K161323 · Baxter Healthcare Corporation · FPA · Nov 30, 2016 · General Hospital
Device Facts
Record ID
K161323
Device Name
Solution Set for Epidural Use
Applicant
Baxter Healthcare Corporation
Product Code
FPA · General Hospital
Decision Date
Nov 30, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
For the administration of fluids from a container into the patient's epidural space with Baxter infusion pumps.
Device Story
Single-use disposable administration set for delivering fluids into epidural space. Components include nonvented spike, 60 drops/mL drip chamber, minidrip adapter, PVC tubing, slide clamp, roller clamp, and male epidural lock connector. Device connects fluid container to Baxter infusion pump; pump controls fluid delivery rate for intermittent or continuous infusion. Healthcare providers use device to manage chronic or post-operative pain. Design incorporates AAMI/CN6:2015 compliant neuraxial connectors to prevent misconnection. Bench testing confirms flow rate accuracy, occlusion performance, and mechanical integrity of bonds and connectors.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including drop volume, cannula pull-out, spike insertion/removal, clamp force, shut-off/leak tests, bond tensile strength, AAMI/CN6:2015 connector testing, and pump-specific performance (flow rate accuracy, occlusion, tubing flexion). Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, systemic toxicity, irritation, sensitization, hemocompatibility, USP physiochemical, and material-mediated pyrogen).
Technological Characteristics
Materials: ABS (spike, connector), PVC (drip chamber, tubing), synthetic polyisoprene/stainless steel (minidrip adapter). Connectivity: Standalone fluid path for use with Baxter infusion pumps. Sterilization: Gamma radiation (ANSI/AAMI/ISO 11137-2). Connector: AAMI/CN6:2015 compliant neuraxial lock. Form factor: Disposable tubing set.
Indications for Use
Indicated for the administration of fluids (analgesic or anesthetic) from a container into the patient's epidural space using Baxter infusion pumps for the reduction of chronic or post-operative pain.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 30, 2016
Baxter Healthcare Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313
Re: K161323
Trade/Device Name: Solution Set for Epidural Use Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: FPA Dated: November 15, 2016 Received: November 16, 2016
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
Tina
Kiang-S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Device Name Solution Set for Epidural Use
Indications for Use (Describe)
For the administration of fluids from a container into the patient's epidural space with Baxter infusion pumps.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement.
# Section 5. 510(k) Summary
November 30, 2016
#### OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
# CONTACT PERSON:
Tiffany Lin Manager, Regulatory Affairs 32650 N Wilson Road Round Lake, IL 60073 Telephone: (224) 270-4343 Fax: (224) 270-4119
### IDENTIFICATION OF THE DEVICE:
Common Name: IV Administration Set Trade Name or Proprietary Name: Solution Set for Epidural Use Classification Panel: 80 General Hospital Classification: Set, Administration, Intravascular (21 CFR 880.5440) Class: Class II Product Code: FPA
#### Table 1. Product Code for Solution Set for Epidural Use
| Code Number | Name |
|-------------|-------------------------------|
| 2E7554 | Solution Set for Epidural Use |
#### PREDICATE DEVICE:
#### Table 2. Predicate Device
| Device | Company | Predicate 510(k) | Clearance Date |
|-----------------------------------|---------------------------------------|------------------|----------------|
| Bard Epidural Spike<br>Tubing Set | C.R. Bard, Inc.<br>(Bard MedSystems)a | K925058 | August 3, 1993 |
a Bard MedSystems, a division of C.R. Bard, Inc., was acquired by Baxter in 1993.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly italicized. The word is likely a logo or brand name.
### DESCRIPTION OF THE DEVICE:
The Solution Set for Epidural Use is a single use disposable device intended for the administration of fluids from a container into the epidural space. The set is used for epidural infusions of fluids (typically analgesic or anesthetic) as an intermittent or continuous infusion to reduce chronic or post-operative pain. The set consists of a nonvented spike, 60 drops per mL drip chamber, minidrip adapter, tubing, fixtured slide clamp, on-off roller clamp, and a male epidural lock connector. The connector is compliant to the provisional AAMI/CN6:2015 Small-bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications.
### INDICATIONS FOR USE:
For the administration of fluids from a container into the patient's epidural space with Baxter infusion pumps.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The proposed device has equivalent technological characteristics as Baxter's legally marketed epidural set cleared under 510(k) premarket notification K925058 (cleared August 3, 1993). The intended use and function of the proposed device are equivalent to the predicate device. Table 3 is a device comparison table outlining the differences between the predicate and proposed devices.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a dynamic appearance.
| | Predicate Device<br>(K925058): | Proposed Device:<br>(K161323) | |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Features | Bard Epidural Spike Tubing Seta | Solution Set for Epidural Useb | Assessment of Difference |
| Intended Use | For the administration of fluids from a<br>container into the patient's epidural<br>space. | Same. | N/A |
| Indications for Use | The Bard Epidural Spike Tubing Set is<br>intended for use with Bard ambulatory<br>infusion pumps to provide epidural<br>delivery of anesthetic or analgesic<br>drugs for periods of up to 96 hours. | For the administration of fluids<br>from a container into the patient's<br>epidural space with Baxter infusion<br>pumps. | The indications for use are being modified<br>from the predicate device to clarify its use and<br>remove reference to a pump that is no longer<br>being marketed. The modification to the<br>indications statement does not impact the<br>safety or efficacy of the proposed device. It<br>represents a clarification only as both devices<br>are intended for the administration of epidural<br>therapy through a disposable solution set, both<br>are intended to administer fluids from a source<br>container to the patient's epidural space, and<br>both are intended for pump administration.<br>The specific pump compatibility information<br>and duration of use is referenced in the<br>product labeling for the proposed device. |
| Sterile | Yes | Same | N/A |
| Non-Pyrogenic | Yes | Same | N/A |
| Single Use | Yes | Same | N/A |
| FLUID PATH MATERIALS: | | | |
| Non-Vented Spike | Acrylonitrile Butadiene Styrene | Same | N/A |
| Features | Predicate Device<br>(K925058):<br>Bard Epidural Spike Tubing Seta | Proposed Device:<br>(K161323)<br>Solution Set for Epidural Useb | Assessment of Difference |
| Drip Chamber | N/A | Polyvinyl Chloride | The predicate device does not have a drip chamber. A drip chamber is included on the proposed device for the end user to verify drops are falling within the drip chamber. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change. |
| Minidrip Adapter | N/A | Synthetic Polyisoprene and<br>Stainless Steel | The predicate device does not have minidrip adapter. A minidrip adapter is included on the proposed device to allow more accurate visibility of drops falling, since epidural infusion tends to be towards the lower side of infusion rates. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change. |
| Tubing | Silicone and Polyvinyl Chloride | Polyvinyl Chloride | The predicate device is used with a Bard pump and has a separate tubing segment made of silicone that interfaces with that pump. The proposed device utilizes PVC tubing throughout the device. A section of this PVC tubing interfaces with Baxter pumps. This same PVC tubing is used in other Baxter IV sets cleared for use with the same Baxter pumps with which the proposed device will be used. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change. |
| | Predicate Device<br>(K925058): | Proposed Device:<br>(K161323) | |
| Features | Bard Epidural Spike Tubing Seta | Solution Set for Epidural Useb | Assessment of Difference |
| Anti-Siphon Valve | Polycarbonate and Silicone | N/A | Not present. The proposed device does not have an anti-siphon valve. This valve has been removed as it was a design specifically for the Bard pump, and to align with the design of other currently marketed Baxter administration sets. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change. |
| Two-Piece Male Luer<br>Lock Connector | Polymethyl Methacrylate | N/A | Not present. The proposed device has an epidural lock connector instead. |
| One-Piece Male Epidural<br>Lock Connector | N/A | Acrylonitrile Butadiene Styrene | The epidural lock is made of the same material type as the Luer lock's body in K142011, but differs in dimensions due to the dimensional requirements of AAMI/CN6:2015. Design control activities confirmed there was no impact to the safety or effectiveness of the device due to this change. |
# Table 3. Device Comparison
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right.
# Table 3. Device Comparison
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif and the letters are closely spaced together.
#### Table 3. Device Comparison
4 Bard MedSystems, a division of C.R. Bard, Inc., was acquired by Baxter in 1993.
b Similar IV sets have been previously cleared in 510(k) premarket notification K142011. These sets have and one of the proposed device, the only difference being the epidural lock connector, which is the same material type as the K142011 Luer lock connector, but just differs in dimension due to the dimensional requirements of AAMI/CN6:2015.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly blurred, suggesting that the image may have been taken out of focus or that the image quality is low. The background is white.
### DISCUSSION OF NONCLINICAL TESTS:
Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the results of these analyses. All test results meet their acceptance criteria and support that the proposed device is appropriately designed for its intended use.
### Performance Data:
The following bench tests were conducted to evaluate the functional performance of the Solution Set for Epidural Use:
- Drop Volume Test
- . Cannula Pull Out Test
- Spike Insertion Test
- Spike Removal Test
- Roller Clamp Force Test
- Roller Clamp Shut-Off and Tubing Leak Test
- Slide Clamp Shut Off Test
- Solvent Bond Tensile Strength Test (for all bonds) ●
- Solvent Bond Air/Water Pressure Test (for all bonds) ●
- AAMI/CN6:2015 Connector Tests on Male Epidural Lock Connectors
- Pump Tests
- -Flow Rate Accuracy
- Upstream/Downstream Occlusion -
- Tubing Crimping/Flexion When Used with Infusion Pumps -
All tests met the acceptance criteria.
# Biocompatibility:
All patient-contacting components meet the category of External Communicating Device, Indirect Blood Path, Prolonged Duration in accordance with ISO 10993-1 and FDA Blue Book Memorandum #G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," as recommended in the IV administration sets guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)].
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly slanted to the right. The word is likely a logo or brand name.
Biocompatibility assessment has been conducted on a representative final, finished device for all materials of the Solution Set for Epidural Use. The following tests were conducted as part of the biocompatibility testing:
- Cytotoxicity
- Systemic Toxicity
- Irritation/Intracutaenous Reactivity
- Sensitization
- Hemocompatibility
- USP Physiochemical ●
- USP Material Mediated Pyrogen
# Sterility:
The Solution Set for Epidural Use is sterilized with gamma radiation. The Minimum Sterilizing Dose (MSD) required to provide a 10th Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility using Method 1 as described in ANSI/AAMI/ISO 11137-2 Sterilization of Health Care Products - Radiation - Part 2: Establishing the Sterilization Dose.
This product is labeled "Sterile, nonpyrogenic." Package verification testing is based on Visual Inspection, ASTM F88 Seal Strength, and ASTM F2096 Bubble Leak.
# Shelf-Life:
Baxter has provided aging tests to support a shelf-life claim of one (1) year. Baxter intends to perform testing to expand the shelf-life up through a period of five (5) years.
# CONCLUSION:
The nonclinical data support the substantial equivalence of the proposed device and demonstrate that the proposed device performs comparably to the predicate device that is legally marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.