# TSK STERIJEKT PREMIUM DISPOSABLE HYPODERMIC NEEDLE (K970370)

_Tsk Laboratory, Japan · FMI · Apr 10, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K970370

## Device Facts

- **Applicant:** Tsk Laboratory, Japan
- **Product Code:** [FMI](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI.md)
- **Decision Date:** Apr 10, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

To inject fluids into, or withdraw fluids from parts of the body below the surface of the skin.

## Device Story

TSK STERUEKT is a sterile, single-use hypodermic needle; consists of stainless steel beveled cannula bonded to plastic hub; compatible with standard luer lock or luer slip syringe tips. Used by clinicians for fluid injection or withdrawal from body tissues. Device provides physical conduit for fluid transfer; ensures patient safety through sterile, disposable design; prevents cross-contamination.

## Clinical Evidence

Bench testing only. Results: Protector removal and needle bond strength met specifications. Biocompatibility testing passed for sensitization, intracutaneous reactivity, hemolysis, systemic toxicity, and cytotoxicity. LAL pyrogen test passed.

## Technological Characteristics

Stainless steel cannula; plastic hub; luer lock/luer slip compatible; sizes 14G-31G; sterile, single-use; removable plastic protector.

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- Becton Dickinson & Company, 'B-D PrecisionGlide' needles

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Attachment VII

510 (k) SUMMARY

K970370

TSK STERUEKT Premium Disposable Hypodermic Needle

APR 10 1997

Manufacturer: Tochigi Seiko KK
2-1-5 Hirayanagi-cho
Tochigi-shi, Tochigi Ken 328
Japan
Telephone: +81 282 27 8426
Fax: +81 282 27 1943

510 (k) submitted by: Norman Shane Johnson
Distribution Manager
TSK Laboratory, Japan
2-1-5 Hirayanagi-cho
Tochigi-shi, Tochigi Ken 328
Japan
Telephone: +81 282 27 8426
Fax: +81 282 27 1943

Date Summary Prepared: January 10, 1997

Device Trade Name: TSK STERUEKT Premium Disposable Hypodermic Needle
Common Name: Hypodermic Needle
Classification Name: Hypodermic Single Lumen Needle (per 21CFR 880.5570)
Predicate Device: Becton Dickinson &amp; Company, 'B-D PrecisionGlide' needles
Description of Device: TSK STERUEKT is a sterile Hypodermic Single Lumen Needle, packaged individually for single patient use. It consists of a stainless steel beveled cannula bonded to a plastic hub. Sizes include 14G-31G and various lengths. The needle is intended to fit any luer lock or luer slip syringe tip.
Intended Use: To inject fluids into, or withdraw fluids from parts of the body below the surface of the skin.
Safety and Efficacy: The following tests were performed to show the safety and efficacy of the device.

|  TEST | RESULT  |
| --- | --- |
|  Protector Removal | Meets Specification  |
|  Needle Bond Strength | Meets Specification  |
|  Sensitization | Pass  |
|  Intracutaneous | Pass  |
|  Hemolysis | Pass  |
|  Systemic Toxicity | Pass  |
|  LAL Pyrogen Test | Pass  |
|  Cytotoxicity | Pass  |

Substantial Equivalence: The TSK Premium Disposable Hypodermic Needle and the predicate device, the B-D PrecisionGlide Needle, have been shown to be substantially equivalent based on the following:

- identical intended use - disposable sterile hypodermic needle, individually packaged.
- equivalent materials - stainless steel cannula and plastic hub.
- equivalent design - metal cannula with bonded plastic hub; removable plastic protector on needle tip; sufficient physical integrity for the intended use; fits Luer Lock or Luer Slip hypodermic tips.
- similar size range - TSK 14G-31G; B-D 16G-30G.
- needle tip selection - both brands are available with several bevels.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K970370](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K970370)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).