← Product Code [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ) · K160592

# SUPERCATH 6 (K160592)

_Togo Medikit Co., Ltd. · FOZ · Aug 12, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K160592

## Device Facts

- **Applicant:** Togo Medikit Co., Ltd.
- **Product Code:** [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ.md)
- **Decision Date:** Aug 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18-22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi

## Device Story

Intravascular catheter for venous/arterial access and fluid administration; single-use; short-term (<30 days). Features metallic introducer needle with side-notch for visual confirmation of vessel entry; built-in check valve in catheter hub reduces blood flashback during needle withdrawal; button-activated spring mechanism retracts needle into hub to prevent needlestick injuries. Includes slide component for catheter insertion assistance. Used by healthcare professionals in clinical settings. Output is vascular access; healthcare provider observes blood flashback in side-notch to confirm placement. Power injection capability (18-22G) allows use with contrast media injectors. Benefits include reduced needlestick risk and improved insertion control.

## Clinical Evidence

Bench testing only. Testing included tensile strength (ISO 10555-1), air/liquid leakage, flow rate, power injection burst pressure, check valve leakage, and wing tensile strength. Biocompatibility testing included cytotoxicity, sensitization, systemic toxicity, irritation, pyrogenicity, implantation, genotoxicity, and hemocompatibility. No clinical data provided.

## Technological Characteristics

Materials: Polyurethane (tube), Copolyester (hub), Polycarbonate (slide/covers), Stainless steel (needle), PVC (wing). Sensing: Side-notch needle for visual flashback. Actuation: Spring-loaded needle retraction. Connectivity: None. Sterilization: Ethylene oxide. Form factor: 14G-24G catheters. Standards: ISO 10555-1.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- SUPERCATH 5 ([K140419](/device/K140419.md))

## Reference Devices

- Introcan Safety® 3 Closed IV Catheter ([K111236](/device/K111236.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 12, 2016

TOGO MEDIKIT CO., LTD. c/o Izuni Maruo MIC International 4-1-17 Hongo Bunkyo-ku, Tokyo 113-0033 JAPAN

Re: K160592

Trade/Device Name: SUPERCATH 6 Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: II Product Code: FOZ Dated: July 13, 2016 Received: July 14, 2016

Dear Izumi Maruo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tina
Kiang -S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160592

Device Name SUPERCATH 6

#### Indications for Use (Describe)

The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18-22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary-K160592

# a. Owner/Company name, address

TOGO MEDIKIT CO., LTD. 17148-6 Aza Kamekawa, Oaza Hichiya Hyuga City, Miyazaki, 883-0062, Japan

Daisuke Nagamizu Manager, Regulatory Affairs Section, Quality Assurance Department Phone: 011-81-982-53-8000 011-81-982-53-8008 Fax: Email: qc@togomedikit.co.jp

# b. Contact/Application Correspondent

Izumi Maruo Senior Consultant MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan

Phone: 011-81-3-3818-8577 Fax: 011-81-3-3818-8573 maruo@mici.co.jp Email:

## c. Date prepared

February 29, 2016

# d. Name of device

| Trade Name:                | SUPERCATH 6                                                          |
|----------------------------|----------------------------------------------------------------------|
| Common Name:               | Intravascular Catheter                                               |
| Classification Name:       | Catheter, Intravascular, Therapeutic<br>Short-term less than 30 days |
| Classification Regulation: | 21 CFR 880.5200                                                      |
| Product Code:              | FOZ                                                                  |
| Classification Panel:      | General Hospital                                                     |

{4}------------------------------------------------

## e. Predicate/reference devices

The SUPERCATH 6 is substantially equivalent to the following legally marketed devices:

| Predicate device |             |
|------------------|-------------|
| 510(k):          | K140419     |
| Trade name:      | SUPERCATH 5 |
| Product code:    | FOZ         |

This submission uses a reference device for the power injection and changes in indications for use statement from the predicate device:

| Reference device |                                       |
|------------------|---------------------------------------|
| 510(k):          | K111236                               |
| Trade name:      | Introcan Safety® 3 Closed IV Catheter |
| Product code:    | FOZ                                   |

### f. Description of the device

The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days) and is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 - 22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi.

The SUPERCATH 6 has technological characteristics as follows;

Its catheter hub has a built-in check valve which together with the healthcare professional's finger pressure on the blood vessel, assists to reduce blood flashback when the metallic introducer needle is withdrawn following blood vessel puncture. The built-in check valve is not intended to stop bleeding completely.

Pressing the button on the needle hub activates the coiled spring in the hub, retracting the metallic introducer needle into the needle hub.

The introducer needle of the SUPERCATH 6 has a side-notch to provide rapid visual confirmation of vessel entry. When the introducer needle is inserted into the vein, blood flows upward into side-notch and returns down along the inside of the catheter tube.

The SUPERCATH 6 has a slide in order to assist insertion of the catheter. The user can choose either to push the catheter hub forward or push the slide forward to insert the catheter. When pushing the slide, the slide is stopped 12 mm forward by a stopper.

The SUPERCATH 6 is available in 14G, 16G, 18G, 20G, 22G and 24G.

## g. Indications for Use

The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days), is

{5}------------------------------------------------

designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 -22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi.

## h. Statement of substantial equivalence

The SUPERCATH 6 is the modified version of the SUPERCATH 5 (K140419). The SUPERCATH 6 is intended to be used with power injectors. The intended use of the SUPERCATH 6 and the SUPERCATH 5 (K140419) differ only in the use for power injection.

Regarding the reference device, the SUPERCATH 6 has a similar intended use to the Introcan Safety® 3 Closed IV Catheter (K111236). Both devices act as an intravascular catheter for short term use (less than 30 days) and the same gauge sizes are suitable for use with power injectors at a maximum pressure of 300 psi.

The SUPERCATH 5 (K140419) and the SUPERCATH 6 have the identical needlestick injury prevention feature, flashback visualization system, side-notch needle, and reduction of blood exposure system.

The comparison table for technological characteristics among the SUPERCATH 6, the SUPERCATH 5 (K140419), and the Introcan Safety® 3 Closed IV Catheter (K111236) is shown below.

| Item                                  | SUPERCATH 6                                                                                                                                                                                                                                                                                                                                                   | The predicate device; SUPERCATH (K140419)                                                                                                                                                                                                                       | The reference device; Introcan Safety® 3 Closed IV Catheter (K111236)                                                                                                                                                                                                                                                                           |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use                          | The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 -22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi. | The SUPERCATH 5 is intended to access a vein or artery and to administer fluids. The SUPERCATH 5 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. | Introcan Safety® 3 Closed IV Catheter is inserted into a patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure or administer fluids and blood intravascularly. The 18-22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi with a luer lock connection only. |
| Needlestick injury prevention feature | The metallic introducer needle is                                                                                                                                                                                                                                                                                                                             | The metallic introducer needle is                                                                                                                                                                                                                               | The passive safety needle-shielding                                                                                                                                                                                                                                                                                                             |

Comparison table for technological characteristics

{6}------------------------------------------------

|                                           | retracted into the<br>needle hub when the<br>activation button on<br>the needle hub is<br>pressed. | retracted into the<br>needle hub when the<br>activation button on<br>the needle hub is<br>pressed. | mechanism of the<br>Introcan Safety® 3 is<br>located inside the<br>catheter hub. Upon<br>withdrawal of the<br>needle, the safety<br>shield engages as the<br>needle passes through<br>the catheter hub and<br>deploys automatically<br>to shield the needle<br>tip. |
|-------------------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Flashback<br>visualization                | Yes                                                                                                | Yes                                                                                                | Yes                                                                                                                                                                                                                                                                 |
| Reduction of<br>blood<br>exposure         | Yes (Check valve*¹)                                                                                | Yes (Check valve*¹)                                                                                | Yes (Septum)                                                                                                                                                                                                                                                        |
| Side-notch needle*²                       | Yes                                                                                                | Yes                                                                                                | No                                                                                                                                                                                                                                                                  |
| Radiopaque                                | Yes                                                                                                | Yes                                                                                                | Yes                                                                                                                                                                                                                                                                 |
| Use<br>with<br>power<br>injectors         | Yes (18-22 gauge)                                                                                  | No                                                                                                 | Yes (18-22 gauge)                                                                                                                                                                                                                                                   |
| Sterilization                             | Ethylene oxide                                                                                     | Ethylene oxide                                                                                     | Ethylene oxide                                                                                                                                                                                                                                                      |
| Packaging                                 | Tear-off pouch                                                                                     | Blister                                                                                            | Unknown                                                                                                                                                                                                                                                             |
| Multiple gauge sizes<br>and needle length | Yes                                                                                                | Yes                                                                                                | Yes                                                                                                                                                                                                                                                                 |

Note:

*1: Check valve assists to reduce blood flashback when the metallic introducer needle is withdrawn following blood vessel puncture together with the healthcare professional's finger pressure.

*2: Side-notch needle is the introducer needle with a side-notch, to provide rapid visual confirmation of vessel entry. When the introducer needle is inserted into the vein, blood flows upward into the side-notch and returns down along the inside of the catheter tube.

The modifications to the SUPERCATH 6 from the SUPERCATH 5 (K140419) include following changes of components and raw materials:

- . All materials of the catheter hubs of the SUPERCATH 6 are different from those of the SUPERCATH 5 (K140419).
- The SUPERCATH 6 has the slide and the slide cap as new components
- Composition of the lubricating oil on the catheter of the SUPERCATH 6 is different from ● that of the SUPERCATH 5 (K140419).
- Formulation of the PVC of the SUPERCATH 6 is different from that of the SUPERCATH 5 (K140419)
- The SUPERCATH 6 does not have an attachment which the SUPERCATH 5 (K140419) has. ●

Following table is a brief comparison of main components and materials between the SUPERCATH

{7}------------------------------------------------

6 and the SUPECATH 5 (K140419);

| Component name         | Materials       |                       |
|------------------------|-----------------|-----------------------|
|                        | SUPERCATH 6     | SUPERCATH 5 (K140419) |
| Inner needle           |                 |                       |
| -Introducer needle     | Stainless steel | Stainless steel       |
| -Outer and Inner cover | Polycarbonate   | Polycarbonate         |
| -Needle hub            | Polycarbonate   | Polycarbonate         |
| -Attachment            | -               | Polypropylene         |
| -Slide                 | Polycarbonate   | -                     |
| -Slide cap             | Polypropylene   | -                     |
| Catheter               |                 |                       |
| -Catheter tube         | Polyurethane    | Polyurethane          |
| -Lubricating oil       | Silicon         | Silicon               |
| -Catheter hub          | Copolyester     | Polycarbonate         |
| -Wing                  | PVC             | PVC                   |

Comparison table for components and materials

The attachment is used just during manufacturing process of the SUPERCATH 5 (K140419). The attachment is used in order to adjust the distance between the bevel of the introducer needle and the tip of the catheter. After the adjustment, the attachment is fixed and users cannot move the attachment. Although the SUPERCATH 6 does not have the attachment, the distance between the bevel of the introducer needle and the tip of the catheter is managed in our quality system. Therefore, the attachment does not affect substantial equivalence between the predicate device and the SUPERCATH 6.

Each gauge size of catheter hubs of the SUPERCATH 5 (K140419) and the SUPERCATH 6 is color-coded. However, all materials of the catheter hubs of the SUPERCATH 6 are different from those of the SUPERCATH 5 (K140419).

The SUPERCATH 6 has the slide and the slide cap as new components. User can choose to push the catheter hub forward or push the slide forward to insert the catheter. When pushing the slide, a stopper stops the slide 12 mm forward.

In order to evaluate any effects of the above changes on substantial equivalence, biocompatibility testing, sterilization, shelf-life testing, and performance testing were conducted. In conclusion, those testing demonstrated that the SUPERCATH 6 is substantially equivalent to the the SUPERCATH 5 (K140419).

#### Bench Testing i.

The following bench tests were performed to verify conformity to the recognized standards and demonstrate substantial equivalence to the SUPERCATH 5 (K140419).

# ● Tensile Strength for the Catheter

The tensile strength of the catheter meets acceptable minimum force until breakage when tested according to ISO 10555-1.

{8}------------------------------------------------

- Air and Liquid Leakage for the Hub Attachment
The catheter hub is impervious to air/liquid infiltration when subjected to positive pressure and aspiration when tested according to ISO 10555-1.

- · Flow rate
The flow rate through catheter meets allowable limits when tested according to ISO 10555-1.

- Power injection
The catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when tested according to ISO10555-1

- Leakage at the Check Valve under pressure
The built-in check valve is impervious to liquid infiltration when subjected to positive pressure according to in-house standard.

- Tensile Strength for the Wing The tensile strength of the wing meets acceptable minimum force until breakage when tested according to in-house standard.
#### j. Biocompatibility Testing

Because the modifications to the SUPERCATH 6 from the SUPERCATH 5 (K140419) include changes of components and raw materials, we performed following biocompatibility tests;

- Cytotoxicity
- · Sensitization
- Systemic (acute) toxicity
- Irritation/ Intracutaneous reactivity
- · Pyrogenicity
- Implantation
- Subchronic and subacute systemic toxicity
- · Genotoxicity
- · Hemocompatibility
- Physicochemical tests

In the biocompatibility testing reports, no biocompatibility concern was raised.

# k. Simulated Clinical Use

The SUPERCATH 6 modified from the SUPERCATH 5 (K140419) has no new sharps injury prevention features. Therefore, this submission does not include the simulated clinical use testing report.

#### l. Conclusion

Based on the above discussion and enclosed sections regarding substantial equivalence to predicate device, TOGO MEDIKIT concludes that the SUPERCATH 6 is substantially equivalent to the SUPERCATH 5 (K140419).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K160592](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K160592)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
