MINIMED SOF-SET ULTIMATE QR INFUSION SET MODELS 315 AND 316
K974163 · Medtronic Minimed · FPA · Feb 2, 1998 · General Hospital
Device Facts
Record ID
K974163
Device Name
MINIMED SOF-SET ULTIMATE QR INFUSION SET MODELS 315 AND 316
Applicant
Medtronic Minimed
Product Code
FPA · General Hospital
Decision Date
Feb 2, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MiniMed Sof-set Ultimate QR is intended for the subcutaneous infusion of medicine, including insulin, from an external infusion pump. The set is not intended nor indicated for use with blood or blood products.
Device Story
Infusion administration set for subcutaneous delivery of insulin/medication from external infusion pump; connects proximally to medicine reservoir via female Luer connector; inserts distally into subcutaneous tissue via 26-gauge stainless steel introducer needle; features 24-gauge FEP Teflon indwelling catheter. Hub includes winged configuration for handling/stability; incorporates antibacterial dressing and adhesive dressing for securement. Used in clinical or home settings by patients or caregivers. Modifications from predicate include flatter/smoother hub/wings, clear catheter, rotated adhesive tab for auto-inserter compatibility, IV 3000 adhesive tape, and higher-density polyethylene tubing liner. Higher-density liner increases sensitivity of pump occlusion alarms. Device facilitates continuous subcutaneous insulin infusion (CSII) therapy, improving patient convenience and alarm reliability.
Clinical Evidence
No clinical data. Substantial equivalence based on technological characteristics and design modifications.
Technological Characteristics
Indwelling catheter: FEP Teflon. Tubing: PVC with high-density polyolefin liner. Introducer needle: 304 stainless steel. Hub: Winged configuration. Adhesive: IV 3000. Connectivity: Mechanical Luer connection to external infusion pump reservoir. Sterilization: Not specified.
Indications for Use
Indicated for subcutaneous delivery of medicine, including insulin, from an external infusion pump to patients requiring such therapy.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
MiniMed Sof-set QR infusion sets, models 115 and 116
Submission Summary (Full Text)
{1}------------------------------------------------
MiniMed Inc. Premarket Notification - 510(k) Sof-set Ultimate QR infusion set, models 315 and 316
## Part E. 510(k) Summary
Submitter: MiniMed® Inc. 12744 San Fernando Rd., Sylmar, California 91342.
Contact: Don Selvey, Regulatory Affairs (818) 362-5958, 3011; (520) 527-0107 (v/t).
Name of Device: MiniMed® Sof-set® Ultimate QR® infusion sets, models 315 and 316.
Predicate Device: MiniMed Sof-set QR infusion sets, models 115 and 116.
Description of the New Device: The MiniMed Sof-set Ultimate QR infusion sets, models 315 and 316, are infusion administration sets, connecting to a medicine reservoir proximally, and inserted in the subcutaneous tissue of a user distally by means of an introducer needle. The reservoir to which the infusion set attaches proximally is inserted into an external infusion pump, such as the MiniMed 507 insulin pump.
The administration set attaches to the reservoir syringe by means of a female Luer connector, and subcutaneously in the user through an indwelling catheter made of FEP Teflon. The tubing is made of polyvinyl chloride (PVC) with a polyolefin liner. This configuration of PVC and polyolefin has been trademarked by MiniMed as Polyfin.
The 24 gauge indwelling catheter is introduced into the subcutaneous tissue by a removable 26 gauge introducer needle made of 304 stainless steel. The needle, indwelling catheter, and tubing share a common hub. The hub incorporates a winged configuration to facilitate handling of the administration set during insertion and stability following insertion.
On the skin-contact side of the wings, around the Teflon catheter, is an antibacterial dressing. Additionally, an adhesive dressing covers the wings of the administration set, securing the subcutaneous catheter and infusion line to the user.
Intended Use of the New Device: The MiniMed Sof-set Ultimate QR is intended for the subcutaneous infusion of medicine, including insulin, from an external infusion pump. The set is not intended nor indicated for use with blood or blood products.
®MiniMed, Sof-set, and QR are registered trademarks of MiniMed Inc.
000009
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for MiniMed. The logo consists of a black square with white lines that form a wave pattern on the left side. To the right of the square is the word "MiniMed" in a serif font, with a trademark symbol in the upper right corner.
MiniMed Inc. Premarket Notification - 510(k) Sof-set Ultimate QR infusion set, models 315 and 316
Comparison of the Technological Features of the New Device and Predicate Device: The modified devices are substantially similar to the lawfully marketed predicate devices. Both are intended for subcutaneous delivery insulin or other appropriately labeled medicines from an infusion pump to the pump user. Both sets attach proximally to a reservoir by means of a Luer connector, and insert distally into the device user's subcutaneous tissues with a removable stainless steel needle and flexible catheter. Both devices are made of substantially similar materials.
Several differences exist between the new and predicate device:
1) changes in the shape/configuration of the hub and wings to be flatter and smoother;
2) the indwelling catheter is clear instead of white;
3) a change in the adhesive tab to be slightly longer and rotated 90°, facilitating tab removal prior to insertion and use with an auto inserter device;
4) a change in the adhesive tape: the infusion set will be secured with IV 3000, a product of Smith & Nephew;
5) a change in the polyethylene plastic in the inner layer of the tubing, from low density to a higher density. This has resulted in a more sensitive occlusion alarm, which will be promoted by MiniMed.
These modifications do not affect the safety or effectiveness of the device.
Signed,
Don Selvey 10-27-97
date
Regulatory Affairs MiniMed Inc.
## 000010
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized eagle emblem. The eagle is depicted with its head facing left and its wings spread, creating a sense of strength and protection. The overall design is simple and professional, reflecting the department's role in promoting the health and well-being of the nation.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 1998
Mr. Don Selvey Requlatory Affairs MiniMed® Incorporated 12744 San Fernando Road Sylmar, California 91342
K974163 Re: MiniMed® Sof-set® Ultimate QR® infusion Trade Name: sets, models 315 and 316 Regulatory Class: II Product Code: FPA Dated: October 27, 1997 Received: November 5, 1997
Dear Mr. Selvey:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895 .... A ... ................................................................................................................... substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does
{4}------------------------------------------------
Page 2 - Mr. Selvey
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Directd Division of Dental, Infection Control and General Hospital Devices 10000000 Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
MiniMed Inc. Premarket Notification - 510(k) Sof-set Ultimate QR infusion set, models 315 and 316
## INDICATIONS FOR USE
510(k) Number:
MiniMed Sof-set Ultimate QR infusion sets, models 315 Device Name: and 316. Indications for Use: The MiniMed Sof-set Ultimate QR infusion set is indicated for the subcutaneous delivery of medicine, including insulin, from an external infusion pump.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patrick Cassinite
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Device 510(k) Number 的
Prescription Use (Per 21 CFR 801.109)
or
Over-the-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.