DISPOSABLE TRANSFER SETS WITH AND WITHOUT SWABBABLE VALVES AND/OR CHECK VALVES
K063114 · Coeur, Inc. · FPK · Feb 9, 2007 · General Hospital
Device Facts
Record ID
K063114
Device Name
DISPOSABLE TRANSFER SETS WITH AND WITHOUT SWABBABLE VALVES AND/OR CHECK VALVES
Applicant
Coeur, Inc.
Product Code
FPK · General Hospital
Decision Date
Feb 9, 2007
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Disposable Transfer Sets with and without Swabbable Valves and/or Check Valves are for use in transferring contrast, saline, or other diagnostic fluids from bulk containers into syringes.
Device Story
Disposable transfer sets consist of tubing, connectors, spikes, clamps, and optional swabbable or check valves. Device connects bulk fluid containers to injector syringes; allows aspiration of contrast, saline, or diagnostic fluids into syringe for subsequent injection. Used in clinical settings; operated by healthcare professionals. Facilitates fluid delivery by providing a sterile pathway between bulk supply and power injector syringe. Benefits include efficient, sterile fluid transfer.
Clinical Evidence
No clinical data. Bench testing only: verification of functional performance, including bond and leak testing, and sterilization cycle validation (SAL 10^-6).
Technological Characteristics
Materials: PVC tubing, various plastic components (spikes, luers, stopcocks, clamps). Connection: ISO 594 Luer. Sterilization: Ethylene Oxide (EtO). Form factor: Tubing assembly with connectors. No software or electronic components.
Indications for Use
Indicated for transferring contrast, saline, or other diagnostic fluids between bulk containers and a syringe.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K022431 — MEDRAD TRANSFER SET · Medrad, Inc. · Sep 5, 2002
K102073 — MED-RX TRANSFER SET, MODELS 10-1227TS, 10-1300TS, 10-1305TS AND 10-1306TS · Benlan, Inc. · Nov 8, 2010
K133147 — BRACCO INJEENERING TRANSFER SET · Bracco Injeneering S.A. · Jun 20, 2014
K041178 — EMPOWER TRANSFER SET, (CAT. NO. 7725) · E-Z-Em, Inc. · Jul 16, 2004
K161723 — ulrich Transfer Set · Ulrich GmbH & Co. KG · Dec 21, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "COEUR" in a bold, serif font, with a stylized heart symbol to the left. The heart symbol is composed of two overlapping hearts and a circle above them. The "R" in "COEUR" has a registered trademark symbol next to it. The text and symbol are all in black and are set against a white background.
K063114
COEUR, INCORPORATED 209 CREEKSIDE DRIVE PO BOX 2125 WASHINGTON, NC 27889 252-946-1963 · FAX: 252-946-2317 E-MAIL: www.coeurinc.com
Coeur, Inc.
9 2007 FEB
Coeur Medical, a division of Coeur, Inc. Disposable Transfer Sets with and without Swabbable Valves and/or Check Valves
# 510(k) Summary
Coeur Medical, a division of Coeur, Inc. Name: Submitter: 1. 209 Creekside Drive Address: Washington, NC 27889 (615) 547-7923 (Corporate Office) Phone: (615) 547-7937 Fax: Debra F. Manning, VP, Q & RA Contact: October 5, 2006 Date:
Trade/Proprietary Name: Disposable Transfer Sets with and without 2. Device: Swabbable Valves and/or Check Valves Transfer Sets Common/Usual Name: Fluid Delivery Tubing (a product included Classification Name: in the definition of Intravascular administration set, 21 CFR 880.5440, Product Code FPK or FPA)
#### Legally Marketed Devices to which Substantial Equivalence is claimed: 3.
Fill Tube (K873597) - Coeur SVTS - Swabable Valve Transfer Set (K031808) - Medrad Transfer Set (K022431) - Medrad Contrast Management System (K961794) – Merit Empower Transfer Set (K041178) - E-Z-EM, INC.
#### Device Description: 4.
The Disposable Transfer Sets with and without Swabbable Valves and/or Check Valves are a combination of connectors and accessories, such as clamps, tubing, and spikes, and are used for fluid delivery. The device is designed, like other legally marketed devices, for one end to connect to the syringe to be filled (which is installed on an injector with the plunger forward) and the other end is connected to the bulk container. The plunger is then retracted and contrast, saline, or other diagnostic fluid is drawn into the syringe. The materials and properties of the device are tabled in Item 6, below.
{1}------------------------------------------------
#### ડ. Intended Use of Device:
יווי
. . .
The Disposable Transfer Sets with and without Swabbable Valves and/or Check Valves are for use in transferring contrast, saline, or other diagnostic fluids from bulk containers into syringes. As the table in Item 6 demonstrates, this is consistent with other legally marketed devices.
### Summary of Technological Characteristics As Compared to Predicate Devices: 6.
| Technological<br>Characteristics | Proposed<br>Device | Coeur Device | Medrad<br>Devices | Merit Device | E-Z-EM,<br>INC. | Rationale for<br>Applicable<br>Differences |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | For use in<br>transferring<br>contrast, saline,<br>or other<br>diagnostic fluids<br>from bulk<br>containers into<br>syringes. | For use in the<br>delivery of<br>contrast media to a<br>syringe. | For use in the<br>delivery of<br>contrast media<br>and saline to a<br>syringe. | For use in the<br>aspiration of<br>contrast media<br>from a<br>container into<br>a syringe and<br>for subsequent<br>injection into<br>the catheter. | Intended to<br>deliver fluid<br>(contrast<br>media or<br>saline) from a<br>container into<br>a CT Power<br>Injector<br>Syringe. | NA - The<br>proposed device is<br>an extension of the<br>existing Coeur<br>device to enable<br>use of contrast,<br>saline, or other<br>diagnostic fluids<br>provided in bulk<br>containers. |
| Connector Tube | Flexible, clear<br>or tinted, PVC | Polyethylene | PVC | Clear, flexible, PVC<br>plastic | PVC | None |
| Sterile | Yes | Yes | Yes | Yes | Yes | NA |
| Sterilization<br>Method | EtO | Same | Same | Same | Same | NA |
| Components | Combinations of<br>spikes, male and<br>female luers,<br>check valves (is<br>used),<br>stopcocks,<br>clamps, various<br>dust caps, and<br>swabbable<br>valves (Halkey-<br>Roberts) | Tubing formed in<br>the shape of a "J" | Vented spike,<br>dust cap,<br>stopcock, pinch<br>clamp<br>And<br>Vented spike,<br>female luer,<br>swabbable<br>threaded valve<br>(Halkey-<br>Roberts) | Spike<br>assembly with<br>stopcock, fixed<br>male luer,<br>burette<br>chamber and<br>check valves | Female luer,<br>clamp, and<br>spike | NA - No new<br>components or<br>materials or<br>intended uses are<br>being proposed<br>with the proposed<br>device |
| Connection<br>Method | ISO 594 Luer | Inner diameter of<br>tubing is placed<br>over syringe luer | ISO 594 Luer | ISO 594 Luer | ISO 594 Luer | NA - Device is<br>connected to<br>syringe via luer<br>connection |
| Multi-Fill Fill<br>Sets | Yes | If required. | Yes | Yes | Yes | NA |
| Packaging | Sealed Pouch | Same | Same | Same | Same | NA |
| Shelf Life | 3 years | Same | 5 years | 5 years | 3 years | There are no<br>significant<br>differences in the<br>materials or<br>components<br>utilized |
{2}------------------------------------------------
If Substantial Equivalence was based on an Assessment of Performance Data, the following information is also provided:
- Nonclinical Tests Submitted: 1. Verification of functional performance has been performed. Bond testing and leak testing were conducted to verify the device performs acceptably. Coeur conducted a product adoption study to verify that the Coeur sterilization cycle is able to sterilize the proposed device with and SAL of 10°.
#### Clinical Tests Submitted: 2. NA
- Conclusions Drawn from Nonclinical and Clinical Tests Submitted: 3. The primary difference between the proposed Coeur devices and other legally marketed predicate devices is that Coeur, who has experience in the production of tubing extension products and has the facilities and equipment for assembly of the proposed devices, will assemble the components to make the proposed devices which will be sterilized in Coeur's sterilization cycle. The results of the testing verify the proposed devices are suitable for their intended use.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white and appears to be a scanned image.
Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850
Ms. Debra F. Manning Vice President, Quality & Regulatory Affairs Coeur, Incorporated 704 Cadet Court Lebanon, Tennessee 37087
FEB 9 2027
Re: K063114
Trade/Device Name: Coeur, Inc. Disposable Transfer Sets with and without Swabbable Valves and/or Check Valves Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPK, FPA Dated: January 24, 2007 Received: January 25, 2007
Dear Ms. Manning:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Manning
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Susan Runnos
Chiu Lin, Ph.D,
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Coeur, Inc. Disposable Transfer Sets with and without Swabbable Device Name: Valves and/or Check Valves
Indications For Use:
for use in transferring contrast, saline, or other diagnostic fluids between bulk containers and a syringe
Prescription Use (Part 21 CFR 801 Subpart D)
יווי
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chas Evans
General Manager,
Local Engines
4-1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.