K230992 · Yangzhou Wei DE LI Trade Co., Ltd. · FPA · Jun 9, 2023 · General Hospital
Device Facts
Record ID
K230992
Device Name
DJF Intravascular Administration Set
Applicant
Yangzhou Wei DE LI Trade Co., Ltd.
Product Code
FPA · General Hospital
Decision Date
Jun 9, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DJF Intravascular Administration Sets are intended for use with Curlin (Moog) Infusion Pump, for the delivery of fluids from a container or bag to the patient's vascular system
Device Story
DJF Intravascular Administration Sets are sterile, single-use, non-DEHP PVC tubing sets designed for use with the Curlin (Moog) Infusion Pump. Components include a universal spike, slide clamp, male luer lock, and optional anti-free flow valve, 0.2 or 1.2 micron filters, yellow cassette, and blue locator pin. The device facilitates fluid delivery from a container/bag to the patient's vascular system. It is used in clinical settings where the Curlin pump is operated by healthcare professionals. The device ensures compatibility with the pump's mechanical requirements via the cassette and locator pin. Clinical benefit is the reliable, controlled administration of intravenous fluids.
Clinical Evidence
Bench testing only. Performance testing included ISO 80369-7 (connectors), ISO 8536-4 (infusion sets), and ASTM D4169 (shipping). Functional testing for Curlin pump compatibility included rate accuracy, set installation, and upstream/downstream occlusion tests. Biocompatibility testing per ISO 10993-1 for prolonged contact (>24hrs to 30 days) included cytotoxicity, hemolysis, irritation, acute systemic toxicity, sensitization, pyrogenicity, sub-acute/sub-chronic systemic toxicity, endotoxin, and particulate testing. Sterilization validated via Gamma Radiation (VDmax25) per ISO 11137-3.
Technological Characteristics
Materials: Non-DEHP PVC tubing, TOTM plasticizer, ABS plastic, 316L stainless steel spring. Components: Universal spike, slide clamp, male luer lock, anti-free flow valve, 0.2/1.2 micron filters, yellow cassette, blue locator pin. Connectivity: None (mechanical). Sterilization: Gamma radiation (VDmax25). Standards: ISO 80369-7, ISO 8536-4, ISO 10993-1, ISO 11137-3, ASTM D4169, ASTM F1886.
Indications for Use
Indicated for use with Curlin (Moog) Infusion Pump for the delivery of fluids from a container or bag to the patient's vascular system.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
June 9, 2023
Yangzhou Wei De Li Trade Co. Ltd % Aaron Compton President Vesco Devices 541 Lakewood Drive Fairview, Texas 75069
Re: K230992
Trade/Device Name: DJF Intravascular Administration Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: March 13, 2023 Received: April 11, 2023
Dear Aaron Compton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloscher
David Wolloscheck, PhD. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230992
Device Name DJF Intravascular Administration Set
Indications for Use (Describe)
DJF Intravascular Administration Sets are intended for use with Curlin (Moog) Infusion Pump, for the delivery of fluids from a container or bag to the patient's vascular system
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K230992- 510K Summary
| Submitter Information | |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By: | Yangzhou Wei De Li Trade Co. Ltd. |
| Address | Li Xin Bridge, Touqiao Township. Yangzhou City,<br>Jiangsu Province, China 225009 |
| Phone Number | +86-514-87897887 |
| Fax Number | +86-514-87889967 |
| Establish Registration No. | 3008808123 |
| Name of Contact Person | Aaron Compton |
| Contact Person Address | 541 Lakewood Dr., Fairview TX 75069 |
| Contact Person Phone No. | 214-533-1416 |
| Contact Person email | acompton@vescodevices.com |
| Date Prepared | June 9th, 2023 |
| Name of Medical Device | |
| Trade or proprietary name | DJF Intravascular Administration Set |
| Common or usual name | Intravascular Administration Set |
| Classification name | Set, Administration, Intravascular |
| Classification panel | General Hospital (80) |
| Regulation | 21 CFR 880.5440 |
| Product Code(s) | FPA |
| Device Classification | II |
| Legally marketed device(s) to<br>which equivalence is claimed | K121803 – Intravascular Administration Set (ACTA Medical<br>LLC) |
| Reason for 510(k) | New devices to offer an alternative infusion set for use with<br>the Curlin (MOOG) infusion pump. |
| Device Description | DJF Administration Sets are sterile/non-pyrogenic, single use,<br>non-DEHP PVC tubing with a Universal Spike, Slide Clamp, Luer<br>Lock (male) with integral anti-free flow valve (AFF) {optional,<br>separately provided}, 0.2 micron or 1.2 micron filters, Yellow<br>Cassette and Blue key (locator pin) for use with the Curlin IV<br>pump. Three (3) new sets are:<br>1. DJFCUR001: Administration set compatible with Curlin<br>Pump<br>2. DJFCUR002: Administration set with 0.2 micron filter<br>compatible with Curlin Pump<br>3. DJFCUR003: Administration set with 1.2 micron filter<br>compatible with Curlin Pump |
| Technological Characteristics | The subject devices have the same technological material,<br>fit, form and functional characteristics with predicate<br>devices. |
| Indications for Use | DJF Intravascular Administration Sets are intended for use<br>with Curlin (Moog) Infusion Pump, for the delivery of fluids<br>from a container or bag to patient's vascular system |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for "WEI DE LI INTERNATIONAL TRADE". The logo consists of a blue circle with a white stylized "W" inside it. To the right of the circle, the text "WEI DE LI" is written in a bold, sans-serif font, with "INTERNATIONAL TRADE" written in a smaller font below it.
Image /page/4/Picture/3 description: The image shows a gray telephone icon in the center of a white circle. A blue arc covers approximately 60% of the circle's circumference, starting from the top and moving clockwise. The telephone icon is a classic rotary phone design, suggesting a vintage or traditional communication method.
F
Image /page/4/Picture/5 description: The image shows a location pin icon inside of a partial circle. The location pin is dark gray, and the circle is a dark blue color. The circle is not complete, and it is open on the right side. The image is simple and clean, with a white background.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo for "WEI DE LI INTERNATIONAL TRADE". The logo consists of a blue circle with a white stylized "W" inside it. To the right of the circle is the text "WEI DE LI" in a larger font, with the words "INTERNATIONAL TRADE" underneath in a smaller font.
## Substantial Equivalence Summary
| Feature | Proposed<br>Device DJF IV<br>Administration<br>Sets, K230992 | Predicate<br>Device K121803 | Discussion |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Indications for<br>use | DJF Intravascular<br>Administration<br>Sets are<br>intended for use<br>with Curlin<br>(Moog) Infusion<br>Pump, for<br>delivery of fluids<br>from a container<br>or bag to<br>patient's<br>vascular system | Intravascular<br>Administration<br>Sets intended<br>for delivery of<br>fluids from a<br>container or bag<br>to patient's<br>vascular system | Indications for<br>use are<br>fundamentally<br>substantially<br>equivalent |
| ABS Container<br>Spike | ABS Container<br>Spike, Non<br>Vented | ABS Container<br>Spike, Universal | Substantially<br>Equivalent |
| PVC Tubing,<br>TOTM | PVC Tubing,<br>TOTM | PVC Tubing,<br>TOTM | Substantially<br>Equivalent |
| PVC Tubing,<br>TOTM, Pump<br>Segment | PVC Tubing,<br>TOTM, Pump<br>Segment | NA | Different<br>See comment<br>#1 |
| Blue, ABS.<br>Locator Pin | Blue. ABS,<br>Locator Pin | NA | Different<br>See Comment<br>#2 |
| Yellow, Flow<br>Stop, ABS Box<br>with 316L Spring | Yellow, Flow<br>Stop, ABS Box<br>with 316L Spring | NA | Different<br>See Comment<br>#3 |
| ABS Male Luer | ABS Male Luer | ABS Male Luer | Substantially<br>Equivalent |
| Slide Clamp | Slide Clamp | Slide Clamp | Substantially<br>Equivalent |
| Sterilization, SAL<br>10-6 | Gamma<br>Radiation,<br>VDmax25 | Gamma<br>Radiation,<br>VDmax25 | SAL Substantially<br>Equivalent |
Image /page/5/Picture/4 description: The image shows a gray telephone icon inside of a white circle. A blue ring surrounds about half of the white circle. The image is cropped on the right side, but it appears to be next to some text.
Image /page/5/Picture/6 description: The image shows a gray location pin icon inside of a white circle. A blue arc surrounds the top left portion of the circle. The location pin is centered inside of the circle.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "WEI DE LI INTERNATIONAL TRADE". The logo consists of a blue circle with a white stylized "W" inside it. To the right of the circle, the text "WEI DE LI" is written in a larger, bolder font, and below it, "INTERNATIONAL TRADE" is written in a smaller font.
### Comment/ Justification.
- 1. Comment #1 = The PVC tubing, TOTM, PUMP segment is different dimensionally from the predicate. However, the pump segment was tested in the performance testing using ISO 8536-4. A reference device was used to support this test method. Compatibility with the Moog Curlin infusion pump was assessed by the functional tests listed below. A reference device, K981816, was used to support these test methods.
- 2. Comment #2 Blue ABS Locator Pin is not present in the predicate device as it's specific to use with the Moog Curlin infusion pump. It's an externally communicating component required for inserting the infusion set into the pump. Compatibility with the Moog Curlin infusion pump was assessed by the functional tests listed below. A reference device, K981816, was used to support these test methods.
- 3. Comment #3 Yellow Flow stop ABS Box with 316L spring is not present in the predicate device as it's specific to use with the Moog Curlin Infusion Pump. It's an externally communicating component required for inserting the infusion set into the pump. Compatibility with the Moog Curlin infusion pump was assessed by the functional tests listed below. A reference device, K981816, was used to support these test methods.
### 1. Performance Testing
The sterile, single use DJF Intravascular Administration Setsintended for use with Curlin (Moog) Infusion Pump, described in this summary were tested and demonstrated to be in conformance with the following FDA recognized standards.
- ISO 80369-7:2016, Small-bore connectors for liquids and gases in healthcare applications- Part 7: Connectors for intravascular or hypodermic application.
- ISO 8536- 4: 2019, Infusion equipment for medical use-Part 4: Infusion sets for single use, gravity feed.
ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
#### ASTM F1886/F1886M: 2016
Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
Image /page/6/Picture/14 description: The image shows a gray telephone icon inside of a circle. The circle is partially filled with a dark blue color, indicating a completion or progress status. To the right of the icon, there is text that is not fully visible.
Image /page/6/Picture/16 description: The image shows a gray location pin icon inside of a circle. The circle is partially filled with a dark blue color on the left side and a light gray color on the right side. The location pin is a common symbol used to represent a specific place or address on a map or in a digital interface.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for "WEI DE LI INTERNATIONAL TRADE". The logo consists of a blue circle with a white stylized "W" inside it on the left. To the right of the circle is the text "WEI DE LI" in a larger font, with the words "INTERNATIONAL TRADE" underneath in a smaller font.
Additional functional testing was conducted to evaluate the administration set performance with Curlin (Moog) pump. The following functional tests were conducted:
- -Rate Accuracy Test
- -Set Installation Test
- -Upstream Occlusion Test
- -Downstream Occlusion Test
-Testing data demonstrates that no additional safety and effectiveness issues were identified in the use of the DJF Administration Sets with Curlin (Moog) pump. DJF sets performed substantially equivalent to reference predicate, OEM Curlin pump sets/
### 2. Biocompatibility Testing
In accordance with ISO 10993-1, the DJF Intravascular Administration Sets for Curlin (Moog) Infusion Set is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hrs to 30days). The following biocompatibility tests were conducted on DJF Curlin Pump compatible devices with and without filter:
- a. Cytotoxicity
- b. Hemolysis
- c. Irritation
- d. Acute Systemic Toxicity
- e. Sensitization
- f. Pyrogenicity
- g. Sub-Acute/Sub-Chronic Systemic Toxicity
- h. Endotoxin
- i. Particulate Testing
## 3. Sterility, Shipping, Shelf Life
DJF Curlin Pump compatible devices have been validated for sterilization by Gamma Radiation, VDmax25 methodology in compliance with ISO 11137-3: 2017, Sterilization of Healthcare Products Shelf Life of 3 years has been assigned by testing product stored at normal storage environments. Shipping integrity has been tested with ASTM D4169-16 and shows no product damage.
## Conclusion
The differences between the DJF Intravascular Administration Set and the predicate device do not raise any new or different questions of safety or effectiveness. The subject device is substantially equivalent with respect to the indications for use and technological characteristics.
Image /page/7/Picture/23 description: The image shows a gray telephone icon inside of a white circle. The circle is partially surrounded by a blue arc. To the right of the circle are the words 'Telephone' and 'Fax' in gray text.
Image /page/7/Picture/25 description: The image shows a gray location pin icon inside of a circle. The circle is partially filled with a dark blue color. The location pin is a common symbol used to represent a specific place or address on a map or in a digital interface.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.