Solution Administration Sets with 0.2 Micron Filter
K153158 · Baxter Healthcare Corporation · FPA · Dec 28, 2015 · General Hospital
Device Facts
Record ID
K153158
Device Name
Solution Administration Sets with 0.2 Micron Filter
Applicant
Baxter Healthcare Corporation
Product Code
FPA · General Hospital
Decision Date
Dec 28, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For the retention of microorganisms and removal of air and particulate matter from infusion fluids.
Device Story
Solution Administration Sets are sterile, single-use disposable devices used to deliver fluids from a container to a patient's vascular system via a needle or catheter. The device incorporates a 0.2 micron polyethersulfone (PES) solution membrane and a 0.1 micron polyvinylidene fluoride (PVDF) air vent membrane within a copolyester housing. These filters function to retain microorganisms and remove air and particulate matter from infusion fluids. The device is used in clinical settings by healthcare professionals. The modification involves replacing the existing hydrophilic PES membrane with an equivalent hydrophilic PES material. The device output is the filtered infusion fluid delivered to the patient, which benefits the patient by reducing the risk of infection and air embolism during intravenous therapy.
Clinical Evidence
Bench testing only. Performance data included air diffusion, bubble point, gravity flow rate, flow rate post-sterile water conditioning, flow rate post-parenteral nutrition conditioning, and bacterial retention tests. All tests met acceptance criteria. Biocompatibility testing was conducted per ISO-10993 and FDA Blue Book Memorandum #G95-1, including cytotoxicity, systemic toxicity, intracutaneous, hemolysis, pyrogen, sensitization, and USP physicochemical tests.
Technological Characteristics
Materials: Copolyester housing, PES solution membrane, PVDF air vent membrane, PVC tubing, stainless steel cannula, ABS spike, synthetic polyisoprene plug. Energy source: None (gravity/pump-driven). Connectivity: None. Sterilization: Sterile. Software: None.
Indications for Use
Indicated for the retention of microorganisms and removal of air and particulate matter from infusion fluids in patients requiring intravenous fluid administration.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Solution Administration Sets with 0.22 Micron Filter (K964850)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other. The profiles are stylized and appear to be made of flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 28, 2015
Baxter Healthcare Corporation Mr. Gary Chumbimune Manager, Regulatory Affairs 32650 N Wilson Road Round Lake. Illinois 60073
Re: K153158
Trade/Device Name: Solution Administration Sets with 0.2 Micron Filter Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: December 2, 2015 Received: December 3, 2015
Dear Mr. Chumbimune:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K153158
#### Device Name
Solution Administration Sets with 0.2 Micron Filter
Indications for Use (Describe)
For the retention of microorganisms and removal of air and particulate matter from infusion fluids.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is colored in a dark blue. The image is simple and only contains the company name.
# Section 5. 510(k) Summary K153158
October 30, 2015
#### OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
### CONTACT PERSON:
Gary Chumbimune Manager, Regulatory Affairs 32650 N Wilson Road Round Lake, IL 60073 Telephone: (224) 270-3312 Fax: (224) 270-4119
### IDENTIFICATION OF THE DEVICE:
Common Name: Intravascular Administration Set Trade Name or Proprietary Name: Solution Administration Sets with 0.2 Micron Filter Classification Panel: 80 General Hospital Classification: Set, Administration, Intravascular (21 CFR 880.5440) Class: Class II
Product Code: FPA
#### Table 1. Product Codes for Solution Administration Sets with 0.2 Micron Filter
| Code Number | Name |
|-------------|------------------------------------------------------------|
| 2C6572 | INTERLINK System CONTINU-FLO Solution Set |
| 2H6480 | INTERLINK System Non-DEHP Solution Set with DUO-VENT Spike |
| 2H8671 | CLEARLINK System Non-DEHP Extension Set |
| 1C8363 | Extension Set |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The color of the text is a dark blue. The text is slightly slanted to the right, giving it a dynamic appearance. The background is plain white, which makes the text stand out.
### PREDICATE DEVICE:
| Device | Company | Predicate 510(k) | Clearance Date |
|---------------------------------------------------------|----------------------------------|------------------|----------------------|
| Solution Administration Sets with 0.22<br>Micron Filter | Baxter Healthcare<br>Corporation | K964850 | February 25,<br>1997 |
#### Table 2. Predicate Device
# DESCRIPTION OF THE DEVICE:
The Solution Administration Sets with a 0.2 micron filter product line consists of sterile, non-pyrogenic, single use disposable devices used for the administration of fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. They are indicated for the retention of microorganisms and removal of air and particulate matter from infusion fluids. The filter consists of a 0.2 micron polyethersulfone (PES) solution membrane and 0.1 micron polyvinylidene fluoride air vent membrane enclosed in a copolyester housing. These devices were previously cleared under 510(k) premarket notification K964850 (cleared February 25, 1997)
The basis for this premarket notification is a modification to the 0.2 micron filter currently used in this product line. The modification consists of a change to the solution membrane material. The solution membrane material is changing from a hydrophilic polyethersulfone (PES) to another hydrophilic polyethersulfone (PES). The solution membrane currently used is no longer available and will be replaced with an equivalent material.
These modifications do not impact the intended use or the fundamental scientific technology of the devices. No other materials of construction are being introduced into these devices as part of this update. The product labels are also being updated to revise statements regarding latex and pump device references, add the indications for use statement of the device, and implement other modifications to comply with Baxter's labeling standards.
#### INDICATIONS FOR USE:
For the retention of microorganisms and removal of air and particulate matter from infusion fluids.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The letters are slightly italicized, giving the word a sense of movement. The color of the text is a vibrant blue, which stands out against the white background.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The proposed devices have equivalent technological characteristics as Baxter's currently legally marketed devices cleared under 510(k) premarket notification K964850 (cleared on February 25, 1997). The intended use, design and function of the proposed devices are equivalent to the predicate devices.
# DEVICE COMPARISON TABLE:
Table 3 is a device comparison table outlining the differences between the current (predicate) devices and the proposed devices.
| Features | | Current Devices<br>(Cleared under K964850) | Proposed Devices |
|--------------------------|----------------------|--------------------------------------------------------------------------------------------------------|------------------|
| Intended Use | | For the retention of microorganisms and removal<br>of air and particulate matter from infusion fluids. | Same |
| Indications for Use | | For the retention of microorganisms and removal<br>of air and particulate matter from infusion fluids. | Same |
| Sterile | | Yes | Same |
| Non-Pyrogenic | | Yes | Same |
| Single Use | | Yes | Same |
| Materials | | | |
| Spike | | Acrylonitrile Butadiene Styrene | Same |
| Blue Plug | | Synthetic Polyisoprene | Same |
| Cannula | | 304 Stainless Steel | Same |
| Drip Chamber | | Polyvinyl Chloride | Same |
| Check Valve | | Polymethyl Methacrylate and Silicone Rubber | Same |
| 0.2 Micron<br>Filter | Housing | Copolyester | Same |
| | Solution<br>Membrane | Polyethersulfone (PES) | Same |
| | Air Vent<br>Membrane | Polyvinylidene (PVDF) | Same |
| Trilayer<br>Tubing | Inner Layer | Linear Low Density Polyethylene | Same |
| | Middle<br>Layer | Polyolefin Adhesive | Same |
| | Outer Layer | Polyvinyl Chloride | Same |
| Tubing | | Polyvinyl Chloride | Same |
| Bushing | | Polyvinyl Chloride | Same |
| Interlink Y-Site | | Copolyester and Synthetic Polyisoprene | Same |
| Clearlink Y-Site | | Polycarbonate and Silicone | Same |
| Male Luer Lock Connector | | Acrylonitrile Butadiene Styrene | Same |
### Table 3. Device Comparison
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Baxter" in a stylized, bold, blue font. The font appears to be italicized, with the letters slightly slanted to the right. The color of the text is a vibrant blue, and the background is white.
### DISCUSSION OF NONCLINICAL TESTS:
Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed devices are appropriately designed for their intended use.
### Performance Data:
The following bench tests were conducted to evaluate the effect of the modification to the 0.2 micron filter:
- Air diffusion test ●
- Bubble point test
- Gravity flow rate test ●
- Flow rate test post sterile water conditioning ●
- Flow rate test post parenteral nutrition conditioning
- Bacterial retention test
All tests met the acceptance criteria.
### Biocompatibility:
The basis for this premarket notification is a modification to the 0.2 micron filter solution membrane material. The solution membrane material is changing from a hydrophilic polyethersulfone (PES) to another hydrophilic polyethersulfone (PES). No other materials of construction are being introduced into these devices as part of this update.
Biocompatibility assessments were conducted in accordance with ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," for prolonged duration, external communicating device, indirect blood path, and FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993" as recommended in the IV Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]''. The battery of testing included the following tests:
- . Cytotoxicity
- Systemic Toxicity ●
- Intracutaneous ●
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement. The background is plain white.
- Hemolysis ●
- Pyrogen ●
- Sensitization ●
- USP Physicochemical ●
All other materials found in these devices, that are the subject of this submission, have been previously cleared under Baxter's 510(k) premarket notifications K123868 (cleared January 8, 2013) and K150860 (cleared April 16, 2015).
# CONCLUSION:
The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate devices that are currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.