← Product Code [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ) · K060341

# SHERLOCK TIP LOCATION SYSTEM (TLS) (K060341)

_C.R. Bard, Inc. · FOZ · Apr 14, 2006 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K060341

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [FOZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ.md)
- **Decision Date:** Apr 14, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Sherlock TLS Detector quickly locates and confirms the position of specially designed, magnet-tipped Peripherally Inserted Central Catheters (PICCS) and Central Venous Catheters (CVCs) during and after initial placement. This device may be used by appropriate caregivers in hospitals, long-term care facilities or home-care settings. The Sherlock TLS Detector provides rapid feedback to the caregiver, but was not designed to replace conventional methods of placement verification. Users are urged to confirm correct placement according to their established institutional protocol and clinical judgment.

## Device Story

System comprises Sherlock TLS Stylet and Sherlock TLS Detector; stylet contains encapsulated passive magnets generating magnetic field; detector senses field to provide real-time feedback on catheter tip location during insertion. Used in hospitals, long-term care, or home-care settings by clinicians to assist in PICC/CVC placement. Provides rapid feedback to supplement, not replace, conventional verification methods (e.g., X-ray). Clinicians use output to guide catheter positioning; aids in confirming tip location; potentially reduces need for multiple placement attempts or post-procedure adjustments.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance evaluated against FDA guidance for guidewires and ISO standards for intravascular catheter introducers, biocompatibility (ISO 10993), and sterilization (ISO 11135).

## Technological Characteristics

System includes stylet with encapsulated passive magnets and magnetic detector. Materials evaluated per ISO 10993-1. Sterilization via ethylene oxide per AAMI/ANSI/ISO 11135. Design provides internal catheter reinforcement and magnetic field generation for tip tracking.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- 5 Fr Dual Lumen Groshong® nXt PICC Catheter ([K023374](/device/K023374.md))
- Poly Per-Q-Cath® PICC ([K034019](/device/K034019.md))
- Sensor Stylet and Sensor Guidewire ([K935380](/device/K935380.md))
- Zortran Detector ([K000997](/device/K000997.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Sherlock™ TLS Traditional 510(k)

Section 5 Sherlock™ Tip Location System 510(k) Summary of Safety and Effectiveness 21 CFR 807.92(a)

んで ົກການປະກວດນາງງາມ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ ແລະ

## 5.1 General Information

| Submitter Name:                  | Bard Access Systems, Inc. (BAS)               |
|----------------------------------|-----------------------------------------------|
|                                  | [Wholly owned Subsidiary of C. R. Bard, Inc.] |
| Address:                         | 5425 W. Amelia Earhart Drive                  |
|                                  | Salt Lake City, UT 84116                      |
| Telephone Number:                | (801) 595-0700 ext. 7136                      |
| Fax Number:                      | (801) 595-5425                                |
| Contact Person:                  | Lynn M. Kirchoff                              |
| Date of Preparation:             | February 9, 2006                              |
| Registration Number:             | 1720496                                       |
| Additional Registration Numbers: |                                               |
| C.R. Bard:                       | 2212754                                       |

## 5.2 Subject Device Information

| Device Name:          | Sherlock™ Tip Location System                                                                                   |
|-----------------------|-----------------------------------------------------------------------------------------------------------------|
| Trade Name:           | Sherlock™ Tip Location System (TLS)                                                                             |
| Common/Usual Name:    | Vascular Access Catheter Accessories                                                                            |
| Classification Name:  | 80 LJS - Accessory to Percutaneous, Implanted, Long-Term<br>Intravascular Catheter<br>21 CFR 880.5970— Class II |
| Classification Panel: | General Hospital                                                                                                |

## 5.3 Predicate Device Information

| Device Name(s):       | 5 Fr Dual Lumen Groshong® nXt PICC Catheter                        |
|-----------------------|--------------------------------------------------------------------|
|                       | Poly Per-Q-Cath PICC                                               |
| Trade Name(s):        | Groshong® nXt, Poly Per-Q-Cath®                                    |
| Common/Usual Name:    | Peripherally Inserted Central Catheter (PICC)                      |
| Classification Name:  | 80 LJS - Percutaneous, Implanted, Long-Term Intravascular Catheter |
|                       | 21 CFR 880.5970—Class II                                           |
| Classification Panel: | General Hospital                                                   |

| Predicate Device Name               | 510(k)  | Clearance Date |
|-------------------------------------|---------|----------------|
| 5 Fr DL Groshong® nXt PICC Catheter | K023374 | 12/18/2002     |
| Poly Per-Q-Cath® PICC               | K034019 | 1/21/2004      |

{1}------------------------------------------------

| Device Name:          | Sensor Stylet and Sensor Guidewire                                                                              |
|-----------------------|-----------------------------------------------------------------------------------------------------------------|
| Trade Name:           | Catheter Locator System                                                                                         |
| Common/Usual Name:    | Vascular Access Catheter Accessories                                                                            |
| Classification Name:  | 80 LJS – Accessory to Percutaneous, Implanted, Long-Term<br>Intravascular Catheter<br>21 CFR 880.5970– Class II |
| Classification Panel: | General Hospital                                                                                                |

| Predicate Device Name              | 510(k)  | Clearance Date |
|------------------------------------|---------|----------------|
| Sensor Stylet and Sensor Guidewire | K935380 | 2/17/1995      |

| Device Name:          | Zortran Detector                                                                                |
|-----------------------|-------------------------------------------------------------------------------------------------|
| Trade Name:           | Zortran Detector                                                                                |
| Common/Usual Name:    | Vascular Access Catheter Accessories                                                            |
| Classification Name:  | FOZ - Accessory to Percutaneous, Implanted, Intravascular Catheter<br>21 CFR 880.5970- Class II |
| Classification Panel: | General Hospital                                                                                |

| Predicate Device Name | 510(k)  | Clearance Date |
|-----------------------|---------|----------------|
| Zortran Detector      | K000997 | 9/7/2000       |

#### 5.4 Intended Use

The intended use of the subject device is the same as the intended use of the predicate devices.

Catheter stylets provide internal reinforcement to aid in catheter placement. The Sherlock™ TLS Stylet contains passive magnets that generate a magnetic field can be detected by the Sherlock™ TLS Detector to provide the placer rapid feedback on catheter tip location.

The Sherlock TLS Detector quickly locates and confirms the position of specially designed, magnet-tipped Peripherally Inserted Central Catheters (PICCS) and Central Venous Catheters (CVCs) during and after initial placement. This device may be used by appropriate caregivers in hospitals, long-term care facilities or home-care settings. The Sherlock TLS Detector provides rapid feedback to the caregiver, but was not designed to replace conventional methods of placement verification. Users are urged to confirm correct placement according to their established institutional protocol and clinical judgment.

#### ર્ડ Indications for Use

The Indications for Use for the Sherlock™ Tip Location System (TLS) is as follows:

Catheter stylets provide internal reinforcement to aid in catheter placement. The Sherlock™ TLS Stylet contains passive magnets that generate a magnetic field can be detected by the Sherlock™ TLS Detector to provide the placer rapid feedback on catheter tip location.

#### Summary of Changes 5.6

The following are modifications made to the currently marketed predicate PICC stylet and Sensor Stylet:

{2}------------------------------------------------

- . The addition of encapsulated passive magnetic material in the distal portion of the stylet to allow for detection when used with a magnetic sensor.
#### 5.7 Device Description

The Sherlock Tip Location System (TLS) consists of the Sherlock TLS Detector and Sherlock TLS Stylet.

The Sherlock TLS Stylet has been developed to aid in the placement of Bard Access Systems catheters using current placement techniques. The stylets are designed to give the catheters added support and stiffness while traversing the patient's venotomy. Also, should the clinician choose to do so, the stylets have been designed to be used in conjunction with Sherlock TLS Detector to allow for rapid feedback of catheter tip placement. The information provided by the Sherlock TLS is not meant to replace conventional methods of catheter placement verification. Clinicians are urged to confirm correct catheter placement according to their established institutional protocol and clinical judgment.

#### ని.8 Technological Comparison to Predicate Devices

The technological characteristics of the Sherlock TLS are substantially equivalent to those of the predicate devices in terms of intended use, application, user population, basic design, performance, labeling, packaging and sterilization method.

#### ર્સ જે 510(k) Substantial Equivalence Decision Tree

New device is compared to Marketed Device? Yes.

Does the new device have the same indication statement and intended use as the predicate? Yes. However, there are minor modifications to the indication verbiage.

Does the new device have the same technological characteristics, e.g. design, materials, etc.? Not in all regards. The principles of operation and basic design are the same as the predicate devices. The main changes in design are the addition of passive magnetic material encapsulated in the distal portion of the stylet.

# Could the new characteristics affect safety or effectiveness?

Yes. The design changes may affect safety or effectiveness of the device.

Do the new characteristics raise new types of safety and effectiveness questions? No. Safety and effectiveness questions are the same as for the predicate devices.

Do accepted scientific methods exist for assessing effects of the new characteristics? Yes. Testing was based on FDA guidance document and standards to evaluate the devices' performance:

- Coronary and Cerebrovascular Guidewire FDA Guidance, dated 1/95 .
- ISO 11070:1998, Sterile, single-use intravascular catheter introducers .
- AAMI/ANSI/ISO 11135:1994, Medical Devices Validation and Routine Control of Ethylene . Oxide Sterilization
- . AAMI/ANSI/ISO 10993-1:1997, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, and the FDA Modified ISO 10993 Test Profile

Image /page/2/Picture/19 description: The image shows the number '600013' in a bold, sans-serif font. The numbers are closely spaced together, and the image has a slightly grainy texture. The number appears to be a scan or a reproduction of a printed document.

{3}------------------------------------------------

# Are performance data available to assess effects of new characteristics?

Yes. Bench testing was based on the above referenced guidance document and standards. All test results confirm that the modified device is substantially equivalent to the predicate devices.

### 5.10 Conclusion

The Sherlock™ Tip Location System (TLS) met all the performance criteria of the tests performed and, based on FDA's decision tree, is substantially equivalent to its predicate devices covered by: K023374, K034019, K935380, and K000997.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines representing its body and wings. The eagle is encircled by the text "U.S. Department of Health and Human Services, USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# APR 1 4 2006

C.R. Bard, Incorporated C/O Ms. Lynn M. Kirchoff Regulatory Affairs Specialist Bard Access Systems, Incorporated 5425 West Amelia Earhart Drive Salt Lake City. Utah 84116

Re: K060341

Trade/Device Name: Sherlock™ Tip Location System (TLS) Regulation Number: 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: February 9, 2006 Received: February 10, 2006

Dear Ms. Kirchoff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Kirchoff

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-fice number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Section 4 Indications for Use

510(k) Number (if known): job6()341

Device Name: SherlockTM Tip Location System (TLS)

Indications for Use:

Catheter stylets provide internal reinforcement to aid in catheter placement. The Sherlock™ TLS Stylet contains passive magnets that generate a magnetic field can be detected by the Sherlock™ TLS Detector to provide the placer rapid feedback on catheter tip location.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Clu is

v. Osneral k

KClO341

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K060341](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FOZ/K060341)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
