← Product Code [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA) · K120548

# HLIC NEEDLE-LESS VALVE (K120548)

_Health Line International Corporation · FPA · May 4, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K120548

## Device Facts

- **Applicant:** Health Line International Corporation
- **Product Code:** [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA.md)
- **Decision Date:** May 4, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The HLIC Needle-Less Valve is intended to allow access to IV administration sets, medication vials, blood donor collection bags, and solution bags with one convenient device without the use of needles or blunt cannulas.

## Device Story

HLIC Needle-Less Valve is a needle-free access device for IV administration sets, medication vials, blood donor collection bags, and solution bags. Device features a silicone stem that remains closed to maintain a sealed fluid path; insertion of a male syringe or secondary line connector forces the stem open to create a fluid path. Upon removal of the connector, the valve body forces the stem shut. Device provides a flat, smooth surface for cleaning and does not require a cap to maintain sterility. It is power injection capable. Used in clinical settings to facilitate medication delivery or fluid access without needles or blunt cannulas, reducing needlestick risks.

## Clinical Evidence

Bench testing only. Performance data from design verification testing demonstrated substantial equivalence to the predicate. Biocompatibility testing was conducted according to ISO-10993 for externally communicating, blood path, indirect, limited contact duration (<24 hours) devices. Power injection testing was performed to verify the new capability.

## Technological Characteristics

Needle-free valve with silicone stem mechanism. Materials meet ISO-10993 biocompatibility standards for limited contact duration. Features a flat, smooth surface for cleaning. Power injection capable. No electronic components or software.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Swabsite Swabbable Valve ([K002689](/device/K002689.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Health Line International Corporation 510(k) Premarket Notification Submission: HLIC Needle-Less Valve

K120548
Page 15 of 86

MAY - 4 2012

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (21 CFR 807.92) for HLIC NEEDLE-LESS VALVE

### SUBMITTER:

Health Line International Corporation 803 N. 1250 W. - STE 1 Centerville, Utah 84014

### ESTABLISHMENT REGISTRATION NUMBER: 3006097687

#### CONTACT:

Nola L. Benstog QA/RA Director Telephone: 801-773-7798 Fax: 801-820-8007 Email: nbenstog@hlic.net

### DATE PREPARED:

February 21, 2012

#### NAME OF MEDICAL DEVICE:

Proprietary Name: Regulation Name: Common/Usual Name:

# HLIC NEEDLE-LESS VALVE

Intravascular Administration Set Needle Free Injection Site

### DEVICE CLASSIFICATION:

Classification Panel: Regulatory Class: Product Code: Regulation Number:

General Hospital Class II FPA 21 CFR 880.5440

### PREDICATE DEVICE:

Proprietary Name: Regulation Name: Common/Usual Name: Classification Panel: Regulator Class: Product Code: Regulation Number:

Swabsite Swabbable Valve Intravascular Administration Set Intravascular Administration Set / Swabbable Valve · General Hospital Class II FPA 21 CFR 880.5440

{1}------------------------------------------------

### DEVICE DESCRIPTION:

The HLIC Needle-Less Valve is a needle free valve that allows the user to add medication into IV sets without the use of a needle. When the valve is in the closed position, it has a flat, smooth surface for cleaning. When the male connector of a syringe or secondary line is pushed into the valve, the silicone stem opens in the middle creating a fluid path. When the male connector is removed from the valve, the body of the valve forces the stem shut and maintains a sealed fluid path. A cap is not required to seal the valve or to maintain sterility.

{2}------------------------------------------------

#### INTENDED USE:

The HLIC Needle-Less Valve is intended to allow access to IV administration sets, medication vials, blood donor collection bags, and solution bags with one convenient device without the use of needles or blunt cannulas.

#### INDICATIONS FOR USE:

The HLIC Needle-Less Valve is intended to allow access to IV administration sets, medication vials, blood donor collection bags, and solution bags with one convenient device without the use of needles or blunt cannulas.

## TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES:

New device is compared to Marketed Device? Yes. It is compared to legally marketed predicates.

Does the new device have the same indication statements? Yes.

Do the differences alter the intended therapeutic/diagnostic/etc. effect (i.e. deciding may consider impact on safety and effectiveness)? No, the differences do not alter the intended use of the device.

Does the new device have the same technological characteristics, e.g. design, material, etc.? Yes. The HLIC NEEDLE-LESS VALVE, is substantially equivalent to the Swabsite Swabbable Valve K002689. The basic fundamental scientific technology of the device has not changed.

Could the new characteristics affect safety or effectiveness? No.

Do the new characteristics raise new types of safety and effectiveness questions? No. The only new characteristic is that the HLIC NEEDLE-LESS VALVE is power injection capable. This is based upon existing test data and does not affect safety or effectiveness.

Do accepted scientific methods exist for assessing effects of the new characteristics? Yes.

Power Injection testing was conducted to assess the power injection capabilities of the HLIC NEEDLE-LESS VALVE.

Biocompatibility requirements according to ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. Test profiles for externally communicating, Blood path, indirect, Limited contact duration (<24 hours) were met.

These and other standards were used to determine the appropriate methods for evaluating the device's performance.

#### CONFIDENTIAL

{3}------------------------------------------------

Are performance data available to assess effects of new characteristics? Yes. Verification testing was performed according to protocols based on the above-referenced guidance document recommendations and additional standards.

Do performance data demonstrate equivalence? Yes. Performance data gathered in design verification testing demonstrated that the HUC NEEDLE-LESS VALVE is substantially equivalent to the noted predicate devices.

# CONCLUSION

The HLIC NEEDLE-LESS VALVE met all established acceptance criteria for performance testing and design verification testing. This testing demonstrated that the HLIC NEEDLE-LESS VALVE is safe and effective for its intended use, and based on FDA's decision tree is substantially equivalent to the following predicate device: Swabsite Swabbable Valve (K002689).

{4}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, services, and people. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Nola Benstog Director QA/RA Health Line International Corporation 803 North 1250 West, Suite 1 Centerville, Utah 84014

MAY - 4 2012

Re: K120548

Trade/Device Name: HLIC NEEDLE-LESS VALVE Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: FPA Dated: April 26, 2012 Received: April 27, 2012

Dear Ms. Benstog:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

### Page 2- Ms. Benstog

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

#### Indications For Use

510(k) Number (if known): 277548

Device Name: HLIC NEEDLE-LESS VALVE

Indications For Use:

The HLIC Needle-Less Valve is intended to allow access to IV administration sets, medication vials, blood donor collection bags, and solution bags with one convenient device without the use of needles or blunt cannulas.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rild C. Chigan 5/4/12
(Division Sign-Off)

Division of Anesthesiology, General Hoapital Infection Control, Dental Devices

510(k) Number: K120548

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K120548](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K120548)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
