SUNDER CENTRAL VENOUS CATHETER KIT, MODELS SD-3L55F30J AND SD-3L70F30J

K024007 · Sunder Biomedical Tech. Co., Ltd. · FOZ · Jul 14, 2003 · General Hospital

Device Facts

Record IDK024007
Device NameSUNDER CENTRAL VENOUS CATHETER KIT, MODELS SD-3L55F30J AND SD-3L70F30J
ApplicantSunder Biomedical Tech. Co., Ltd.
Product CodeFOZ · General Hospital
Decision DateJul 14, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5200
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The triple lumen central venous Catheter is intended for vascular access infusion and withdrawal of blood, blood products, and fluids, central venous blood pressure monitoring (CVP) plasma pheresis, acute hyperalimentation, continuous or int4ermittent drug infusion.

Device Story

Triple-lumen polyurethane central venous catheter (5.5 and 7.0 French); features three independent non-communicating lumens, extension lines, luer hubs, and slide clamps. Placed in a vein with distal end in vena cava. Used for vascular access, infusion/withdrawal of blood/fluids, CVP monitoring, and drug delivery. Operated by clinicians in clinical settings. Kit includes catheter in PET tray, sealed with Tyvek, and sterilized. Provides multi-lumen access for patients requiring complex intravenous therapy or monitoring.

Clinical Evidence

Bench testing only. Performance testing included tensile strength, elongation, anti-corrosive properties, flexural fatigue, stiffness, flow rate, burst pressure, collapse, leakage, and ink adhesion. Microbiological testing included sterilization process validation and E.O. residual testing. Biocompatibility testing was performed. No clinical data provided.

Technological Characteristics

Triple-lumen polyurethane catheter; 5.5 and 7.0 French sizes. Features three independent non-communicating lumens, extension lines, luer hubs, and slide clamps. Sterilized via E.O. gas. Packaged in PET tray with Tyvek lid.

Indications for Use

Indicated for patients requiring multiple IV access sites, lacking usable peripheral sites, requiring central venous pressure monitoring, total parenteral nutrition, incompatible medications, multiple fluid/medication/chemotherapy infusions, frequent blood sampling, blood transfusions, or hypertonic/hyperosmolar/divergent pH infusions.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K024007 ## Section 1. -510 (k) Summary - a) Submitter: SUDER Sunder Biomedical Tech. Co., Ltd Contact person: Tony Hung Address: 10F-1. 1-67, Wu-Chuan Rd., Taichung, Taiwan 403 R.O.C. 886-4-23755650 Phone: Fax: 886-4-23755651 Email: tonv@sunder.com.tw Date Summary prepared: 11/15/2002 - b) Device | Device trade name: | Sunder Central Venous Catheter Kit | |-----------------------------|----------------------------------------------------------------------------------| | Device common name: | Central Venous Catheter (CVC) | | Device Classification name: | Short-term catheters (less than 30 days) -Percutaneous<br>Intravascular catheter | c) Legally marketed device to which the device is substantially equivalent ARROW g* and blue Plus™ Multi-Lumen Central Venous Catheter 510(k) No.: K993691 - d) Description of the device The medical device made by Sunder Biomedical Tech, CO., LTD are triple-lumen. polyurethane catheter, 5.5 and 7.0 French in size, with three independent non-communication lumens, extension lines, luer hubs and slide clamps, placed within a vein and whose distal end is intended to be located within the vena cava (inferior of superior). Central Venous Catheter (CVC) - a tubular device with single lumen or rnulti-lumen catheter, or cannula, placed within a vein and whose distal end is intended to be located within the vena cava (inferior of superior). The catheterization kit components are configured in a Polyester Terephthalate (PET) tray and sealed with a Tyvek lid stock, and sterilized. - e) Intended use of the device The triple lumen central venous Catheter is intended for vascular access infusion and withdrawal of blood, blood products, and fluids, central venous blood pressure monitoring (CVP) plasma pheresis, acute hyperalimentation, continuous or int4ermittent drug infusion. - f) Performance tests: The following performance tests are included in the submission: - 1. Functional testing - ﺳﺮ ﺍ Tensile strength and elongation test - Anti-corrosive - - , ( Catheter-flexural fatigue tolerance test {1}------------------------------------------------ K0241007 - Stiffness test ﺮ - ア Flow rate - ア Catheter burst pressure test. Catheter collapse test - تر Leakage test - تر Ink adhesion test - 2. Validation (Microbiological test) - ア Sterilization Process Validation - ア E.O. residual test - 3. Biocompatibility (Biological tests) The result of the laboratory tests demonstrate that the device is as safe as and is effective with the legally marketed predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and features an abstract image of an eagle. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the eagle in a circular fashion. The logo is black and white. Public Health Service JUL 1 4 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Tony Hung President Sunder Biomedical Tech. Co. Ltd. 10F-1, 1-67, Wu-Chung Road Taichung City 403 TAIWAN R.O.C. Re: K024007 Trade/Device Name: Sunder Central Venous Catheter Kits (SD-3L70F30J and SD-3L55F30J) Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: May 26, 2003 Received: May 28, 2003 Dear Mr. Hung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Hung Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Susan Runner Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Attachment 4. 510(k) Number (if known): 서스2400 7 Device Name: SUNDER CENTRAL VENOUS CATHETER KITS (SD-3L70F30J AND SD-3L55F30J) Indications for Use: The triple lumen central venous Catheter is intended for vascular access infusion and withdrawal of blood, blood products, and fluids, central venous blood pressure monitoring and continuous or intermittent drug infusion. - Patients requiring multiple sites for IV access. - Patients lacking useable peripheral IV sites. - Patients requiring central venous pressure monitoring. - Patients requiring total parenteral nutrition. - Patients receiving incompatible medications. - Patients requiring multiple infusions of fluids, medications, or chemotherapy. - . Patients subject to frequent blood sampling or receiving blood transfusions. - Patients receiving infusions that are hypertonic, hyperosmolar or infusions that have divergent pH values. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) 1 Pollara Cuceniti 510(k) Number: K024007
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...