← Product Code [FRN](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN) · K060479

# MODIFICATION TO BODYGUARD INFUSION PUMP SYSTEM (K060479)

_Caesarea Medical Electronics , Ltd. · FRN · May 12, 2006 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN/K060479

## Device Facts

- **Applicant:** Caesarea Medical Electronics , Ltd.
- **Product Code:** [FRN](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN.md)
- **Decision Date:** May 12, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The BodyGuard infusion Pump system is designed for infusion and The pump is intended for infusion of medications or fluids monitoring. requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, enteral, in close proximity to nerves, and into an intraoperative site (soft tissue/body cavity/surgical wound site). Monitoring accessory currently includes Pulse Oximeter, used for non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The system is intended for patients who require maintenance medications, analgesics, PCA therapy, parenteral and enteral nutrition fluids, chemotherapeutic agents and general fluids therapy. The system is indicated for use with adult, pediatric and neonatal patients in hospital and home care environments.

## Device Story

BodyGuard Infusion Pump System provides precisely-controlled delivery of fluids/medications; supports continuous or intermittent infusion. Inputs include user-programmed infusion parameters; optional inputs from pulse oximeter, drop sensor, and bolus cable. System comprises infusion pump (single/dual channel), docking station, and accessories. Used in hospital and home care environments by clinicians or patients. Output includes fluid delivery and physiological monitoring data (SpO2, pulse rate). Healthcare providers use output to manage therapy, adjust infusion rates, and monitor patient status. Benefits include accurate medication administration and integrated patient monitoring.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Infusion pump system with single/dual channel models. Includes optional pulse oximeter, drop sensor, and bolus cable. Connectivity via docking station. Intended for various clinical routes including IV, subcutaneous, and epidural. Software-controlled infusion rates.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of the HHS symbol, which is a stylized representation of a human figure, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" arranged in a circular fashion around the symbol. The symbol is composed of three curved lines that resemble a stylized human figure in motion. The text is in uppercase letters and is evenly spaced around the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# MAY 1 2 2006

Mr. Gali Tzury Quality Assurance Manager Caesarea Medical Electronics, Limited Caesarea Industrial Park - 16 Shacham Street P.O. Box 4294 Caesarea, ISRAEL 38900

Re: K060479

Trade/Device Name: BodyGuard Infusion Pump System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN, DOA Dated: February 13, 2006 Received: February 23, 2006

### Dear Mr. Tzury:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

## Page 2 - Mr. Tzury

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Auy Sulom for

Chiu Lin Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications For Use

## 510(k) Number:

BodyGuard Infusion Pump System Device Name:

#### Indications for Use:

The BodyGuard infusion Pump system is designed for infusion and The pump is intended for infusion of medications or fluids monitoring. requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, enteral, in close proximity to nerves, and into an intraoperative site (soft tissue/body cavity/surgical wound site). Monitoring accessory currently includes Pulse Oximeter, used for non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The system is intended for patients who require maintenance medications, analgesics, PCA therapy, parenteral and enteral nutrition fluids, chemotherapeutic agents and general fluids therapy. The system is indicated for use with adult, pediatric and neonatal patients in hospital and home care environments.

The BodyGuard Infusion system includes: Infusion Pump (one/dual channel models) Docking Station Bolus cable (optional) Drop sensor (optional) Pulse Oximeter (optional)

Prescription Use (Part 21 CFR 801 Subpart D)

i firmber:_

AND/ OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

of Anesthesiology, General Hospital,
on Control, Dental Devices
Pa

Page 1 Of 1

· Caesarea Medical Electronics Ltd. · Traditional 510(k) Premarket Submission · BodyGuard Infusion Pump System · Confidential

7

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN/K060479](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FRN/K060479)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
