MEDFUSION MODEL 2001 SYRINGE INFUSION PUMP

K955231 · Medex, Inc. · FRN · Jul 18, 1996 · General Hospital

Device Facts

Record IDK955231
Device NameMEDFUSION MODEL 2001 SYRINGE INFUSION PUMP
ApplicantMedex, Inc.
Product CodeFRN · General Hospital
Decision DateJul 18, 1996
DecisionST
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended medical application of Medex Medfusion 2000 Series syringe pumps is to produce controlled movement of the plunger of a syringe to inject a set amount of therapeutic fluid into a patient within a hospital setting at a set rate and at set times.

Device Story

Medfusion 2001, 2010, & 2010i syringe pumps; hospital-use infusion devices. Function: controlled plunger movement for precise therapeutic fluid delivery. Modification: replacement of microprocessor/memory chip with drop-in microprocessor-EPROM daughter board; associated software memory address updates. Operation: identical to predicate models. Output: controlled fluid infusion. Benefit: consistent, timed medication delivery.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Syringe pump; plastic exterior; no fluid/drug contacting surfaces. Energy source: same as predicate. Microprocessor/EPROM daughter board hardware. Software: modified memory addresses to accommodate hardware change.

Indications for Use

Indicated for patients requiring controlled infusion of therapeutic fluids via syringe pump in a hospital setting.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K955231 medexinc JUL 18 1996 # 510(k) SUMMARY FOR MEDEX MEDFUSION 2001, 2010, & 2010i WITH NEW MICROPROCESSOR/MEMORY SYSTEM The following summary of safety and effectiveness information is provided as required by Section 807.92(c) of the Safe Medical Devices Act of 1990. | Submitter | Medex, Inc. | | --- | --- | | Contact person | James W. Lewis | | Summary prepared | 22 April 1996 | | Trade names | Medfusion 2001, 2010, & 2010i | | Common name | Syringe pump | | Classification name | Infusion pump (per 21 CFR 880.5725) | | Predicate devices | The legally marketed devices to which these new devices are equivalent are the previous products marketed by Medex under the same trade names | | Device description | The new model Medex Medfusion Series 2000 pumps are identical to their predicate models with the exception that the microprocessor/memory chip is replaced by a drop-in replacement microprocessor–EPROM daughter board and that associated memory addresses in the software are modified to accommodate the hardware change. The exterior surfaces are made of plastics. There are no fluid or drug contacting surfaces. | | Intended use | The intended medical application of Medex Medfusion 2000 Series syringe pumps is to produce controlled movement of the plunger of a syringe to inject a set amount of therapeutic fluid into a patient within a hospital setting at a set rate and at set times. | | Technological characteristics | The basic design, material, chemical composition, and energy sources are the same as for the current corresponding Medex Medfusion model pumps. They are identical in operation, function, features, and form as their predicate devices and represent no technological differences. | 11360 Technology Circle • Duluth, GA 30155 USA • (770) 623-9809 • FAX: (770) 497-0736
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...