HYDRO-TEMP Neonatal Skin Surface Temperature Sensor, HYDRO-TEMP Neonatal Skin Surface Temperature Sensor with Interface Cable
K200319 · Deroyal Industries, Inc. · FMT · Apr 2, 2020 · General Hospital
Device Facts
Record ID
K200319
Device Name
HYDRO-TEMP Neonatal Skin Surface Temperature Sensor, HYDRO-TEMP Neonatal Skin Surface Temperature Sensor with Interface Cable
Applicant
Deroyal Industries, Inc.
Product Code
FMT · General Hospital
Decision Date
Apr 2, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5130
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The HYDRO-TEMP Neonatal Skin Surface Temperature Sensor is a non-sterile device intended for connection to an infant radiant warmer to monitor the neonatal patient's skin surface temperature.
Device Story
Device consists of wire set with distal thermistor chip and proximal connector; interfaces with infant radiant warmer to monitor neonatal skin temperature. Thermistor passively modifies electrical current based on temperature; signal transmitted to warming bed for display. Used in NICU settings by clinical staff. Adhesive pad secures sensor to patient skin. Modification involves replacing molded plastic cap/epoxy glue encapsulation with UV-cured adhesive. Device provides continuous temperature feedback to warming bed, enabling automated thermal regulation of the neonatal environment; helps maintain patient normothermia.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation) per ISO 10993. Electrical safety and electromagnetic compatibility testing per IEC 60601-1 and IEC 60601-1-2. Performance testing per IEC 60601-2-21 and ISO 80601-2-56 confirmed accuracy within ±0.2°C and appropriate time response.
Technological Characteristics
Passive thermistor-based temperature sensor. Materials: ceramic thermistor, copper wire with PVC insulation, plastic-molded connector, UV-cured adhesive encapsulation. Non-sterile. Compatible with various infant radiant warmer brands (Datex-Ohmeda, Draeger, Atom, GE). Compliant with ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-21, and ISO 80601-2-56.
Indications for Use
Indicated for monitoring skin surface temperature in neonatal patients using an infant radiant warmer.
Regulatory Classification
Identification
The infant radiant warmer is a device consisting of an infrared heating element intended to be placed over an infant to maintain the infant's body temperature by means of radiant heat. The device may also contain a temperature monitoring sensor, a heat output control mechanism, and an alarm system (infant temperature, manual mode if present, and failure alarms) to alert operators of a temperature condition over or under the set temperature, manual mode time limits, and device component failure, respectively. The device may be placed over a pediatric hospital bed or it may be built into the bed as a complete unit.
Special Controls
*Classification.* Class II (Special Controls):(1) The Association for the Advancement of Medical Instrumentation (AAMI) Voluntary Standard for the Infant Radiant Warmer;
(2) A prescription statement in accordance with § 801.109 of this chapter (restricted to use by or upon the order of qualified practitioners as determined by the States); and
(3) Labeling for use only in health care facilities and only by persons with specific training and experience in the use of the device.
Predicate Devices
HYDRO-TEMP NICU and Skin Surface Temperature Sensor (K925006)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
April 2, 2020
DeRoyal Industries, Inc. Sarah Bennett Senior Regulatory Affairs Specialist 200 DeBusk Lane Powell, Tennessee 37849
Re: K200319
Trade/Device Name: HYDRO-TEMP Neonatal Skin Surface Temperature Sensor, HYDRO-TEMP Neonatal Skin Surface Temperature Sensor with Interface Cable Regulation Number: 21 CFR 880.5130 Regulation Name: Infant Radiant Warmer Regulatory Class: Class II Product Code: FMT Dated: February 5, 2020 Received: February 7, 2020
Dear Sarah Bennett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tina Kiang, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K200319
Device Name
HYDRO-TEMP Neonatal Skin Surface Temperature Sensor
Indications for Use (Describe)
The HYDRO-TEMP Neonatal Skin Surface Temperature Sensor is a non-sterile device intended for connection to an infant radiant warmer to monitor the neonatal patient's skin surface temperature.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for DeRoyal. The logo consists of a stylized blue swoosh above the text "DeRoyal" in blue. The swoosh appears to be a curved line that starts thin and becomes thicker as it curves over the text.
DeRoyal Industries, Inc. Traditional 510(k) Submission – HYDRO-TEMP Neonatal Skin Surface Temperature Sensor
### 510(k) Summary
| Date prepared: | February 5, 2020 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | DeRoyal Industries, Inc.<br>200 DeBusk Lane<br>Powell, TN 37849<br>Owner/Operator #1044833 |
| 510(k) Contact: | Sarah Bennett<br>Senior Regulatory Affairs Specialist<br>P: 865-362-6112<br>F: 865-362-3741<br>sabennett@deroyal.com |
| Contract Manufacturer: | DeRoyal Cientifica de Latinoamerica<br>S.R.L.<br>Global Park<br>Box 180-3006, 602 Parkway<br>La Aurora, Heredia Heredia, Costa<br>Rica 146 |
| | DeRoyal Industries, Inc.<br>185 Richardson Way<br>Maynardville, TN 37807 |
| Trade Name: | HYDRO-TEMP Neonatal Skin<br>Surface Temperature Sensor |
| | HYDRO-TEMP Neonatal Skin<br>Surface Temperature Sensor with<br>Interface Cable |
| Common Name: | Skin Surface Temperature Sensor |
| Classification Name: | Warmer, Infant Radiant |
| Device Product Code: | FMT |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 880.5130 |
| Classification Panel: | General Hospital |
| K200319 | February 5, 2020<br>Page 1 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DeRoyal. The logo consists of a blue swoosh-like graphic on the left, followed by the word "DeRoyal" in blue text. The "R" in Royal has a registered trademark symbol next to it.
DeRoyal Industries, Inc. Traditional 510(k) Submission – HYDRO-TEMP Neonatal Skin Surface Temperature Sensor
Predicate Devices:
HYDRO-TEMP NICU and Skin Surface Temperature Sensor [K925006]
# Device Description
The HYDRO-TEMP Neonatal Skin Surface Temperature Sensor consists of a probe used to monitor a neonate's skin surface temperature used as an accessory while the patient is on an infant radiant warmer. The probe is a wire set with a thermistor chip at the distal end that passively modifies the electrical current traveling through the device. A colored connector at the proximal end of the wire set interfaces with the warming bed or a cable that interfaces with the bed. The probe is packaged with an adhesive pad used to affix the sensor to the patient's skin surface.
The device is non-sterile and individually packaged in a resealable plastic bag. The probe covers that are packaged with the probe are cleared under K191425. The probes are compatible with the following warming beds: Datex-Ohmeda (Ohmeda, Care Plus, and IWS brands), Draeger (Babytherm, Isolette, Babyleo, Caleo, Resuscitaire, ICS, and Globe-Trotter brands), Atom (Transcapsule, Infa Warmer, INCU 1, and Sunflower brands), and GE (Giraffe and Panda brands).
# Indications for Use
The HYDRO-TEMP Neonatal Skin Surface Temperature Sensor is a non-sterile device intended for connection to an infant radiant warmer to monitor the neonatal patient's skin surface temperature.
## Differences to Predicate Device
The encapsulation method of the subject device is different that the predicate device. The predicate device, the current encapsulation method utilizes a molded plastic cap that is filled with epoxy glue, and the thermistor is inserted into this cap. In the subject device, these materials would be replaced with a UV-cured adhesive. The thermistor is dipped into the adhesive and cured under UV light. This is the only component change in the device. The thermistor cap directly contacts the patient's intact skin.
#### Summary of Technological Characteristics
| Characteristic | HYDRO-TEMP Neonatal<br>Skin Surface<br>Temperature Sensor | HYDRO-TEMP NICU<br>and Skin Surface<br>Temperature Sensor<br>Probe (K925006) |
|----------------|-----------------------------------------------------------|------------------------------------------------------------------------------|
|----------------|-----------------------------------------------------------|------------------------------------------------------------------------------|
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DeRoyal. The logo consists of a blue swoosh-like graphic on the left, followed by the text "DeRoyal" in a blue, stylized font. The "R" in "DeRoyal" has a registered trademark symbol next to it.
## DeRoyal Industries, Inc. Traditional 510(k) Submission – HYDRO-TEMP Neonatal Skin Surface Temperature Sensor
| Prescription Only | Yes | Yes |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Use Environment | NICU | NICU |
| Design | A wire set packaged with an adhesive pad. The wire set has a thermistor chip at the distal end and a colored connector at the proximal end. | A wire set packaged with an adhesive pad. The wire set has a thermistor chip at the distal end and a colored connector at the proximal end. |
| Materials | Probe Cover – Adhesive pad with polyester film backing and release liner<br>Wire – Copper with PVC insulation<br>Thermistor – Ceramic<br>Connector – Plastic-molded metal connector<br>Strain Relief – PVC<br>Cap – UV-curing Adhesive | Probe Cover – Adhesive pad with polyester film backing and release liner<br>Wire – Copper with PVC insulation<br>Thermistor – Ceramic<br>Connector – Plastic-molded metal connector<br>Strain Relief – PVC<br>Cap – PVC and epoxy glue |
| Accuracy | $±0.2°C$ | $±0.2°C$ |
| Sterility | Non-Sterile | Non-Sterile |
| Biocompatibility | ISO 10993-1 compliant | ISO 10993-1 compliant |
| Shelf life | 1 year (intent to extend to 5 years once testing completed) | 3 years |
## Summary of Performance Tests
The proposed device has been tested and/or evaluated according to the following standards as applicable: ISO 10993-1, IEC 60601-2-21, ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2.
The following biocompatibility tests were performed on the final, finished proposed device in accordance with ISO 10993 and FDA guidance: cytotoxicity, sensitization, and irritation. These results show the proposed device is biologically safe for its intended use.
Accuracy testing was performed to IEC 60601-2-21 demonstrating the tolerance range of temperature readings with the proposed change. Additionally, a time response test in accordance with ISO 80601-2-56 also was performed. The results of this testing demonstrate proposed modification to the thermistor encapsulation does not impact the device's performance of its intended use. Evaluation and testing according to IEC 60601-1 and IEC 60601-1-2 also was performed to ensure the change to the thermistor encapsulation did not affect the electrical safety of the device.
All testing was performed on final, finished product manufactured with the proposed modification. The test results met the requirements of the aforementioned standards and demonstrate that
K200319
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DeRoyal. The logo features a stylized blue swoosh above the company name. The company name is also in blue and in a sans-serif font.
DeRoyal Industries, Inc. Traditional 510(k) Submission – HYDRO-TEMP Neonatal Skin Surface Temperature Sensor
the proposed modification does not impact the safet or efficacy of the device.
#### Conclusion
The results of performance testing demonstrate the HYDRO-TEMP Neonatal Skin Surface Temperature Sensor is substantially equivalent to the predicate device. The difference between the predicate device and subject device is change to the thermistor encapsulation method and all other features are the same.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.