The Hospira GemStar® Infusion Pump System with Hospira GemStar Connect™ Software is intended for use in intravenous arterial, subcutaneous, short term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products. The indications for use include hospital, ambulatory, and home care environments. The pump must be used with sterile, dedicated, GemStar® administration sets.
Device Story
Hospira GemStar is a single-channel, software-controlled, electromechanical infusion pump using a volumetric, piston-driven, fluid displacement principle. A stepper motor meters fluids through sterile, dedicated, non-pyrogenic, latex-free GemStar administration sets. Power options include AC adapter, rechargeable battery pack, or disposable AA alkaline/lithium batteries. Used in hospital, ambulatory, and home care environments by healthcare practitioners. User interface allows programming of fluid delivery via weight-based or medication-based units. Pump displays provide status of operations, alarms, and flow parameters. Functions as both pole-mounted and ambulatory device. GemStar Connect software provides remote communication capabilities. Output allows clinicians to manage therapy protocols, ensuring accurate fluid delivery to patients.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics, performance specifications, and intended use to the predicate devices.
Technological Characteristics
Volumetric, piston-driven, fluid displacement principle; stepper motor actuation. Single-channel configuration. Power: AC mains, rechargeable battery, or disposable AA batteries. Materials: Sterile, dedicated, non-pyrogenic, latex-free administration sets. Connectivity: GemStar Connect software for remote communication. Software-controlled electromechanical system.
Indications for Use
Indicated for patients requiring intravenous, arterial, subcutaneous, or short-term epidural infusion of fluids, medications, nutritional fluids, or blood products. Suitable for use in hospital, ambulatory, and home care settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{1}------------------------------------------------
Confidential
# 510(K) Summary
1. Name of Submitter:
Hospira, Incorporated 275 North Field Drive Lake Forest, Illinois 60045
Owner/Operator # 9063339
2. Manufacturer and Establishment Registration Number:
Hospira, Inc. - Morgan Hill 755 Jarvis Drive Morgan Hill, CA 95037
Establishment Registration # 2921482
3. Proprietary or Trade Name of Proposed Device: Hospira GemStar I.V. Infusion Pump
- 4. Common Name: Infusion Pump
5. Device Classification, Pancode and ProCode: Class II, 80-FRN (Infusion Pump) Class II, 80-FPA (Administration Sets)
- 6. Performance Standards: No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for intravenous infusion pumps. Infusion pumps are listed in 21 CFR 880.5725.
## 7. Intended Use:
The Hospira GemStar IV Infusion Pump is intended for use in intravenous, arterial, short-term epidural, and parenteral administration of general I.V. fluids, medications, nutritionals fluids, and blood/blood products to patients.
#### 8. Indications for Use:
The Hospira GemStar® Infusion Pump System with Hospira GemStar Connect™ is intended for use in Intravenous (central line or peripheral access), arterial, subcutaneous, short term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products.
The indications for use include hospital, ambulatory, and home care environments. The pump must be used with sterile, dedicated, GemStar® administration sets.
#### 9. Proposed Device Description:
The Hospira GemStar" Infusion Pump Systems are a family of single channel, software controlled, electromechanical infusion pumps that operate on a volumetric, piston driven, fluid displacement principle. An in-line cassette is used to meter IV fluids through sterile dedicated administration sets designed to be used exclusively with GemStar infusers. Power options included an AC main adaptor, a rechargeable battery pack, a docking station, and two disposable AA alkaline or lithium batteries. The
{2}------------------------------------------------
user interface allows the healthcare practitioner to program fluid delivery through a variety of weight and medication based units. The pump displays provide visible indication of several functions, including active pump operations, alarm and program status, and the parameters of fluid flow. The infusers function as both pole mounted and ambulatory infusion pumps.
As of May 03, 2004, both the infusers and the dedicated GemStar® sets are manufactured and distributed by Hospira Incorporated, formerly the Hospital Products Division of Abbott Laboratories.
All Hospira GemStar® I.V. Infusion Pumps are single channel pumps that are available in the following configurations:
| | Overview of GemStar I.V. Infusion Pump Therapies and Configurations | |
|----------------------------------|---------------------------------------------------------------------|----------------------|
| 7 Therapy Pump | 6 Therapy Pump | Pain Management Pump |
| List #: 13000-04 | List #: 13100-04 | List #: 13150-04 |
| TPN (Total Parenteral Nutrition) | TPN (Total Parenteral Nutrition) | |
| Pain Management | Intermittent | |
| Intermittent | Continuous | |
| Continuous | Weight-Dosed | Pain Management Only |
| Weight-Dosed | mL/hr Only | |
| Variable Time | Variable Time | |
| ML/hr Only | | |
### 10. Predicate Device Information:
Infusion pumps cleared for commercial distribution and determined to be appropriate for use as predicates are summarized in the following table.
| 510(k)# | Product Name | Clearance Date |
|---------|--------------------------------------|----------------|
| K023062 | Abbott GemStar® Infusion Pump System | 09/30/2002 |
| K000821 | GemStar® I.V. Infusion Pump | 05/24/2000 |
#### 11. Statement of Substantial Equivalence:
The Hospira GemStar® Infusion Pump System with Hospira GemStar Connect® is substantially equivalent to the predicate Abbott GemStar® Infusion Pump Systems and Abbott GemStar Connect™ Software based on the following characteristics.
#### Similarities:
- 1) Same intended use and indications for use.
- 2) Same fundamental scientific technology.
- 3) Same physical, operational, and performance specifications,
- 4) Same features for programmability of therapy protocols.
{3}------------------------------------------------
# 12. Comparison to Legally Marketed Device(s)
| Factors | Subject Device(s)<br>Hospira GemStar I.V. Infusion Pump System with<br>Hospira GemStar Connect | Predicate Device(s)<br>Abbott GemStar<br>Infusion Pump System |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Intended Use | Intended for use in intravenous arterial, subcutaneous,<br>short term epidural infusion and parenteral administration<br>of general I.V. fluids, medications, nutritional fluids, and<br>blood/blood products to a patient. | Same |
| Indications for Use | Hospital, ambulatory, and home care environments using<br>sterile, dedicated, GemStar® administration sets. | Same |
| Operating Principle | Volumetric, piston driven, fluid displacement principle.<br>Stepper motor with in-line cassette meters IV fluids<br>through sterile dedicated administration sets.<br>Programmable fluid delivery through a variety of weight<br>and medication based units. Visible indication of several<br>functions, including active pump operations, alarm and<br>program status, and the parameters of fluid flow. | Same |
| Administration Sets and<br>Fluid Contact Materials | Sterile, dedicated, non-pyrogenic, latex-free "GemStar"<br>administration sets. | Same |
| Physical Features | Materials, Size, Weight, Input Lines, Output Lines,<br>Power Sources, Battery Type, Power Cord | Same |
| Environmental Features | Operating Temperature, Storage Temperature, Relative<br>Humidity, Pressure | Same |
| Performance Features | Delivery Rates, VTBI Range, Dose Units, Delivery<br>Accuracy, Delivery Modes, Therapies, Distal Occlusion<br>Limits, Proximal Occlusion Limits, Alarm Types and<br>Conditions, Default Drug Library. | Same |
| BioMed Settings | Configuration settings available for customization. | Same |
| Accessories (Optional) | GemStar Connect™ Remote Communication Software<br>(Clinician Kit, Patient Kit), Docking Station, Bolus<br>Cord, Pole Clamps (2), Battery Pack, Lockboxes (3), AC<br>Mains Adapters (2), Carrying Cases (4) and Carrier,<br>Serial Cable | Same |
## 13. Summary of Substantial Equivalence
Summary of Subtainar Digar Micheles
The Hospira GemStar® I.V. Infusion Pump System with Hospira GemStar Connect" as described in This submission is substantially equivalent to the predicate Abbott GemStar® Infusion Pump System
this submission is substantially equivalent to the predicate Abbott GemStar"
- 1) intended use,
- 2) indication for use,
- 3) fundamental technology and operating principle,
- 4) physical, environmental and performance features, and
- 5) materials of construction for all infuser components and administration sets.
{4}------------------------------------------------
# 14. Statement of Safety and Effectiveness
The Hospira GemStar® I.V. Infusion Pump System with Hospira GemStar Connect™ meets the functional claims and intended use as described in the product labeling, and is as safe and effective in terms of substantial equivalence as the predicate Abbott GemStar® Infusion Pump System with Abbott GemStar Connect™.
Prepared and submitted on October 28, 2004 by:
Patricia Miderki
Patricia Melerski Manager Global Device Regulatory Affairs Hospira, Inc. 275 North Field Drive Lake Forest, IL 60045 Phone: 224/212-4880 Fax: 224/212-5401
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol of three wavy lines, which are meant to represent the human form.
Public Health Service
NOV 1 7 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Patricia Melerski Manager, Global Regulatory Affairs Hospira, Incorporated Dept. 389, Bldg. H-2 275 North Field Drive Lake Forest, Illinois 60045-5045
Re: K042980
Trade/Device Name: Hospira GemStar® Infusion Pump System Hospira GemStar Connect™ Software Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: October 28, 2004 Received: October 29, 2004
Dear Ms. Melerski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Melerski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Clive
Shia-Lirn, Ph.D.
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
| Indications for Use Statement | |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K042980 |
| Device Name: | Hospira GemStar <sup>®</sup> Infusion Pump System<br>Hospira GemStar Connect <sup>™</sup> Software |
| Indications for Use | The Hospira GemStar <sup>®</sup> Infusion Pump System with Hospira GemStar Connect <sup>™</sup> Software is intended for use in intravenous arterial, subcutaneous, short term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products.<br>The indications for use include hospital, ambulatory, and home care environments. The pump must be used with sterile, dedicated, GemStar <sup>®</sup> administration sets. |
| Prescription Use _X_<br>(Part 21 801 Subpart D) | AND/OR | Over-The Counter Use<br>(Part 21 CFR 807 Subpart C) |
|-------------------------------------------------|--------|-----------------------------------------------------|
|-------------------------------------------------|--------|-----------------------------------------------------|
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
A
(Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control. Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.