← Product Code [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA) · K123874

# INTERLINK SYSTEM BURETROL SOLUTION SETS, INTERLINK SYSTEM BURETROL SOLUTION SETS, CONTINUED, CLEARLINK SYSTEM BURETROL S (K123874)

_Baxter Healthcare Corp · FPA · Jan 10, 2013 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K123874

## Device Facts

- **Applicant:** Baxter Healthcare Corp
- **Product Code:** [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA.md)
- **Decision Date:** Jan 10, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Buretrol Solution Sets are intended for use in the administration of fluids from a container into the patient's vascular system through a vascular access device.

## Device Story

Buretrol Solution Sets are sterile, single-use disposable IV administration sets; feature a burette chamber for mixing supplementary medication in measured diluent amounts. Device allows for either metered volume (intermittent) or continuous solution administration; conversion between modes achieved by closing air vent at top of burette. Used in clinical settings by healthcare providers to deliver fluids to patients via vascular access devices. Benefits include precise medication delivery and flexible administration modes.

## Clinical Evidence

No clinical data. Bench testing only. Performance evaluated via spike leakage, burette squeeze/pressure tests, scale legibility, roller clamp shut-off, check valve backflow, air vent flow, and ISO Luer tests. All tests met acceptance criteria.

## Technological Characteristics

Sterile, single-use IV administration set. Features burette chamber, roller clamps, and valve configurations (ball valve, valveless, or drip chamber filter valve). Materials evaluated per ISO 10993 and Blue Book Memorandum G95-1. Connectivity: None. Energy source: None (gravity/manual).

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Modified Buretrol Solution Sets ([K984381](/device/K984381.md))
- Burette Set with Air Control Lever - 2C0147 ([K780970](/device/K780970.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Special 510(k) Premarket Notification Buretrol Solution Sets

# 5. 510(K) SUMMARY

December 14, 2012

Owner:

Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015

#### Contact person:

Nanette Hedden Associate Director, Global Regulatory Affairs 36250 N. Wilson Rd. Round Lake, IL 60073 Telephone: (224) 270-4871 Fax: (224) 270-4900

## DEVICE NAME: IV Administration Sets

Trade Name: Buretrol Solution Sets

| Code Number | Name                                                                                                                                                                                                                      |
|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1C8441      | Interlink System Buretrol Solution Set, 77" (2.0 m), 150 mL Burette Ball Valve                                                                                                                                            |
| 2H8862      | Clearlink System Non-DEHP Buretrol Solution Set, 115" (2.9 m), 150 mL Valveless Burette                                                                                                                                   |
| 3C0128      | Interlink System Buretrol Solution Set, 105" (2.7 m), 150 mL Burette Drip Chamber Filter Valve<br>Large Bore 4-Way Stopcock Extension Set, 34" (87 cm), Vol. 4.6 mL<br>Minivolume Extension Set, 2.9" (7 cm), Vol. 0.1 mL |

# Table 5-1 Representative Product Codes for Buretrol Solution Sets

Section 5, 510(k) Summary Page 1 of 3

# K123874

JAN 1 0 2013

35

{1}------------------------------------------------

# Common Name: Solution Set with Burette Chamber

Classification Name: IV Administration Set (21 CFR 880.5440, Product Code FPA)

#### PREDICATE DEVICE:

# Table 5-2 Previous 510(k)s

| Device                                         | Company                   | Previous 510(k) | Clearance<br>Date    |
|------------------------------------------------|---------------------------|-----------------|----------------------|
| Modified Buretrol Solution Sets                | Baxter Healthcare         | K984381         | February 19,<br>1999 |
| Burette Set with Air Control Lever -<br>2C0147 | Travenol<br>Laboratories1 | K780970         | July 27, 1978        |

Travenol Laboratories was renamed Baxter Healthcare in 1988.

## DESCRIPTION OF THE DEVICE:

The Buretrol Solution Sets are sterile, single use disposable devices indicated for use in the administration of fluids from a container into the patient's vascular system through a vascular access device. The sets contain a burette chamber which can be used to mix supplementary medication in a measured amount of diluent from the main container. The sets can be adjusted for either metered volume solution administration (intermittent) or continuous solution administration. The sets can be converted from intermittent to continuous administration by closing the air vent at the top of the burette and allowing continuous infusion from the main container. There are different configurations of the burette, based on presence of and differences in valve type.

## STATEMENT OF INTENDED USE:

The Buretrol Solution Sets are intended for use in the administration of fluids from a container into the patient's vascular system through a vascular access device.

## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The proposed Buretrol Solution Sets are equivalent to Baxter's currently legally marketed Modified Buretrol Solution Sets cleared February 19, 1999 (K984381). The modification to the labeling does not impact the intended use or the fundamental scientific technology of the device. . The intended use, basic design, function and the materials for the proposed device are equivalent to the predicate device.

{2}------------------------------------------------

# DISCUSSION OF NONCLINICAL TESTS:

Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the results of these analyses. All test results meet their acceptance criteria, and support that the devices are appropriately designed for their intended use.

Performance Data: The following tests were conducted to evaluate the functional performance of the Buretrol Sets:

- . Spike Leakage Test Over Time
- Burette Squeeze Test .
- Burette Pressure Test .
- Burette Scale Legibility .
- Roller Clamp Immediate Shut-Off .
- Roller Clamp Shut-Off Over Time .
- Check Valve Backflow Evaluation .
- Regulating Roller Clamp Cold Flow Test .
- Regulating Roller Clamp Shut-Off Test .
- Air Vent Air Flow Test ●
- ISO Luer Tests on Male Luer lock .

All tests met the acceptance criteria.

Biocompatibility: No new materials of construction are being introduced into these devices as part of this Special 510(k) Premarket Notification. Biocompatibility assessment of the Buretrol Solution Sets has been conducted based on ISO 10993, Biological Evaluation of Medical Devices for prolonged duration, external communicating, indirect blood path and Blue Book Memorandum G95-1 "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," as recommended in the IV Administration sets guidance.

## CONCLUSION:

The Buretrol Solution Sets are substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Silver Spring, MD 20993-0002

## January 10, 2013

Ms. Nanette Hedden Associate Director, Global Regulatory Affairs Baxter Healthcare Corporation 36250 North Wilson Road ROUND LAKE IL 60073

Re: K123874

Trade/Device Name: Buretrol Solutions Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: December 14, 2012 Received: December 17, 2012

Dear Ms. Hedden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Hedden

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{5}------------------------------------------------

# 4. INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

Device Name: Buretrol Solution Sets

Indication(s) for Use:

The Buretrol Solution Sets are intended for use in the administration of fluids from a container into the patient's vascular system through a vascular access device.

D AND/OR Prescription Use:

Over-the-Counter Use: T

21 CFR 801 Subpart D

21 CFR Subpart C

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

weihong Cu -5 emment, ou=HHS, ou=FDA, ou c=US. o=US. Go cn=Weihong Gu -S, 0.9.2342.19200300.100.1.1 2013.01.11 11:51:01 -05 00

(Division Sign-Off) (Division of Anestheslology, General Hospital Infection Control, Dental Devices

510(k) Number:

રેન્દ્રન

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K123874](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K123874)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
