← Product Code [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ) · K123267

# SURFFLASH(R) PLUS SAFETY I.V. CATHETER (K123267)

_Kofu Factory of Terumo Corp. · FOZ · Nov 2, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K123267

## Device Facts

- **Applicant:** Kofu Factory of Terumo Corp.
- **Product Code:** [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ.md)
- **Decision Date:** Nov 2, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The SURFLASH® Plus Safety I.V. Catheter is inserted into the patient's vascular system for short term use (<30 days) to withdraw blood samples, administer fluid intravenously, or monitor blood pressure by attaching a monitoring line. The needle shaft cover and tip shield feature aids in the prevention of needle stick injuries. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. The 18 to 22 gauge catheters are suitable for use with power injectors rated for a maximum of 325 psi.

## Device Story

The SURFLASH® Plus Safety I.V. Catheter is a manually operated, over-the-needle peripheral catheter designed for short-term vascular access. It features a notched stainless steel cannula for visual flashback confirmation of vessel entry. Upon needle withdrawal, a passive safety mechanism (shutter and needle shaft cover) automatically shields the needle tip to prevent sharps injuries. The catheter hub includes a valve and seal to minimize blood leakage. Used by clinicians in clinical settings, the device allows for blood sampling, fluid administration, and pressure monitoring. The 18-22G models support power injection up to 325 psi. The device benefits patients by providing secure vascular access while reducing the risk of accidental needle sticks to healthcare workers.

## Clinical Evidence

No clinical trials were conducted. Evidence consists of bench testing and a simulated use study. Bench testing verified compliance with ISO 10555-1, ISO 10555-5, and ISO 23908, including tensile strength, burst pressure, flow rate, and biocompatibility (ISO 10993). The simulated use study confirmed that healthcare workers could effectively activate the sharps injury prevention feature and follow the instructions for use without prior training. All safety features activated as intended with no adverse events.

## Technological Characteristics

Over-the-needle peripheral catheter; polyurethane catheter body; stainless steel cannula; passive needle-shielding mechanism (shutter/cylinder design); integrated valve/seal in hub; radio-opaque; ethylene oxide sterilized (SAL 10^-6). Complies with ISO 10555-1, ISO 10555-5, and ISO 23908. Manual operation; no electronic components or software.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- TERUMO® Surshield®-PUR Safety I.V. Catheter ([K100282](/device/K100282.md))
- TERUMO® SURFLASH® I.V. Catheter ([K991406](/device/K991406.md))
- BD Insyte Autoguard BC shielded I.V. Catheter ([K110443](/device/K110443.md))
- BD Insyte Autoguard shielded I.V. Catheter ([K971339](/device/K971339.md))
- Saf-T-Intima Closed IV Catheter System ([K923702](/device/K923702.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K123267

Terumo Corporation TERUMO® SURFLASH® Plus Safety I.V. Catheter 510k II. 510(k) Summary

### 510(k) SUMMARY

NOV 2 2012

Prepared by Sandi Hartka Manager Regulatory Affairs Terumo Medical Corporation Tel. 410 392-7243 Fax 410 398 6079

Date prepared : September 11, 2012

Prepared for : Owner/Operator

> TERUMO CORP. 44-1, 2-Chome, Hatagaya Shibuya-Ku, Tokyo, JAPAN 151-0072 Registration Number: 8010026

### Manufacturer and Sterilization Facility

KOFU FACTORY OF TERUMO CORP. 1727-1, Tsuijiarai, Showa-Cho Nakakoma-Gun, Yamanashi, JAPAN 409-3853 Registration Number: 9681835

### Contact Person

SANDI HARTKA Regulatory Affairs Manager TERUMO MEDICAL CORP. 950 Elkton, Blvd. Elkton, MD 21921 Phone: 410 3927243 Ext Fax: 410 3986079 Sandi.hartka@terumomedical.com

{1}------------------------------------------------

erumo Corporation RUMO® SURFLASH® Plus Safety I.V. Catheter 510k II. 510(k) Summary

### A. DEVICE NAME

### Proprietary Name

SURFLASH® Plus Safety I.V. Catheter

### Classification Name

Intravascular Catheter

Intravascular Catheter (880.5200) Product Code: FOZ Panel: General Hospital Classification: Class II

### Common Name

Intravascular catheter with needle protection device

### B. INTENDED USE

The SURFLASH® Plus Safety I.V. Catheter is inserted into the patient's vascular system for short term use (<30 days) to withdraw blood samples, administer fluid intravenously, or monitor blood pressure by attaching a monitoring line. The needle shaft cover and tip shield feature aids in the prevention of needle stick injuries. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. The 18 to 22 gauge catheters are suitable for use with power injectors rated for a maximum of 325 psi.

# C. DEVICE DESCRIPTION (SUMMARY OF TECHNOLOGICAL CHARACTERISTICS)

### Principle of Operation/Technology

The SURFLASH® Plus Safety I.V. Catheter is manually operated and contains a passive safety feature that automatically activates as the needle is withdrawn from the catheter.

{2}------------------------------------------------

Terumo Corporation TERUMO® SURFLASH® Plus Safety I.V. Catheter 510k II. 510(k) Summary

### Design / Construction

The SURFLASH® Plus Safety I.V. Catheter are devices consisting of catheter assembly (catheter, caulking pin and catheter hub containing valve, seal component and plug), needle assembly (needle, needle housing, transparent flash chamber with filter and needle protector) and a passive needle-shielding mechanism (shutter and needle shaft cover consisting of inner cylinder, outer cylinder, junction cylinder).

The SURFLASH® Plus Safety I.V. Catheter has a valve and a seal component inside the catheter hub to minimize the possibility of blood leakage at catheter hub after needle removal. The catheter hub is also provided with a plug inside which penetrates the valve to create a fluid path when a connector is inserted in the catheter hub. The fluid path is permanently opened once a secure luer connection is made.

The devices are an over-the needle, peripheral catheter made of a slender, flexible, radio-opaque plastic with a hub that is inserted into the patient's vascular system for short term (<30 days) use to withdraw blood samples, administer fluids intravenously, or monitor blood pressure by attaching a monitoring line.

Whole length cannula including the sharp end of the inner needle is covered by the needle shielding mechanism as the needle is withdrawn from catheter's hub to aid in the prevention of needle stick injuries.

The grooved cannula which allows the clinician to visualize flashback detection through the groove indicating that there is confirmation of vessel entry is the same as the TERUMO® Surshield® PUR Safety I.V. Catheter (K100282) and TERUMO® SURFLASH® I.V. Catheter (K991406).

### Material

The stainless steel cannula is placed in the catheter to maintain rigidity and is withdrawn after the catheter is placed in the vascular system.

The catheter that is advanced into the vessel is made of polyurethane which is the same as the TERUMO® Surshield®-PUR Safety I.V. Catheter (K100282) and TERUMO® SURFLASH® I.V. Catheter (K991406).

{3}------------------------------------------------

### Specifications

The SURFLASH® Plus Safety I.V. Catheter are available in 6 sizes with combination of 18, 20, 22 and 24 gauge catheters diameters and 19, 25 and 32 mm catheters length.

### D. NON-CLINICAL TESTS

Performance testing was conducted to ensure the safety and effectiveness of the SURFLASH® Plus Safety I.V. Catheter throughout the shelf life, to verify conformity to the applicable part of ISO standards and demonstrate substantial equivalence to the predicate devices as mentioned in the table on next pages.

No new issues of safety and effectiveness were raised with the testing performed. Performance testing demonstrates that the SURFLASH® Plus Safety 1.V. Catheter conforms to the recognized consensus ISO standards, is substantially equivalent to the predicate devices and acceptable for clinical use throughout the shelf life.

A simulated use study was conducted in accordance with FDA's Guidance For Industry and Staff, Medical Devices with Sharps Injury Prevention Features issued on August 9. 2005. The objectives of this study were designed to confirm that study participants are able to read and follow the Instructions for Use to safely and effectively operate the safety feature of the device. The objectives were:

- 1. Verify that proper activation of the sharps injury prevention feature of the device can be accomplished in simulated clinical environments by healthcare workers who use these types of safety I.V. catheters.
- 2. Determine if the Instructions for Use are adequate for proper activation of the sharps injury prevention feature.

All safety features activated effectively and there were no de-activations after discarding. All participants were able to read and activate the catheter safety feature without further explanation or training. No adverse events occurred.

Conclusion: The safety feature of the device operates safely and effectively as intended. Users with various levels of experience can read the IFU and operate the device safely and effectively.

{4}------------------------------------------------

# Terumo Corporation TERUMO® SURFLASH® Plus Safety I.V. Catheter Response to Comments dated September 20, 2012

Attachment 2

| Performance test                                                                                                                                         | Testing           |              |              |                                    |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--------------|--------------|------------------------------------|
|                                                                                                                                                          | Compliance to ISO |              |              |                                    |
|                                                                                                                                                          | ISO 10555-1       | ISO 10555-5  | ISO 23908    | Testing by<br>internal<br>standard |
| Force to needle breaking shutter (Puncture resistance of needle<br>shield)                                                                               | Not required      | Not required | X            | Required by<br>SO                  |
| Tensile strength of needle shaft and needle housing (Break<br>strength of safety mechanism)                                                              | Not required      | Not required | X            | Required by<br>ISO                 |
| Force to release shutter and reactive force when the shutter<br>opens to cover the cannula tip (Reaction force generated by the<br>activation mechanism) | Not required      | Not required | Not required | X                                  |
| Force to detach needle shaft cover from catheter hub (Force to<br>activate safety feature)                                                               | Not required      | Not required | X            | Required by<br>ISO                 |
| Simulated maximum pressure                                                                                                                               | Not required      | Not required | Not required | X                                  |
| Burst pressure ( whole device)                                                                                                                           | X                 | Not required | Not required | Required by<br>ISO                 |
| Collapse (under negative pressure)                                                                                                                       | X                 | Not required | Not required | Required by<br>ISO                 |
| Catheter to catheter hub tensile strength (Force at break of<br>catheter / hub)                                                                          | X                 | Not required | Not required | Required by<br>ISO                 |
| Strength of union between transparent flash chamber (needle<br>connecting part) and needle                                                               | Not required      | X            | Not required | Required by<br>ISO                 |
| Needle attachment to catheter (Initial sliding friction)                                                                                                 | Not required      | Not required | Not required | X                                  |
| Detection of flashback at catheter tip                                                                                                                   | Not required      | Not required | Not required | X                                  |
| Detection of flashback at transparent flash chamber (needle<br>connecting part)                                                                          | Not required      | X            | Not required | Required by<br>ISO                 |
| Flow rate                                                                                                                                                | Not required      | X            | Not required | Required by<br>ISO                 |
| Vent fitting                                                                                                                                             | Not required      | X            | Not required | Required by<br>ISO                 |
| Protector attachment (Drop test)                                                                                                                         | Not required      | Not required | Not required | X                                  |
| Conical fittings of catheter hub                                                                                                                         | X                 | Not required | Not required | Required by<br>ISO                 |
| Pressure monitoring                                                                                                                                      | Not required      | Not required | Not required | X                                  |
| Blood leakage at catheter hub                                                                                                                            | Not required      | Not required | Not required | X                                  |
| Measurement of valve penetration resistance to catheter hub                                                                                              | Not required      | Not required | Not required | X                                  |
| Simulated use study                                                                                                                                      | Not required      | Not required | X            | Required by<br>ISO                 |
| Corrosion resistance                                                                                                                                     | X                 | Not required | Not required | Required by<br>ISO                 |
| Catheter body tensile strength (Force at break)                                                                                                          | X                 | Not required | Not required | Required by<br>ISO                 |
| Stiffness                                                                                                                                                | Not required      | Not required | Not required | X                                  |
| Elongation                                                                                                                                               | Not required      | Not required | Not required | X                                  |
| Flexural fatigue tolerance                                                                                                                               | Not required      | Not required | Not required | X                                  |
| Radio-detectatbility                                                                                                                                     | Not required      | X            | Not required | Required by<br>ISO                 |
| Surface                                                                                                                                                  | X                 | Not required | Not required | Required by<br>ISO                 |
| Catheter unit                                                                                                                                            | Not required      | X            | Not required | Required by<br>ISO                 |
| Needle point                                                                                                                                             | Not required      | X            | Not required | Required by<br>ISO                 |

X: Confirmed the conformance to the standard.

. "

Image /page/4/Picture/4 description: The image shows the numbers 00015 written on top of 00000. The numbers are handwritten and appear to be written with a thick marker. The numbers are aligned vertically, with the ones place aligned on the right.

{5}------------------------------------------------

The SURFLASH® Plus Safety I.V Catheter is classified as

-Catheter and lubricant: Externally Communicating Device, Circulating Blood, Prolonged Exposure (24hrs - 30 days).

-All other materials contacting with patient's body: Externally Communicating Devices, Blood path indirect, Prolonged Exposure (24 hrs to 30 days)

As mentioned below table, the device's materials contacting with patient's body were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO - 10993, "Biological Evaluation of Medical Devices part 1: Evaluation and Testing." Screening tests were performed on accelerated aged whole devices to show that the biocompatibility is maintained throughout the shelf life of the product. Results of the testing demonstrate that the materials contacting with patient's body are biocompatible throughout the shelf life of the product.

| Test                               |
|------------------------------------|
| Cytotoxicity                       |
| Sensitization                      |
| Intracutaneous reactivity          |
| Systemic toxicity (acute)          |
| Pyrogen                            |
| Genotoxicity                       |
| Implantation / Subchronic toxicity |
| Hemolysis                          |
| Physicochemical                    |
| Characterization                   |

### E. ADDITIONAL SAFETY INFORMATION

The sterility of the device is assured using a sterilization method validated in accordance with Method C: Half-cycle method in Annex B of ANSI/AAMI/ISO 11135: Sterilization of health care products -- ethylene oxide. The product is sterilized to provide a Sterility Assurance Level (SAL) of 10°.

{6}------------------------------------------------

### F. SUBSTANTIAL EQUIVALENCE

The SURFLASH® Plus Safety I.V. Catheter manufactured by Terumo Corporation is substantially equivalent to:

- 1. K100282 TERUMO® Surshield®-PUR Safety I.V. Catheter (Terumo Corporation)
- TERUMO® SURFLASH® I.V. Catheter (Terumo Corporation) 2. K991406
- BD Insyte Autoguard BC shielded I.V. Catheter (Becton Dickinson) 3. K110443
- BD Insyte TM Autoguard TM shielded I.V. Catheter (Becton Dickinson) 4. K971339
- Saf-T-Intima Closed IV Catheter System (Becton Dickinson) 4. K923702

Comparison of the intended use/indication for use statements and technological characteristics are summarized in the following pages.

The minor differences of intended use and the technological differences do not impact the safety and effectiveness of the device in clinical settings.

## G. CONCLUSION

The SURFLASH® Plus Safety I.V. Catheter is substantially equivalent to the predicate devices in intended use and technological characteristics.

{7}------------------------------------------------

Terumo Corporation
TERUMO® SURFLASH® Plus Safety I.V. Catheter 5
II. 510(k) Summary

omparison of Intended Use/Indications for Use statements
 SURFLASH® Plus Safety I.V. Cather intentions for use statements are different from those of the predicate devices as below.

| FLASH®Plus<br>afety I.V.<br>eter (subject<br>is 510k) | TERUMO®<br>Surshield®<br>-PUR Safety I.V.<br>Catheter                                                                                                                                                                        | TERUMO®SURFLASH®<br>I.V. Catheter<br>K991406                                                                                                                                                                                                                 | BD Insyte<br>Autoguard BC<br>shielded I.V.<br>Catheter<br>K110443                                                                                                                                                                            | BD Insyte<br>Autoguard<br>shielded<br>I.V.Catheter<br>K971339                                                                                                                 | BD Saf-T-Intima<br>Closed I.V.<br>Catheter System<br>K923702                                                                                                                                                                         |                                                                                                                                                                                                                                       |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                       | K100282                                                                                                                                                                                                                      | Is a device consisting of a<br>slender, flexible,<br>radiopaque, plastic<br>catheter with hub.<br>The stainless steel cannula,<br>placed in the catheter to<br>maintain rigidity,<br>is withdrawn after the<br>catheter is placed in the<br>vascular system. |                                                                                                                                                                                                                                              |                                                                                                                                                                               | The device<br>consists of a<br>slender, flexible,<br>radiopaque, plastic<br>catheter with a<br>hub... (see below<br>section)<br>The steel needle<br>(stylet) placed in<br>the catheter to<br>maintain rigidity is<br>withdrawn after |                                                                                                                                                                                                                                       |
| Summary<br>Section                                    | SURFLASH®Plus<br>Safety I.V.<br>Catheter (subject<br>of this 510k)                                                                                                                                                           | TERUMO®<br>Surshield®<br>-PUR Safety I.V.<br>Catheter<br>K100282                                                                                                                                                                                             | TERUMO®SURFLASH®<br>I.V. Catheter<br>K991406                                                                                                                                                                                                 | BD Insyte<br>Autoguard BC<br>shielded I.V.<br>Catheter<br>K110443                                                                                                             | BD Insyte<br>Autoguard<br>shielded<br>I.V.Catheter<br>K971339                                                                                                                                                                        | BD Saf-T-Intima<br>Closed I.V.<br>Catheter System<br>K923702                                                                                                                                                                          |
| II                                                    | Inserted into<br>patient's vascular<br>system for short<br>term use (<30<br>days) to withdraw<br>blood samples,<br>administer fluid<br>intravenously, or<br>monitor blood<br>pressure by<br>attaching a<br>monitoring device | Inserted into the<br>patient's vascular<br>system for short<br>term use (<30<br>days) to<br>withdraw blood<br>samples,<br>administer fluid<br>intravenously, or<br>monitor blood<br>pressure by<br>attaching a<br>monitoring line.                           | The plastic catheter is<br>inserted into the patient's<br>vascular system for short<br>term use to withdraw<br>samples, administer fluid<br>intravenously or through<br>which to place monitoring<br>equipment such as pressure<br>monitors. | The BD Insyte<br>Autoguard BC<br>catheter is<br>inserted into a<br>patient's<br>vascular system<br>to sample blood,<br>monitor blood<br>pressure, or<br>administer<br>fluids. | As indicated in<br>21CFR 870.1200,<br>to sample blood,<br>to monitor blood<br>pressure, or to<br>introduce<br>substances into<br>the heart and<br>vessels.                                                                           | the catheter is<br>placed in the<br>vascular system<br>...that is inserted<br>into the patient's<br>vascular system<br>for short-term use<br>to sample blood,<br>monitor blood<br>pressure, or<br>administer fluids<br>intravenously. |
| Section<br>Summary                                    | SURFLASH®Plus<br>Safety I.V.<br>Catheter (subject<br>of this 510k)                                                                                                                                                           | TERUMO®<br>Surshield®<br>-PUR Safety I.V.<br>Catheter<br>K100282                                                                                                                                                                                             | TERUMO®SURFLASH®<br>I.V. Catheter<br>K991406                                                                                                                                                                                                 | BD Insyte<br>Autoguard BC<br>shielded I.V.<br>Catheter<br>K110443                                                                                                             | BD Insyte<br>Autoguard<br>shielded<br>I.V.Catheter<br>K971339<br>the classification that<br>applies to the proposed<br>device and the predicate<br>devices in this subject<br>510k.                                                  | BD Saf-T-Intima<br>Closed I.V.<br>Catheter System<br>K923702                                                                                                                                                                          |
|                                                       | The needle shaft<br>cover & tip shield<br>feature aids in the<br>prevention of<br>needle stick<br>injuries                                                                                                                   | The needle shield<br>feature aids in the<br>prevention of<br>needle stick<br>injuries.                                                                                                                                                                       |                                                                                                                                                                                                                                              |                                                                                                                                                                               |                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                       |
|                                                       | These catheters<br>may be used for<br>any patient<br>population with<br>consideration give                                                                                                                                   | These catheters<br>may be used for<br>any patient<br>population with<br>consideration                                                                                                                                                                        |                                                                                                                                                                                                                                              |                                                                                                                                                                               | These catheters<br>may be used for<br>any patient<br>population with<br>consideration give                                                                                                                                           |                                                                                                                                                                                                                                       |
| AI                                                    |                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                              |                                                                                                                                                                               |                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                       |
| ASH®Plus                                              | TERUMO®<br>Surshield®<br>(subject to 510k)                                                                                                                                                                                   | TERUMO®<br>-PUR Safety I.V.<br>Catheter<br>K100282                                                                                                                                                                                                           | TERUMO®SURFLASH®                                                                                                                                                                                                                             | BD Insyte<br>Autoguard BC<br>shielded I.V.<br>Catheter<br>K110443                                                                                                             | BD Insyte<br>Autoguard<br>shielded<br>I.V.Catheter<br>K971339                                                                                                                                                                        | BD Saf-T-Intima<br>Closed I.V.<br>Catheter System<br>K923702                                                                                                                                                                          |
|                                                       |                                                                                                                                                                                                                              | given to<br>adequacy of<br>anatomy                                                                                                                                                                                                                           |                                                                                                                                                                                                                                              |                                                                                                                                                                               | to patient size,<br>appropriateness<br>for the solution<br>being infused, and<br>duration of<br>therapy.                                                                                                                             |                                                                                                                                                                                                                                       |
|                                                       |                                                                                                                                                                                                                              | adequacy of<br>vascular anatomy<br>and<br>appropriateness<br>for the solution<br>being infused<br>and duration of<br>therapy                                                                                                                                 |                                                                                                                                                                                                                                              |                                                                                                                                                                               |                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                       |
|                                                       |                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                              |                                                                                                                                                                               |                                                                                                                                                                                                                                      | The catheters<br>maybe used with<br>power injectors<br>for which the<br>maximum rated<br>pressure is 300<br>psi.                                                                                                                      |
|                                                       |                                                                                                                                                                                                                              | 22 G<br>are<br>for use<br>power<br>is rated for<br>maximum of 325                                                                                                                                                                                            |                                                                                                                                                                                                                                              |                                                                                                                                                                               |                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                       |

{8}------------------------------------------------

Terumo Corporation
TERUMO® SURFLASH® Plus Safety I.V. Catheter 51
II. 510(k) Summary

{9}------------------------------------------------

Terumo Corporation
TERUMO® SURFLASH® Plus Safety I.V. Catheter 5100
11. 510(k) Summary________________________________________________________________________________________

{10}------------------------------------------------

Terumo Corporation
TERUMO® SURFLASH® Plus Safety I.V. Catheter 516
II. 510(k) Summary

11:00 PM

{11}------------------------------------------------

## Discussion of Sections of the Intended Use/Indications for Use

## I. Device Descriptive section (Full physical description of the device):

A full physical description of the device is not necessary for a discussion in the Intended Use/Indications for Use section. Omission of a physical description in the Intended Use/Indications for use of the proposed device subject of this 510k is not relevant to a comparison of the device's intended use/indication for use. Comparison of the physical description is presented in the comparison of technological characteristics of this 510k. The omission of a physical description in the Intended Use/Indication for use has no impact on the safety and effectiveness. The physical attributes of the proposed device have been tested and compared to the appropriate predicate that has the same/similar physical attribute.

### II. Basic Device description and use as prescribed in 21CFR 880.5200

The proposed device and the predicate devices are classified under 21CFR880.5200 which states:

" An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 davs) to sample blood, monitor blood pressure or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials. "

The proposed device and the predicate devices use a similar statement (underlined portion) regarding the actual use of the device.

Note: BD Insyte Autoguard states in the 510k that the device is classified under 21CFR 870.1200, however, FDA has classified this 510k (K971339) under 21CFR880.5200 which is identical to the proposed device subject of this 510k and the predicate devices.

{12}------------------------------------------------

## III. Description of Sharps Injury Protection

Terumo has included the description of the sharps injury protection feature in the Intended Use/Indications for Use statement for the proposed device and for one predicate device (TERUMO® Surshield®-PUR Safety I.V. Catheter (K100282). Although three of the other predicate devices (K110443, K971339 & K923702) contain a sharps injury protection feature, the manufacturers have not included this description in their Intended Use/Indication for Use statement of their cleared 510k (this feature is discussed in the Technological Characteristics section). Even though the descriptive statement does not appear in the Intended Use/Indication for Use for those predicates, the feature is cleared under the respective 510ks.

Note: The predicate cleared under K991406 does not include a sharps injury protection feature as discussed in the Technological Characteristics section and, therefore, does not have this feature described herein.

### IV. Discussion of usage in populations

The proposed device and two of the predicate devices (K100282 and K971339) clarify that these devices may be used in any population based on medical judgment. There is no restriction placed on the use of devices in any specific population and no specific indication of use in a specific population, but rather the statement indicates these devices can be used wherever medical judgment deems appropriate. These devices are general medical use devices and can be used in any population deemed acceptable by a medical professional. These devices are "Rx only."

These clarifying statements have no impact on the general use of the products and do not affect the safety or effectiveness nor the determination of substantial equivalence.

### V. Use with Power Injectors

The proposed device and one of the predicate devices (K971339) include an indication for use with Power Injectors. The other three predicate devices do not include this indication. Appropriate testing has been included in this 510k to address the new issues of safety and effectiveness raised by this indication. The test results demonstrate the substantial equivalence of this indication with the product covered under K971339. This indication in

{13}------------------------------------------------

Terumo Corporation TERUMO® SURFLASH® Plus Safety I.V. Catheter 510k II. 510(k) Summary

the 510(k) subject device has no impact on the other indications for use for which substantial equivalence is being claimed. This indication can effectively be reviewed separately regarding substantial equivalence.

{14}------------------------------------------------

Terumo Corporation
TERUMO® SURFLASH® Plus Safety 1.V. Catheter 5 I
II. 510(k) Summary

It - Torinal Partier State Plan Safety I V. Catheter are similar to the predicate devices in terms of principle

|                                               | Trade Name                                                     | Subject of this 510(k)                                | Manufacturer                                          | Operation principle                                                                                                                                                                                         | Design /Construction                  | Needle Shape/Material                 | Flashback                  | II. 510(k) Summary<br>Trade Name                                                         | SURFLASH® Plus<br>Safety I.V.<br>Catheter                                                | TERUMO®<br>Surshield®-PUR<br>Safety I.V.<br>Catheter                                     | TERUMO®<br>SURFLASH®<br>I.V. Catheter                                          | BD Insyte<br>Autoguard BC<br>shielded I.V.<br>Catheter | BD Insyte<br>Autoguard<br>shielded I.V.<br>Catheter | Saf-T-Intima<br>Closed IV<br>Catheter System |
|-----------------------------------------------|----------------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------|----------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------|-----------------------------------------------------|----------------------------------------------|
| SURFLASH® Plus Safety I.V. Catheter           |                                                                | Terumo Corporation                                    | Conduit as fluid pathway Operated manually.           | a. Catheter assembly<br>b. Needle assembly<br>c. Needle shield assembly<br>d. Filter<br>e. Needle protector<br>f. Valve<br>g. Seal component<br>h. Plug                                                     | Notched cannula Stainless steel       | Visible through                       | Subject of this 510(k)     | Subject of this 510(k)                                                                   | K100282                                                                                  | K991406                                                                                  | K110443                                                                        | K971339                                                | K923702                                             |                                              |
| TERUMO® Sursbield®-PUR Safety I.V. Catheter   | K100282                                                        | Terumo Corporation                                    | Conduit as fluid pathway Operated manually.           | a. Catheter assembly<br>b. Needle assembly<br>c. Needle shield assembly<br>d. Filter cap with filter<br>e. Protector                                                                                        | Notched cannula Stainless steel       | Visible through                       | Catheter material          | ditch on the cannula surface.<br>Polyurethane<br>Biocompatibility of material confirmed. | ditch on the cannula surface.<br>Polyurethane<br>Biocompatibility of material confirmed. | ditch on the cannula surface.<br>Polyurethane<br>Biocompatibility of material confirmed. | on the cannula surface.<br>Polyurethane                                        | on the cannula surface.<br>Polyurethane                | needle lumen to tube.<br>Polyurethane               |                                              |
| TERUMO® SURFLASH® I.V. Catheter               | K991406                                                        | Terumo Corporation                                    | Conduit as fluid pathway Operated manually.           | a. Catheter assembly<br>b. Needle assembly<br>c. Filter cap with filter<br>d. Protector                                                                                                                     | Notched cannula. Stainless steel      | Visible through                       | Radiopaque medium          | Yes (Barium sulfate)                                                                     | Yes (Barium sulfate)                                                                     | Yes (Barium sulfate)                                                                     | Yes (material not identified)                                                  | Yes (material not identified)                          | Yes (material not identified)                       |                                              |
| BD Insyte Autoguard BC shielded I.V. Catheter | K110443                                                        | Becton Dickinson                                      | Conduit as fluid pathway Operated manually.           | a. Catheter assembly<br>b. Needle assembly<br>c. Needle shield assembly<br>d. Filter<br>e. Needle protector<br>f. Valve (called Septum)<br>g. plug*                                                         | Holed cannula. Stainless steel        | Visible through hole                  | Safety mechanism           | Passive needle shielding                                                                 | Passive needle shielding                                                                 | None                                                                                     | Passively needle shielding                                                     | Passively needle shielding                             | Passive needle shielding                            |                                              |
| BD Insyte Autoguard shielded I.V. Catheter    | K971339                                                        | Becton Dickinson                                      | Conduit as fluid pathway Operated manually.           | a. Catheter assembly<br>b. Needle assembly<br>c. Needle shield assembly<br>d. Filter<br>e. Needle protector                                                                                                 | Holed cannula. Stainless steel        | Visible through hole                  | Valve                      | YES to reduce the blood leakage at catheter hub in view of blood exposure risk           | No                                                                                       | No                                                                                       | YES to reduce the blood leakage at catheter hub in view of blood exposure risk | No                                                     | No                                                  |                                              |
| Saf-T-Intima Closed IV Catheter System        | K923702                                                        | Becton Dickinson                                      | Conduit as fluid pathway Operated manually            | a. Catheter assembly with tube and I or Y adaptor<br>b. Needle assembly<br>c. Needle shield<br>d. Filter cap with air filter<br>e. Needle cover<br>f. IV access port<br>g. Wings available Notched cannula. | Stainless steel                       | Visible through                       | Color code of catheter hub | In accordance with ISO 10555-5                                                           | In accordance with ISO 10555-5                                                           | In accordance with ISO 10555-5                                                           | In accordance with ISO 10555-5                                                 | In accordance with ISO 10555-5                         | In accordance with ISO 10555-5                      |                                              |
| Package                                       | Sterility barrier:<br>Blister package<br>Unit box<br>Cardboard | Sterility barrier:<br>Casing<br>Unit box<br>Cardboard | Sterility barrier:<br>Casing<br>Unit box<br>Cardboard | Sterility barrier:<br>Blister package                                                                                                                                                                       | Sterility barrier:<br>Blister package | Sterility barrier:<br>Blister Package |                            |                                                                                          |                                                                                          |                                                                                          |                                                                                |                                                        |                                                     |                                              |
| Sterilization method                          | Ethylene oxide                                                 | Ethylene oxide                                        | Ethylene oxide                                        | Ethylene oxide                                                                                                                                                                                              | Ethylene oxide                        | Ethylene oxide                        |                            |                                                                                          |                                                                                          |                                                                                          |                                                                                |                                                        |                                                     |                                              |

{15}------------------------------------------------

Terumo Corporation
TERUMO® SURFLASH® Plus Safety I.V. Catheter 510
L. 510/k) Summary

{16}------------------------------------------------

Terumo Corporation
TERUMO® SURFLASH® Plus Safety I.V. Catheter 5
II. 510(k) Summary
II. 510(k) Summary
Conclusion

SURFLASH® Plus Safety I.V. Catheter and the predicate devices have the same Technological Characteristics excep

- 
- e predicate cleared under K991406 does not include a sharps injury protection teature as mentioned in the above table.
ne predicates cleared under K100282, K991406, K971339

{17}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/17/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a person embracing or supporting another, conveying a sense of care and assistance. The overall design is simple and professional, reflecting the department's role in public health and human services.

### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Kofu Factory of Terumo Corporation C/O Terumo Medical Corporation Ms. Sandi Hartka Regulatory Affairs Manager 950 Elkton Boulevard Elkton, Maryland 21921

November 2, 2012

Re: K123267

Trade/Device Name: SURFLASH® Plus Safety I.V. Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: October 18, 2012 Received: October 19, 2012

Dear Ms. Hartka:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{18}------------------------------------------------

### Page 2 - Ms. Hartka

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

### Sincerely yours.

Digitally signed by Anthony D. Watson DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Anthony D. Watson, 0.9.2342.19200300.100.1.1=1300092402 Date: 2012.11.02 13:08:22 -04/00'

Anthony D. Watson, B.S., M.S., M.B.A.

Director Division of Anesthesiology, General Hospital,

Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

### Enclosure .

{19}------------------------------------------------

# Indications for Use Form

# Indications for Use

510(k) Number (if known): _KIL 326 7

SURFLASH® Plus Safety I.V. Catheter Device Name:

Indications for Use:

The SURFLASH® Plus Safety I.V. Catheter is inserted into the patient's vascular system for short term use (<30 days) to withdraw blood samples, administer fluid intravenously; or monitor blood pressure by attaching a monitoring line. The needle shaft cover and tip shield feature aids in the prevention of needle stick injuries. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. The 18 to 22 gauge catheters are suitable for use with power injectors rated for a maximum of 325 psi.

Prescription Use X (Part 21 CFR 801 Subpart D)

.

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Khd. Chay 11/1/2012

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Anest. Dental Devices

**510(k) Number:**

Page of

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K123267](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K123267)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
