SERVO-CONTROL WALL-MOUNT COSYCOT INFANT WARMER MODEL IW960, MANUAL CONTROL WALL-MOUNT COSYCOT INFANT WARMER MODEL IW970
K972885 · Fisher & Paykel Electronics , Ltd. · FMT · Oct 30, 1997 · General Hospital
Device Facts
Record ID
K972885
Device Name
SERVO-CONTROL WALL-MOUNT COSYCOT INFANT WARMER MODEL IW960, MANUAL CONTROL WALL-MOUNT COSYCOT INFANT WARMER MODEL IW970
Applicant
Fisher & Paykel Electronics , Ltd.
Product Code
FMT · General Hospital
Decision Date
Oct 30, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Fisher & Paykel Healthcare IW960 Servo-Control Wall-Mount CosyCot Infant Warmer and IW970 Manual-Control Wall-Mount CosyCot Infant Warmer are infant radiant warmers (as per 80 FMT, CFR §880.5130) containing an infrared heating element mounted above a built-in pediatric bed assembly, in order to maintain an infant's body temperature by means of controlled radiant heat. The IW960 and IW970 are designed to provide warmth to babies in the first few weeks of life, when an infant's self-thermoregulation capacity may be reduced, or if external thermal support is required or desirable. This may include new-born babies in delivery room applications, including premature / low birth-weight infants, and support of critically ill babies in neonatal intensive care units (NICU's) or special care baby units (SCBU's). Situations which necessitate unobstructed access to an infant, including during resuscitation or surgical procedures, may indicate the need for a radiant heat source instead of equivalent support devices such as infant incubators.
Device Story
Wall-mounted infant radiant warmer; provides thermal support via infrared heating element. Input: skin temperature via dual-thermistor probe (IW960) or manual power settings (IW960/IW970). Operation: microprocessor-controlled circuitry adjusts heater power to maintain set temperature (Servo mode) or constant power (Manual/Prewarm modes). Output: radiant heat to infant; visual/audible alarms for temperature, power, or fault conditions. Used in delivery rooms, NICUs, SCBUs by hospital personnel. Benefits: stabilizes infant temperature to minimize metabolic rate, supporting growth/healing; allows unobstructed access for clinical interventions. Healthcare providers monitor LED displays and adjust settings via control panel. Safety features include dual-thermistor cross-checking and independent thermal output limits.
Clinical Evidence
Clinical verification studies demonstrated the ability of the warmers to warm infants to a stable set temperature accurately and maintain that temperature in a stable environment. Bench testing confirmed compliance with IEC 601-2-21, including mattress surface temperature accuracy, distribution, and irradiance levels.
Technological Characteristics
Heater: single rod infrared element with parabolic reflector. Construction: aluminum extrusion support column, thermoplastic enclosures, fabric-on-aluminum bassinet. Sensing: dual-thermistor skin probe. Control: microprocessor-based. Connectivity: standalone. Power: electrical. Mounting: dual wall-brackets. Standards: IEC 601-2-21.
Indications for Use
Indicated for newborn infants, including premature and low birth-weight infants, requiring thermal support in delivery rooms, NICUs, or SCBUs, particularly when unobstructed access for resuscitation or surgery is required.
Regulatory Classification
Identification
The infant radiant warmer is a device consisting of an infrared heating element intended to be placed over an infant to maintain the infant's body temperature by means of radiant heat. The device may also contain a temperature monitoring sensor, a heat output control mechanism, and an alarm system (infant temperature, manual mode if present, and failure alarms) to alert operators of a temperature condition over or under the set temperature, manual mode time limits, and device component failure, respectively. The device may be placed over a pediatric hospital bed or it may be built into the bed as a complete unit.
Special Controls
*Classification.* Class II (Special Controls):(1) The Association for the Advancement of Medical Instrumentation (AAMI) Voluntary Standard for the Infant Radiant Warmer;
(2) A prescription statement in accordance with § 801.109 of this chapter (restricted to use by or upon the order of qualified practitioners as determined by the States); and
(3) Labeling for use only in health care facilities and only by persons with specific training and experience in the use of the device.
Predicate Devices
Ohmeda, Ohio Infant Warmer System, Model 3150 (K921766)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "HEALTHCARE" in all caps. The font is a simple sans-serif style. There is a horizontal line above the word, and some blurred text above the line.
Fisher & Paykel Electronics Limited 25 Carbine Road, Panmure, Auckland, New Zealand P.O. Box 14-348. Panmure, Auckland, New Zealand Tel: +64-9-574-0100 Fax:
K972885
OCT 30 1997
30 July, 1997
## 510(k) Summary of Safety and Effectiveness Information
IW960 Servo-Control Wall-Mount CosyCot™ Infant Warmer Model No. / Name: IW970 Manual-Control Wall-Mount CosyCot™ Infant Warmer Classification Name: Warmer, Infant Radiant - 80 FMT General Hospital Devices, 21 CFR §880.5130 (Class III) Ohmeda, Ohio Infant Warmer System, Model 3150, K921766 Predicate Devices: Fisher & Paykel, RD1000 Infant Resuscitator, K892885
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The Fisher & Paykel Healthcare IW960 Servo-Control and IW970 Manual-Control Wall-Mount CosyCot Infant Warmers consist of a heater assembly, controller unit, support column, built-in bed assembly, and wall mounting hardware components.
The heater assembly includes a single rod infrared healing clement housed inside a parabolic reflector. An observation lamp is mounted at the back of the heater unit. The thermoplastic enclosure is of similar cross-sectional shape to the reflector and is approx. 125mm high × 197mm wide × 625mm deep. It can be rotated to either side of the warmer, clear of the infant bed. A metal grill on the underside of the heater assembly prevents contact with the element.
The heater assembly is mounted on top of the support column, which consists of a single aluminum extrusion section with thermoplastic front panel sections attached. Instruction / warning labeling is located on the back at the top of the support column, with a duplicate copy of label instructions provided. Identification / specification labeling is located on the front panel at the base of the support column, with the power inlet and auxiliary power outlet sockets. The dimensions of the support column are approx. 1448mm high × 193mm wide × 90mm deep (104mm deep at controller section top). The column section is supported by dual wall-mounting brackets, attached to the top and bottom of the column extrusion.
{1}------------------------------------------------
The controller unit is built into the top of the support column. The transformer and Power PCB are mounted to the aluminum extrusion back section. The Control PCB is mounted on the inside of the thermoplastic front panel. This panel contains the control buttons, displays, main power switch and temperature sensor socket. Controls consist of buttons to select operating modes, timer functions and lamp operation. A control knob selects temperature or power level. LED displays include indicators for operating mode, alarms, timer, lamp and heater power, and 3-digit displays for temperature and timer readings.
The bassinet assembly is supported by a four-bar link system attached to the support column extrusion. It consists of a stretched fabric surface over an aluminum extrusion frame. Removable barriers on each side may be folded down for access to an infant, and latch into clips in the bassinet frame. The bassinet may be tilted continuously through +10° to -10° to achieve Trendelenburg and Fowler positions. A spring-loaded cable brake system is used to change the tilt position. An x-ray tray module can be mounted underneath the bassinet for placement of an x-ray cassette. The bassinet assembly is approx. 650mm square × 86mm high, with side barrier panels 127mm above the bassinet surface.
A variety of accessory mounting and storage options are available. Infant resuscitator and oxygen flowmeter modules may be mounted in column front panels. A gas supply module may be mounted to the back of the support column extrusion. The column extrusion features a channel mounting system in either side to support mounting blocks for further accessories, which may be fixed at the required height. These include short and long mounting poles, side shelves, gas supply and venturi suction unit mounting blocks and accessory hooks. Storage trays and bins may also be mounted under the bassinet area using this mounting system.
In Baby mode, the IW960 provides stable control of the baby's skin temperature by automatically adjusting the heater power to compensate for varying metabolic and environmental conditions. In Manual mode, both the IW960 and IW970 provide useradiustable heater power. In Prewarm mode, the IW960 and IW970 maintain power at a constant level of 25% ready for use.
A double thermistor sensor probe measures the baby's skin temperature, and audible and visual alarms alert the user to high or low temperature situations, equipment fault, power failure and periodic reminders to reassess the baby's clinical condition, depending on the control mode being used. Various independent safety features are included to control maximum output and avoid thermal injury to the infant.
page 2-5
{2}------------------------------------------------
The intended use of Infant Radiant Warmers is to provide thermal support for newborn babies in the first few weeks of life. This may include in the delivery room in the period immediately after birth, and in the neonatal intensive care unit for critically ill babies which may require frequent intervention from hospital personnel. New-born babies (including low birth-weight or premature infants) and critically ill babies may have a reduced self-thermoregulation capacity. Body heat can be lost through the mechanisms of conduction, convection, radiation and evaporation. Low quantities of internal energy and insulating fat, and a high surface area-to-mass ratio can also be contributing factors. In these cases, or when thermal support is required or desirable, radiant heat may be provided to prevent the various clinical consequences of excessive heat loss.
Infant Radiant Warmers contain an infrared heating element intended to be placed over an infant in a pediatric hospital bed, to maintain an infant's body temperature by means of controlled radiant heat. Heat energy is absorbed through an infant's skin, increasing local blood flow which transfers heat to the rest of the body by blood convection and tissue conduction.
Situations which necessitate unobstructed access to an infant, including during resuscitation or surgical procedures, may indicate the need for a radiant heat source instead of equivalent support devices such as infant incubators.
The object of providing controlled radiant heat is to stabilize the infant's temperature at the level where metabolic rate is at a minimum. At this state the infant's internal energy sources are used primarily for growth and healing, and not trying to keep warm or cool.
The technological characteristics of the IW960 and IW970 Wall-Mount CosyCot Infant Warmers are equivalent to those of the predicate device. The built-in bed assembly and wall-mounting configurations, location of heater and controller sections, and capacity to swing aside the heater section are equivalent. Both feature a single-bar radiant heating element with a parabolic metal reflector design. Heater output is regulated by microprocessor-controlled circuitry. The modes of operation, alarm configurations and user controls are very similar between the devices. Both devices use self-monitoring software and hardware options to ensure device faults are detected and do not result in hazardous states. The power used and irradiance levels achieved on a mattress surface by the two devices are very similar in quantity.
Improvements in safety and effectiveness include the use of a dual-thermistor skin temperature sensor, the comparison of readings from which allows the equipment to detect any variation in the sensor performance, and ensure accurate temperature measurement. The temperature controlling system and software used to regulate this allow for very accurate and stable control of an infant's skin temperature, hence providing optimal environmental clinical conditions.
{3}------------------------------------------------
Performance testing for the IW960 and IW970 has been carried out in the areas of functional verification, temperature control, irradiance distribution patterns and clinical verifications.
This testing demonstrates the safety, uniform distribution, accuracy and absolute accuracy of temperatures achieved on an infant bed, and the qualitative nature of the irradiance distribution pattern on the mattress, including irradiance in specific regions of the infra-red spectrum. Clinical verification studies demonstrated the ability of the warmers to warm up babies to a stable desired set temperature level accurately in a short period of time, and the ability to control the set temperature very accurately for a stable situation. The proposed devices meet specific aspects of performance required by the standard for Infant Radiant Warmers, IEC 601-2-21, including:
- temperatures achieved on the mattress surface for different materials.
- temperature distribution and variance across the mattress surface.
- accuracy of temperature control in the servo-controlled mode.
- absolute accuracy of temperature measurement against an external comparison.
- maximum irradiance levels for overall IR and near IR spectrum regions.
The product testing carried out for the IW960 and IW970 Wall-Mount CosyCot Infant Warmers indicate that they meet their design and performance functional requirements. Clinical verification studies demonstrate the successful use of the warmers and their ability to provide accurate and stable warming of infants. The proposed devices also meet the requirements of the international standard for Infant Radiant Warmers, IEC 601-2-21.
signed:
Chris Mander Fisher & Paykel Healthcare
te: 30 July 1997
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes forming its body and wings, symbolizing health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Chris Mander Requlatory Affairs Engineer Fisher & Paykel Electronics Ltd. Healthcare Division 25 Carbine Road Panmure, Auckland New Zealand
OCT 30 1997
Re : K972885 Trade Name: Servo-Control Wall-Mount Cosycot Infant Warmer Model IW960 and IW970 Regulatory Class: II Product Code: FMT Dated: July 30, 1997 Received: August 5, 1997
Dear Mr. Mander:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਚੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note:
{5}------------------------------------------------
Page 2 - Mr. Mander
this response to your premarket notification submission does. not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy J. Ulatowski
Timothy I Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
- [510(k)] Number:
HEALTHCARE
Fisher & Paykel Flectronics | imited 25 Carbine Road, Panmure, Auckland, New Zealand P.O. Box 14-348. Panmure, Auckland, New +64-9-574-0100 +64-9-574-0158
30 July, 1997
## Fisher & Paykel IW960 and IW970 Wall-Mount CosyCot Infant Warmers
## PREMARKET NOTIFICATION 510(k) INDICATIONS FOR USE STATEMENT
The Fisher & Paykel Healthcare IW960 Servo-Control Wall-Mount CosyCot Infant Warmer and IW970 Manual-Control Wall-Mount CosyCot Infant Warmer are infant radiant warmers (as per 80 FMT, CFR §880.5130) containing an infrared heating element mounted above a built-in pediatric bed assembly, in order to maintain an infant's body temperature by means of controlled radiant heat.
The IW960 and IW970 are designed to provide warmth to babies in the first few weeks of life, when an infant's self-thermoregulation capacity may be reduced, or if external thermal support is required or desirable. This may include new-born babies in delivery room applications, including premature / low birth-weight infants, and support of critically ill babies in neonatal intensive care units (NICU's) or special care baby units (SCBU's). Situations which necessitate unobstructed access to an infant, including during resuscitation or surgical procedures, may indicate the need for a radiant heat source instead of equivalent support devices such as infant incubators.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patricia Cusciola
(Division Sign-Off)
Division of Dental, Infection Control, and General Hospital Devices
510(k) Number 4972885
Prescription Use (Per 21 CFR §801.109)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.