K090736 · ZOLL Medical Corporation · FRN · May 20, 2009 · General Hospital
Device Facts
Record ID
K090736
Device Name
POWER INFUSER, MODEL M100B-3A
Applicant
ZOLL Medical Corporation
Product Code
FRN · General Hospital
Decision Date
May 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Power Infuser® Model M100B-3A is intended for continuous or intermittent administration of therapeutic and clinically appropriate intravenous fluids, blood and packed red blood cells through clinically acceptable access points. The device is intended for use by medical, paramedical and EMT personnel in the field and in pre-hospital and hospital environments. When used with the Crystalloid/Colloid Cartridge the device is intended to deliver crystalloid and colloid resuscitative fluids. It is not intended to support the infusion of blood or blood products. When used with the Blood Cartridge the device is intended to deliver crystalloid and colloid resuscitative fluids, whole blood and packed red blood. The device is not intended to support the delivery of any pharmaceutical or other medications.
Device Story
Power Infuser Model M100B-3A is a miniature, battery-operated infusion pump; delivers IV fluids, whole blood, and packed red blood cells. Used in field, pre-hospital, and hospital environments by medical, paramedical, and EMT personnel. Operates with specific cartridges: Crystalloid/Colloid Cartridge for resuscitative fluids; Blood Cartridge for fluids, whole blood, and packed red blood cells. Device does not support pharmaceutical or medication delivery. Provides clinicians with controlled fluid administration to restore volume and blood pressure. Benefits patients by enabling rapid, portable resuscitation in emergency and clinical settings.
Clinical Evidence
No clinical data; bench testing only. The submission is limited to an update of the Intended Use statement. No hardware or software changes were made to the device since the predicate (K030739), and performance remains unchanged.
Technological Characteristics
Miniature, battery-operated infusion pump. Mechanical design unchanged from predicate. No software or hardware modifications. Intended for use with specific fluid/blood cartridges.
Indications for Use
Indicated for patients requiring continuous or intermittent delivery of IV fluids, whole blood, or packed red blood cells. Intended for use by medical, paramedical, and EMT personnel in field, pre-hospital, and hospital settings. Contraindicated for the delivery of pharmaceutical or other medications.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
David Clark Pressure Infuser for an I.V. bag (K820159)
Submission Summary (Full Text)
{0}------------------------------------------------
K0907BL
Image /page/0/Picture/1 description: The image shows the logo for ZOLL Medical Corporation. The logo features the word "ZOLL" in large, bold, black letters. Below the company name is the tagline "Advancing Resuscitation. Today." in a smaller, italicized font. There is a curved line below the tagline.
MAY 20 2009
ZOLL Medical Corporation Worldwide Headguarters 269 Mill Road Chelmsford, MA 01824 U.S.A
## 510(k) Summary:
Submitter's Name and Address:
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655
Contact Person: Eileen M. Boyle (978) 421-9171
| Date Summary Prepared: | March 10, 2009 |
|------------------------|--------------------------------------------|
| Device: | Power Infuser |
| Classification: | Infusion Pump: Class II (21 CFR 880.5725) |
| Reason for submission: | Change in the Intended Use |
## Description:
The Power Infuser model M100B-3A, reviewed and cleared under (K030739) is a miniature battery-operated infusion pump designed to deliver IV fluids and blood products to patients in the field, in pre-hospital transport, or in the hospital. When used with the Crystalloid/Colloid Cartridge, the device infuses crystalloid and colloid IV fluids, the most common initial therapy used to restore a patient's blood pressure and intravascular volume in both pre-hospital and hospital settings. When the device is used with the Blood Cartridge, it is intended to deliver crystalloid and colloid resuscitative fluids as well as whole blood and packed red blood cells. The device is not intended to support the delivery of any pharmaceutical or other medications.
The current indication for use is to restore volume and blood pressure to patients experiencing clinical shock, hypotension, and hypoperfusion states as a result of hemorrhagic blood loss, occult hemorrhage, neurogenic shock and septic shock. The indication is being expanded to administer fluids to any patient requiring continuous or intermittent delivery of IV fluids, whole blood, or packed red blood cells. The device is intended for use by medical, paramedical and EMT personnel in the field and in pre-hospital and hospital environments. The device is not intended to support the delivery of any pharmaceutical or other medications.
{1}------------------------------------------------
A previous 510(k) submissions demonstrated substancial equivalence to the David Clark Pressure Infuser for an I.V. bag (K820159 as required by 21 CFR Part 880.5420). The purpose of this submission is to more closely align the Intended Use of the Power Infuser to the predicate device as described below.
#### Intended Use:
The Power Infuser® Model M100B-3A is intended for continuous or intermittent administration of therapeutic and clinically appropriate intravenous fluids, blood and packed red blood cells through clinically acceptable access points.
The device is intended for use by medical, paramedical and EMT personnel in the field and in pre-hospital and hospital environments.
When used with the Crystalloid/Colloid Cartridge the device is intended to deliver crystalloid and colloid resuscitative fluids. It is not intended to support the infusion of blood or blood products.
When used with the Blood Cartridge the device is intended to deliver crystalloid and colloid resuscitative fluids, whole blood and packed red blood.
The device is not intended to support the delivery of any pharmaceutical or other medications.
#### Substantial Equivalence:
The features and functions of the Power Infuser M100B-3A are substantially equivalent to the current features and functions of the Power Infuser.
#### Comparison of Technological Characteristics
The Power Infuser design characteristics have not been changed from the original design, which was reviewed and cleared by FDA as a result of this update.
## Performance Testing:
Since the design and functionality of the Power Infuser M100B-3A has not changed, performance testing was not conducted. This submission is focused on the change to the Intended Use Statement.
## Conclusion
Performance of the Power Infuser M100B-3A has not changed and remains substantially equivalent to those of the indicated commercially distributed predicate devices with regard to performance, safety and effectiveness. No software or hardware changes have been implemented into the device since the predicate (K030739) as a result of this update.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in a bold, dark color.
Public Health Service
MAY 2 0 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Eileen M. Boyle Regulatory Affairs Specialist Zoll Medical Corporation 269 Mill Road Chelmsford, Massachusetts 01824
Re: K090736
Trade/Device Name: Power Infuser Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: March 10, 2009 Received: March 19, 2009
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Ms. Boyle
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) régulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Susan Runner
Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# SECTION 4 - Indications for Use
510(k) Number (if known): _ KO90 736
Device Name: Power Infuser
#### Intended Use:
The Power Infuser® Model M100B-3A is intended for continuous or intermittent administration of therapeutic and clinically appropriate intravenous fluids, blood and packed red blood cells through clinically acceptable access points.
The device is intended for use by medical, paramedical and EMT personnel in the field and in pre-hospital and hospital environments.
When used with the Crystalloid/Colloid Cartridge the device is intended to deliver crystalloid and colloid resuscitative fluids. It is not intended to support the infusion of blood or blood products.
When used with the Blood Cartridge the device is intended to deliver crystalloid and colloid resuscitative fluids, whole blood and packed red blood.
The device is not intended to support the delivery of any pharmaceutical or other medications.
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
1090736 510(k) Number:
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.