← Product Code [FMI](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI) · K972469

# SELFSHIELD HN SELF SHEATHING HYPODERMIC NEEDLE (K972469)

_Beech Medical Products, Inc. · FMI · Aug 27, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K972469

## Device Facts

- **Applicant:** Beech Medical Products, Inc.
- **Product Code:** [FMI](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI.md)
- **Decision Date:** Aug 27, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The SELFSHIELD HN™ Self Sheathing Hypodermic Needle is indicated for non patient contacting and non-skin penetrating applications such as fluid transfer from a drug vial or ampoule to an IV administration set injection site (Y-site) or an IV catheter intermittent injection cap (heparin lock).

## Device Story

SelfShield HN is a spring-loaded, self-sheathing hypodermic needle designed for fluid transfer; adapts to standard luer-fitting piston syringes. Operation: user presses shield against rubber injection site, rotates 40° clockwise, and pushes to penetrate; upon removal, shield automatically advances to cover needle tip. Used in clinical settings by healthcare workers for non-skin applications (e.g., vial-to-IV catheter). Benefits: reduces accidental needlesticks via automatic shielding; provides more robust transport protection than predicate. Output: protected needle tip for safe disposal. Evaluated via simulated clinical use (548 successful transfers) and comparative bench testing (tensile/compressive loads).

## Clinical Evidence

Simulated clinical study with experienced users (n=548 cycles). Primary endpoints: sharps injury incidence and shield activation reliability. Results: 0 sharps injuries, 0 activation failures. User evaluation: 93% rated device acceptable (score ≥8), 98% recommended over predicate for fluid transfer. Bench testing: tensile loads for shield separation exceeded clinical requirements; peak compressive load to expose needle tip ~10x higher than predicate.

## Technological Characteristics

Standard hypodermic needle with spring-loaded plastic shield; luer fitting. Supplied sterile in Tyvek-nylon blister package. Gamma sterilization. Mechanical activation via rotation and compression.

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- Sherwood Monoject® Safety Syringe (1, 3, 6 and 12 cc sizes)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "BEECH" in a stylized, bold font. The letters are enclosed within a rounded rectangular shape. There is a registered trademark symbol to the bottom right of the letter H.

K 77 2467

## 510(k) Summary

Date: 30 June 1997

AUG 27 1997

Contact Name: John Romano, President

Submitter: Beech Medical Products, Inc. 241 Lackland Drive Middlesex, NJ 08846 tel: 908-563-0303 fax: 908-563-1115

Proprietary Name: SelfShield HNTM Self Sheathing Hypodermic Needle

Common Name: Shielded Hypodermic Needle

Classification Name: Hypodermic Single Lumen Needle

Predicate Device: Sherwood Monoject® Safety Syringe (1, 3, 6 and 12 cc sizes)

The SelfShield HN™ device incorporates a standard hypodermic needle which adapts to piston syringe standard luer fittings. The device embodies a spring loaded needle shield which automatically covers the needle tip prior to and after each use. It helps reduce the occurrence of accidental needlesticks to healthcare workers when the product is used as directed. The user activates the device by placing the tip of the shield against a rubber injection site, clockwise rotating the device by at least 40° and pushing the shield back until the needle tip penetrates into and beyond the injection site rubber plug. Once fluid, drug, agent or blood is drawn, the unit is removed from the injection site. The needle shield automatically advances covering and protecting the sharp tip for transport. The device is reactivated again against another injection site until disposed in an approved sharps container. The device is supplied sterile in a peelable Tyvek-nylon blister package.

Both the SelfShield HNTM device and the predicate device use the same hypodermic needle and are gamma sterilized , single patient use, and disposable devices. However, the needle shield of the Self Shield HN™ device automatically extends forward over the needle tip into a terrippary locked position after each use whereas the predicate device's needle shield must be manually extenced forward to the temporary locked position. The automatic return of the shield to the temporary locked position does not require a conscious effort by the user whereas the predicate device does. However, since the needle tip is not exposed so that it can be used to pierce the skin under visual control, the Self Shield HN™ indications are limited to non-skin transfers such as from a drug vial to the injection cap on an IV catheter. On the other hand, the predicate device can be used for both skin and non-skin administration of drugs or direct blood draw since the needle is initially exposed to view.

The results from comparative bench tests indicate that tensile loads needed to separate the shield from either the SelfShield HNTM unit or the predicate device are more than adequate in reducing the potential for failure by pulling since the loads are well above the force that can be reasonably applied by hand in a clinical setting. The SelfShield HN™ device's peak compressive load needed to expose the needle tip is approximately 10 times higher than the predicate device while in the temporary locked position.

Page 4 of 17

{1}------------------------------------------------

Therefore, when transporting a filled syringe the SelfShield HN™ unit requires substantially more force to override the lock compared to the predicate device, which relies on a less robust detent feature to hold the shield in position during transport. The SelfShield HN™ device was tested in a simulated clinical test by experienced users of the predicate device. It performed as indicated in its labeling 548 times without any sharps injuries reported or failure of the shield activation feature and therefore meets the test requirements of the FDA guidance document on devices with sharps injury prevention features. The SelfShield HNTM needle was rated as very effective (score =10) by 50% and is generally rated as acceptable (score =8 and above) by 93% of the user-evaluators, who were positive about the safety and ease of use of the device. They indicated that it will repeatedly transfer fluids aseptically until easily disposed of in a sharps container and that users can be easily trained on the technique. Analysis of the data demonstrates that it can be used effectively by either left or right handed users. Among users who currently use the predicate device, the SelfShield HNTM device was rated as easier and requiring less time to use by the majority of user-evaluators and recommending it 98% of the time than the predicate device for fluid transfer applications.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three stripes extending from its head, representing the department's commitment to health, services, and human well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

电网 27 ppg

Mr. John Romano President Beech Medical Products, Incorporated ----241 Lackland Drive Middlesex, New Jersey 08846

K972469 Re : Selfshield HN Self Sheathing Hypodermic Trade Name: Needle Requlatory Class: II Product Code: FMI Dated: June 30, 1997 Received: July 1, 1997

Dear Mr. Romano:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸੈ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Mr. Romano

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to beqin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

| 510(k) Number (if known): | K 972469       |
|---------------------------|----------------|
| Device Name:              | Self Shield HN |
| Indications For Use:      |                |

ﯿﺴﮯ

The SELFSHIELD HN™ Self Sheathing Hypodermic Needle is indicated for non patient contacting and non-skin penetrating applications such as fluid transfer from a drug vial or ampoule to an IV administration set injection site (Y-site) or an IV catheter intermittent injection cap (heparin lock).

IPLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                                                 | Patricia Crucente |
|---------------------------------------------------------------------|-------------------|
| Division of Dental, Infection Control, and General Hospital Devices |                   |
| 510(k) Number                                                       | K978469           |

Prescription Use______________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)

14:

+ '

t

ોન

Over-The-Counter Use________

(Optional Format 1-2-96)

. ﺍ

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K972469](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K972469)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
