The Baxter Colleague™ 2 and 3 Volumetric Infusion Pumps are designed to meet the fluid delivery needs of today’s evolving health care environment. The pump can be utilized for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of fluid spaces applications. Fluid delivery applications include: parenteral fluids, drugs and electrolytes (e.g. cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.); and whole blood and blood products.
Device Story
Colleague 2 and 3 are multichannel volumetric infusion pumps for clinical use. Devices accept Baxter 's' suffix standard sets with keyed on/off slide clamps and various reservoirs (bags, bottles, syringes). Operation involves a shuttle pumping mechanism to deliver fluids at rates from 0.1 to 1200 mL/hr. Powered by AC, external 12 VDC, or internal rechargeable battery. Clinicians program infusion parameters; device provides flow control, air-in-line detection, and occlusion monitoring. Output includes fluid delivery and status alerts. Benefits include precise, automated fluid/drug administration across diverse care settings (ICU, OR, home, etc.), reducing manual titration errors and supporting complex dosing regimens (e.g., mg/kg/hr).
Clinical Evidence
Bench testing only. Hemolysis testing performed on aged packed red blood cells using a shuttle mechanism similar to the subject device; results showed minimal hemolysis (0.0000% to 0.114%). Flow accuracy testing conducted in accordance with IEC 601-2-24 standards.
Indicated for patients requiring continuous or intermittent delivery of parenteral fluids, drugs, electrolytes, whole blood, or blood products via IV, IA, subcutaneous, epidural, or irrigation routes. Suitable for use across the continuum of care, including hospital, home, and subacute settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
K242115 — Plum Duo Precision IV Pump · Icu Medical, Inc. · Apr 2, 2025
K063696 — COLLEAGUE VOLUMETRIC INFUSION PUMP MODEL# 2M8151,8153,8161,8163,9161,9163 · Baxterhealthcare, Inc. · Feb 27, 2007
K110901 — SYMBIQ INFUSION AND ADMINSTRATION SET · Hospira, Inc. · Mar 5, 2012
K063288 — SIGNATURE EDITION GOLD INFUSION SYSTEM, MODEL 7X3X · Cardinal Health 303,Inc. · Nov 28, 2006
K223607 — Plum Duo Infusion System · Icu Medical, Inc. · Aug 24, 2023
Submission Summary (Full Text)
{0}
510(k) Premarket Notification
Volumetric Infusion Pumps
K961703
JUL - 8 1996
## APPENDIX I.1
## SUMMARY OF SAFETY AND EFFECTIVENESS
## BAXTER VOLUMETRIC INFUSION PUMPS
s:\510k\..\SMDA
APR 30 1996
784
{1}
510(k) Premarket Notification
Volumetric Infusion Pumps
# SUMMARY OF SAFETY AND EFFECTIVENESS
## BAXTER VOLUMETRIC INFUSION PUMPS
### Submitted by:
Tamima Itani, Ph.D.
Manager, Regulatory Affairs
Baxter Healthcare Corporation
I.V. Systems Division
Rte. 120 and Wilson Road
Round Lake, IL 60073
**Date Prepared:**
April 30, 1996
**Proposed Device:**
Colleague™ 2 Volumetric Infusion Pump
Colleague™ 3 Volumetric Infusion Pump
**Predicate Devices:**
- The Colleague™ Volumetric Infusion Pump, by Baxter, cleared under K953098 on December 22, 1995
- The Flo-Gard® 6301 Volumetric Infusion Pump, by Baxter, cleared under K915522 on February 28, 1995.
**Device Description and Statement of Intended Use:**
The Baxter Colleague™ 2 and 3 Volumetric Infusion Pumps are designed to meet the fluid delivery needs of today’s evolving health care environment. The pump can be utilized for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of fluid spaces applications.
Fluid delivery applications include:
* parenteral fluids, drugs and electrolytes (e.g. cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.); and
* whole blood and blood products.
s:\510k\..\SMDA
APR 30 1996
785
{2}
510(k) Premarket Notification
Volumetric Infusion Pumps
The Baxter Volumetric Infusion Pumps are designed to travel the continuum of care, following the patient into a variety of care areas, including, but not limited to:
| Hospital: | Post Anesthesia/Recovery | Blood Centers |
| --- | --- | --- |
| General Floor | Cardiac Catheter Lab | Nuclear Medicine |
| Medical/Surgical | Emergency Room | Hospice |
| Critical/Intensive Care Areas | Burn/ Trauma Units | Subacute Facilities |
| Pediatrics/Neonatal | Oncology | Outpatient/Surgical Centers |
| Labor/Delivery/Post Partum | Mobile Intensive Care | Long Term Care |
| OR/Anesthesia | | Nursing Homes |
The Baxter Colleague™ 2 and 3 Volumetric Infusion Pumps are designed to accept a wide range of Baxter "s" suffix standard sets equipped with a keyed on/off slide clamp. The uni-directional on/off slide clamp on the sets facilitates appropriate set loading. The pump will accept a variety of currently marketed drug reservoirs, including glass bottles, plastic containers and syringes.
The Baxter Colleague™ 2 and 3 Volumetric Infusion Pump will operate on 90 - 260 VAC, 50/60 Hz. The device can also operate on an optional 12 VDC power adapter for an external power source. Alternatively, power may be supplied from a rechargeable battery integral to the device.
None of the pump components or materials come in contact with the fluid path. All materials used to build the Baxter pump meet the requirements of standards applicable to infusion pumps and electronic devices.
The Baxter Colleague™ 2 and 3 Volumetric Infusion Pump will operate from 0.1 to 1200 mL/hr. Flow rates will be less than or equal to +/- 10% error at rates less than 1.0 mL/hr, and less than or equal to +/- 5% error at rates greater than or equal to 1.0 mL/hr, over any hour, or a collection volume of 0.5 mL increment, whichever is the greater volume.
Flow and rate accuracy will not degrade outside of the specified accuracy over a period of 72-hours using the same set and catheter at 100 mL/hr. The devices will operate at all programmable flow rates with a head height ranging from -36 inches to +48 inches.
As with any electronic infusion device, flow profile is dependent on flow rate. Fluid will not stop flowing for a period greater than 40 seconds at a rate of 0.1 mL/hr; 20 seconds at a rate of 0.2 mL/hr; and 10 seconds at a rate of 0.4 mL/hr. For each subsequent doubling of rate, up to 99.9 mL/hr, the no flow period halves - i.e., 5 seconds at a rate of 0.8 mL/hr. The no flow period will be less than or equal to 0.038 seconds for rates above 100 mL/hr.
The device will operate within the specified accuracy when subjected to static back pressures of 5.8 psi (+300 mm Hg) to -1.9 psi (-100 mm Hg). The device will deliver with a maximum
s:\510k\..\SMDA
APR 30 1996
786
{3}
510(k) Premarket Notification
Volumetric Infusion Pumps
drop-off of less than or equal to 1% incremental change per psi with up to the maximum operating back pressure of 15 psi (775 mm Hg).
Environmental criteria for the operation of the Baxter pump are 59°F to 100°F; 20% to 95% relative humidity, non-condensing; 700 hPa to 1060 hPa air pressure.
The marketed product will meet the requirements of the following standards:
- UL 2601-1, Medical Electrical Equipment, Part 1: General Requirements for Safety
- IEC 601-2-24, Biological Evaluation of Medical Devices
- IEC 601-1-2, Medical Electrical Equipment, Part 1.
- NFPA 99. Standard for Health Care Facilities, National Fire Protection Association. dated 02/12/93, ANSI/NFPA
It will also meet the requirements of applicable portions of the following standards:
- Electromagnetic Compatibility Standard for Medical Devices, MDS-201-0004
- Electromagnetic Emissions and Susceptibility Requirements for the Control of Electromagnetic Interference, MIL-STD-461C
- Limits and Methods of Measurement of Radio Interference Characteristics of Industrial, Scientific and Medical (ISM) Radio-Frequency equipment (excluding surgical diathermy apparatus), CISPR 11
- Electromagnetic Compatibility for Industrial Process Measurement and Control Equipment, IEC 801-1
- Electrostatic Discharge Requirements, IEC 801-2
- Radiated Electromagnetic Field Requirements, IEC 801-3
- Electrical Fast Transient/Burst Requirements, IEC 801-4
- Radio Interference Suppression of Radio Frequency for Equipment for Industrial, Scientific, and Medical and Similar Purposes, VDE 0871
## Summary of Technological Characteristics of New Device to Predicate Devices
A table, comparing the Baxter Volumetric Infusion Pump to predicate devices is attached.
s:\510k\..\SMDA
APR 30 1996
787
{4}
510(k) Premarket Notification
Volumetric Infusion Pumps
# Discussion of Non Clinical Tests; Conclusions Drawn from Nonclinical Tests
Blood hemolysis testing was performed on aged packed red blood cells using an infusion pump with a shuttle mechanism similar to the one used for the proposed volumetric infusion pump. Aged packed red blood cells were tested because our previous experience with infusion pump indicates that they represent the worst case scenario relative to possible hemolysis. The results range from a low of 0.0000% to a high of 0.114% hemolysis.
Accuracy data was generated in accordance with the testing methodology defined by the IEC 601-2-24 standard.
s:\510k\..\SMDA
APR 30 1996
788
{5}
510(k) Premarket Notification
Volumetric Infusion Pumps
# Comparison of the Baxter Multichannel Volumetric Infusion Pumps to Other Large Volume Infusion Pumps
| Feature | BAXTER Colleague™ 2 | BAXTER Colleague™ 3 | BAXTER Colleague™ | BAXTER Flo-Gard® 6301 |
| --- | --- | --- | --- | --- |
| Pump Mechanism | Shuttle | Shuttle | Shuttle | Linear Peristaltic |
| Number of Channels | 2 | 3 | 1 | 2 |
| Set Used | Standard Baxter “s” Sets | Standard Baxter “s” Sets | Standard Baxter “s” Sets | Standard Baxter “s” Sets |
| Rate Range (mL/hr) | 0.1 - 99.9
1.0 - 1,200 | 0.1 - 99.9
1.0 - 1,200 | 0.1 - 99.9
1.0 - 1,200 | 1.0 - 99.9
1.0 - 1,999 |
| VTBI (mL) | 0.1 - 9,999 | 0.1 - 9,999 | 0.1 - 9,999 | 1 - 9,999 |
| Adjustable Maximum Rate Limit | ● | ● | ● | ● |
| Adjustable Maximum VTBI | ● | ● | ● | ● |
| Source Containers
Bags | ● | ● | ● | ● |
| Bottles | ● | ● | ● | ● |
| Syringes | ● | ● | ● | ○ |
| Stated Accuracy | ≤ 10% at rates less than 1.0 mL/hr; ≤ ± 5% otherwise | ≤ 10% at rates less than 1.0 mL/hr; ≤ ± 5% otherwise | ≤ 10% at rates less than 1.0 mL/hr; ≤ ± 5% otherwise | ≤ ± 10% |
| Motor Resolution | 3414 parts per cc | 3414 parts per cc | 3414 parts per cc | 4000 parts per cc |
| Air-in-line Detection | ●
4 Settings | ●
4 Settings | ●
4 Settings | ●
2 Settings |
| Upstream Occlusion | ● | ● | ● | ● |
| Occlusion Pressure | Minimum, Moderate, Maximum, Rate Dependent 2 PSI - 15 PSI | Minimum, Moderate, Maximum, Rate Dependent 2 PSI - 15 PSI | Minimum, Moderate, Maximum, Rate Dependent 2 PSI - 15 PSI | 7 - 17 PSI
Adjustable to 7, 12, 17 PSI |
● Available
○ Not Available
- Unknown
s:\510k\...\SMDA
APR 30 1996
789
{6}
510(k) Premarket Notification
Volumetric Infusion Pumps
| Feature | BAXTER Colleague™ 2 | BAXTER Colleague™ 3 | BAXTER Colleague™ | BAXTER Flo-Gard® 6301 |
| --- | --- | --- | --- | --- |
| Auto Restart | ● | ● | ● | ● |
| Free Flow Protection | ● | ● | ● | ● |
| RS232 | ● | ● | ● | ● |
| Computer Monitoring | ● | ● | ● | ● |
| Computer Control | ○ | ○ | ○ | ● |
| Pump Configuration Utility | ● | ● | ● | ● |
| Panel Lockout | ● | ● | ● | ● |
| Set Removal Lockout | ● | ● | ● | ○ |
| Automatic Piggybacking | ● | ● | ● | ● |
| Flow Check | ● | ● | ● | ● |
| Adjustable Alarm/Alert Intervals | ● | ● | ● | ● |
| Secondary Complete Alert | ● | ● | ● | ● |
| Blood Infusion | ● | ● | ● | ● |
| Epidural Infusion | ● | ● | ● | ● |
| Event Recorder | ● | ● | ● | ○ |
| Change Rate while running | ● | ● | ● | ● |
| Power Sources | Internal Rechargeable | Internal Rechargeable | Internal Rechargeable | Internal Rechargeable |
| Back Light | ● | ● | ● | ● |
| Battery Life Indicator | ● | ● | ● | ○ |
| Battery Capacity | 5 hr at 100 mL/hr | 4 hr at 100 mL/hr | 5 hr at 100 mL/hr | 6 hr at 1400 mL/hr |
| Multidirectional Pole Clamp | ○ | ○ | ● | ○ |
| Automatic Tube Loading | ● | ● | ● | ○ |
| Tube Misload Detection | ● | ● | ● | ● |
● Available
○ Not Available
- Unknown
s:\510k\..\SMDA
APR 30 1996
790
{7}
510(k) Premarket Notification
Volumetric Infusion Pumps
| Feature | BAXTER Colleague™ 2 | BAXTER Colleague™ 3 | BAXTER Colleague™ | BAXTER Flo-Gard® 6301 |
| --- | --- | --- | --- | --- |
| Infusion Modes | | | | |
| mL/hr | ● | ● | ● | ● |
| mg/hr | ● | ● | ● | ○ |
| mg/kg/hr | ● | ● | ● | ○ |
| mcg/hr | ● | ● | ● | ○ |
| mcg/kg/hr | ● | ● | ● | ○ |
| units/hr | ● | ● | ● | ○ |
| units/kg/hr | ● | ● | ● | ○ |
| mcg/m²/hr | ● | ● | ● | ○ |
| mg/min | ● | ● | ● | ○ |
| mg/kg/min | ● | ● | ● | ○ |
| mcg/min | ● | ● | ● | ○ |
| mcg/kg/min | ● | ● | ● | ○ |
| units/min | ● | ● | ● | ○ |
| units/min | ● | ● | ● | ○ |
| mcg/m²/hr | ● | ● | ● | ○ |
| units/m²/hr | ● | ● | ● | ○ |
| Volume - Time | ● | ● | ● | ● |
| Ramping | ● | ● | ● | ● |
| KVO (mL/hr) | 0.1 - 5 | 0.1 - 5 | 0.1 - 5 | 1.0 - 5.0 |
| Device Diagnostics | ● | ● | ● | ● |
| Label Library | ● | ● | ● | ○ |
| Drop Sensor | ● | ● | ● | ○ |
| | Optional | Optional | Optional | |
| Volume History | ● | ● | ● | ● |
| Nurse call Via External Adapter | ● | ● | ● | ● |
| 12 VDC External Cable Connector (accessory) | ● | ● | ● | ○ |
| Pump Personality | ● | ● | ● | ○ |
● Available
○ Not Available
- Unknown
s:\510k\..\SMDA
APR 30 1996
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.