K041178 · E-Z-Em, Inc. · FPK · Jul 16, 2004 · General Hospital
Device Facts
Record ID
K041178
Device Name
EMPOWER TRANSFER SET, (CAT. NO. 7725)
Applicant
E-Z-Em, Inc.
Product Code
FPK · General Hospital
Decision Date
Jul 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Empower Transfer Set is intended to deliver fluid (contrast media or saline) from a container into a CT power Injector syringe.
Device Story
Empower Transfer Set is a disposable fluid delivery device used to transfer contrast media or saline from a container to a CT power injector syringe. The device consists of a vented spike, medical-grade PVC tubing, a pinch clamp, and a female luer connector. It is operated manually by clinical staff in a radiology or clinical setting. The device facilitates the filling of injector syringes, ensuring fluid delivery for CT imaging procedures. It is a passive mechanical device; no electronic or software components are involved.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1 and 10993-7.
Technological Characteristics
Disposable fluid transfer set. Components: Vented spike, 6" medical-grade PVC tubing, pinch clamp, ISO 594 female luer connector. Sterilization: Ethylene Oxide (EtO). Biocompatibility: ISO 10993-1 and 10993-7 compliant.
Indications for Use
Indicated for the transfer of contrast media or saline from a container into a CT power injector syringe.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K022431 — MEDRAD TRANSFER SET · Medrad, Inc. · Sep 5, 2002
K173913 — MEDRAD® Imaging Bulk Package Transfer Set · Bayer U.S., LLC · May 4, 2018
K063114 — DISPOSABLE TRANSFER SETS WITH AND WITHOUT SWABBABLE VALVES AND/OR CHECK VALVES · Coeur, Inc. · Feb 9, 2007
K133147 — BRACCO INJEENERING TRANSFER SET · Bracco Injeneering S.A. · Jun 20, 2014
K161723 — ulrich Transfer Set · Ulrich GmbH & Co. KG · Dec 21, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 1 6 2004
Image /page/0/Picture/1 description: The image shows a handwritten alphanumeric string. The string appears to be "K041178" and is written in black ink on a white background. The numbers are slightly slanted to the right, and the overall impression is that of a quick, informal note.
Visualize a bealthier world
## Empower Transfer Set 510(k) Summary
| Submitter's Information: | E-Z-EM, Inc.<br>717 Main Street<br>Westbury, NY 11590 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone Number: | (516) 333-8230, ext. 2345 |
| Fax Number: | (516) 333-8278 |
| Contact: | Steven Hartman, Senior Project Engineer, CT Injectors |
| Date Prepared: | June 28, 2004 |
| Trade Name: | Empower™ Transfer Set |
| Common Name: | Transfer Set |
| Classification Name: | Tubing, Fluid Delivery [21 CFR 880.5440]<br>Product Code: FPK<br>Regulatory Class: II |
| Predicate Device: | Venusa Flu Ven Tr-1 (K791131) |
| Indications for Use: | The Empower Transfer Set is intended to deliver fluid<br>(contrast media or saline) from a container into a CT<br>power Injector syringe. |
| Device Description: | The Empower Transfer Set is a disposable set used to<br>transfer contrast media or saline from a container that<br>can be spiked to a CT power Injector syringe. |
E-Z-EM Empower Transfer Set 510(k) Submission Summary Page 1 of 2
{1}------------------------------------------------
## Comparative Summary Tables
| Parameter | Proposed E-Z-EM Empower<br>Transfer Set | Predicate Venusa Flu Ven Tr-1 (K791131) |
|--------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Indication For Use | It is intended to be used in the delivery<br>of fluid (contrast media or saline) into a<br>syringe. | The device has been categorized for use of<br>I.V. fluid transfer. |
| Female Luer | Polycarbonate, Clear | Same |
| Clamp | Pinch clamp | Same |
| Spike | Vented Spike | Same |
| Tubing | 6" Length<br>PVC - Medical Grade | 7" Length<br>Same |
| Connection method | ISO 594 Luer | Same |
| Sterility | Ethylene Oxide (EtO) sterilized | Same |
| Biocompatibility | Tested to ISO 10993-1 and 10993-7 | Same |
A summary of the proposed device as compared to the predicate device is as follows:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white circular logo. The logo features a symbol of three stylized lines resembling a bird in flight. The symbol is enclosed within a circle, and the text "U.S. Public Health Service - USA" is written around the perimeter of the circle. The text is arranged to follow the curve of the circle.
Public Health Service
JUL 1 6 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Steven Hartman Senior Project Engineer E-Z-EM, Incorporated 717 Main Street Westbury, New York 11590
Re: K041178
Trade/Device Name: Empower Transfer Set Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPK Dated: May 4, 2004 Received: May 5, 2004
Dear Mr. Hartman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 – Mr. Hartman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsma/dsmamain.html
Sincerely vours.
Ken Morley
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): _K041178
Device Name: Empower Transfer Set
Indications For Use:
The Empower Transfer Set is intended to deliver fluid (contrast media or saline) from a container into a CT power Injector syringe.
Prescription Use __ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Amore Nareem for Abdul
Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number:________________________________________________________________________________________________________________________________________________________________
Empower Transfer Set 510k Submission
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.