K161681 · Nipro Medical Corporation · FOZ · Mar 8, 2017 · General Hospital
Device Facts
Record ID
K161681
Device Name
Nipro SafeTouch IV Catheter
Applicant
Nipro Medical Corporation
Product Code
FOZ · General Hospital
Decision Date
Mar 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nipro SafeTouch IV Catheter is a passive anti-needle stick intra-vascular catheter that is inserted into the patient's vascular system for short term use for the infusion of fluids, and/or blood components. 16-24 gauge catheters may be used with power injectors for which the maximum pressure setting is 300psi
Device Story
Nipro SafeTouch IV Catheter is an over-the-needle, radiopaque polyurethane catheter; features passive safety mechanism engaging upon removal from patient to prevent accidental needlestick injuries. Available in 16-24 gauges, lengths 3/4"–2", with or without wings. Used by clinicians for short-term venous or arterial access; enables infusion of fluids/blood components. Compatible with power injectors up to 300psi. Device is single-use, sterile (EtO), and color-coded for gauge identification. Passive safety mechanism reduces risk of sharps injury to healthcare providers during catheter removal.
Clinical Evidence
No clinical data provided. Bench testing only. Performance validated via simulated clinical study of safety mechanism (zero failures), dimensional/performance testing, biocompatibility (ISO 10993), sterilization validation (ISO 11135), and shelf-life testing (ASTM F1980).
Technological Characteristics
Materials: Polyurethane (PUR), Polypropylene (PP), Polycarbonate (PC), Stainless Steel. Passive safety mechanism. Sterilization: Ethylene Oxide (ISO 11135). Biocompatibility: ISO 10993. Packaging: ISO 11607-2. Shelf-life: 5 years (ASTM F1980). Connectivity: None. Energy source: None.
Indications for Use
Indicated for patients requiring short-term vascular access for the infusion of fluids and/or blood components via an intravascular catheter.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2017
Nipro Medical Corporation % Jessica Oswald-McLeod Director, Quality Assurance & Regulatory Affairs 3150 NW 107th Ave. Miami, Florida 33172
Re: K161681
Trade/Device Name: Nipro SafeTouch IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: June 14, 2016 Received: June 17, 2016
Dear Jessica Oswald-McLeod:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Lori A. Wiggins -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
#### K161681
Device Name Nipro SafeTouch IV Catheter
#### Indications for Use (Describe)
The Nipro SafeTouch IV Catheter is a passive anti-needle stick intra-vascular catheter that is inserted into the patient's vascular system for short term use for the infusion of fluids, and/or blood components. 16-24 gauge catheters may be used with power injectors for which the maximum pressure setting is 300psi
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary: Nipro SafeTouch IV Catheter K161681
807.92(a)(1)
Applicant:
| Applicant: | Nipro Medical Corporation<br>3150 NW 107th Ave.<br>Miami FL 33172<br>Tel: 305-599-7174 |
|---------------------|----------------------------------------------------------------------------------------|
| Establishment Reg.: | 1056186 |
| Contact Person: | Jessica Oswald-McLeod<br>Director, Quality Assurance & Regulatory Affairs |
Date of summary preparation: March 7, 2017
#### 807.92(a)(2)
Trade Name: Nipro SafeTouch IV Catheter Common Name: Safety IV Catheter Classification Name: catheter, intravascular Regulation Number: 21 CFR 880.5200 Panel: 80 Product Code: FOZ
#### 807.92(a)(3)
Legally marketed substantial equivalent device: B Braun Introcan Safety IV Cath (K020785)
#### 807.92(a)(4)
#### Description of device:
Nipro SafeTouch IV Catheter consists of an over-the-needle, catheter made of radiopaque polyurethane with a passive safety mechanism that will engage upon removal from the patient in order to prevent accidental needlestick injuries. The device is available in gauges 16-24 and in catheter lengths of ¾" – 2". It is also available with or without wings on the hub. Gauges are color coded for easy differentiation. Nipro SafeTouch IV Catheter is sterilized by Ethylene Oxide gas (EtO). It is packaged in a peel blister that ensures the sterility of the device until package is opened and labeled for single use only. A sleeve of SafeTouch Cath contains 5 units of blister packages. An inner box contains 10 sleeves (50 units). An outer box contains 40 inner boxes (3000 units). Its shelf life period is 5 years.
#### 807.92(a)(5)
Indications for Use:
The Nipro SafeTouch IV Catheter is a passive anti-needle stick intra-vascular catheter that is inserted into the patient's vascular system for short term use for the infusion of fluids, and/or blood components. 16-24 gauge catheters may be used with power injectors for which the maximum pressure setting is 300psi
{5}------------------------------------------------
## 807.92(a)(6)
Comparison of technological characteristics:
The catheter is substantially equivalent to the predicate device in the following technological characteristics:
| Element of Comparison | Nipro SafeTouch IV Catheter | Predicate |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Materials | PUR, PP, PC, Stainless Steel | Available in two materials:<br>1. FEP, PP, ABS, chrome-<br>nickel steel<br>2. PUR, PP, ABS, chrome-<br>nickel steel |
| Design: Operational mode | Mechanical - Manually<br>operated IV catheter with a<br>passive safety mechanism | Same |
| Design: Physical<br>characteristics | Safety IV Catheter, with<br>passive anti-stick safety<br>mechanism. | Same |
| Design: Device<br>Configurations | Available in both winged and non-winged Needle gauges 16-24 catheter lengths of ¾" – 2" Color coded hubs for easy identification distal end: Luer connector proximal end: Cannula and catheter | Same |
| Intended Use | The Nipro SafeTouch IV<br>Catheter is a passive anti-<br>needle stick intra-vascular<br>catheter that is inserted into<br>the patient's vascular system<br>for short term use for the<br>infusion of fluids, and/or<br>blood components. 16-24<br>gauge catheters may be used<br>with power injectors for<br>which the maximum pressure<br>setting is 300psi | Passive anti-needle stick<br>device for venous or arterial<br>access for the infusion of<br>fluids, drugs, and/or blood<br>components. 14-22 gauge<br>catheters may be used with<br>power injectors for which<br>the maximum pressure<br>setting is 300psi |
| Sterilization and Expiry<br>dating | Method: Ethylene Oxide<br>Shelf-life: 5 years | Same |
| Energy Source | NA | NA |
Table 4: Comparison of technological characteristics
{6}------------------------------------------------
Non-clinical tests submitted are for dimensional and performance, sterilization and biocompatibility:
- 1. Disengagement force between components (internal test method)
- Penetration force (internal test method) 2.
- 3. Air leakage (internal test method)
- 4. Flow rate (internal test method)
- 5. Safety mechanism performance (simulated clinical study)
- Package integrity/microbial barrier (ISO 11607-2) 6.
- 7. Chemical testing / acid, alkalinity and extractable metals (ISO 7864:1993)
- Corrosion resistance (ISO 10555-1:1995) 8.
- Sterilization validation (ISO 11135-1:2007, ISO 11135-2:2008) 9.
- 10. Eto residuals (ISO 10993-7)
- 11. Transportation tests (ASTM-d4169-09)
- 12. Biocompatibility (ISO 10993)
- a. Cytotoxicity: MEM Elution
- b. Sensitization: Kligman Maximization
- ﻥ Intracutaneous Irritation
- d. Acute Systemic Toxicity: Systemic Injection
- e. Rabbit Pyrogen
- f. Hemolysis –Rabbit Blood – direct and indirect contact
- Complement Activation SC5b9 g.
- h. Complement Activation - СЗа
- i. Dog Thrombogenicity
- j. ASTM Partial Thromboplastin Time
- Genotoxicity: Bacterial Reverse Mutation Study k.
- 1. Mouse Peripheral Blood Micronucleus Study
- m. Two Week Rat Study, Repeated Parenteral Administration of Two Extracts
- n. Genotoxicity: Mouse Lymphoma Assay
- o. ISO Muscle Implantation Study 2 Week
- p. LAL testing
- q. Particulate matter testing
- 13. Pyrogen testing (USP 35-NF30:2012, <85>)
- 14. Shelf-life testing (ASTM F1980-07)
All of the above tests were performed within the specifications of the standards listed above.
#### 807.92(b)(2)
Clinical tests:
This submission does not warrant any clinical testing, therefore no clinical testing performed for or provided in this submission.
A simulated clinical test was performed to evaluate the effectiveness of the safety mechanism. This test passed with zero failures.
### 807.92(b)(3)
Conclusions drawn from non-clinical and clinical tests:
The results of the performance testing and the comparison of technological characteristics with the predicate device demonstrate that the Nipro SafeTouch IV Catheter performs equivalent to the predicate device when used as intended.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.