← Product Code [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA) · K112893

# CLEARLINK LUER ACTIVATED VALVE, CLEARLINK SYSTEM NON-DEHP CATHETER EXTENSION SETS (K112893)

_Baxter Healthcare Corporation · FPA · Oct 18, 2011 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K112893

## Device Facts

- **Applicant:** Baxter Healthcare Corporation
- **Product Code:** [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA.md)
- **Decision Date:** Oct 18, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

For use with a vascular access device for the administration of drugs and solutions. The Clearlink Luer Activated Valve is an in-line injection site, which can be connected to standard male Luer adapters (e.g., syringes or sets) for continuous or intermittent fluid administration or the withdrawal of fluids.

## Device Story

Clearlink Luer Activated Valve is an in-line injection site for IV administration and extension sets; enables needleless access to fluid/vascular paths via standard male Luer adapters (syringes/sets). Used in clinical settings for continuous or intermittent fluid administration or withdrawal. Device is a single-use disposable component. Healthcare providers connect male Luer adapters to the valve to facilitate medication delivery or fluid management. Design modification to Y-site port optimizes manufacturing assembly process. Benefits include secure, needleless vascular access for patient therapy.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included visual inspection, pressure differential, lipid resistance, back pressure (open/closed), low water pressure seepage, and ultrasonic weld joint performance. All results met acceptance criteria.

## Technological Characteristics

Single-use disposable IV administration and extension sets. Features Clearlink Luer Activated Valve as an in-line injection site. Materials and basic design are equivalent to predicate. Manufacturing assembly optimized via minor Y-site design change. No electronic components, software, or energy sources.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Modification to Solution Administration Set with Capped Luer Activated Device (Clearlink) ([K003225](/device/K003225.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Special 510(k) Premarket Notification Clearlink Luer Activated Valve, IV Administration and IV Extension Sets with Clearlink Luer Activated Valve Section 5, 510(k) Summary

OCT 1 8 2011

# K112 893

#### 5. 510(k) SUMMARY

September 29, 2011

#### OWNER:

。

Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015

#### CONTACT PERSON:

Nanette Hedden Senior Manager, Regulatory Affairs 1620 Waukegan Rd. MPGR-AL McGaw Park, IL 60085 Telephone: (847) 270-4871 Fax: (847) 785-5116

#### DEVICE NAME:

Trade name: Clearlink Luer Activated Valve, Intravenous (IV) administration and IV extension sets with the Clearlink Luer Activated Valve

# Clearlink Luer Activated Valve, Representative Intravenous (IV) administration and IV extension sets with the Clearlink Luer Activated Valve

Table 5-1.

| Code number | Name                                                                                     |
|-------------|------------------------------------------------------------------------------------------|
| 2N8399      | Clearlink System Luer Activated Valve for IV Access, Vol. 0.25 mL                        |
| 2N8378      | Clearlink System Non-DEHP Catheter Extension Set, 7.6" (19.3 cm),<br>Vol. 0.9 mL         |
| 2N8377      | Clearlink System Non-DEHP Y-Type Catheter Extension Set, 6.5'' (16.5<br>cm), Vol. 1.2 mL |
| 2N8374      | Clearlink System Non-DEHP Catheter Extension Set, 8.2" (21.0 cm),<br>Vol. 0.5 mL         |
| 2N8371      | Clearlink System Non-DEHP Y-Type Catheter Extension Set, 6.0" (15.2<br>cm), Vol. 1.0 mL  |
| 2H8401      | Clearlink System Non-DEHP Solution Set, 76" (1.9 m)                                      |
| 2C8425s     | Clearlink System Solution Set, 100" (2.5 m)                                              |

{1}------------------------------------------------

Special 510(k) Premarket Notification Clearlink Luer Activated Valve, IV Administration and IV Extension Sets with Clearlink Luer Activated Valve

Common name: Clearlink Luer Activated Valve, Intravenous (IV) administration and IV extension sets with the Clearlink Luer Activated Valve

Classification name: IV Administration Set (21 CFR 880.5440, Product Code FPA)

. .

#### PREDICATE DEVICE:

Table 5-2. Previous 510(k)s

| Device                                                                                           | Company              | Previous<br>510(k) | Clearance<br>date   |
|--------------------------------------------------------------------------------------------------|----------------------|--------------------|---------------------|
| Modification to Solution<br>Administration Set with Capped Luer<br>Activated Device (Clearlink). | Baxter<br>Healthcare | K003225            | October 19,<br>2000 |

#### DESCRIPTION OF THE DEVICE:

The proposed devices, which are the subject of this Special 510(k) Premarket Notification, consist of the Clearlink Luer Activated Valve, Intravenous (IV) administration and IV extension sets with the Clearlink Luer Activated Valve. They are single use disposable devices intended for use with a vascular access device for continuous or intermittent fluid administration or withdrawal of fluids. These devices are the same as the current marketed devices, previously cleared on October 19, 2000 under 510(k) premarket notification "Modification to Solution Administration Set with Capped Luer Activated Device (Clearlink)" by Baxter Healthcare Corporation (K003255).

The Clearlink Luer Activated Valve is an in-line access site and can be connected to male Luer adapters (e.g., syringes or sets) to allow needleless access to the fluid or vascular path.

The IV administration sets with the Clearlink Luer Activated Valve are used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. They contain the Clearlink Luer Activated Valve Y-Site port that can be used for the administration of secondary medication.

. .

{2}------------------------------------------------

Special 510(k) Premarket Notification Clearlink Luer Activated Valve, IV Administration and IV Extension Sets with Clearlink Luer Activated Valve

The IV extension sets with the Clearlink Luer Activated Valve are used for the administration and withdrawal of fluids. They consist of the Clearlink Luer Activated Valve connected to the IV extension set.

The basis for this premarket notification is a proposed minor design change to Baxter's Clearlink Luer Activated Valve Y-Site design that is integral to the IV administration sets.

#### STATEMENT OF INTENDED USE:

For use with a vascular access device for the administration of drugs and solutions. The Clearlink Luer Activated Valve is an in-line injection site, which can be connected to standard male Luer adapters (e.g., syringes or sets) for continuous or intermittent fluid administration or the withdrawal of fluids.

#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The proposed devices consist of the Clearlink Luer Activated Valve, Intravenous (IV) administration and IV extension sets with the Clearlink Luer Activated Valve. These devices are the same as the current marketed devices, previously cleared on October 19, 2000 under 510(k) premarket notification "Modification to Solution Administration Set with Capped Luer Activated Device (Clearlink)" by Baxter Healthcare Corporation (K003255). A minor design change will be made to the Clearlink Luer Activated Valve Y-Site design that is integral to the IV administration sets in order to optimize the manufacturing assembly process. The intended use, the basic design, function and the materials for the proposed device are equivalent to the predicate device.

{3}------------------------------------------------

Section 5, 510(k) Summary

Special 510(k) Premarket Notification Clearlink Luer Activated Valve, IV Administration and IV Extension Sets with Clearlink Luer Activated Valve

ﺎ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ

# DISCUSSION OF NONCLINICAL TESTS:

Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet the acceptance criteria, and support that the devices are appropriately designed for their intended use. The following bench tests were conducted to evaluate the effect of the design modification on the functional performance of the Clearlink Luer Activated Valve:

- Visual inspection ●
- . Pressure differential test
- Lipid resistance test ●
- . Back pressure test (closed position)
- Back pressure test with male Luer in (open position) ●
- 6 psi low water pressure seepage test .
- Ultrasonic weld joint performance .

#### CONCLUSION:

The proposed devices are substantially equivalent to the predicate device.

{4}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Nanette Hedden Senior Manager, Regulatory Affairs Baxter Healthcare Corporation 1620 Waukegan Road McGaw Park, Illinois 60085

OCT 1 8 2011

Re: K112893

Trade/Device Name: Clearlink Luer Activated Valve, Intravenous (IV) administration and IV extension sets with the Clearlink Luer Activated Valve Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: September 29, 2011 Received: October 3, 2011

Dear Ms. Hedden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Hedden

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{6}------------------------------------------------

# INDICATIONS FOR USE

# 510(k) Number (if known): K11289 3

# Device Name: Clearlink Luer Activated Valve, Intravenous (IV) administration and IV extension sets with the Clearlink Luer Activated Valve

Indications for Use:

For use with a vascular access device for the administration of drugs and solutions. The Clearlink Luer Activated Valve is an in-line injection site, which can be connected to standard male Luer adapters (e.g., syringes or sets) for continuous or intermittent fluid administration or the withdrawal of fluids.

| Prescription Use            | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |

AND/OR

| Over-The-Counter Use   |  |
|------------------------|--|
| (21 CFR 807 Subpart C) |  |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) 14: 1 : 1 : 1 : 1 :

13

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rill C 10/7/4

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K112893

: :

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K112893](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K112893)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
