← Product Code [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA) · K971500

# ACCESS SYSTEMS NEEDLELESS VIAL ADAPTER (K971500)

_Biological & Environmental Control Laboratories · FPA · Aug 18, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K971500

## Device Facts

- **Applicant:** Biological & Environmental Control Laboratories
- **Product Code:** [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA.md)
- **Decision Date:** Aug 18, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Needleless Injection Site is intended to provide additional protection against inadvertent "needlestick" injuries to health care providers during the administration of fluids and medications. The Needleless Vial Adapter provides for the entry into a stoppered vial without the need of a sharp needle by allowing for the use of a blunt cannula.

## Device Story

Device consists of needleless injection site (tubular body with male luer slip) bonded to female luer lock of Medimop Vial Access device; includes enlarged section with retaining ring compressing pre-slit natural rubber or polyisoprene septum. Used by healthcare providers in clinical settings to access stoppered vials; replaces sharp needles with blunt cannulas (e.g., BEC, MED-NET, or IMED shrouded plastic cannulas). Septum re-seals upon cannula removal. Primary benefit is protection against inadvertent needlestick injuries during medication/fluid administration.

## Clinical Evidence

No clinical data. Bench testing and performance data for the injection site were leveraged from the IMED predicate device (K944320).

## Technological Characteristics

Materials: natural rubber or polyisoprene septum. Design: tubular body with male luer slip, female luer lock, and septum retaining ring. Sterilization: validated ethylene oxide process. Connectivity: none (mechanical device).

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- IMED predicate device ([K944320](/device/K944320.md))

## Reference Devices

- Baxter Healthcare needleless vial adapters
- Abbott Laboratories needleless vial adapters
- McGaw, Inc. needleless vial adapters

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K97/500" in a handwritten style at the top. Below this, the text "AUG 18 1967" is printed in a typewriter-like font. The date indicates that the image was taken on August 18, 1967. The text is all in black ink, and the background is white.

Image /page/0/Picture/1 description: The image shows a logo for "bec laboratories, inc." The logo consists of a black circle with the letters "bec" in white inside the circle. Below the letters, the words "laboratories, inc." are written in a smaller font. The logo has a horizontal line design on either side of the circle.

## 510(k) Summary

Headquarters/ Laboratory 705 front street toledo, ohio 43605 phone: (419) 693-5307 fax: (419) 691-0418

Environmental Laboratory 1632 enterprise parkway twinsburg, ohio 44087 phone: (216) 425-8200 fax: (216) 425-1349

Saies Office 129 dixie way south south bend, indiana 46637 phone: (219) 272-2200

> Sales Office 77 west elmwood drive suite 123 dayton, ohio 45459 phone: (513) 438-1711

## Medical Products Division

615 front street toledo, ohio 43605 phone: (419) 693-5307 fax: (419) 691-1227 Re: Trade Name: Common Name:

Access Systems Needleless Vial Adapter Needleless Vial Access Device Classification Name: Tubes, Vials, Systems, Serum Separators, Blood Collection 75JKA

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92 and DSMA 1995.

The B.E.C. Needleless Vial Adater consists of a Needleless Injection Site which is a tubular body with a male luer slip attachment at one end. This site is bonded to the female luer lock of the Medimop Vial Access device. The other end of the injection sites tubular body is provided with an enlarged section adapted to receive a cylindrical retaining ring which surrounds and compresses a pre-slit, natural rubber or polyisoprene septum. The septum retaining ring and enlarged section of the tubular body are adapted to receive and latch with the BEC. MED-NET or IMED shrouded plastic cannula. The pre-slit septum re-seals upon removal of the cannula.

The Needleless Injection Site is intended to provide additional protection against inadvertent "needlestick" injuries to health care providers during the administration of fluids and medications. The Needleless Vial Adapter provides for the entry into a stoppered vial without the need of a sharp needle by allowing for the use of a blunt cannula.

This needleless vial adapter is similar to those marketed by Baxter Healthcare, Abbott Laboratories, and McGaw, Inc. Technological data and performance data for the injection site employed were submitted for the IMED predicate device in 510(k) #K944320.

{1}------------------------------------------------

## 510(k) Summary (Continued)

Packaging of the device is either performed in-house or under contract by a Sterilization is performed in-house using a registered device establishment. validated ethylene oxide process. Both packaging and sterilization procedures are consistent with those generally used by the medical device industry.

Contact Person:

James P. Kulla, President

4/22/97
Date Prepared

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 8 1997

Mr. James P. Kulla President Biological & Environmental Control Laboratories 705 Front Street -----Toledo, Ohio 43605

Re : K971500 Trade Name: Access Systems Needleless Vial Adapter Regulatory Class: II Product Code: FPA Dated: Auqust 1, 1997 Received: August 4, 1997

Dear Mr. Kulla:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Mr. Kulla

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fida.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

Device Name: Needleless Vial Adapter

ations For Use:

The risk to health care providers of "needlestick" injuries has become a major public health and worker safety concern. The Needleless Vial Adapter is intended to provide additional protection against inadvertent "needlestick" injuries to health care providers during the administration of fluids and medications. The Needleless Vial Adapter provides for the entry into a stoppered vial without the need of a sharp needle by allowing for the use of a blunt cannula.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Patricia Cucente

(Division Sig Division of De and General Hoopita 510(k) Number

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K971500](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K971500)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
