← Product Code [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA) · K973916

# NP MEDICAL CAPLESS LUER ACTIVATED VALVE (K973916)

_Np Medical, Inc. · FPA · Mar 9, 1998 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K973916

## Device Facts

- **Applicant:** Np Medical, Inc.
- **Product Code:** [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA.md)
- **Decision Date:** Mar 9, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Capless Luer Activated Valve, incorporating a luer activated valve, is intended for use in facilitating needleless fluid delivery and may be swabbed with antiseptic just prior to use, thereby eliminating the need for capping between uses.

## Device Story

Capless Luer Activated Valve facilitates needleless fluid delivery; eliminates need for capping between uses. Device consists of two-part assembly: Gland Housing and Gland/Center Post. Gland acts as swabbable valve/gasket; Center Post provides mechanical support and high-pressure seal. Resting state: Center Post flush with housing, fluid path closed. Activation: Syringe/male luer connector forces Gland open and pushes Center Post down, establishing fluid path. Used in clinical settings; operated by healthcare professionals. Swabbable surface allows disinfection per hospital protocol. Benefits include reduced risk of contamination and simplified fluid access.

## Clinical Evidence

No clinical data. Evidence consists of bench testing: mechanical performance testing, biocompatibility testing per ISO 10993, and microbial challenge testing to verify fluid pathway sterility.

## Technological Characteristics

Two-part assembly: Gland Housing and Gland/Center Post. Gland: silicone valve/gasket with slit. Center Post: mechanical support and high-pressure seal. Materials: silicone and plastic. Mechanical press-fit assembly. Sterility maintained via closed fluid path design.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- ICU Medical Clave Needleless Connector ([K915571](/device/K915571.md) / [K941190](/device/K941190.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K973916

#### 510(K) SUMMARY

MAR - 9 1998

# 1. SUBMITTER:

NP Medical, Inc. 101 Union Street Clinton, MA 01510 Telephone: 978-365-2500

Contact: Mike Jones, Quality Assurance Manager Date Prepared: October 9, 1997

## 2. DEVICE:

Classification Name: Set, Administration, Intravascular Trade Name: NP Medical Capless Luer Activated Valve Class II per 21 CFR 880.5440 The Product Code is 80FPA

## 3. PREDICATE DEVICE:

The predicate device used to determine substantial equivalence for the NP Medical Capless Needleless Luer Connector is the ICU Medical Clave Needleless Connector (K915571 / K941190).

## 4. DEVICE DESCRIPTION:

The Capless Luer Activated Valve is a two-part device consisting of a Gland Housing Assembly and a Gland/Center Post Assembly, as described below:

The Gland is a valve/gasket which provides a seal against the syringe/luer connector when the device is being utilized. The Gland incorporates a slit to accept the syringe/luer connector. The Gland is also the swabbable surface of the Capless Luer Activated Valve. The valve can be easily swabbed per hospital protocol before each connection.

The Center Post mechanically supports the Gland and serves as the primary high pressure seal to keep the fluid path closed during the resting state.

The Gland and the Center Post are mechanically press-fit together to form the Gland/Center Post Assembly.

{1}------------------------------------------------

In the resting state, the Center Post is flush with the walls of the Gland Housing, ensuring that there is no fluid path. As the device is activated by a syringe/male luer connector, the flexible Gland is forced open and the Center Post is pushed down. As the Center Post is forced further down into the Gland Housing, the fluid path is established.

#### 5. INTENDED USE:

The Capless Luer Activated Valve, incorporating a luer activated valve, is intended for use in facilitating needleless fluid delivery and may be swabbed with antiseptic just prior to use, thereby eliminating the need for capping between uses.

## 6. COMPARISON OF CHARACTERISTICS:

The Capless Luer Activated Valve and the currently marketed Clave Needleless Connector are very similar in design: Both devices consist of a silicone valve encapsulated within a plastic housing. Both devices have the same Intended Use.

#### 7. PERFORMANCE DATA:

The following performance data was provided in support of the substantial equivalence determination:

- 1. Mechanical Testing:
Mechanical Testing has been conducted on the Capless Luer Activated Valve to demonstrate that the device's performance characteristics are adequate for the intended use.

The results of the Mechanical Testing demonstrate that the NP Medical Capless Luer Activated Valves meet their performance requirements.

- 2. Biocompatibility Testing:
Full biocompatibility testing has been performed on the materials being proposed in this 510(k) submission, per ISO 10993. The materials passed all of the biocompatibility tests.

- 3. Microbial Challenge Testing:
Microbial Challenge testing has been conducted on the Capless Luer Activated Valve. The results of the testing demonstrate that, in response to excessive microbial challenge conditions, the sterility of the fluid pathway was maintained.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 9 1998

Mr. Paul A. Caracciolo Ouality Assurance Manager N.P. Medical Incorporated 101 Union Street Clinton, Maryland 01510

Re : K973916 NP Medical Capless Luer Activated Valve Trade Name: Requlatory Class: II Product Code: FPA Dated: January 23, 1998 Received: January 28, 1998

Dear Mr. Caracciolo:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਬ substantially equivalent determination assumes compliance with annon the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Mr. Caracciolo

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours

Timo By A Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K973916

.

# INDICATIONS FOR USE STATEMENT

The Capless Luer Activated Valve, incorporating a luer activated valve, is intended for use in facilitating needleless fluid delivery and may be swabbed with antiseptic just prior to use, thereby eliminating the need for capping between uses.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Patara Cucente

(Division Sign-Off Division of Dental, Info and General Hospita 510/k) Number

Prescription Use 2 OR

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K973916](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K973916)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
