BD Nexiva and BD Nexiva Diffusics Closed IV Catheter Systems

K161777 · Becton Dickinson Infusion Therapy Systems, Inc. · FOZ · Aug 29, 2016 · General Hospital

Device Facts

Record IDK161777
Device NameBD Nexiva and BD Nexiva Diffusics Closed IV Catheter Systems
ApplicantBecton Dickinson Infusion Therapy Systems, Inc.
Product CodeFOZ · General Hospital
Decision DateAug 29, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5200
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

BD Nexiva closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed. BD Nexiva Diffusics closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa) when access ports not suitable for use with power injectors are removed.

Device Story

Over-the-needle peripheral intravascular catheter systems; include radiopaque Vialon catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer adapter, and vent plug. Nexiva Diffusics variant includes side holes at catheter tip to optimize power injection. Features BD Instaflash technology for immediate blood visualization. Closed system design contains blood during insertion; septum wipes needle surface upon withdrawal to reduce exposure risk; passive needle tip protection reduces needlestick injury risk. Used in clinical settings by healthcare providers for vascular access. Output is physical access to patient's vascular system for fluid/medication delivery, blood sampling, or pressure monitoring. Benefits include reduced blood exposure, improved insertion visualization, and power injection capability.

Clinical Evidence

Bench testing only. Performance testing included air leakage, liquid leakage, fluid leakage at venous pressure, septum assembly failure/separation at max pressure, cannula/hub bond strength, needle hub/tube union strength, and needle removal force (system drag). All pre-determined acceptance criteria were met.

Technological Characteristics

Over-the-needle peripheral IV catheter. Materials: Vialon catheter, copolyester septum canister, polyisoprene septum. Features: integrated extension set, passive needle shielding, Instaflash flashback visualization. Diffusics model includes tip side holes for power injection. Connectivity: None. Sterilization: Not specified. Software: None.

Indications for Use

Indicated for any patient population requiring short-term vascular access for blood sampling, blood pressure monitoring, or fluid administration. Suitable for power injection (up to 300 psi for Nexiva; up to 325 psi for Nexiva Diffusics) when appropriate ports are used.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth. The graphic is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 29, 2016 Becton Dickinson Infusion Therapy Systems Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street North West Buffalo, Minnesota 55313 Re: K161777 Trade/Device Name: BD Nexiva™ and BD Nexiva™ Diffusics™ Closed IV Catheter Systems Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II, Product Code: FOZ Dated: August 12, 2016 Received: August 16, 2016 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Michael J. Ryan -S for Tina Kiang, Ph.D Acting Division Director Science and Policy Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br><b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | |-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number ( <i>if known</i> ) | | | Device Name | BD Nexiva Closed IV Catheter System | | Indications for Use (Describe) | BD Nexiva closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to<br>sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with<br>consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of<br>therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300<br>psi (2068 kPa) when access ports not suitable for use with power injectors are removed. | | Type of Use (Select one or both, as applicable) | <div> <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </div> <div> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 {3}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | |-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| |-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| **Indications for Use** | 510(k) Number (if known) | | |--------------------------------|-----------------------------------------------| | Device Name | BD Nexiva Diffusics Closed IV Catheter System | | Indications for Use (Describe) | | BD Nexiva Diffusics closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa) when access ports not suitable for use with power injectors are removed. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label><br><label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | FORM FDA 3881 (8/14) | Page 1 of 1 | |----------------------|-------------| |----------------------|-------------| 40 {4}------------------------------------------------ bd.com Image /page/4/Picture/2 description: The image shows the BD logo. On the left is an orange circle with a white star-like shape inside. To the right of the circle are the letters "BD" in blue. The letters are large and bold. ## 510(k) Summary (21 CFR §807.92) ## BD Nexiva™ and BD Nexiva™ Diffusics™ Closed IV Catheter Systems | Submitter<br>Information | Submitter Name:<br>Submitter Address: | Becton Dickinson Infusion Therapy Systems Inc.<br>9450 South State Street<br>Sandy, UT 84070 | Device<br>Description<br>( <i>BD Nexiva</i> ) | BD Nexiva closed IV catheter systems are over-the-needle, intravascular<br>catheters. These devices have a radiopaque BD Vialon™ catheter, needle, needle<br>shield, septum, stabilization platform, integrated extension tubing, clamp, Luer<br>adapter, vent plug and, with dual port configurations, a pre-attached BD Q-Syte<br>Luer Access Split Septum. The needle and catheter are protected by a needle<br>cover. For dual port configurations, a BD Q-Syte device or an end cap with<br>protective cover is provided in the unit package. | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Contact Person: | Kimberly Geisler<br>Staff Regulatory Affairs Specialist<br>kimberly.geisler@bd.com<br>(801) 565-2422 (phone) | | The closed system is designed to keep blood contained within the device<br>throughout the insertion process. The septum is designed to wipe visible blood<br>from the needle surface as the needle is withdrawn from the catheter, further<br>reducing the risk of blood exposure. The needle tip is passively protected when<br>the needle is removed, reducing the risk of accidental needlestick injury. | | | Date of Preparation: | July 28, 2016 | | These devices have BD Instaflash™ Needle Technology, allowing for immediate<br>visualization of blood along the catheter. Continuous blood return is seen in the<br>extension tubing. The vent plug prevents blood leakage from the extension tubing<br>during insertion. Both the stabilization platform and Luer connector are color coded<br>to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue,<br>20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green). | | Subject Device<br>(BD Nexiva) | Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | BD Nexiva™ Closed IV Catheter System<br>Peripheral Intravascular or IV Catheter<br>21 CFR §880.5200<br>Intravascular Catheter<br>II<br>FOZ<br>General Hospital | Device<br>Description<br>( <i>BD Nexiva<br/>Diffusics</i> ) | BD Nexiva Diffusics closed IV catheter systems are over-the-needle, intravascular<br>catheters. These devices have a radiopaque BD Vialon™ catheter with side holes<br>located near the tip of the catheter which are designed to optimize power injection<br>procedures. These devices also have a needle, needle shield, septum,<br>stabilization platform, integrated extension tubing, clamp, Luer connector, and vent<br>plug. The Luer connector displays gauge-specific maximum flow rate and the<br>maximum power injector pressure limit setting. The needle and catheter are<br>protected by a needle cover. An end cap with protective cover is provided in the<br>unit package. | | Predicate<br>Device<br>(BD Nexiva) | Trade Name:<br>510(k) Reference:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | BD Nexiva™ Closed IV Catheter System<br>K102520<br>Peripheral Intravascular or IV Catheter<br>21 CFR §880.5200<br>Intravascular Catheter<br>II<br>FOZ<br>General Hospital | | The closed system is designed to keep blood contained within the device<br>throughout the insertion process. The septum is designed to wipe visible blood<br>from the needle surface as the needle is withdrawn from the catheter, further<br>reducing the risk of blood exposure. The needle tip is passively protected when<br>the needle is removed, reducing the risk of accidental needlestick injury. | | Subject Device<br>(BD Nexiva<br>Diffusics) | Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | BD Nexiva™ Diffusics™ Closed IV Catheter System<br>Peripheral Intravascular or IV Catheter<br>21 CFR §880.5200<br>Intravascular Catheter<br>II<br>FOZ<br>General Hospital | | These devices have BD Instaflash™ Needle Technology, allowing for immediate<br>visualization of blood along the catheter. Continuous blood return is seen in the<br>extension tubing. The vent plug prevents blood leakage from the extension tubing<br>during insertion. The stabilization platform and Luer connector are color coded to<br>indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA<br>(1.1 mm)=Pink, 18GA (1.3 mm)=Green). | | Predicate<br>Device<br>(BD Nexiva<br>Diffusics) | Trade Name:<br>510(k) Reference:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | BD Nexiva™ Diffusics™ Closed IV Catheter System<br>K123734<br>Peripheral Intravascular or IV Catheter<br>21 CFR §880.5200<br>Intravascular Catheter<br>II<br>FOZ<br>General Hospital | | | {5}------------------------------------------------ bd.com {6}------------------------------------------------ bd.com | Indications for<br>Use<br>(BD Nexiva) | Modifications were made to the Indications for Use for the purpose of aligning the<br>Indications for Use across the BD Nexiva family of products and for clarification.<br><br>Modifications include: 1) removal of the regulation reference as it is not relevant to<br>the healthcare practitioner; 2) removal of device features and performance<br>information which does not determine intended use; 3) removal of "less than 30<br>days" since this statutory regulation language for 21 CFR §880.5200 regarding the<br>device classification could be confusing to end users as the duration of use is<br>dependent on a number of factors, including patient status, duration of treatment,<br>and institutional protocol; 4) addition of "fluids being infused" and "duration of<br>therapy" as considerations for clinicians to evaluate prior to use of these catheters<br>for a particular patient; and 5) minor wording changes. | |---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| None of these changes are critical to the intended use and do not raise new questions of safety and effectiveness of the devices when used as indicated. | SUBJECT | PREDICATE (K102520) | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | BD Nexiva™ Closed IV Catheter System | BD Nexiva™ Closed IV Catheter System | | Indications for Use | Indications for Use | | BD Nexiva closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed. | As indicated in 21 CFR Part 880.5200, The Nexiva™ intravascular catheter is inserted into a patient's vascular system for a short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature and luer access port, aid in the prevention of needle-stick injuries. Blood is contained within the device during the catheter insertion process aiding in the prevention of blood exposure. This catheter may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure. The 18-22 gauge Nexiva™ catheters are suitable for use with power injectors rated for a maximum of 300 psi when the luer access port(s) is removed and a direct connection is made. | {7}------------------------------------------------ | bd.com | | |--------|--| | | | | | Indications for<br>Use<br>(BD Nexiva<br>Diffusics) | Modifications were made to the Indications for Use for the purpose of aligning the<br>Indications for Use across the BD Nexiva family of products and for clarification.<br>Modifications include: 1) addition of "short term use" in alignment with the device<br>classification; 2) addition of the intended patient population, which is consistent<br>with the predicate device, which does not indicate a specific patient population<br>and is intended for general use; 3) removal of the flow rate information as it<br>pertains to device performance and does not determine intended use (Note: flow<br>rate information is provided in the Instructions for Use); and 4) minor wording<br>changes. None of these changes are critical to the intended use and do not raise<br>new questions of safety and effectiveness of the devices when used as indicated. | | | |--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--| | | SUBJECT<br>BD Nexiva™ Diffusics™ Closed IV<br>Catheter System<br>Indications for Use | PREDICATE (K123734)<br>BD Nexiva™ Diffusics™ Closed IV<br>Catheter System<br>Indications for Use | | | | | BD Nexiva Diffusics closed IV catheter<br>systems are intended to be inserted into a<br>patient's vascular system for short term use<br>to sample blood, monitor blood pressure, or<br>administer fluids. These devices may be<br>used for any patient population with<br>consideration given to adequacy of vascular<br>anatomy, procedure being performed, fluids<br>being infused, and duration of therapy.<br>These devices are suitable for use with<br>power injectors set to a maximum pressure<br>of 325 psi (2240 kPa) when access ports not<br>suitable for use with power injectors are<br>removed. | The BD Nexiva™ Diffusics™ intravascular<br>catheter is inserted into a patient's<br>vascular system to sample blood, monitor<br>blood pressure or administer fluids. The<br>BD Nexiva Diffusics catheters are suitable<br>for use with power injectors when a direct<br>connection is made. The maximum flow<br>rate and maximum power injector pressure<br>setting for each catheter size are listed in<br>the table below: | | | | | | Max Flow Rate<br>(mL/sec) | Max Injector Setting<br>(psi) | | | | 24 GA x 0.75 IN | 3.0 | 325 | | | | 22 GA x 1.00 IN | 6.5 | 325 | | | | 20 GA x 1.00 IN | 10.0 | 325 | | | | 20 GA x 1.25 IN | 10.0 | 325 | | | | 18 GA x 1.25 IN | 15.0 | 325 | | | | | | | | | | Technological<br>Characteristics | Technological characteristics of the subject and predicate devices are equivalent.<br>The subject BD Nexiva™ and BD Nexiva™ Diffusics™ Closed IV Catheter<br>Systems achieve their intended use based on the same technology and principles<br>of operation as the predicate BD Nexiva™ and BD Nexiva™ Diffusics™ Closed IV<br>Cather Systems, respectively. Design modifications were made to the subject<br>devices to reduce the force required to withdraw the needle from the catheter<br>assembly. Performance testing performed to support the design modifications<br>included:<br>• Air Leakage<br>• Liquid Leakage<br>• Fluid Leakage at Venous Pressure<br>• Septum Assembly Failure / Separation at Max Pressure<br>• Cannula / Hub Bond Strength<br>• Strength of Union between Needle Hub and Needle Tube<br>• Force to Remove Needle from Catheter System (Average System Drag)<br>• System Drag - Peak Offset Force<br>All of the pre-determined acceptance criteria were met. | | | {8}------------------------------------------------ ## bd.com | Attribute | SUBJECT<br>BD Nexiva™ Closed IV Catheter System | PREDICATE (K102520)<br>BD Nexiva™ Closed IV Catheter System | |--------------------------------------------|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Classification | Same as predicate | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter | | Fundamental<br>Scientific<br>Technology | Same as predicate | Closed peripheral intravascular catheter<br>system designed with an integrated extension<br>set incorporating either a single port or Y<br>(dual)-port injection site. Incorporates BD<br>Instaflash™ technology to assist with<br>flashback visualization. | | Components /<br>Materials | Septum Canister:<br>Copolyester + grey<br>colorant* | Septum Canister:<br>Copolyester + white<br>colorant | | | All other subject component materials are the same as the predicate device | | | Physical /<br>Mechanical<br>Specifications | Same as predicate | Catheter Diameters<br>Catheter Lengths<br>18G, 20G, 22G, 24G<br>0.56", 0.75", 1.00",<br>1.25", 1.75" | *The subject BD Nexiva™ Closed IV Catheter System and subject BD Nexiva™ Diffusics™ Closed IV Catheter System septum canister materials are identical. As such, the Nexiva™ Closed IV Catheter System (K123734) serves as the reference device supporting this modification to the BD Nexiva™ Closed IV Catheter System non-patient contacting septum canister material. | Attribute | SUBJECT<br>BD Nexiva™ Diffusics™ Closed IV<br>Catheter System | PREDICATE (K123734)<br>BD Nexiva™ Diffusics™ Closed IV Catheter<br>System | |----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Classification | Same as predicate | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter | | Fundamental<br>Scientific<br>Technology | Same as predicate | Closed peripheral intravascular catheter<br>system designed with an integrated extension<br>set incorporating either a single port or Y<br>(dual)-port injection site. Incorporates BD<br>Instaflash™ technology to assist with<br>flashback visualization.<br>In addition, the catheter tip includes diffuser<br>holes to reduce the velocity of contrast media<br>exiting the catheter tip during CT scans. | | Components /<br>Materials | Septum Canister: Copolyester + grey<br>colorant (modified<br>resin to colorant ratio<br>compared to predicate)<br><br>Septum: Polyisoprene (no<br>lubricant) | Septum Canister: Copolyester + grey<br>colorant<br><br>Septum: Polyisoprene + Silicone<br>Lubricant | | All other subject component materials are the same as the predicate device | | | | Physical /<br>Mechanical<br>Specifications | Same as predicate | Catheter Diameters Catheter Lengths<br>18G, 20G, 22G, 24G 0.75", 1.00", 1.25" | {9}------------------------------------------------ bd.com | Summary of<br>Performance<br>Tests | The following performance testing was performed to demonstrate substantial<br>equivalence of the subject BD NexivaTM and BD NexivaTM DiffusicsTM Closed IV<br>Catheter Systems to the predicate BD NexivaTM Closed IV Catheter System<br>(K102520) and BD NexivaTM DiffusicsTM Closed IV Catheter System (K123734),<br>respectively, and applicable standards: | |------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Air Leakage Liquid Leakage Fluid Leakage at Venous Pressure Septum Assembly Failure / Separation at Max Pressure Cannula / Hub Bond Strength Strength of Union between Needle Hub and Needle Tube Force to Remove Needle from Catheter System (Average System Drag) System Drag - Peak Offset Force All of the pre-determined acceptance criteria were met. | | Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and performance<br>testing, the subject BD NexivaTM Closed IV Catheter System and BD NexivaTM<br>DiffusicsTM Closed IV Catheter System are demonstrated to be substantially<br>equivalent to the predicate BD NexivaTM Closed IV Catheter System (K102520)<br>and BD NexivaTM DiffusicsTM Closed IV Catheter System (K123734), respectively. |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...