← Product Code [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ) · K110443

# BD INSYTE AUTOGUARD BC (K110443)

_Becton, Dickinson & CO · FOZ · Jul 19, 2011 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K110443

## Device Facts

- **Applicant:** Becton, Dickinson & CO
- **Product Code:** [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ.md)
- **Decision Date:** Jul 19, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The BD Insyte™ Autoguard™ BC catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids.

## Device Story

Over-the-needle peripheral intravascular catheter; incorporates spring-activated needle-shielding technology to prevent needlestick injuries. Features integrated blood control septum in catheter hub; septum automatically activates upon needle removal to restrict blood flow until luer connection is secured. Includes BD Instaflash™ Needle Technology (20, 22, 24 gauge) for vessel entry confirmation via side-notch. Clinician-operated; button-activated needle retraction into shielding barrel; non-reusable. Used in clinical settings for vascular access. Benefits include reduced clinician blood exposure and enhanced safety via needle shielding.

## Clinical Evidence

Prospective, randomized, unblinded healthy volunteer study conducted to validate blood control feature design. Primary endpoints: clinical acceptability of blood control feature, clinician perception of blood exposure during insertion, and frequency of blood leakage from catheter hub.

## Technological Characteristics

Peripheral IV catheter; BD Vialon™ polyurethane material; radiopaque. Features: spring-activated needle-shielding, integrated septum for blood control, luer slip/lock hub. Available in 16G, 18G, 20G, 22G, 24G. Sterilization: Ethylene Oxide. Manual operation.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- BD Insyte™ Autoguard ([K013800](/device/K013800.md))
- SUPERCATH 5 ([K093546](/device/K093546.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K110443

BD Insyte 73 Autoguard™ BC Premarket Notification- Traditional 510(k) Section 5 - 510(k) Summary

# Section 5 - 510(k) SUMMARY

JUL 1 9 2011

# SUMMARY OF SAFETY AND EFFECTIVENESS FOR

BD Insyte™ Autoguard™ BC

### 1. Submitted by:

Becton Dickinson Infusion Therapy Systems Inc. 9450 South State Street Sandy, UT 84092

Contact Information: Rand Pugmire Manager, Regulatory Affairs Phone: (801) 565-2550 Fax: (801) 565-2749

Date Summary Prepared: February 11, 2011

### 2. Device Name:

Trade Name: Common Name: Classification: CFR Reference: Classification Panel:

BD Insyte TM Autoguard TM BC Peripheral Intravascular Catheter or IV Catheter 80 FOZ - Intravascular Catheter 21CFR 880.5200 - Class II General Hospital

### 3. Predicate Devices:

Substantial equivalence is being claimed to the following legally marketed devices:

| Trade Name:             | BD Insyte™ Autoguard™           |
|-------------------------|---------------------------------|
| Common Name:            | Intravascular Catheter          |
| Classification:         | 80 FOZ - Intravascular Catheter |
| CFR Reference:          | 21CFR 880.5200 - Class II       |
| Classification Panel:   | General Hospital                |
| Premarket Notification: | K013800                         |

Trade Name: Common Name: Classification: CFR Reference: Classification Panel: Premarket Notification:

SUPERCATH 5 Intravascular Catheter 80 FOZ - Intravascular Catheter 21CFR 880.5200 - Class II General Hospital K093546

### 4. Product Description:

The purpose of the premarket notification is to introduce a new intravascular catheter. the BD Insyte™ Autoquard™ BC Shielded IV Catheter. This new catheter is identical to the BD Insyle™ Autoguard device (K013800) with the addition of a septum and actuator (blood control features) integrated into the catheter hub.

The BD Insyte Autoquard BC catheter's, catheter hub has a built in blood control septurn. The blood control feature is a single-use septum that automatically activates 

{1}------------------------------------------------

to stop the blood flow in the catheter hub when the needle is removed from the catheter during initial insertion by the clinician. Blood flow from the catheter hub will be restricted immediately after the needle retraction until a secure luer connection is made. The flow path is permanently opened once a secure luer connection has been made.

BD Insyte ™ Autoguard™ BC device is an over-the-needle, peripheral intravascular catheter that incorporates a spring-activated needle-shielding technology. Each catheter needle-shielding component inciudes a cylindrical needle-shielding barrel, a spring, a needle hub with flash chamber, and a needle.

The catheter component that is advanced into the vasculature is a BD Vialon™ polyurethane material. The catheter hub accommodates a luer slip or luer lock connection. It is available either with a straight or a winged catheter hub.

The BD Insyte Autoguard BC catheter needle-shielding technology provided the clinician with a user-activated button, the clinician pushes it to initiate the needle's retraction into the needle-shielding barrel. Once the button has been pushed and the needle retracted, the user cannot override the shielding mechanism to re-expose the needle tip.

The BD Insvte Autoquard BC catheter's needle incorporates BD Instaflash™ Needle Technology: this feature allows the clinician to visualize flashback through a notch in the side of the needle indicating that there is confirmation of vessel entry. This is available on the 20, 22, 24 gauge catheters only.

The BD Insyte Autoguard BC catheter is available in 16GA, 18GA, 20GA, 22GA, and 24GA sizes.

### 5. Indications for Use:

The BD Insyte™ Autoguard™ BC catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids.

#### Statement of substantial equivalence: 6.

The characteristic of the BD Insyte Autoguard BC catheter are similar to those of the predicate devices. The similarities are:

| Factor/Component                              | BD Insyte<br>Autoguard<br>BC | BD Insyte<br>Autoguard<br>(K013800) | SUPERCATH<br>5<br>(K093546) |
|-----------------------------------------------|------------------------------|-------------------------------------|-----------------------------|
| Same Intended use                             | Yes                          | Yes                                 | Yes                         |
| Same type of catheter material (Polyurethane) | Yes                          | Yes                                 | Yes                         |
| Radiopaque catheter                           | Yes                          | Yes                                 | Yes                         |
| Flashback visualization                       | Yes                          | Yes                                 | Yes                         |
| Needle-shielding prevention feature           | Yes                          | Yes                                 | Yes                         |
| Ethylene Oxide sterilization                  | Yes                          | Yes                                 | Yes                         |
| Single Use, sterile device                    | Yes                          | Yes                                 | Yes                         |
| Multiple gauge sizes and needle lengths       | Yes                          | Yes                                 | Yes                         |
| Side-notch needle (certain gauges)            | Yes                          | Yes                                 | Yes                         |

{2}------------------------------------------------

|  |  |                                |  |  |  | the fact of the countent controllation for the property of the concentration and commendent of the reserved the real memor of the reserved the reserved the really with and an |  |  |  |  |  |  |
|--|--|--------------------------------|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
|  |  | Blood Control feature - Septum |  |  |  |                                                                                                                                                                                |  |  |  |  |  |  |
|  |  |                                |  |  |  | The real production and and sell sells sell seller inter least the find like for will least of the first of the first of the first of the first of the first of the first of t |  |  |  |  |  |  |

The BD Insyte Autoguard BC catheter has the same intended use and similar technological characteristics and the BD Insyte Autoguard (KOl3800) and the SUPERCATH 5 (K093546). Similar components and materials are used in both the BD Insyte Autoquard (predicate) and the BD Insyte Autoguard BC catheter.

Verification and validation activities were designed and performed to demonstrate that the BD Insvte Autoquard BC catheter met predetermined product specifications. See Section. 9, 18, 19. & 20.

# 7. Risk Analysis:

The BD Insyte Autoquard BC catheter was evaluated in accordance with ISO 14971:2009. No new types of safety or efficacy questions were identified for the BD Insyte Autoguard BC device. See Section 21.

#### 8. Bench Testing:

Bench testing were performed to ensure the safety and effectiveness of the BD Insyte Autoquard BC catheter, to verify conformity to the standards listed in this application and demonstrate substantial equivalence to the predicates devices used in this application. No new issues of safety and effectiveness were raised with the testing performed, so therefore, the BD Insyte Autoguard BC is substantially equivalent. See Section 18.

# 9. Clinical Use:

A prospective, randomized, unblinded healthy volunteer (subjects) clinical study was conducted by BD to validate the design of the BD Insyte Autoguard BC catheter with the primary objective to evaluate the overall clinical acceptability of the new blood control feature, the clinician's perception of blood exposure during the IV Catheter insertion process, and the frequency of blood leakage observed from the IV Catheter hub. See section 20.

# 10. Conclusions:

Conclusions drawn from performance testing and clinical testing were conducted in accordance with consensus standards and design control requirements. Test results and technological characteristics of like gauge size catheters were shown to be equivalent between the subject device and the predicates. The differences amonq the devices do not raise any issues of safety or effectiveness. The subject BD Insyte Autoguard BC catheter met the minimum requirements and is substantially equivalent in design, materials, sterilization, principles of operations and intended use to the predicate.

Based on the above summary and the enclosed sections of this premarket notification regarding substantial equivalence to the predicate devices, Becton Dickinson Infusion Therapy Systems Inc. concludes that the BD Insyte Autoguard BC catheter is substantially equivalent to the BD Insyte Autoguard device, (K013800), and the SUPERCATH 5 device, (K093546) and does not raise any new questions regarding safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Becton Dickinson Infusion Therapy Systems, Incorporated Mr. Rand Pugmire Managers Regulatory Affairs Becton Dickinson & Company 9450 South State Street Sandy, Utah 84070

JUL 1 9 2011

Re: K110443

Trade/Device Name: BD Insyte™ Autoguard BCTM IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: July 7, 2011 Received: July 8, 2011

Dear Mr. Pugmire:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Pugmire

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm 1 1 5809 htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony O. Matson
B.S., M.S.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): Unknown

BD Insyte™ Autoguard BC™ IV Catheter

Indications for Use:

Device Name:

The BD Insyte TM Autoguard™ BC catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Richard C. Chayen 7/18/4

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K110443

Confidential & Proprietary

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K110443](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K110443)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
