K980751 · Sabratek Corp. · FRN · Feb 24, 1999 · General Hospital
Device Facts
Record ID
K980751
Device Name
SABRATEK 4040 INFUSION PUMP
Applicant
Sabratek Corp.
Product Code
FRN · General Hospital
Decision Date
Feb 24, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Provide controlled delivery of fluids, blood and blood products and can be used in enteral, epidural, subcutaneous, arterial and intravenous applications. Prescription Device. Federal Law (US) restricts this device to sale by or on the order of a physician.
Device Story
Linear peristaltic infusion pump; delivers fluids, blood, and blood products. Inputs: user-programmed flow rates (0.1-999 ml/hr) and infusion volumes. Operates via linear peristaltic mechanism; includes pressure sensors and ultrasound air-in-line detection. Outputs: controlled fluid delivery; audible and LCD alarms for air-in-line, occlusion, battery status, and malfunction. Used in clinical settings by healthcare professionals. Benefits: precise, controlled administration of intravenous/arterial/subcutaneous therapies; reduces risk of free-flow via anti-free-flow clamp. External communication capability included.
Clinical Evidence
Bench testing only. Included accuracy testing and comparison against stated performance claims. No clinical data provided.
Technological Characteristics
Linear peristaltic pump. Materials: PVC tubing, ABS spikes/luer locks, silicone rubber tubing, modified acrylic/supor filters. Power: 90-132/138-264Vac. Features: pressure sensors, ultrasound air-in-line detection, external communication. Software: embedded control for flow rates, volume limits, and alarm monitoring.
Indications for Use
Indicated for controlled delivery of fluids, blood, and blood products via enteral, epidural, subcutaneous, arterial, or intravenous routes. Contraindicated for use with blood sets not specifically designated and manufactured by Sabratek Corporation.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{1}------------------------------------------------
2/04/99
510(k) Submission, Sabratek 4040 Infusion Pump
K980751
Sabratek, Niles, IL 60714
The following is a Summary of the Sabratek 4040 Infusion Pump substantial equivalence and safety and efficacy.
| Classification Name: | Infusion Pump |
|------------------------|------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Infusion Pump |
| Proprietary Name: | Sabratek 4040 Infusion Pump |
| Classification: | Class II Medical Device |
| Performance Standards: | No applicable performance standards<br>Have been issued under Section 514 of<br>The Food, Drug and Cosmetic Act. |
| Predicate Device: | Sabratek 6060 HOMERUN® Volumetric Infusion Pump |
K941984 and Sabratek 3030 Infusion Pump, K914581.
| ELEMENT OF<br>COMPARISON | SABRTATEK 4040 PUMP | SABRATEK 6060 PUMP | SABRATEK 3030 PUMP |
|-------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|--------------------------------|
| 510(k)# | | K941984 | |
| Pump type | Linear Peristaltic | Rotary Peristaltic | Linear Peristaltic |
| Intended use(s) | Intravenous, Intra-arterial,<br>subcutaneous & intracavity<br>infusions | Intravenous, Intra-arterial,<br>subcutaneous & intracavity<br>infusions | Linear Peristaltic |
| Administration Sets<br>-Captive Standard Set | Yes | Yes | No |
| -Captive 1.2 Micron Filter | Yes | Yes | No |
| -Captive .22 Micron Filter | Yes | Yes | No |
| -Standard Sabraset | Yes | No | Yes |
| -Approved Competitive Sets | Yes | No | Yes |
| Free Flow Clamp<br>-Anti Free Flow Clamp (Set) | Yes | Yes | No |
| -Free Flow Clamp (Pump) | Yes | No | Yes |
| Specific Drug, biologic use:<br>-Blood/Blood Products | Blood/Blood Products | No | No |
| Components<br>-Pumping Mechanism | Yes | Yes | Yes |
| -Battery | Yes | Yes | Yes |
| -Pressure Sensor | Yes | Yes | Yes |
| -Ultrasound Air In Line | Yes | Yes | Yes |
| Pumping mechanism | Linear Peristaltic | Rotary Peristaltic | Linear Peristaltic |
| Power Requirements | Switchable<br>90-132/138-264Vac 50/60<br>hz | 12 - 18 Volts DC, 50 ma<br>maximum | 110V AC, 10 watts, 50/60<br>hz |
| Battery Life | 6-6.5 hrs at 125 ml/hr | 25 hrs at 125 ml/hr | 7-8 hrs at 125 ml/hr |
| Recharge Time | 24 hrs | N/A | 24 hrs |
| | | | |
| Fluid Path Material: | | | |
| -Tubing | PVC | PVC | PVC |
| -Spike | ABS | ABS | ABS |
| -Check Valve | N/A | N/A | N/A |
| -Silicone Rubber Tubing | Silicone Rubber | Silicone Rubber | Silicone Rubber |
| -1.2 or .22 Micron Filter | Modified acrylic/supor | Modified acrylic/supor | Modified acrylic/supor |
| -MLL (Male Luer Lock) | ABS | ABS | ABS |
| -Connecting Barbs | Acrylic | Acrylic | Acrylic |
| -Y-Site | Acrylic | Acrylic | Acrylic |
| -Fluid Bag | N/A | EVA | N/A |
| -Solvent bonding Solution | Cyclohexanone | Cyclohexanone | Cyclohexanone |
| Infusion Parameters: | | | |
| -Flow Rate Range | 0.1 - 999 ml/hr | 0.1 400 ml/hr | 1- 999 ml/hr |
| -Flow Accuracy, % | +/-5% | +/-6% | +/-5% |
| -KO rate | 0.1 - 10 ml/hr | 0.1- 10 ml/hr | 1 - 9 ml/hr |
| -Infusion volume limit | 0.1 - 999.9 ml | 1.1 - 9999 ml | 1 - 999 ml |
| -Prime/purge mode | No | Yes | No |
| Profiles | | | |
| -Continuous | Yes | Yes | Yes |
| -Auto-Ramp | Yes | Yes | Yes |
| -Intermittent | Yes | Yes | Yes |
| -25 Period | Yes | Yes | Yes |
| -PCA | No | Yes | No |
| Alarm Type: | | | |
| -Air-In-Line | Audible, LCD | Audible, LCD | Audible, LCD |
| -Cassette Not Installed | N/A | Audible, LCD | N/A |
| -Check Internal 9V Batteries | N/A | Audible, LCD | N/A |
| -Complete XX ml (or mg) | Audible, LCD | Audible, LCD | Audible, LCD |
| -Door Open | Audible, LCD | Audible, LCD | Audible, LCD |
| -Down Occlusion | Audible, LCD | Audible, LCD | Audible, LCD |
| -Empty Bag or Up Occlusion | Audible, LCD | Audible, LCD | Audible, LCD |
| -Empty Battery | Audible, LCD | Audible, LCD | Audible, LCD |
| -Hold | Audible, LCD | Audible, LCD | Audible, LCD |
| -Low Bag | None | Audible, LCD | None |
| -Low Battery | Audible, LCD | Audible, LCD | Audible, LCD |
| -Malfunction | Audible, LCD | Audible, LCD | Audible, LCD |
| -Move Tubing | Audible, LCD | None | Audible, LCD |
| -Nicad Depleted | Audible, LCD | Audible, LCD | Audible, LCD |
| -Rate Exceeds Microset Limits | N/A | Audible, LCD | N/A |
| -Release/Remove PCA Cord | N/A | Audible, LCD | N/A |
| -Reprogram | Audible, LCD | Audible, LCD | Audible, LCD |
| -Stuck Key | Audible, LCD | Audible, LCD | Audible, LCD |
| -High Pressure | Audible, LCD | Audible, LCD | Audible, LCD |
| -Low Pressure | Audible, LCD | Audible, LCD | Audible, LCD |
| External Communication<br>Capability | Yes | Yes | Yes |
{2}------------------------------------------------
## 510(k) Submission, Sabratek 4040 Infusion Pump
## Sabratek, Niles, IL 60714
:
{3}------------------------------------------------
| Indications | Provides controlled delivery of fluids, blood and blood<br>products and can be used in enteral, epidural, subcutaneous,<br>arterial and intravenous applications. |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contraindications | The Sabratek 4040 Infusion Pump is contraindicated for use<br>with a blood set, other than those specifically designated as<br>such and manufactured by Sabratek Corporation. |
| Non-Clinical Tests | Non-clinical tests included accuracy testing and comparison<br>with Sabratek's stated accuracy claim and recording of data. |
| Conclusions | The Sabratek 4040 Infusion Pump is equivalent in safety and<br>efficacy to its predicate devices. |
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 4 1999
Mr. Charles P . Gill Manager, Regulatory Affairs Sabratek Corporation 8111 North St.Louis Avenue Skokie, Illinois 60076
K980751 Re : Sabratek 4040 Infusion Pump Trade Name: Requlatory Class: II Product Code: FRN Dated: November 13, 1998 December 3, 1998 Received:
Dear Mr. Gill
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal You may, therefore, Food, Drug, and Cosmetic Act (Act). market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಕ್ಕೆ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{5}------------------------------------------------
Page 2 - Mr. Gill
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdag.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Directbr Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
ﻴﺔ
510(k) Number (if Known)
Device Name: Sabratek 4040 Infusion Pump
Provide controlled delivery of fluids, blood and blood products and can be Indications for use: used in enteral, epidural, subcutancous, arterial and intravenous applications.
Prescription Device. Federal Law (US) restricts this device to sale by or on the order of a physician.
## (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use -----
Labacia Cucente (Optional Format 1-2-96)
(Division)
(Division Sign-Off) Division of Denial. Infection Control, and General Hospital Devices
510(k) Number K980751
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.