K123729 · Emed Technologies Corporation · FPA · Dec 21, 2012 · General Hospital
Device Facts
Record ID
K123729
Device Name
VERSARATE FLOW RATE CONTROLER
Applicant
Emed Technologies Corporation
Product Code
FPA · General Hospital
Decision Date
Dec 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The VersaRate™ Flow Rate Controller is intended for use in the intravascular infusion of fluids to be delivered to the patient in a precise manner for no longer than 72 hours.
Device Story
VersaRate™ Flow Rate Controller is a disposable, manual IV flow control device. It connects to standard administration sets and fluid sources via luer lock connectors. Operation involves user rotation of a flow dial to adjust the internal fluid path dimensions, thereby regulating the fluid flow rate (5–230 mL/hr). Used in clinical settings to manage fluid delivery; provides precise, controlled infusion. Benefits include extended duration of use (up to 72 hours) and high pressure tolerance (up to 25 psi) compared to predicate. Healthcare providers manually set the rate; output is the physical regulation of fluid flow to the patient.
Clinical Evidence
Bench testing only. Testing compared flow rate control (5–230 mL/hr), residual volume (<0.25 mL), duration of use (72 hours), and pressure tolerance (25 psi) against the predicate. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemocompatibility) and ASTM 756, with all tests passing.
Technological Characteristics
Materials: PVC (Luer, Tubing), Polycarbonate and styrene-ethylene-butylene (Regulator). Principle: Manual mechanical flow restriction via dial-adjusted fluid path dimensions. Form factor: 16 cm length. Connectivity: None (standalone). Sterilization: Ethylene Oxide. Shelf life: 4 years.
Indications for Use
Indicated for patients requiring intravascular infusion of fluids in a precise manner for a duration not exceeding 72 hours.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Section 5. 510(k) Summary
K123729 K Number
Submission Date:
11/6/12
Class II
General Information
Classification
Trade Name
VersaRate™ Flow Rate Controller
Common Name:
I.V. Flow Controller
Classification Name and Reference:
Intravascular Administration Set
21 CFR §880.5440
Submitter
Peter Kollings EMED Technologies Corporation 1264 Hawks Flight Ct., Ste. 200 El Dorado Hills, Ca 95762 916.932.0071 x114 Tel: 916.932.0074 Fax:
#### Intended Use
The VersaRate™ Flow Rate Controller is intended for use in the intravascular infusion of fluids to be delivered to the patient in a precise manner for no longer than 72 hours.
#### Predicate Device(s)
Baxter Extension Set w/Flow Regulator (K890489)
#### Device Description
The EMED VersaRate™ Flow Rate Controller is a disposable device allowing users to adjust the flow rate of fluids. The EMED VersaRate™ Flow Rate Controller can be connected to commercially available administration sets and fluid sources utilizing standard luer lock connectors.
#### Materials and Characteristics
The EMED VersaRate™ Flow Rate Controller is equivalent in performance, physical properties, using similar materials, and having the same indications for use as the
# DEC 2 1 2012
{1}------------------------------------------------
predicate device. Therefore no new issues of safety or effectiveness are introduced by the minimal differences in design.
Table 5-1 below provides a comparison of technological and other characteristics of the EMED VersaRate™ Flow Rate Controller and the predicate.
| | EMED VersaRate™<br>Flow Rate Controller | Baxter Extension Set<br>w/Flow Regulator |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The VersaRate™ Flow<br>Rate Controller is<br>intended for use in the<br>intravascular infusion of<br>fluids to be delivered to<br>the patient in a precise<br>manner for no longer than<br>72 hours. | The Extension set with<br>Flow Regulator is<br>intended for use in the<br>administration of<br>intravenous fluids to be<br>delivered to the patient in<br>a precise manner over a<br>specified period of time. |
| Flow Control<br>Material/Components | Biocompatible, non-toxic<br>materials widely used in<br>medical products, such as:<br>Luer: PVC<br>Tubing: PVC<br>Regulator: Polycarbonate<br>and styrene-ethylene-<br>butylene | Biocompatible, non-toxic<br>materials widely used in<br>medical products. |
| Method of<br>Sterilization | Ethylene Oxide | Radiation |
| Principle of Flow Rate<br>Control | The fluid path of the<br>VersaRate™ Flow Rate<br>Controller is regulated by<br>rotating the flow dial to<br>alter dimensions of the<br>internal fluid path, thereby<br>altering the flow rate. | Fluid flow is regulated by<br>rotating the diaphragm<br>holder to create a<br>restrictive path of varying<br>depth thereby altering the<br>rate of fluid<br>administration. |
| Length | 16 cm | 46 cm |
Table 5-1
{2}------------------------------------------------
# Performance
Table 5-2 below summarizes testing results performed to establish conformance of the VersaRate™ Flow Rate Controller to internal product specifications and requirements, as well as equivalence to the predicate device.
## Table 5-2
| | VersaRate™ Flow Rate<br>Controller | Baxter Extension Set<br>w/Flow Regulator |
|----------------------------|----------------------------------------------------------|----------------------------------------------------------|
| Flow Rate Control<br>Range | 5 – 230 mL/hr at 80 cm<br>head height | 5 – 250 mL/hr |
| Residual Volume of<br>Set | < 0.25 ml | 2.9 mL |
| Duration of Use | Performance remains<br>within tolerance up to 72<br>hrs. | Performance remains<br>within tolerance up to 24<br>hrs. |
| Pressure | Up to 25 psi | Up to 3 psi |
The outcomes of these tests further indicate that the VersaRate™ Flow Rate Controller is substantially equivalent to the predicate in performance, effectiveness, and safety.
.
。
{3}------------------------------------------------
### Biocompatibility
In accordance with ISO 10993, studies were performed including cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, and hemocompatibility. Table 5-3 presents a summary of testing and results indicating compliance with biocompatibility standards.
| Table 5-3 | |
|-----------|--|
|-----------|--|
| Test Performed | Standard | Test Name | Test<br>Result | Other Name |
|------------------|--------------|----------------------------|----------------|-----------------------------------------------|
| Biocompatibility | ISO 10993-5 | Cytotoxicity | Pass | Neutral Red<br>Uptake |
| Biocompatibility | ISO 10993-10 | Sensitization | Pass | Kligman<br>Maximization |
| Biocompatibility | ISO 10993-10 | Irritation | Pass | Intracutaneous<br>Injection |
| Biocompatibility | ISO 10993-11 | Acute systemic<br>toxicity | Pass | Systemic Injection |
| Biocompatibility | ISO 10993-11 | Pyrogenecity | Pass | Rabbit Pyrogen |
| Biocompatibility | ISO 10993-4 | Hemocompatibility | Pass | Unactivated Partial<br>Thromboplastin<br>Time |
| Biocompatibility | ASTM 756 | Hemocompatibility | Pass | Hemolysis<br>(complete) |
#### Sterility, Shelf-life, and Packaging
The VersaRate™ Flow Rate Controller will be sterilized to a sterility assurance level (SAL) of 10 of 10 and with a shelf life of 4 years.
#### Summary of Substantial Equivalence
The EMED Technologies Corporation VersaRate™ Flow Rate Controller is substantially equivalent to the commercially available predicate device in terms of function, safety, performance, intended use, technology/principles and mechanical properties. Differences between the EMED VersaRate™ Flow Rate Controller and the predicate do not raise any new issues of safety or effectiveness.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the text "Public Health Service" in a bold, serif font. The text is stacked vertically, with "Public Health" on the first line and "Service" on the second line. The text is black against a white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### December 21, 2012
EMED Technologies Corporation C/O Mr. Morten S. Christensen Staff Engineer Underwriters Laboratories, Incorporated 455 East Trimble Road SAN JOSE CA 95131-1230
Re: K123729
Trade/Device Name: VersaRate™ Flow Rate Controller Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA . Dated: November 29, 2012 Received: December 5, 2012
#### Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Christensen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony Lo m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Section 4. Indications for Use Statement
510(k) Number (if known):
Device Name:
Indications for Use:
# [TBD] K123729
VersaRate™ Flow Rate Controller
The VersaRate™ Flow Rate Controller is intended for use in the intravascular infusion of fluids to be delivered to the patient in a precise manner for no longer than 72 hours.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Concurrence of CDRI
12/19/12
Division Sign-Off) (Division Sign-con)
Division of Anesthesiology, General Hospital Division Control, Dental Devices
:10(k) Number: K123729
Section 4, Page 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.