K030323 · Mammendorfer Institute For Physics & Medicine GmbH · FRN · Apr 2, 2003 · General Hospital
Device Facts
Record ID
K030323
Device Name
MRI-CADDY
Applicant
Mammendorfer Institute For Physics & Medicine GmbH
Product Code
FRN · General Hospital
Decision Date
Apr 2, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended medical application of MRI-Caddy with three 2000-Series syringe pumps is to produce controlled movement of the plunger of a syringe to inject a set amount of therapeutic fluid into a patient within a hospital setting at a set rate and at set times. The MRI-Caddy is designed for use in an MR-environment at a maximum magnetic field strength of 20mT.
Device Story
MRI-Caddy is a mobile cart housing three Medex 2000-series syringe pumps and a power supply. Designed for use in MR-environments, the chassis incorporates Faraday cage screening to prevent interference. The device functions by mechanically driving a syringe plunger to deliver therapeutic fluids at controlled rates and times. Operated by clinicians qualified in MRI and infusion procedures, the device allows for fluid administration during MR imaging. The system provides the same infusion functionality as standard Medex syringe pumps, enabling patient treatment within the MR suite while mitigating risks associated with magnetic field exposure.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and adherence to safety standards (EN 60601-1 / IEC 601-1).
Technological Characteristics
Cart-based infusion system; integrates three Medex 2000-series syringe pumps; Faraday cage screening for MR compatibility (max 20mT); electrical power supply; mechanical plunger actuation; conforms to EN 60601-1/IEC 601-1 safety standards.
Indications for Use
Indicated for patients requiring controlled infusion of therapeutic fluids in a hospital setting, specifically within an MR-environment (up to 20mT).
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Medex Model 3001 Syringe Infusion Pump
Medex Model 2001 Syringe Infusion Pump
Related Devices
K021988 — MEDRAD CONTINUUM MR COMPATIBLE INFUSION SYSTEM · Medrad, Inc. · Aug 29, 2002
K040899 — MEDEX 3000 SERIES MR SYRINGE INFUSION PUMP · Medex, Inc. · Apr 30, 2004
K061128 — MODIFICATION TO MEDRAD CONTINUUM MR INFUSION SYSTEM · Medrad, Inc. · May 22, 2006
K191910 — SpaceStation MRI · B. Braun Melsugen AG · Mar 12, 2020
K050301 — MRIDIUM 3850 MRI PUMP SYSTEM · Iradimed Corporation · Mar 24, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Number: _______________________________________________________________________________________________________________________________________________________________ Date:
Page 1 of 2
APR - 2 2003
K030323
# 510(k) Summary
#### Introduction
This summary is intended to comply with requirements of the SMDA and 21CFR§807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.
## 510(k) Submitted by
Mammendorfer Institut für Physik und Medizin GmbH Oskar-von-Miller-Strasse 6 82291Mammendorf / Munich, Germany
## 510(k) Correspondent
Robert N. Clark, President and Senior Consultant Medical Device Regulatory Advisors 13605 West 7th Ave., Golden, CO USA Tel: 303-234-9412 / Fax: 303-234-9413
# Date Prepared
January 24, 2003
#### Trade Name of Device
MRI-Caddy
### Common Name of Device
Infusion Pump
#### Classification Name
Infusion Pump
### 510(k) Classification
880.5725 Class II
### Device Description and Intended Use
The MRI-Caddy® is a cart with an integrated chassis, which is equipped with three Medex syringe pumps and a power supply. The MRI-Caddy's mechanical construction makes it possible to position the system with the MR-room. For this purpose the system was equipped with adequate screening following the principle of a Faraday cage. The function principle corresponds to that of the Medex syringe pumps already on the market.
{1}------------------------------------------------
| 510(k) Number: | |
|----------------|--|
| Date: | |
Page 2 of 2
# Predicate Devices
Syringe Infusion Pump. Model 3001 Syringe Infusion Pump. and Medfusion Model 2001 Syringe Infusion Pump manufactured by Medex, Inc.
The basic design, material, chemical composition of the Medex pump series 2000 used in the MRI-Caddy are the same as for the current corresponding Medex Medfusion model pumps. They are identical in operation, features, and form as their predicate devices, and represent no technological differences.
# Non-Clinical Testing
The requirements of the following standards have been used in part to establish substantial equivalence:
EN 60601-1 / IEC 601-1 "Medical Electrical Equipment - Part 1: General Requirements for Safety", including Amendment 1 (1991) & Amendment 1 (1995)
The company did not conduct, nor depend on, clinical studies in order to establish substantial equivalence.
# Risk Management
This device has been designed to either completely eliminate or mitigate known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program.
The user must be qualified in MRI and infusion pump procedures, and must be familiar with all labeling and instructions for use associated with the device. The company believes many device health hazards are due to user error, or failure to follow instructions for use.
Mammendorfer Institut für Physik und Medizin GmbH believes that the MRI-Caddy is safe and effective when used as instructed by knowledgeable and trained personnel, and is substantially equivalent to the legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's emblem, which is a stylized representation of a human figure. The emblem is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 2 2003
Mammendorfer Institut Für Physik und Medizin GmbH C/O Mr. Robert N. Clark Medical Device Regulatory Advisors 13605 West 7th Avenue Golden, Colorado 80401
Re: K030323
Trade/Device Name: MRI-Caddy® Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: January 24, 2003 Received: January 30, 2003
Dear Mr. Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Clark
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Rusner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
510(k) Number: K030323
Device Name:
MRI-Caddy®
Indications for Use:
The intended medical application of MRI-Caddy with three 2000-Series syringe pumps is to produce controlled movement of the plunger of a syringe to inject a set amount of therapeutic fluid into a patient within a hospital setting at a set rate and at set times. The MRI-Caddy is designed for use in an MR-environment at a maximum magnetic field strength of 20mT.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patuxn Cusenite
510(k) Number: K030323
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.