MRidium 3860+ MRI infusion Pump/ Monitoring System
K143369 · Iradimed Corporation · FRN · Dec 15, 2016 · General Hospital
Device Facts
Record ID
K143369
Device Name
MRidium 3860+ MRI infusion Pump/ Monitoring System
Applicant
Iradimed Corporation
Product Code
FRN · General Hospital
Decision Date
Dec 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Iradimed Corporation's MRidium 3860+ MRI Infusion Pump/Monitoring System is intended for: - General hospital or clinical use by medical professionals whenever it is required to infuse patients with subcutaneous, intra-venous or intra-arterial fluids before, during, or after Magnetic Resonance Imaging (MRI) scans, functioning while either in a stationary or mobile position. - The system is useful in the administration of fluids requiring precisely controlled infusion rates. The system can operate in either continuous, intermittent, or bolus delivery mode. - The Infusion Pump can be used inside the MRI room mounted outside the (1 Tesla line (1 Tesla line), and with shielded magnets of field strength of 3.0 Tesla or less. - This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications. - The Pulse Oximeter is used to measure, display, and record functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric, and infant patients in an MR environment. Testing of the oximeter was performed in MR conditional environments at 1.5T and 3.0T. It is indicated for spot checking and/or continuous monitoring of patients who are well or poorly perfused in the MRI.
Device Story
MRidium 3860+ is an MRI-compatible infusion pump and monitoring system for use in MR scan rooms (up to 3.0 Tesla). Device utilizes an ultrasonic (non-magnetic) motor for peristaltic pumping; supports primary/secondary channels and remote display. Input includes user-programmed infusion parameters (volume, dose, concentration, weight, time) via Dose Rate Calculator and DERS library card; SpO2 data via fiberoptic sensors. System provides continuous, intermittent, or bolus fluid delivery; monitors SpO2 and pulse rate. Healthcare providers use the interface to set infusion protocols and monitor patient vitals. DERS feature allows facility-defined drug libraries with hard/soft limits to reduce medication errors. Benefits include safe fluid administration and patient monitoring within high-magnetic field environments without image interference or magnetic attraction hazards.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed design requirements for software version 3.5.3, including DERS library functionality and pulse oximeter performance. Human factors studies and safety assurance cases were conducted per FDA guidance to demonstrate substantial equivalence.
Technological Characteristics
MRI-compatible infusion pump; ultrasonic (non-magnetic) motor; peristaltic pump channel. Operates in magnetic fields up to 10,000 Gauss (3.0 Tesla). Fiberoptic SpO2 sensors (no conductive components). Connectivity via SD memory card for drug library updates. Software version 3.5.3 includes Dose Error Reduction System (DERS).
Indications for Use
Indicated for adult, pediatric, and infant patients requiring fluid infusion (subcutaneous, IV, or intra-arterial) or SpO2/pulse rate monitoring before, during, or after MRI scans. Contraindicated for use on the inlet side of ECMO systems where negative pressure exceeds -100 mmHg.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
MRidium 3860 MRI Infusion Pump/Monitoring System (K090087)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 15, 2016
Iradimed Corporation Mr. Francis X Casey Vice-President, QA & Regulatory Affairs 1025 Willa Springs Drive Winter Springs, Florida 32708
Re: K143369
Trade/Device Name: MRidium 3860+ MRI Infusion Pump/Monitoring System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: November 8, 2016 Received: November 9, 2016
Dear Mr. Francis Casey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina Kiang -
s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K143369
#### Device Name
MRidium 3860+ MRI Infusion Pump/Monitoring System
#### Indications for Use (Describe)
The Iradimed Corporation's MRidium 3860+ MRI Infusion Pump/Monitoring System is intended for:
- General hospital or clinical use by medical professionals whenever it is required to infuse patients with subcutaneous, intra-venous or intra-arterial fluids before, during, or after Magnetic Resonance Imaging (MRI) scans, functioning while either in a stationary or mobile position.
- The system is useful in the administration of fluids requiring precisely controlled infusion rates. The system can operate in either continuous, intermittent, or bolus delivery mode.
- The Infusion Pump can be used inside the MRI room mounted outside the (1 Tesla line (1 Tesla line), and with shielded magnets of field strength of 3.0 Tesla or less.
- This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications.
- The Pulse Oximeter is used to measure, display, and record functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric, and infant patients in an MR environment. Testing of the oximeter was performed in MR conditional environments at 1.5T and 3T. It is indicated for spot checking and/or continuous monitoring of patients who are well or poorly perfused in the MRI.
The infusion pump is contraindicated for use on the inlet side of Extracorporeal Membrane Oxygenation (ECMO) systems where the negative pressure is greater than -100 mmHG as the high negative pressures can result in uncontrolled fluid flow.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
# 510(k) SUMMARY
## IRadimed's MRidium 3860+ MRI Infusion Pump/Monitoring System
Submitter Information:
Iradimed Corporation 1025 Willa Springs Dr. Winter Springs, FL 32708
407-677-8022 x106 Contact: Mr. Francis Casey
Date Prepared: December 7, 2016
#### Product Name:
Proprietary: MRidium 3860+ MRI Infusion Pump/Monitoring System Common: Infusion Pump Classification: Class II (see 21 CFR 880.5725) Product Code: FRN
## Primary Predicate/Legally Marketed Device to Which Equivalence is Claimed:
MRidium 3860 MRI Infusion Pump/Monitoring System: K090087 (Marketed as MRidium 3860+ MRI Infusion Pump/Monitoring System)
## Device Description:
The MRidium 3860+ MRI Infusion Pump/Monitoring System with software version 3.5.3 is an MRI compatible IV pump intended for use within the MR Scan room. The device operates to full specification in magnetic fields of up to 10,000 Gauss and has RF emissions at Larmor frequencies up to and including 3.0 Tesla MR scanners (132 MHz) such that image signal to noise is not visibly affected with the pump within 1 foot from the MRI bore opening. The magnetic content of the device is such that there is no hazard of magnetic attraction.
The pump unit is designed with an integral single peristaltic pump channel utilizing an ultrasonic (nonmagnetic) motor. This integral channel is vertically oriented to the right side of the main assembly of the pump unit shall contain the controls, display, power supply, battery, processor/memory functions suitable to meet the complete and expanded system requirements.
The left side of the unit is designed to accommodate attachment of an optional second peristaltic pump channel (Model 3861) which is controlled through the main pump assembly's display and controls. A remote display (Model 3865) is also available for independent viewing and control from the adjacent MRI Control areas.
{4}------------------------------------------------
The Dose Rate Calculator feature, allows the user to set up a patient's infusion rate based upon user selected parameters, including volume to be infused, dose, concentration, weight, and/or time.
The Dose Rate Calculator feature also provides a Drug Library, allowing the user to program a patient's infusion protocol from selected parameters, including volume to be infused, dose, concentration, weight and/or time.
The Drug Library includes a small drug library consisting of 5 drugs (four common medications most frequently infused during MRI procedures, and one customizable drug) and with the DERS library card, can hold up to 50 customizable drugs. The optional DERS library card also includes the ability of setting hard and soft limits for each drug. This Drug Library feature can only be activated/de-activated with a service-related, limited user access menu.
The Dose Error Reduction System (DERS) option (P/N 1145) to the Dose Rate Calculator feature allowing user facilities to program custom drug names, doses, and limits for use in the Dose Rate Calculator. A user programmable drug library memory card stores the specific infusion protocols established by the hospital facility. This library is accessed using the pump's Dose Rate Calculator Menu. The drug library card supports a number of separate user-programmed infusion protocols for either primary and/or primary/bolus infusions, retrievable by drug/protocol name. The library can be programmed with nominal starting values for: Dose, Concentration, and Time. Also, hard limits (maximum and minimum) and soft limits (high and low limits that require a user confirmation to exceed) for: Dose, Concentration, Time and Patient Weight can be programmed. Programming the custom medications and limits is performed with the pump and a dedicated drug library SD Memory Card in the limited access service mode.
The Pulse Oximeter feature of the 3860+ MRI Infusion Pump/Monitoring System is used in measuring, displaying, and recording functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric, and infant patients in an MR environment. It is indicated for continuous monitoring of patients in the MRI who are well or poorly perfused. The 3860+ System displays the patient's pulse rate and SpO2 values, as well as applicable alarm limits. The 3860+ System includes adjustable and visual pulse rate and oxygen saturation (SpO2) alarms. It also includes a variety of additional features, including low battery alarms, status indicators and sensor-related alarms. The 3860+ System only uses fiberoptic SpO2 sensors. These fiber optic sensors, cables, and associated finger wrap accessories contain no conductive components; they can be applied safely to the patient while inside an MR (magnetic resonance) environment. The 3860 model number is used for the same infusion device with no SpO2 module.
## Indications for Use:
General hospital or clinical use by medical professionals whenever it is required to infuse patients with subcutaneous, intra-venous or intra-arterial fluids before, during, or after Magnetic Resonance Imaging (MRI) scans, functioning while either in a stationary or mobile position.
The system is useful in the administration of fluids requiring precisely controlled infusion rates. The system can operate in either continuous, intermittent, or bolus delivery mode.
{5}------------------------------------------------
The Infusion Pump can be used inside the MRI room mounted outside the 10,000 Gauss line (1 Tesla line), and with shielded magnets of field strength of 3.0 Tesla or less.
This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications.
The Pulse Oximeter is used to measure, display, and record functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric, and infant patients in an MR environment. Testing of the Oximeter was performed in MR conditional environments at 1.5T and 3T. It is indicated for spot checking and/or continuous monitoring of patients who are well or poorly perfused in the MRI.
The infusion pump is contraindicated for use on the inlet side of Extracorporeal Membrane Oxygenation (ECMO) systems where the negative pressure is greater than -100 mmHg as the high negative pressures can result in uncontrolled fluid flow.
## Comparison to Primary Predicate/Reason for 510(k) Submission:
The Iradimed Corporation MRidium 3860+ MRI Infusion Pump/Monitoring System is the same as the primary predicate, which is the MRidium 3860 MRI Infusion Pump/Monitoring System (K090087) with the two modifications:
- 1. Updated version of software: The Infusion Pump in this submission includes software version 3.5.3. whereas the Infusion Pump in the K090087 submission included software version 2.0. The changes in the software include:
- . Addition of an optional Drug Error Reduction System library card. The primary predicate has a small drug library consisting of 5 drugs, with one of these being customizable; where the DERS library card can hold up to 50 customizable drugs.
- The optional DERS library card includes the ability of setting hard and soft limits for . each drug.
- 2. Pulse Oximeter: The subject device utilizes OEM Masimo Set Technology (K053269) for monitoring hemoglobin (SpO2) and Pulse Rate, with IRadimed 1170 fiberoptic SpO2 sensor. This employs the same fundamental scientific technology as the primary predicate, which utilized OEM Nonin Medical (K071415) for monitoring hemoglobin (SpO2) and Pulse Rate, with Nonin's fiberoptic SpO2 sensor. The differences between the two include:
- Different OEM supplier for the pulse Oximeter monitor, however all performance ● specifications are identical between the two, and both have been cleared for market through a Premarket Notification.
- Different manufacturer for fiberoptic cable: Nonin vs Iradimed. Both use fine diameter ● glass fibers arranged in two bundles, red/infrared Leds and photodiodes to transmit data to the monitoring unit within the pump.
Other than these two differences, the subject device is identical to the primary predicate device. The table below is a side-by-side comparison for the two modifications:
{6}------------------------------------------------
| Characteristic | Subject Device | Primary Predicate Device MRidium |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MRidium 3860+ MRI Infusion | 3860 MRI Infusion Pump/Monitoring System |
| | Pump/Monitoring System with SW Version | with SW Version 2.0 K090087 |
| | 3.5.3 | |
| Product Code | FRN | FRN |
| Regulation Number | 21 CFR 880.5725 | 21 CFR 880.5725 |
| Regulation Name | Infusion Pump | Infusion Pump |
| Regulatory Class | II | II |
| Indications for Use | General hospital or clinical use by medical<br>professionals whenever it is required to infuse<br>patients with subcutaneous, intra-venous or intra-<br>arterial fluids before, during, or after Magnetic<br>Resonance Imaging (MRI) scans, functioning while<br>either in a stationary or mobile position.<br>The system is useful in the administration of fluids<br>requiring precisely controlled infusion rates. The<br>system can operate in either continuous, intermittent,<br>or bolus delivery mode.<br>The Infusion Pump can be used inside the MRI room<br>mounted outside the 10,000 Gauss line (1 Tesla line),<br>and with shielded magnets of field strength of 3.0<br>Tesla or less.<br>This device is available for sale only upon the order<br>of a physician or other related licensed medical<br>professional, and not intended for any home use<br>applications.<br>The Pulse Oximeter is used to measure, display, and<br>record functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate of adult, pediatric,<br>and infant patients in an MR environment. Testing of<br>the Oximeter was performed in MR conditional<br>environments at 1.5T and 3T. It is indicated for spot<br>checking and/or continuous monitoring of patients<br>who are well or poorly perfused in the MRI.<br>The infusion pump is contraindicated for use on the<br>inlet side of Extracorporeal Membrane Oxygenation<br>(ECMO) systems where the negative pressure is<br>greater than -100 mmHg as the high negative<br>pressures can result in uncontrolled fluid flow. | General hospital or clinical use by medical<br>professionals whenever it is required to infuse patients<br>with subcutaneous, intra-venous or intra-arterial fluids<br>before, during, or after Magnetic Resonance Imaging<br>(MRI) scans, functioning while either in a stationary<br>or mobile position.<br>The system is useful in the administration of fluids<br>requiring precisely controlled infusion rates. The<br>system can operate in either continuous, intermittent,<br>or bolus delivery mode.<br>The Infusion Pump can be used inside the MRI room<br>mounted outside the 10,000 Gauss line (1 Tesla line),<br>and with shielded magnets of field strength of 3.0<br>Tesla or less.<br>This device is available for sale only upon the order<br>of a physician or other related licensed medical<br>professional, and not intended for any home use<br>applications.<br>The Pulse Oximeter is used to measure, display, and<br>record functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate of adult, pediatric,<br>and infant patients in an MR environment. Testing of<br>the Oximeter was performed in MR conditional<br>environments at 1.5T and 3T. It is indicated for spot<br>checking and/or continuous monitoring of patients<br>who are well or poorly perfused in the MRI.<br>The infusion pump is contraindicated for use on the<br>inlet side of Extracorporeal Membrane Oxygenation<br>(ECMO) systems where the negative pressure is<br>greater than -100 mmHg as the high negative<br>pressures can result in uncontrolled fluid flow. |
| Configuration | 3860+ Infusion Pump<br>3861 Sidecar (Channel B)<br>3865 Remote | 3860 Infusion Pump<br>3861 Sidecar (Channel B)<br>3865 Remote |
| Dose Rate Calculator | Yes | Yes |
| Drug Library | Yes | Yes |
| # Drugs in Library | up to 50 with DERS* Drug Library<br>(customizable by customer) | 5 (1 customizable by customer) |
| Ability to Set Hard/Soft | Yes* | No |
| Limits for Drugs | | |
| Pulse Oximeter | Yes | Yes |
| Mfg of Pulse Oximeter | Masimo (K053269) | Nonin (K071415) |
| Mfg of Pulse Oximeter | IRadimed p/n 1170 | Nonin |
| Cable (fiberoptic) | | |
*Only with addition of optional Dose Error Reduction System (DERS) Library Card, without optional DERS Library Card, there is no difference between the two software versions for the Drug Library
{7}------------------------------------------------
| Performance Testing to Demonstrate Equivalence | | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Characteristic | Standard/Test/FDA Guidance | Result |
| Human Factors/Usability | "Applying Human Factors and Usability Engineering to Medical<br>Devices: Guidance for Industry and Food and Drug Administration<br>Staff", issued Feb 3, 2016 | Pass |
| Software | Software Verification and Validation:<br>• Code Review<br>• Static Analysis<br>• Unit Testing<br>• Regression Testing<br>• System Validation | Pass |
| Risk Management | | |
| Safety Assurance Case | "Infusion Pumps Total Product Life Cycle: Guidance for Industry and<br>FDA Staff", issued Dec 2, 2014 | Complete |
| Risk Analysis | ISO 14971 | Complete |
## Summary of Performance Testing:
Verification and validation activities were performed to ensure that the MRidium 3860+ MRI Infusion Pump/Monitoring System with software version 3.5.3 meets design input and safety requirements. Verification and validation testing was conducted and confirmed that the new feature design requirements were met. Additionally, pre-existi…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.