← Product Code [FRN](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN) · K100704

# ACCU-CHEK FLEXLINK PLUS INFUSION SET, ACCU-CHECK LINKASSIST PLUS INSERTION DEVICE (K100704)

_Disetronic Medical Systems AG · FRN · May 14, 2010 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN/K100704

## Device Facts

- **Applicant:** Disetronic Medical Systems AG
- **Product Code:** [FRN](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN.md)
- **Decision Date:** May 14, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

ACCU-CHEK® FlexLink Plus is an infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps. ACCU-CHEK® LinkAssist Plus is an insertion device, which is intended specifically for placement of compatible ACCU-CHEK® FlexLink Plus infusion sets.

## Device Story

ACCU-CHEK FlexLink Plus is a disconnectable infusion set with a soft cannula perpendicular to an adhesive patch; used for subcutaneous insulin delivery from external micro-dosage pumps. Includes a needle protection feature that automatically retracts the introducer needle post-insertion. ACCU-CHEK LinkAssist Plus is an optional, non-invasive, non-sterile, reusable mechanical insertion aid designed to facilitate the placement of the FlexLink Plus infusion set. Both devices are used by patients with insulin-dependent diabetes mellitus. The infusion set interfaces with standard insulin pumps; the insertion device is operated by the patient to ensure proper cannula placement. The system benefits patients by providing a reliable, disconnectable method for insulin delivery with integrated needle safety features to prevent accidental sticks.

## Clinical Evidence

No clinical studies were performed. Evidence consists of in vitro functional testing, biocompatibility testing, and human factors usability testing. Usability evaluations involved patients who routinely use insulin pumps and infusion sets; results confirmed that customer requirements were met, primary operating functions performed as intended, and no needle stick injuries occurred during use.

## Technological Characteristics

Infusion set: soft cannula, flexible catheter, adhesive patch, luer connector, automatic needle retraction mechanism. Insertion device: mechanical, non-sterile, reusable, includes safety catch. Materials: biocompatible. Connectivity: interfaces with commercially available insulin infusion pumps. No software or electronic components.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- ACCU-CHEK Ultraflex infusion set
- ACCU-CHEK LinkAssist insertion device

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KUUZOY

····

; .

ﺑ

MAY 1 4 2010

| 510(k) Summary      |                                                                                                                                                                                                                                                                                                                                                                           |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Introduction        | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence.                                                                                                                                                                                        |
| Submitter           | Submitted by:<br>Disetronic Medical Systems AG<br>Kirchbergstrasse 190, Postfach<br>CH-3401 Burgdorf, Switzerland                                                                                                                                                                                                                                                         |
|                     | United States Contact Person:<br>Scott Thiel<br>Roche Diagnostics<br>9115 Hague Road<br>Indianapolis, Indiana 46250<br>317-521-3362<br>scott.thiel@roche.com<br>Date Prepared: April 12, 2010                                                                                                                                                                             |
| Device name         | Proprietary name: ACCU-CHEK® FlexLink Plus infusion set<br>Common name: subcutaneous infusion set<br>Classification name: intravascular administration set<br>Product Code: FPA<br><br>Proprietary name: ACCU-CHEK® LinkAssist Plus insertion device<br>Common name: infusion set insertion device<br>Classification name: syringe needle introducer<br>Product Code: KZH |
| Predicate<br>device | We claim substantial equivalence to the current legally marketed ACCU-<br>CHEK® Ultraflex infusion set and ACCU-CHEK® LinkAssist insertion device.                                                                                                                                                                                                                        |

Continued on next page

- --

{1}------------------------------------------------

# 510(k) Summary, Continued

| Device description | The ACCU-CHEK® FlexLink Plus is a disconnectable infusion set with soft             cannula perpendicular to the adhesive, for transfusion of insulin into the             subcutaneous tissue. The unit is designed to interface with commercially             available insulin infusion pumps with suitable connections. The insulin             infusion pump systems are designed to control the delivery of insulin as             prescribed by a health care professional. The system (infusion set, insulin             infusion pump, and insulin) is indicated for patients with insulin dependent             diabetes mellitus. |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                    | The ACCU-CHEK® LinkAssist Plus is an insertion aid for automatic             application of ACCU-CHEK® infusion sets equipped with a compatible             adapter. The ACCU-CHEK® LinkAssist device is non-invasive, non-sterile             and intended for multiple uses by the same patient.                                                                                                                                                                                                                                                                                                                                           |
| Intended use       | ACCU-CHEK® FlexLink Plus is an infusion set for the subcutaneous             infusion of insulin administered with micro dosage insulin pumps.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                    | ACCU-CHEK® LinkAssist Plus is an insertion device, which is intended             specifically for placement of compatible ACCU-CHEK® FlexLink Plus             infusion sets.                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                    | Continued on next page                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{2}------------------------------------------------

### 510(k) Summary, Continued

The ACCU-CHEK® FlexLink Plus infusion set was compared to the ACCU-Device CHEK® Ultraflex infusion set. The ACCU-CHEK® FlexLink Plus is Comparisons substantially equivalent to this product by having the same intended use, same storage conditions, same luer connector, same operating conditions, a flexible catheter and needle for insertion into the subcutaneous tissue and separate extension tubing with detachable connector. Both sets have an adhesive patch that secures the headset to the skin. Prior to infusion, both sets require removal of the introducer needle. The ACCU-CHEK® FlexLink Plus incorporates a needle protection feature that retracts the needle automatically following insertion of the needle and cannula. The optional ACCU-CHEK® LinkAssist Plus insertion device was compared to the optional ACCU-CHEK® LinkAssist insertion device. The ACCU-CHEK® FlexLink Plus is substantially equivalent to this product by having the same intended use, same storage conditions, same operating conditions, and same general functionality. The overall dimensions of the ACCU-CHEK® FlexLink Plus are slightly wider and longer than the ACCU-CHEK® FlexLink device. Both devices include a safety catch; the ACCU-CHEK® FlexLink Plus incorporates the safety catch in the base of the device

such that it is depressed when placing the device into position on the body.

Continued on next page

{3}------------------------------------------------

### 510(k) Summary, Continued

#### Summary of Functional Testing:

In vitro functional testing of the ACCU-CHEK® FlexLink Plus infusion set and the optional ACCU-CHEK® LinkAssist Plus insertion device was conducted. Biocompatibility testing was performed on the materials used in both devices.

#### Clinical Studies:

Human clinical studies were not deemed necessary to evaluate the safety or effectiveness of the ACCU-CHEK® FlexLink Plus infusion set or the optional ACCU-CHEK® LinkAssist Plus insertion device.

System validation testing included human factors usability testing of the customer requirements and primary operating function of the ACCU-CHEK® FlexLink Plus infusion set and the optional ACCU-CHEK® LinkAssist Plus insertion device. Individuals who participated in the evaluation included patients who routinely use an insulin pump and infusion set. The usability evaluations determined the customer requirements were met and that the primary operating function operated as intended. No needle stick injuries were reported associated with unintentional ejection.

Study Conclusions

Studies

#### Functional Testing:

The results of the testing conducted indicate the ACCU-CHEK® FlexLink Plus infusion set and the optional ACCU-CHEK® LinkAssist Plus insertion device functioned according to specifications and the materials used in the devices are biocompatible.

### Human factors usability testing:

The design validation of ACCU-CHEK® FlexLink Plus & ACCU-CHEK® LinkAssist Plus confirmed that the devices fulfill its intended use, customer requirements and primary operating functions.

Based upon these results, the product is considered acceptable for human use.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Disetronic Medical System AG C/O Mr. Scott Thiel Regulatory Affairs Program Manager Roshe Diagnostics 9115 Hague Road Indianapolis, Indiana 46250-0457

MAY 1 4 2010

Re: K100704

Trade/Device Name: ACCU-CHEK® FlexLink Plus infusion set with optional ACCU- CHEK® LinkAssist Plus insertion device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FRN Dated: April 14, 2010 Received: April 16, 2010

Dear Mr. Thiel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Mr. Thiel

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,

Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known): K100704

Device Name: ACCU-CHEK® FlexLink Plus infusion set with optional ACCU-CHEK® LinkAssist Plus insertion device

Indications for Use:

ACCU-CHEK® FlexLink Plus is an infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps.

ACCU-CHEK® LinkAssist Plus is an insertion device, which is intended specifically for placement of compatible ACCU-CHEK® FlexLink Plus infusion sets.

Prescription Use_ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K100704

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN/K100704](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN/K100704)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
