# IMM HYPODERMIC NEEDLE SET (K070801)

_Innovative Medical Manufacturing Co. · FPA · May 24, 2007 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K070801

## Device Facts

- **Applicant:** Innovative Medical Manufacturing Co.
- **Product Code:** [FPA](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA.md)
- **Decision Date:** May 24, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The IMM Hypodermic Needle Set is to be used for sampling blood from or infusing fluid into body system underneath the skin.

## Device Story

IMM Hypodermic Needle Set consists of stainless steel needle, needle hub, extension tubing, clamp, and female 6% luer conical fitting. Device connects to male luer fitting for fluid administration or blood sampling. Used by healthcare professionals in clinical settings for subcutaneous access. Provides physical conduit for fluid transfer between patient and external medical devices.

## Clinical Evidence

Bench testing only. Performance testing demonstrated compliance with ISO 594 (conical fittings) and ISO 7864 (sterile hypodermic needles). Biocompatibility testing confirmed compliance with ISO 10993 for limited exposure, indirect blood path, external communicating devices.

## Technological Characteristics

Stainless steel needle, needle hub, extension tubing, clamp, female luer fitting with end cap. Complies with ISO 594 (fittings) and ISO 7864 (needles). Biocompatibility per ISO 10993 for limited exposure, indirect blood path, external communicating device.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Nipro Scalp Vein Set ([K955053](/device/K955053.md))
- SURFLO Winged Infusion Set ([K771204](/device/K771204.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070801

# MAY 2 4 2007

PAGE 1 of 2

Pre-market Notification for IMM Hypodermic Needle Set

# IMM Hypodermic Needle Set

## 510(k) Summary of Safety and Effectiveness

# 1. Submitter's Name:

INNOVATIVE MEDICAL MAUFACTURING COMPANY 107, 181 Lane, Sect. 1 Yong Jane Road Chunan, Miaoli, 350 TAIWAN (ROC)

Contact: J. P. Lee, General Manager

## 2. Name of Device

| Common/Usual Name:   | Hypodermic Needle Set            |
|----------------------|----------------------------------|
| Proprietary Name:    | IMMTM Hypodermic Needle Set      |
| Classification Name: | Needle, hypodermic, single lumen |

## 3. Predicate Device

| Trade Name                 | 510(k) Number | Decision Date |
|----------------------------|---------------|---------------|
| Nipro Scalp Vein Set       | K955053       | 01/04/1996    |
| SURFLO Winged Infusion Set | K771204       | 07/14/1977    |

#### 4. Device Description

The IMM™ Hypodermic Needle Set consists of a conventional stainless steel needle, needle hub, extension tubing, clamp, and standard female 6% luer conical fitting for connection to a device contains a male luer fitting.

## 5. Indication for Use

The IMM™ Hypodermic Needle Set is to be used for sampling blood from or infusing fluid into body system underneath the skin.

# 6. Technological Characteristics

The IMMTM Hypodermic Needle Set has a conventional stainless steel needle, needle hub, extension tubing, clamp, female luer fitting with an end cap. The end cap can be removed to allow for connection of the female luer fitting to a device contains a male luer fitting for sampling or administering fluid.

{1}------------------------------------------------

1000 - 1000

.

:

and the state of the state

:

. . . .

1000

1. 1.

:

:

# 7. Performance Summary

: 上一篇:

. . . . . . . .

:

陆儿

1

2007 - 100 - 100 - 100 - 100

:

The functional and performance tests demonstrated that the IMM™ Hypodermic Needle Set meets specific requirements established in voluntary standards: ISO 594, and ISO 7864. Biocompatibility tests indicated that the device meets the requirements per ISO 10993 for "Iimited exposure, indirect blood path, external communicating device". 

and the comments of the comments of

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Innovative Medical Manufacturing Company C/O Mr. Joseph J. Chang, PhD, P.E. Innomedtech LLC 7128 Staffordshire Street Houston, Texas 77030

MAY 2 4 2007

Re: K070801

Trade/Device Name: IMM™ Hypodermic Needle Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: March 16, 2007 Received: March 23, 2007

Dear Dr. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Ko70801

# 510(k) Number (if known):

Device Name: IMM™ Hypodermic Needle Set

Indications For Use:

The IMM Hypodermic Needle Set is to be used for sampling blood from or infusing fluid into body system underneath the skin.

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Chas D. mint

(k) Number _ Kong841

i Sign-Off) of Anesthesiology, General Hospital, tion Control, Dental Devices

Page 1 of

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K070801](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FPA/K070801)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).