← Product Code [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ) · K993691

# ARROWG +ARD BLUE PLUS ANTIMICROBIAL MULTI-LUMEN CENTRAL VENOUS CATHETER (K993691)

_Arrow Intl., Inc. · FOZ · Mar 8, 2000 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K993691

## Device Facts

- **Applicant:** Arrow Intl., Inc.
- **Product Code:** [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ.md)
- **Decision Date:** Mar 8, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The multiple-lumen catheter permits venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. The ARROWg ard Blue Plus™ antimicrobial catheter is intended to help provide protection against catheter-related infections. It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use.

## Device Story

Triple-lumen, 7 French polyurethane central venous catheter; features antimicrobial coating of chlorhexidine acetate and silver sulfadiazine on external surface and internal lumens. Used for central venous access via femoral, jugular, or subclavian veins; provides protection against catheter-related infections. Operated by clinicians in clinical settings. Output is venous access for fluid/medication delivery, blood sampling, and pressure monitoring. Benefits patient by reducing risk of catheter-related infections during short-term therapy.

## Clinical Evidence

Bench testing included in vitro efficacy (zone of inhibition, internal lumen adherence, release rate), tensile testing, fatigue life, stability, and biocompatibility. In vivo testing included a swine safety study, half-life study, and delayed inoculum study. Results demonstrate the device is as safe as and more effective than the predicate.

## Technological Characteristics

Polyurethane triple-lumen catheter, 7 French. Antimicrobial agents: chlorhexidine acetate and silver sulfadiazine. Impregnated on external surface and internal lumens (catheter body, extension lines, and hubs).

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- ARROWg+ard Blue® Antimicrobial Multi-Lumen Central Venous Catheter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K993691

510(k) Premarket Notification ARROWg†ard Blue Plus™ Antimicrobial Central Venous Catheter Page 52 of 70

## Section 13 - 510(k) Summary

### a) Submitter

ARROW International. Inc. 2400 Bernville Road Reading, PA 19605

Contact Person: Thomas D. Nickel Vice President, Regulatory Affairs and Quality Assurance Direct phone: 610-478-3137 Fax: 610-478-3172 e-mail: tom nickel@arrowintl.com

October 28, 1999 Date summary prepared:

#### b) Device

ARROWg ard Blue Plus™ Multi-Lumen Central Venous Catheter Trade Name:

Antimicrobial Central Venous Catheter and Antimicrobial CVC Common Name:

Classification Name: Class II at 21 CFR 880.5200, Intravascular Catheter

#### c) Legally marketed device to which the device is substantially equivalent

ARROWq*ard Blue® Antimicrobial Multi-Lumen Central Venous Catheter

#### d) Description of the device

The device is a triple-lumen, polyurethane catheter, 7 French in size, with three independent non-communicating lumens, extension lines, Luer hubs and slide clamps. It is identical in appearance and function to the ARROW predicate catheter except for the increased amount of chlorhexidine acetate and silver sulfadiazine on the external surface and the addition of the internal lumen chlorhexidine and chlorhexidine acetate impregnation.

#### e} Intended use of the device

The multiple-lumen catheter permits venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. The ARROWg ard Blue Plus™ antimicrobial catheter is intended to help provide protection against catheter-related infections. It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use.

#### Technological characteristics f)

The device has the same exact technological characteristics as the predicate, with the only differences being the increased concentration of chlorhexidine acetate and silver sulfadiazine to the catheter body outer surface and the impregnation of chlorhexidine and chlorhexidine acetate to the catheter body, extension line, and extension line hubs internal lumens.

{1}------------------------------------------------

## 510(k) Premarket Notification ARROWg*ard Blue Plus™ Antimicrobial Central Venous Catheter Page 53 of 70

The performance tests included in the submission include:

- In vitro efficacy -- zone of inhibition 1)
- In vitro efficacy internal lumen adherence 2)
- 3) In vitro performance release rate test
- 4) Tensile tests
	- a) blue tip to catheter body
	- b) catheter body tensile
	- c) catheter body to juncture hub
	- d) extension line to juncture hub
	- e) extension line tensile
	- extension line to extension line hub f) f f f f f f f f f f f
- 5) Fatigue life testing
- 6) Stability tests
- 7) Biocompatibility tests

The in vivo performance tests included in the submission include:

- 1) In vivo swine safety study
- 2) In vivo half-life study
- In vivo delayed inoculum study 3)

The results of the laboratory tests demonstrate that the device is as safe as and is more effective than the legally marketed predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol. The logo is presented in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

**MAR - 8 2000**

Mr. Thomas D. Nickel Vice President, Regulatory Affairs and Quality Assurance Arrow International, Incorporated 2400 Bernville Road Reading, Pennsylvania 19605

K993691 Re:

> Trade Name: ARROWg+ard Blue Plus™ Antimicrobial Multi-Lumen Central Venous Catheter Kit Regulatory Class: II Product Code: FOZ Dated: January 3, 2000 Received: January 4, 2000

Dear Mr. Nickel:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the You may, therefore, market the device, subject to the Act). general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Please note: If you purchase your device Device Amendments. components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal

{3}------------------------------------------------

Page 2 - Mr. Nickel

Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, the Food and Drug Administration (FDA) action. may publish further announcements concerning your device in Please note: this response to your the Federal Register. premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

In addition, we have determined that your device kit contains the following: three swabstick foil pack 10% povidone-iodine and one 5 mL ampule HCL, 1% Lidocaine solution which are subject to regulation as drugs.

Our substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:

Director, Division of Druq Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the 

{4}------------------------------------------------

Page 3 - Mr. Nickel

regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on nour responsibilities under the Act, may be obtained from the your reopeners as Manufacturers Assistance at its toll free bivieror (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/qsma/dsmamain.html".

Sincerely,

Timothy A. Ulatowski

Timo thv Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

P.O. Box 12888 Reading, PA 19612

Image /page/5/Picture/1 description: The image shows the logo for Arrow International. The word "ARROW" is in large, bold, sans-serif font. Below it, in a smaller font, is the word "INTERNATIONAL". The logo is in black and white.

2400 Bernville Road Reading, PA 19605

(610) 378-0131 FAX: (610) 374-5360

# Section 14 -- Indications for Use

The ARROWg ard Blue Plus™ antimicrobial catheter is indicated in the short-term (<30 day) The ARROWN are Blue Frice - Blue Frice - and access, including, but not limited to the following:

- multiple peripheral sites for IV access ●
- lack of usable peripheral IV sites ●
- central venous pressure monitoring ●
- total parenternal nutrition (TPN) �
- incompatible medications ●

11 993691

- multiple infusions of fluids, medications, or chemotherapy .
- frequent blood sampling or receiving blood transfusions/blood products .
- infusions that are hypertonic, hyperosmolar, or infusions that have divergent pH values

There are no specific CDC guidelines for maximum indwelling times or catheter exchanges. There are the Spoolite Ober guing or should be exchanged per hospital protocol.

Vick Hilliard for Pat Ciarrocchi

(Division Sign-0 Division of Dental, Infection Control, and General Hospital D . Olk) Number K

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K993691](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K993691)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
