← Product Code [FMI](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI) · K231720

# Sterile Hypodermic Needles for Single Use (K231720)

_Wepon Medical Technology Co., Ltd. · FMI · Sep 8, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K231720

## Device Facts

- **Applicant:** Wepon Medical Technology Co., Ltd.
- **Product Code:** [FMI](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI.md)
- **Decision Date:** Sep 8, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Sterile Hypodermic Needles for Single Use are intended to be used with a luer lock or luer slip syringe and injection devices for general purpose fluid injection/aspiration.

## Device Story

Sterile, single-use hypodermic needles; consist of stainless steel needle tube, polypropylene hub, and protective cover. Used with luer lock or luer slip syringes for fluid injection or aspiration. Operated manually by medical professionals or trained caregivers in hospitals and clinics. Device facilitates drug delivery or fluid extraction. Benefits include sterile, reliable fluid path for clinical procedures. Color-coded per ISO 6009 for gauge identification.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified via biocompatibility (ISO 10993-1), particulate matter (USP <788>), sterilization validation (ISO 11135), and mechanical performance testing per ISO 7864, ISO 9626, and ISO 80369-7. All acceptance criteria met.

## Technological Characteristics

Materials: Stainless steel (needle tube), polypropylene (hub and protective cover). Energy: Manual. Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 7864 (needle specs), ISO 9626 (tubing), ISO 80369-7 (luer connectors), ISO 10993 (biocompatibility), ASTM F88/F1929/F1140/F2096 (packaging).

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- Sterile Hypodermic Needles for Single Use ([K180417](/device/K180417.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 8, 2023

Wepon Medical Technology Co., Ltd. % Esther Zhang Official Correspondent Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District Shanghai, Shanghai 201102 China

Re: K231720

Trade/Device Name: Sterile Hypodermic Needles for Single Use Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: June 8, 2023 Received: June 13, 2023

# Dear Esther Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Alan Sthe

CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231720

Device Name Sterile Hypodermic Needles for Single Use

Indications for Use (Describe)

The Sterile Hypodermic Needles for Single Use are intended to be used with a luer slip syringe and injection devices for general purpose fluid injection/aspiration.

Type of Use (Select one or both, as applicable)

| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------|---------------------------------------------|
|------------------------------------------------|---------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5 510(k) summary

## l Submitter

Device submitter: Wepon Medical Technology CO., LTD. Floor 4, Building A, No. 58, Jinhu Road, Chengdong Street, Wenling Zhejiang, CN 317500

Contract manufacturer: Zhejiang Kangkang Medical-Devices CO., Ltd. Longwang Industrial District, Chumen Town, Yuhuan, Zhejiang, 317605, China Registration number: 3015042030

Contact person: Di Zhao Deputy General Manager Phone: 928-5922380 Email: dizhao@wepon-ppe.com

#### II Correspondent

Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District, Shanghai, P.R.China Contact: Esther ZHANG Email: Esther.zhang@llins-tech.com

#### III Device

Trade Name of Device: Sterile Hypodermic Needles for Single Use Common Name: Hypodermic Single Lumen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product code: FMI Review Panel: General Hospital

#### IV Predicate Devices

| Trade name:             | Sterile Hypodermic Needles for Single Use (used as the predicate device); |
|-------------------------|---------------------------------------------------------------------------|
| Common name:            | Hypodermic single lumen needle                                            |
| Classification:         | Class II, 21 CFR 880.5570                                                 |
| Product Code:           | FMI                                                                       |
| Premarket Notification: | K180417                                                                   |

{4}------------------------------------------------

Manufacturer:

## V Device description

The Sterile Hypodermic Needles for Single Use is composed of a needle hub, protective cover, needle tube and jointing, connected with a luer lock or luer slip syringe and injection devices for general purpose fluid injection. The Sterile Hypodermic Needles for Single Use is for single use only. After opening the inner packaging and taking out the product, tighten the needle to the syringe and remove the protective cover, extract the drug liquid, exhaust the air and then inject. It is provided sterile. The sterilization method is EO sterilization and the sterilization assurance level is 10-6.

| Gauge<br>Length | 30G | 27G | 26G | 25G | 24G | 23G | 22G | 21G | 20G | 19G | 18G |
|-----------------|-----|-----|-----|-----|-----|-----|-----|-----|-----|-----|-----|
| 1/2"            | ●   | ●   | ●   |     |     |     |     |     |     |     |     |
| 5/8"            |     |     | ●   | ●   | ●   |     |     |     |     |     |     |
| 1"              |     |     | ●   | ●   |     | ●   | ●   | ●   | ●   | ●   | ●   |
| 1 1/4"          |     |     |     |     |     | ●   | ●   | ●   | ●   | ●   | ●   |
| 1 1/2"          |     |     |     |     |     | ●   | ●   | ●   | ●   | ●   | ●   |

Table 1 specification of proposed device

### VI Indications for use

The Sterile Hypodermic Needles for Single Use are intended to be used with a luer lock or luer slip syringe and injection devices for general purpose fluid injection/aspiration.

# VII Comparison of technological characteristics with the predicate devices

The Sterile Hypodermic Needles for Single Use have intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the Sterile Hypodermic Needles for Single Use and predicate devices do not alter suitability of the proposed device for its intended use.

| Device<br>feature               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device<br>K180417                                                                                                                                                                                                 | Comments                                    |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Indications for<br>use          | The Sterile Hypodermic<br>Needles for Single Use<br>are intended to be used<br>with a luer lock or luer slip<br>syringe and injection<br>devices for general<br>purpose fluid<br>injection/aspiration.                                                                                                                                                                                                                                                                                                                    | The Sterile Hypodermic<br>Needles for<br>Single Use are intended to<br>be used with a luer slip or<br>luer slip syringe and<br>injection devices for<br>general purpose fluid<br>injection/aspiration.                      | Same                                        |
| Product code                    | FMI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | FMI                                                                                                                                                                                                                         | Same                                        |
| Regulation                      | 21 CFR 880.5570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 21 CFR 880.5570                                                                                                                                                                                                             | Same                                        |
| Device<br>feature               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device<br>K180417                                                                                                                                                                                                 | Comments                                    |
| number                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                             |                                             |
| Class                           | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | II                                                                                                                                                                                                                          | Same                                        |
| Principle of<br>operation       | For manual use only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | For manual use only                                                                                                                                                                                                         | Same                                        |
| Intended user                   | Medical professionals and<br>trained care givers                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Medical professionals and<br>trained care givers                                                                                                                                                                            | Same                                        |
| Environment<br>of use           | Hospitals and clinics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Hospitals and clinics                                                                                                                                                                                                       | Same                                        |
| Needle gauge                    | 30G, 27G, 26G, 25G,<br>24G, 23G, 22G, 21G,<br>20G, 19G, 18G                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 14G, 15G, 16G, 17G,<br>18G, 19G, 20G, 21G,<br>22G, 23G, 24G, 25G,<br>26G, 27G, 29G, 30G                                                                                                                                     | Difference                                  |
| Length                          | 1/2", 5/8", 1", 1¼", 1½"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 6-60mm                                                                                                                                                                                                                      |                                             |
| Type of wall                    | normal wall or thin wall                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | not provide                                                                                                                                                                                                                 |                                             |
| blade angle                     | short bevel or long bevel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | not provide                                                                                                                                                                                                                 |                                             |
| Needle hub                      | Needle hub<br>Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Needle hub<br>Polypropyle ne                                                                                                                                                                                                | Same                                        |
| main structure<br>and materials | Needle tube<br>Stainless steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Needle<br>Stainless steel                                                                                                                                                                                                   |                                             |
|                                 | protective<br>cover<br>Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | protective<br>cap<br>Polypropyle ne                                                                                                                                                                                         |                                             |
| Needle hub                      | Color-coded per ISO 6009                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Color-coded per ISO 6009                                                                                                                                                                                                    | Same                                        |
| Single use                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                         | Same                                        |
| Performance<br>specifications   | Complies with ISO 7864;<br>ISO 9626; ISO 80369-7                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Complies with ISO 7864;<br>ISO 9626; ISO 80369-7                                                                                                                                                                            | Same                                        |
| Sterilization                   | EO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | EO                                                                                                                                                                                                                          | Same                                        |
| SAL                             | 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 10-6                                                                                                                                                                                                                        | Same                                        |
| Pyrogen                         | Non-pyrogenic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Non-pyrogenic                                                                                                                                                                                                               | Same                                        |
| Biocompatibility                | The<br>biocompatibility<br>evaluation for the subject<br>device was conducted in<br>accordance with the<br>International Standard ISO<br>10993-1 "Biological<br>Evaluation of Medical<br>Devices - Part 1:<br>Evaluation and Testing<br>Within a Risk                                                                                                                                                                                                                                                                     | Comply with ISO 10993.<br>The test is as follows:<br>The devices meet<br>biocompatibility endpoints<br>for cytotoxicity, irritation,<br>sensitization, systemic<br>toxicity, and<br>hemolysis material-mediated<br>pyrogens | Same                                        |
| Device<br>feature               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device<br>K180417                                                                                                                                                                                                 | Comments                                    |
|                                 | Management Process," as<br>recognized by FDA and<br>the "Use of International<br>Standard ISO 10993-1<br>"Biological evaluation of<br>medical devices- Part 1:<br>Evaluation and testing<br>within a risk management<br>process", June 16, 2016.<br>The syringe of testing<br>included the following<br>tests:<br>Cytotoxicity;<br>Skin sensitization;<br>Hemolysis;<br>Intracutaneous reactivity;<br>Acute systemic toxicity;<br>Pyrogenicity.<br>The evaluation of the<br>above testing items meets<br>the requirements |                                                                                                                                                                                                                             |                                             |
|                                 | Labeling                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Meet the requirements of<br>21 CFR Part 801                                                                                                                                                                                 | Meet the requirements of<br>21 CFR Part 801 |

{5}------------------------------------------------

{6}------------------------------------------------

Different 1

The needle gauge and length of subject devices are different from the predicate device, and the type of wall and needle bevel are unknown. The subject device specifications are within the range of the predicate device. This difference does not affect intended use and does not raise different questions of safety and effectiveness.

# VIII Performance data

The following performance data were provided in support of the substantial equivalence determination.

# Biocompatibility testing

Biocompatibility of the Sterile Hypodermic Needles for Single Use were evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External

{7}------------------------------------------------

communication device - Blood path indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended:

| Cytotoxicity              | ISO 10993-5: 2009  |
|---------------------------|--------------------|
| Skin sensitization        | ISO 10993-10: 2010 |
| Hemolysis                 | ISO 10993-4: 2017  |
| Intracutaneous reactivity | ISO 10993-10: 2010 |
| Acute systemic toxicity   | ISO 10993-11: 2017 |
| Pyrogenicity              | ISO 10993-11: 2017 |

All evaluation acceptance criteria were met.

Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.

#### Sterilization and shelf life testing

The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters. The shelf life of the Sterile Hypodermic Needles for Single Use is determined based on stability study which includes ageing test. The shelf-life of the Sterile Hypodermic Needles for Single Use is five (5) years.

| Sterilization Evaluation                                     | ISO11135: 2014                                                                                                                               |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| EO residue                                                   | ISO 10993-7:2008                                                                                                                             |
| ECH residue                                                  | ISO 10993-7:2008                                                                                                                             |
| Bacterial Endotoxin testing                                  | USP42-NF37<85>                                                                                                                               |
| Sterile barrier packaging testing &<br>Shelf Life Evaluation | Seal Strength ASTM F88/F88M-15<br>Dye Penetration ASTM F1929-15<br>Creep/Burst Testing<br>ASTMF1140/F1140M-13<br>Gross Leakage ASTM F2096-11 |

Antibacterial Testing DIN 58953-6:2010

#### Performance testing

Freedom from defects, Lubricant)

Performance testing is performed according to the following standards:

| ➤ | ISO 7864: 2016                    |                                   |
|---|-----------------------------------|-----------------------------------|
|   | Cleanliness                       | Clause 4.3 of ISO 7864: 2016      |
|   | Limits for acidity or alkalinity  | Clause 4.4 of ISO 7864: 2016      |
|   | Limits for extractable metals     | Clause 4.5 of ISO 7864: 2016      |
|   | Tubular needle designation        | Clause 4.6 of ISO 7864: 2016      |
|   | Colour coding                     | Clause 4.7 of ISO 7864: 2016      |
|   | Needle hub                        | Clause 4.8 of ISO 7864: 2016, ISO |
|   |                                   | 80369-7 and ISO 6009              |
|   | Needle cap                        | Clause 4.9 of ISO 7864: 2016      |
|   | Needle tube (Tolerance on length, | Clause 4.10 of ISO 7864: 2016     |

{8}------------------------------------------------

|   | Needle Point                                        | Clause 4.11 of ISO 7864: 2016     |
|---|-----------------------------------------------------|-----------------------------------|
|   | Bond between Tube and Hub                           | Clause 4.12 of ISO 7864: 2016     |
|   | Patency of Lumen                                    | Clause 4.13 of ISO 7864: 2016     |
| > | ISO 9626:2016                                       |                                   |
|   | Surface finish and visual appearance                | Clause 5.2 of ISO 9626:2016       |
|   | Cleanliness                                         | Clause 5.3 of ISO 9626:2016       |
|   | Limits for acidity and alkalinity                   | Clause 5.4 of ISO 9626:2016       |
|   | Size designation                                    | Clause 5.5 of ISO 9626:2016       |
|   | Dimensions                                          | Clause 5.6 of ISO 9626:2016       |
|   | Stiffness                                           | Clause 5.8 of ISO 9626:2016       |
|   | Resistance to breakage                              | Clause 5.9 of ISO 9626:2016       |
|   | Resistance to corrosion                             | Clause 5.10 of ISO 9626:2016      |
| > | ISO 80369-7:2016                                    |                                   |
|   | Dimensional requirements for luer<br>connectors.    | Clause 5 of ISO 80369-7: 2021     |
|   | Fluid leakage (Positive pressure liquid<br>leakage) | Clause 6.1.3 of ISO 80369-7: 2021 |
|   | Sub-atmospheric pressure air leakage                | Clause 6.2 of ISO 80369-7: 2021   |
|   | Stress cracking                                     | Clause 6.3 of ISO 80369-7: 2021   |
|   | Resistance to separation from axial load            | Clause 6.4 of ISO 80369-7: 2021   |
|   | Resistance to separation from                       | Clause 6.5 of ISO 80369-7: 2021   |
|   | unscrewing                                          |                                   |
|   | Resistance to overriding                            | Clause 6.6 of ISO 80369-7: 2021   |

## IX Conclusion

The Sterile Hypodermic Needles for Single Use are substantially equivalent to its predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K231720](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMI/K231720)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
