← Product Code [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA) · K230528

# Extension Set (K230528)

_Medcaptain Life Science Co., Ltd. · FPA · Dec 1, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K230528

## Device Facts

- **Applicant:** Medcaptain Life Science Co., Ltd.
- **Product Code:** [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA.md)
- **Decision Date:** Dec 1, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Extension Sets are intended to be used with a vascular access device for direct injection, intermittent infusion, continuous infusion or aspiration. The Extension Sets may be used with power injector procedures to a maximum pressure of 200 kPa (2 bar).

## Device Story

Extension set is a single-use, sterile, non-pyrogenic add-on device for intravascular catheters. It functions as a conduit for fluid delivery (gravity or pressure infusion) or aspiration. Components include PVC tubing, luer lock connectors, clamps (slide/pinch), flow regulators, filters, check valves, anti-siphon valves, and needle-free injection ports. Used in clinical settings by healthcare providers to extend reach or provide additional access points for infusion therapy. The device connects to a vascular access device hub; it facilitates medication or fluid administration. Benefits include increased flexibility in infusion setup and patient access. It is compatible with power injection up to 200 kPa.

## Clinical Evidence

No clinical data. Performance demonstrated through extensive bench testing, including biocompatibility (ISO 10993), mechanical integrity (tensile strength, leakage), flow performance, and component-specific testing (check valves, filters, needle-free ports) per ISO 8536 and ISO 80369-7 standards.

## Technological Characteristics

Materials: PVC (TOTM), PC, Silicone, PES, PTFE, MABS, PP, POM, ABS, HDPE, SBC. Components: Luer lock, check/anti-siphon valves, filters (0.2/1.2/5.0um), clamps, needle-free ports. Standards: ISO 8536 series, ISO 80369-7, ISO 10993, ASTM F838-20, ANSI/AAMI CN27. Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: None (mechanical).

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Extension Set ([K153293](/device/K153293.md), B.Braun Medical Inc.)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 1, 2023

Medcaptain Life Science Co., Ltd. Alan Tao Official Correspondent 601, Building C, Jinweiyuan Industrial Park, Pingshan District Shenzhen, Guangdong 518118 China

Re: K230528

Trade/Device Name: Extension Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: October 31, 2023 Received: November 1, 2023

Dear Alan Tao:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Porsche Bennett

Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230528

Device Name Extension Set

Indications for Use (Describe)

Extension Sets are intended to be used with a vascular access device for direction, intermittent infusion, continuous infusion or aspiration. The Extension Sets may be used with power injector procedures to a maximum pressure of 200 kPa (2 bar).

| Type of Use (Select one or both, as applicable) |                                                                                                            |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------|
|                                                 | <span> <span style="font-size: 100%;"> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> |
|                                                 | <span> <span style="font-size: 100%;"> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

MEDCAPTAIN LIFE SCIENCE CO., LTD. 601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen, Guangdong, CN 518118. Tel.: +86-755-28380626 Email: disposable(@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

## 510(k) Summary- K230528

### I. Submitter/510(k) Holder

| Submission:     | Traditional 510(k) Premarket Notification                                                          |
|-----------------|----------------------------------------------------------------------------------------------------|
| Submitter:      | MEDCAPTAIN LIFE SCIENCE CO., LTD.                                                                  |
| Address:        | 601, Building C, Jinweiyuan Industrial Park, Pingshan<br>District, Shenzhen, Guangdong, CN 518118. |
| Contact Person: | Alan Tao                                                                                           |
| Telephone:      | +86-755-28380626                                                                                   |
| Telefax:        | +86-755-84517910                                                                                   |
| Email:          | alan.tao@medcaptain.com                                                                            |
| Date prepared:  | December 1, 2023                                                                                   |

### II. Device

| Device Trade Name:  | Extension Set                      |
|---------------------|------------------------------------|
| Device Common Name: | Intravascular Administration Set   |
| Regulatory Name:    | Intravascular Administration Set   |
| Regulation Number:  | 21 CFR 880.5440                    |
| Product Code:       | FPA                                |
| Product Code Name:  | Set, Administration, Intravascular |
| Regulatory Class:   | Class II                           |
| Review Panel:       | General Hospital                   |

### III. Predicate Devices

The Extension Set is substantially equivalent to the following legally marketed predicate device: Extension Set (K153293, B.Braun Medical Inc.) cleared on April 6, 2016.

Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA

## IV. Device Description

The subject device, Extension sets are single use, sterile, non-pyrogenic, add-on devices used for direct injection, intermittent infusion, continuous infusion or aspiration. The device is to connect the infusion device through the luer lock connector to add length and provide clamping capacities, or added to an intravascular catheter

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Medcaptain" in a bold, sans-serif font. The "M" in "Medcaptain" is stylized with two purple triangles below it. The rest of the letters in the word are black. The word appears to be a logo or brand name.

MEDCAPTAIN LIFE SCIENCE CO., LTD. 601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen. Guangdong. CN 518118. Tel.: +86-755-28380626 Email: disposable@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

hub as a conduit for flow to and from the catheter. It is available in various lengths and tube dimensions, and may be comprised of various generic components such as connectors, clamps, flow regulators, filters, check valves and needleless injection ports. It can be used for gravity infusion and pressure infusion with pressure up to a maximum of 200 kPa (2 bar).

#### V. Indications for Use

Extension Sets are intended to be used with a vascular access device for direct injection, intermittent infusion, continuous infusion or aspiration. The Extension Sets may be used with power injector procedures to a maximum pressure of 200 kPa (2 bar).

#### VI. Comparison of Technological Characteristics with the Predicate Device

The subject device, Extension Set, and the predicate device, Extension Sets (K153293), are substantially equivalent in that these devices have the same intended use and patient contact category/duration, similar indications for use, patient population and technological characteristics. They have the same characteristics including technology or design of PVC tube, luer lock connector, slide clamp, check valve, filters, pinch clamp and needle free injection port, etc. The differences between the subject device and the predicate device include the maximum pressure resistance, components of anti-siphon valve and protective cap, and materials. The maximum pressure of the subject device is less than that of the predicate device, which can be acceptable in clinical use to achieve the intended use. Performance testing has been conducted on the anti-siphon valve and protective cap components and have been demonstrated to meet the requirements. The materials of subject device are biocompatible based on biocompatibility tests. The differences between the subject and predicate device do not raise any new or different questions of safety and/or effectiveness.

Similarities and differences in technology characteristics are captured in the substantial equivalence comparison between the subject device and the predicate device, which are provided in Table 1.

| Description                          | Extension Sets<br>(Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Extension Set<br>(Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison to<br>predicate device                                                                                                 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                        | K153293                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | K230528                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | N/A                                                                                                                               |
| Regulation Number                    | 21 CFR 880.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21 CFR 880.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                         |
| Regulatory Name                      | Intravascular Administration Set                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Intravascular Administration Set                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                         |
| Product Code                         | FPA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | FPA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                         |
| Description                          | Extension Sets<br>(Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Extension Set<br>(Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison to<br>predicate device                                                                                                 |
| Regulatory Class                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                                                                                                         |
| Intended Use                         | B.Braun Extension Sets are<br>intended for direct injection,<br>intermittent infusion,<br>continuous infusion or<br>aspiration.                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Extension Sets are intended for<br>direct injection, intermittent<br>infusion, continuous infusion or<br>aspiration.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                         |
| Indications for use                  | B.Braun Extension Sets may<br>be used for direct injection,<br>intermittent infusion,<br>continuous infusion or<br>aspiration of fluids,<br>medications, blood and<br>blood products. Select sets<br>may be used with power<br>injector procedures to a<br>maximum pressure of 400<br>psi at a maximum flow rate<br>of 15mL per second.<br>B.Braun's optional<br>stabilization component on<br>an extension set is intended<br>to provide stability to the<br>patient connector, which is<br>attached to an intravascular<br>catheter. B.Braun Extension<br>Sets may be used for any<br>patient population. | Extension Sets are intended to be<br>used with a vascular access device<br>for direct injection, intermittent<br>infusion, continuous infusion or<br>aspiration. The Extension Sets<br>may be used with power injector<br>procedures to a maximum<br>pressure of 200 kPa (2 bar).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Different<br><br>Difference does not<br>raise new or different<br>questions of safety or<br>effectiveness.<br>See Justification 1 |
| Mode of Fluid<br>Delivery            | Gravity, or power injection<br>to a maximum pressure of<br>400 psi.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Gravity, or pressure injection to a<br>maximum pressure of 200kPa.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Different<br><br>Difference does not<br>raise new or different<br>questions of safety or<br>effectiveness.<br>See Justification 2 |
| Prescription/<br>Over-the-Counter    | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                         |
| Patient Population                   | Any patient population                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Adults and pediatrics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Different                                                                                                                         |
| Description                          | Extension Sets<br>(Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Extension Set<br>(Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison to<br>predicate device                                                                                                 |
|                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Difference does not<br>raise new or different<br>questions of safety or<br>effectiveness.<br>See Justification 3                  |
| Patient Contact<br>category/duration | Externally communicating,<br>Blood path indirect,<br>prolonged exposure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Externally communicating, Blood<br>path indirect, prolonged exposure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                         |
| Device Components                    | Configured: PVC tubing,<br>Luer,<br>Slide clamp, Stabilization<br>component, Spin-lock<br>connector (patient<br>connector/male luer adapter).<br>May be comprised of various<br>generic components:<br>Stopcocks, Clamps (e.g.<br>roller clamp, on-off clamp),<br>Injection sites, Connectors,<br>Manifolds, Filter (0.2um,<br>1.2um, 5.0um), Check valve,<br>Needleless connector.                                                                                                                                                                                                                         | PVC tube, Luer lock connector,<br>Check valve, Anti-siphon valve,<br>Filters (0.2um, 1.2um, 5.0um),<br>Clamp (slide or pinch), Flow<br>regulator, Needle free injection<br>port, Protective cap.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Different<br>Difference does not<br>raise new or different<br>questions of safety or<br>effectiveness.<br>See Justification 4     |
| Summary of Non-<br>clinical Tests    | • Stabilization component<br>performance<br>• Visual<br>• Catheter Angle<br>• Flow Rate – No Catheter<br>• Flow-rate – With Catheter<br>• Tape Removal<br>• Occlusion<br>• Negative Pressure<br>• Positive Pressure<br>• Clamp and Positive<br>Pressure<br>• Tensile Strength<br>• Power Injection<br>• Mechanical Hemolysis –<br>Aspiration and injection                                                                                                                                                                                                                                                  | • Appearance<br>• Dimension<br>• Particulate Contamination<br>• Leakage<br>• Tensile Strength<br>• Flow Rate<br>• Luer Connector (Size, Fluid<br>leakage, Sub-atmospheric<br>pressure air leakage, Stress<br>cracking, Resistance to<br>separation from axial load,<br>Resistance to separation from<br>unscrewing, Resistance to<br>overriding)<br>• Check Valve (Counter flow<br>pressure resistance, Flow rate,                                                                                                                                                                                                                                                                                                                                                                                                                 | Different<br>Difference does not<br>raise new or different<br>questions of safety or<br>effectiveness.<br>See Justification 5     |
| Description                          | Extension Sets<br>(Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Extension Set<br>(Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison to<br>predicate device                                                                                                 |
|                                      | • Luer Connection<br>• Gauging<br>• Liquid and Air leakage<br>• Separation Force<br>• Stress Cracking<br>• Collar Retention<br>• Joint Qualification<br>• Particulate Contamination                                                                                                                                                                                                                                                                                                                                                                                                                         | Blocking performance, Opening<br>pressure)<br>• Anti-siphon Valve (Counter<br>flow pressure resistance, Flow<br>rate, Blocking performance,<br>Opening pressure)<br>• Liquid Medicine Filter<br>• Needle Free Injection Port<br>(Flow rate, Test for exposure to<br>IPA, Resistance to separation<br>from axial load, Resistance to<br>separation from unscrewing,<br>Resistance to overriding,<br>Backpressure (un-activated),<br>Positive pressure fluid leakage<br>(activated), Sub-atmospheric<br>pressure air leakage (un-<br>activated), Sub-atmospheric<br>pressure air leakage (activated),<br>Duration of activation, Number<br>of activation)<br>• Storage Volume<br>• Clamp (Construction) and Flow<br>Regulator (Construction, Flow<br>rate)<br>• Protective Cap<br>• Chemical performance<br>• Biological performance |                                                                                                                                   |
| Materials                            | PVC, LDPE, HDPE, PC, PP,<br>ABS, Methylene Chloride<br>(MC), Tetrahydrofuran<br>(THF)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | PVC (TOTM), PC, Silicone, PES,<br>PTFE, MABS, PP, POM, ABS,<br>HDPE, Styrene-Butadiene<br>Copolymer (SBC).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different<br><br>Difference does not<br>raise new or different<br>questions of safety or<br>effectiveness.<br>See Justification 6 |
| Biocompatibility                     | ISO 10993                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ISO 10993                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                         |
| Sterilization                        | Ethylene Oxide, SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Ethylene Oxide, SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                                                                                                         |
| Sterile                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                         |
| Description                          | Extension Sets<br>(Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Extension Set<br>(Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison to<br>predicate device                                                                                                 |
| Single Use                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                         |

Table 1 Substantial Equivalence Comparison

{6}------------------------------------------------

MEDCAPTAIN LIFE SCIENCE CO., LTD.

601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen, Guangdong, CN 518118. Tel.: +86-755-28380626 Email: disposable@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

{7}------------------------------------------------

MEDCAPTAIN LIFE SCIENCE CO., LTD. 601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen, Guangdong, CN 518118. Tel.: +86-755-28380626 Email: disposable@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

{8}------------------------------------------------

MEDCAPTAIN LIFE SCIENCE CO., LTD.

601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen, Guangdong, CN 518118. Tel.: +86-755-28380626 Email: disposable@medcaptain.com

Fax.: +86-755-84517910

Website: www.medcaptain.com

{9}------------------------------------------------

MEDCAPTAIN LIFE SCIENCE CO., LTD. 601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen. Guangdong. CN 518118. Tel.: +86-755-28380626 Email: disposable@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

### Justification 1: Indications for Use

The subject device, Extension Set, is intended to be used with a vascular access device for direct injection, intermittent infusion, continuous infusion or aspiration. The predicate device was described to be added to an intravascular catheter hub as a conduit for flow to and from the catheter as per K153293, and the subject device is same to it. Extension Sets may be used with pressure injection to a maximum pressure of 200kPa, and the differences do not raise new or different questions of safety or effectiveness according to Justification 2. The subject device does not contain a stabilization component which aims to fix the set, and the component has no influence to achieve the intended use. The patient population of the subject device is adults and pediatrics, the difference does not raise new or different questions of safety or effectiveness according to Justification 3. Based on the Justification 2 and 3, the differences in the "Indications for Use" of the subject device do not result in a new intended use. So the differences do not raise new or different questions of safety or effectiveness.

### Justification 2: Mode of Fluid Delivery

The subject device is intended for gravity use or pressure injection. The gravity use is same to the predicate device. For pressure injection, the subject device is intended for pressure injection to a maximum pressure of 200kPa, while the predicate device for power injection to a maximum pressure of 400 psi. To achieve the maximum pressure of 200kPa, the subject device needs to connect with pressure infusion equipment. The pressure resistance testing of leakage is verified according to ISO 8536-9: 2015 Infusion equipment for medical use - Part 9: Fluid lines for single use with pressure infusion equipment and the connection part of the subject device is standard components meeting the requirements of ISO 80369-7: 2021 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications. So the pressure injection of 200kPa is acceptable to achieve the intended use. The difference does not raise new or different questions of safety or effectiveness.

### Justification 3: Patient Population

The patient population of the subject device is "adults and pediatrics", while the patient population of the predicate device is "any patients". The subject device has demonstrated to perform to specification and is biocompatible, independent of the patient population. The difference does not raise new or different questions of safety or effectiveness.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "Medcaptain" in a bold, sans-serif font. The word is black, and there are two purple triangles below the "M" in "Medcaptain". The triangles are pointing to the right and are slightly offset from each other.

MEDCAPTAIN LIFE SCIENCE CO., LTD. 601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen. Guangdong. CN 518118. Tel.: +86-755-28380626 Email: disposable@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

#### Justification 4: Device Components

The subject device and predicate device have the similar components, such as PVC tube, luer lock connector, slide clamp, check valve, filters, pinch clamp and needle free injection port. The pinch clamp of the subject device is similar to the on-off clamp of the predicate device. The subject device has the components of anti-siphon valve and protective cap, which are different from the predicate device. The anti-siphon valve is the same as the check-valve structure except the difference of opening pressure, and meets the performance requirements after testing according to ISO 8536-12:2021 Infusion equipment for medical use - Part 12: Check valves for single use. The protective cap is designed to protect the connectors and is secure and easily removable after testing. The differences of configurations between the subject device and predicate device do not raise different questions of safety and effectiveness. Both achieve the same intended use.

#### Justification 5: Summary of Non-Clinical Tests

The subject device has the components of PVC tube, luer lock connector, check valve, anti-siphon valve, filters (0.2um, 1.2um, 5.0um), clamp (slide or roller), flow regulator, needle free injection port, protective cap, no stabilization component. The different components compared with the predicate device include anti-siphon valve and protective cap. For all components, the verification testing was performed on the subject device and met the requirements of applicable parts of ISO 8536 series, ISO 80369-7: 2021 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications and FDA guidance. For example, the luer lock connector met the requirements of ISO 80369-7, check valve and anti-siphon valve met the requirements of ISO 8536-12: 2021 Infusion equipment for medical use - Part 12: Check valves for single use, filters met the requirements of ASTM F838-20 Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration or ISO 8536-4: 2019 Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed, clamp and flow regulator met the requirements of ISO 8536-14: 2016 Infusion equipment for medical use - Part 14: Clamps and flow regulators for transfusion and infusion equipment without fluid contact, and needle free injection met the requirements of ISO 80369-7: 2021 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications, ANSI/AAMI CN27:2021 General requirements for Luer activated valves (LAVs) incorporated into medical devices for intravascular applications, and FDA guidance. In addition, chemical and biological performances, including Reducing matter, metal ions, titration acidity or alkalinity, residue on evaporation, UV absorption, EO residual, ECH residual, sterility and endotoxin were tested as well. Verification results confirm that the differences in the components do not raise new questions of safety and effectiveness.

{11}------------------------------------------------

MEDCAPTAIN LIFE SCIENCE CO., LTD. 601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen. Guangdong. CN 518118. Tel.: +86-755-28380626 Email: disposable@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

### Justification 6: Materials

The liquid pathway of the subject device is PVC (TOTM) which is a DEHP-free PVC and similar to the material of the predicated device. And they have same PC, ABS, HDPE, PP materials. The materials used by the subject device are common medical materials with no significant clinical safety problems reported. In addition, a biological evaluation was conducted on the subject device according to ISO 10993-1: 2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, including cytotoxicity, skin sensitization, intracutaneous reactivity, pyrogenicity, acute systemic toxicity, subacute toxicity and hemolysis. It has been demonstrated that the materials used for the device is biocompatible. Based on bench testing of Extension Set, the performance of the subject device meets the requirement of ISO 8536 series, ISO 80369-7, etc. The materials differences have no influence to the product performance. So the differences of materials do not raise new or different questions of safety or effectiveness.

### VII. Performance Data

The subject device, Extension Set, was subjected to the following applicable testing to ensure reliable design and performance under the specified testing parameters:

### Biocompatibility Testing:

Per ISO 10993-1: 2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA guidance, the following tests were performed to ensure the biocompatibility of the subject device.

- In vitro cytotoxicity, per ISO 10993-5: 2009 Biological evaluation of ● medical devices - Part 5: Tests for in vitro cytotoxicity
- Skin sensitization, per ISO 10993-10: 2021 Biological evaluation of . medical devices - Part 10: Tests for skin sensitization
- Intracutaneous reactivity, per ISO 10993-23: 2021 Biological evaluation of . medical devices - Part 23: Tests for irritation
- Pyrogenicity, per ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity and USP <151> Pyrogenicity Test
- Acute systemic toxicity, per ISO 10993-11: 2017 Biological evaluation of ● medical devices - Part 11: Tests for systemic toxicity
- Subacute toxicity, per ISO 10993-11: 2017 Biological evaluation of medical ● devices - Part 11: Tests for systemic toxicity
- . Hemolysis, per ISO 10993-4: 2017 Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood and ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials

{12}------------------------------------------------

MEDCAPTAIN LIFE SCIENCE CO., LTD. 601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen. Guangdong. CN 518118. Tel.: +86-755-28380626 Email: disposable@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

### Bench Testing:

Per ISO 8536-4: 2019 Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed, ISO 80369-7: 2021 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications, FDA guidance "Intravascular Administration Sets Premarket Submission Notifications (FDA; July 11, 2008)", etc., the following tests were performed to ensure performance/functionality of the subject device.

- Appearance, per ISO 8536 series. .
- Dimensions, per FDA guidance, similar devices and product characteristics. ●
- Particulate contamination, per ISO 8536 series and USP <788> Particulate ● Matter in Injections
- Leakage, per ISO 8536 series and FDA guidance. ●
- . Tensile strength, per ISO 8536 series and FDA guidance.
- Flow rate, per FDA guidance. ●
- Luer connector: Size, Fluid leakage, Sub-atmospheric pressure air leakage, ● Stress cracking, Resistance to separation from axial load, Resistance to separation from unscrewing, Resistance to overriding, per ISO 80369-7: 2021 and FDA guidance.
- . Check valve: Counter flow pressure resistance, Flow rate, Blocking performance, Opening pressure, per ISO 8536-12: 2021 Infusion equipment for medical use - Part 12: Check valves for single use
- Anti-siphon valve: Counter flow pressure resistance, Flow rate, Blocking ● performance, Opening pressure, per ISO 8536-12: 2021.
- Liquid medicine filter 0.2um: Bacterial interception test, per ASTM F838-20 Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration
- Liquid medicine filter 1.2um: Retention of latex particles, per ISO 8536-4: . 2019 and Candida albicans interception.
- Liquid medicine filter 5.0um: Retention of latex particles, per ISO 8536-4: ● 2019.
- Needle free injection port: Flow rate, Test for expose to IPA, Resistance to ● separation from axial load, Resistance to separation from unscrewing, Resistance to overriding, Backpressure (unactivated), Positive pressure fluid leakage (activated), Sub-atmospheric pressure air leakage (unactivated), Subatmospheric pressure air leakage (activated), Duration of activation, Number of activation, per ANSI/AAMI CN27:2021 and ISO 80369-7: 2021.
- Storage volume, per ISO 8536-8: 2015 and FDA guidance. ●
- Clamp, per ISO 8536-14: 2016 Infusion equipment for medical use Part 14: ● Clamps and flow regulators for transfusion and infusion equipment without fluid contact
- Flow regulator, per ISO 8536-14: 2016. ●

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "Medcaptain" in a bold, sans-serif font. The word is black, and there are two purple triangles below the "M" in "Medcaptain". The triangles are oriented so that their points are facing to the left.

MEDCAPTAIN LIFE SCIENCE CO., LTD. 601Building C, Jinweiyuan Industrial Park, Pingshan District, Shenzhen. Guangdong. CN 518118. Tel.: +86-755-28380626 Email: disposable(@medcaptain.com Fax.: +86-755-84517910 Website: www.medcaptain.com

- Protective cap, per ISO 8536-4: 2019. ●
- Chemical performance, per ISO 8536-4: 2019. ●
- Sterility, per ISO 11135: 2014 Sterilization of health-care products Ethylene . oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
- Bacterial endotoxin, per ANSI/AAMI ST72: 2019 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
- ISTA 3A: 2018, Packaged-Products for Parcel Delivery System Shipment ●
- Package performance, per ISO 11607-1: 2019 Packaging for terminally . sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems and ISO 11607-2: 2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- Shelf Life Validation, per ASTM F1980-21 Standard Guide for Accelerated . Aging of Sterile Barrier Systems and Medical Devices

#### Clinical Tests:

Clinical tests were not required to demonstrate performance of Extension Set. Product functionality has been adequately assessed by non-clinical tests.

#### Animal Tests:

Animal tests were not required to demonstrate performance of Extension Set. Product functionality has been adequately assessed by non-animal tests.

#### VIII. Conclusions

The results of these tests confirm that the Extension Set meets the design input requirements based on the intended use and support the conclusion that this device is as safe and effective as the legally marketed predicate device Extension Sets (K 153293, B.Braun Medical Inc.) and therefore, substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K230528](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K230528)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
