1998 NOV
Mr. Edward W. Numainville Vice President, Regulatory Affairs and Quality Systems SIMS Deltec, Incorporated 1265 Grey Fox Road St. Paul, Minnesota 55112
K982838 Re : CADD-Legacy™ 1 Ambulatory Infusion Trade Name: System, Model 6400 Requlatory Class: II Product Code: FRN Dated: August 12, 1998 August 12, 1998 Received:
Dear Mr. Numainville:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
### Page 2 - Mr. Numainville
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
i.
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known):
CADD-Legacy™ 1 Model 6400 Ambulatory Infusion Pump Device Name:
Indications For Use:
"The CADD-Legacy™ pump is suitable for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion."
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use L (Per 21 CFR 801.109)
Over-The-Counter Use
(Optional Format 1/2/96)
Palace Crescenta
(Division Sign-On)
Division of Denal, Infection Control, and General Hospital Devices
510(k) Number 4982858
M:\510(k)ISDSubmissions1998\CADD-LegacyNIndicationsLegacy1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.