← Product Code [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA) · K161323

# Solution Set for Epidural Use (K161323)

_Baxter Healthcare Corporation · FPA · Nov 30, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K161323

## Device Facts

- **Applicant:** Baxter Healthcare Corporation
- **Product Code:** [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA.md)
- **Decision Date:** Nov 30, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

For the administration of fluids from a container into the patient's epidural space with Baxter infusion pumps.

## Device Story

Single-use disposable administration set for delivering fluids into epidural space. Components include nonvented spike, 60 drops/mL drip chamber, minidrip adapter, PVC tubing, slide clamp, roller clamp, and male epidural lock connector. Device connects fluid container to Baxter infusion pump; pump controls fluid delivery rate for intermittent or continuous infusion. Healthcare providers use device to manage chronic or post-operative pain. Design incorporates AAMI/CN6:2015 compliant neuraxial connectors to prevent misconnection. Bench testing confirms flow rate accuracy, occlusion performance, and mechanical integrity of bonds and connectors.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including drop volume, cannula pull-out, spike insertion/removal, clamp force, shut-off/leak tests, bond tensile strength, AAMI/CN6:2015 connector testing, and pump-specific performance (flow rate accuracy, occlusion, tubing flexion). Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, systemic toxicity, irritation, sensitization, hemocompatibility, USP physiochemical, and material-mediated pyrogen).

## Technological Characteristics

Materials: ABS (spike, connector), PVC (drip chamber, tubing), synthetic polyisoprene/stainless steel (minidrip adapter). Connectivity: Standalone fluid path for use with Baxter infusion pumps. Sterilization: Gamma radiation (ANSI/AAMI/ISO 11137-2). Connector: AAMI/CN6:2015 compliant neuraxial lock. Form factor: Disposable tubing set.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Bard Epidural Spike Tubing Set ([K925058](/device/K925058.md))

## Reference Devices

- IV sets ([K142011](/device/K142011.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 30, 2016

Baxter Healthcare Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313

Re: K161323

Trade/Device Name: Solution Set for Epidural Use Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: FPA Dated: November 15, 2016 Received: November 16, 2016

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv.

Tina
Kiang-S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Device Name Solution Set for Epidural Use

Indications for Use (Describe)

For the administration of fluids from a container into the patient's epidural space with Baxter infusion pumps.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement. 

# Section 5. 510(k) Summary

November 30, 2016

#### OWNER:

Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015

# CONTACT PERSON:

Tiffany Lin Manager, Regulatory Affairs 32650 N Wilson Road Round Lake, IL 60073 Telephone: (224) 270-4343 Fax: (224) 270-4119

### IDENTIFICATION OF THE DEVICE:

Common Name: IV Administration Set Trade Name or Proprietary Name: Solution Set for Epidural Use Classification Panel: 80 General Hospital Classification: Set, Administration, Intravascular (21 CFR 880.5440) Class: Class II Product Code: FPA

#### Table 1. Product Code for Solution Set for Epidural Use

| Code Number | Name                          |
|-------------|-------------------------------|
| 2E7554      | Solution Set for Epidural Use |

#### PREDICATE DEVICE:

#### Table 2. Predicate Device

| Device                            | Company                               | Predicate 510(k) | Clearance Date |
|-----------------------------------|---------------------------------------|------------------|----------------|
| Bard Epidural Spike<br>Tubing Set | C.R. Bard, Inc.<br>(Bard MedSystems)a | K925058          | August 3, 1993 |

a Bard MedSystems, a division of C.R. Bard, Inc., was acquired by Baxter in 1993.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly italicized. The word is likely a logo or brand name.

### DESCRIPTION OF THE DEVICE:

The Solution Set for Epidural Use is a single use disposable device intended for the administration of fluids from a container into the epidural space. The set is used for epidural infusions of fluids (typically analgesic or anesthetic) as an intermittent or continuous infusion to reduce chronic or post-operative pain. The set consists of a nonvented spike, 60 drops per mL drip chamber, minidrip adapter, tubing, fixtured slide clamp, on-off roller clamp, and a male epidural lock connector. The connector is compliant to the provisional AAMI/CN6:2015 Small-bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications.

### INDICATIONS FOR USE:

For the administration of fluids from a container into the patient's epidural space with Baxter infusion pumps.

# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The proposed device has equivalent technological characteristics as Baxter's legally marketed epidural set cleared under 510(k) premarket notification K925058 (cleared August 3, 1993). The intended use and function of the proposed device are equivalent to the predicate device. Table 3 is a device comparison table outlining the differences between the predicate and proposed devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a dynamic appearance. 

|                                           | Predicate Device<br>(K925058):                                                                                                                                                                      | Proposed Device:<br>(K161323)                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Features                                  | Bard Epidural Spike Tubing Seta                                                                                                                                                                     | Solution Set for Epidural Useb                                                                                           | Assessment of Difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Intended Use                              | For the administration of fluids from a<br>container into the patient's epidural<br>space.                                                                                                          | Same.                                                                                                                    | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Indications for Use                       | The Bard Epidural Spike Tubing Set is<br>intended for use with Bard ambulatory<br>infusion pumps to provide epidural<br>delivery of anesthetic or analgesic<br>drugs for periods of up to 96 hours. | For the administration of fluids<br>from a container into the patient's<br>epidural space with Baxter infusion<br>pumps. | The indications for use are being modified<br>from the predicate device to clarify its use and<br>remove reference to a pump that is no longer<br>being marketed. The modification to the<br>indications statement does not impact the<br>safety or efficacy of the proposed device. It<br>represents a clarification only as both devices<br>are intended for the administration of epidural<br>therapy through a disposable solution set, both<br>are intended to administer fluids from a source<br>container to the patient's epidural space, and<br>both are intended for pump administration.<br>The specific pump compatibility information<br>and duration of use is referenced in the<br>product labeling for the proposed device. |
| Sterile                                   | Yes                                                                                                                                                                                                 | Same                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Non-Pyrogenic                             | Yes                                                                                                                                                                                                 | Same                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Single Use                                | Yes                                                                                                                                                                                                 | Same                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| FLUID PATH MATERIALS:                     |                                                                                                                                                                                                     |                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Non-Vented Spike                          | Acrylonitrile Butadiene Styrene                                                                                                                                                                     | Same                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Features                                  | Predicate Device<br>(K925058):<br>Bard Epidural Spike Tubing Seta                                                                                                                                   | Proposed Device:<br>(K161323)<br>Solution Set for Epidural Useb                                                          | Assessment of Difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Drip Chamber                              | N/A                                                                                                                                                                                                 | Polyvinyl Chloride                                                                                                       | The predicate device does not have a drip chamber. A drip chamber is included on the proposed device for the end user to verify drops are falling within the drip chamber. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change.                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Minidrip Adapter                          | N/A                                                                                                                                                                                                 | Synthetic Polyisoprene and<br>Stainless Steel                                                                            | The predicate device does not have minidrip adapter. A minidrip adapter is included on the proposed device to allow more accurate visibility of drops falling, since epidural infusion tends to be towards the lower side of infusion rates. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change.                                                                                                                                                                                                                                                                                                                                                                       |
| Tubing                                    | Silicone and Polyvinyl Chloride                                                                                                                                                                     | Polyvinyl Chloride                                                                                                       | The predicate device is used with a Bard pump and has a separate tubing segment made of silicone that interfaces with that pump. The proposed device utilizes PVC tubing throughout the device. A section of this PVC tubing interfaces with Baxter pumps. This same PVC tubing is used in other Baxter IV sets cleared for use with the same Baxter pumps with which the proposed device will be used. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change.                                                                                                                                                                                                            |
|                                           | Predicate Device<br>(K925058):                                                                                                                                                                      | Proposed Device:<br>(K161323)                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Features                                  | Bard Epidural Spike Tubing Seta                                                                                                                                                                     | Solution Set for Epidural Useb                                                                                           | Assessment of Difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Anti-Siphon Valve                         | Polycarbonate and Silicone                                                                                                                                                                          | N/A                                                                                                                      | Not present. The proposed device does not have an anti-siphon valve. This valve has been removed as it was a design specifically for the Bard pump, and to align with the design of other currently marketed Baxter administration sets. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change.                                                                                                                                                                                                                                                                                                                                                                           |
| Two-Piece Male Luer<br>Lock Connector     | Polymethyl Methacrylate                                                                                                                                                                             | N/A                                                                                                                      | Not present. The proposed device has an epidural lock connector instead.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| One-Piece Male Epidural<br>Lock Connector | N/A                                                                                                                                                                                                 | Acrylonitrile Butadiene Styrene                                                                                          | The epidural lock is made of the same material type as the Luer lock's body in K142011, but differs in dimensions due to the dimensional requirements of AAMI/CN6:2015. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change.                                                                                                                                                                                                                                                                                                                                                                                                                                            |

# Table 3. Device Comparison

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right. 

# Table 3. Device Comparison

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif and the letters are closely spaced together.

#### Table 3. Device Comparison

4 Bard MedSystems, a division of C.R. Bard, Inc., was acquired by Baxter in 1993.

b Similar IV sets have been previously cleared in 510(k) premarket notification K142011. These sets have and one of the proposed device, the only difference being the epidural lock connector, which is the same material type as the K142011 Luer lock connector, but just differs in dimension due to the dimensional requirements of AAMI/CN6:2015.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly blurred, suggesting that the image may have been taken out of focus or that the image quality is low. The background is white.

### DISCUSSION OF NONCLINICAL TESTS:

Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the results of these analyses. All test results meet their acceptance criteria and support that the proposed device is appropriately designed for its intended use.

### Performance Data:

The following bench tests were conducted to evaluate the functional performance of the Solution Set for Epidural Use:

- Drop Volume Test
- . Cannula Pull Out Test
- Spike Insertion Test
- Spike Removal Test
- Roller Clamp Force Test
- Roller Clamp Shut-Off and Tubing Leak Test
- Slide Clamp Shut Off Test
- Solvent Bond Tensile Strength Test (for all bonds) ●
- Solvent Bond Air/Water Pressure Test (for all bonds) ●
- AAMI/CN6:2015 Connector Tests on Male Epidural Lock Connectors
- Pump Tests
	- -Flow Rate Accuracy
	- Upstream/Downstream Occlusion -
	- Tubing Crimping/Flexion When Used with Infusion Pumps -

All tests met the acceptance criteria.

# Biocompatibility:

All patient-contacting components meet the category of External Communicating Device, Indirect Blood Path, Prolonged Duration in accordance with ISO 10993-1 and FDA Blue Book Memorandum #G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," as recommended in the IV administration sets guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)].

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly slanted to the right. The word is likely a logo or brand name.

Biocompatibility assessment has been conducted on a representative final, finished device for all materials of the Solution Set for Epidural Use. The following tests were conducted as part of the biocompatibility testing:

- Cytotoxicity
- Systemic Toxicity
- Irritation/Intracutaenous Reactivity
- Sensitization
- Hemocompatibility
- USP Physiochemical ●
- USP Material Mediated Pyrogen

# Sterility:

The Solution Set for Epidural Use is sterilized with gamma radiation. The Minimum Sterilizing Dose (MSD) required to provide a 10th Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility using Method 1 as described in ANSI/AAMI/ISO 11137-2 Sterilization of Health Care Products - Radiation - Part 2: Establishing the Sterilization Dose.

This product is labeled "Sterile, nonpyrogenic." Package verification testing is based on Visual Inspection, ASTM F88 Seal Strength, and ASTM F2096 Bubble Leak.

# Shelf-Life:

Baxter has provided aging tests to support a shelf-life claim of one (1) year. Baxter intends to perform testing to expand the shelf-life up through a period of five (5) years.

# CONCLUSION:

The nonclinical data support the substantial equivalence of the proposed device and demonstrate that the proposed device performs comparably to the predicate device that is legally marketed for the same intended use.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K161323](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K161323)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
