K152514 · Ypsomed AG · FMI · Nov 18, 2015 · General Hospital
Device Facts
Record ID
K152514
Device Name
Clickfine AutoProtect Pen Needle
Applicant
Ypsomed AG
Product Code
FMI · General Hospital
Decision Date
Nov 18, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Clickfine AutoProtect Pen Needle is intended for use with pen injector devices for the injection of fluids, including insulin and exenatide. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.
Device Story
The Clickfine AutoProtect Pen Needle is a sterile, single-use, hypodermic needle designed for attachment to commercially available pen injectors. The device features a passive safety shield that automatically covers and locks over the needle upon withdrawal from the skin, minimizing accidental needle-stick injuries. The user attaches the needle to the pen injector, removes the protective cap, and performs the injection. During insertion, the shield retracts into the housing; upon withdrawal, the shield extends and locks, indicated by a red visual indicator. The device is intended for use by patients or healthcare providers in home or clinical settings. It provides a physical barrier to prevent needle exposure before and after use, enhancing patient safety during fluid administration.
Clinical Evidence
No clinical data was required to establish substantial equivalence. Evidence is based on bench testing and design verification activities confirming compliance with ISO 11608-2, ISO 9626, ISO 7864, and ISO 23908 standards.
Indicated for use with pen injector devices for the subcutaneous injection of fluids, including insulin and exenatide, in patients requiring such therapy. Includes a passive safety shield to prevent accidental needle sticks.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the first profile being the most prominent.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2015
Ypsomed AG Mr. Stephan Affolter Head of Quality System & Regulatory Affairs Brunnmattstrasse 6, CH- 3401 Burgdorf SWITZERLAND
Re: K152514
Trade/Device Name: Clickfine AutoProtect Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: II Product Code: FMI Dated: October 20, 2015 Received: October 22, 2015
Dear Mr. Affolter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows a signature and the name "Tina Kiang". There is a large, looping signature to the left of the name. Below the name, there is a small "-S".
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152514
#### Device Name Clickfine AutoProtect Pen Needle
Indications for Use (Describe)
The Clickfine AutoProtect Pen Needle for use with pen injector devices for the injection of fluids, including insulin and exenatide.
Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 8 510(k) Summary for K152514
| Submitted By/<br>Contact Person: | Stephan Affolter<br>Head of Quality System & Regulatory Affairs<br>Ypsomed AG<br>Brunnmattstrasse 6<br>CH - 3401 Burgdorf<br>Switzerland<br>Tel. 0041-34 424 3382<br>Fax 0041-34 424 4122<br>E-mail: stephan.affolter@ypsomed.com |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Alternative Contact: | Lee Leichter<br>President<br>P/L Biomedical<br>10882 Stonington Avenue<br>Fort Myers, FL 33913<br>USA<br>Tel. (239) 244-1448<br>Fax. (815) 550-0162<br>E-mail: leichter@plbiomedical.com |
| Date Prepared: | September 23, 2015 |
| 8.1 Trade/Proprietary Name: | Clickfine AutoProtect Pen Needle |
Injection Pen Needle
8.3 Classification Name:
II 1 . . 1 1
8.3 Classification Name:
8.2 Common/Usual Name:
Hypodermic single lumen needle
8.4 Classification: Regulation 21 CFR Part 880.5570 - Hypodermic single lumen needle was published in the Federal Register on October 21, 1980 after review by the General Hospital and Personal Use Devices Classification, classifying Hypodermic single lumen needles in Class II.
> Panel: 80 Procode: FMI
8.5 Predicate Device: K131123 Clickfine AutoProtect Pen Needle
8.6 Purpose of Submission: To widen the range of the needle lengths and include a 5 mm Clickfine AutoProtect pen needle.
{4}------------------------------------------------
### 8.7 Device Description
The pen needles are sterile, non-pyrogenic, single use needles designed to be used with commercially available pen-injectors for the administration of fluids. Each needle is individually packaged in a sealed protective container with a peel tab.
The pen needle is used by peeling back the peel tab and snapping or screwing the housing onto the threaded end of the pen-injector. The back end of the cannula punctures the septum of the drug reservoir in the pen-injector. The outer protective cap is then removed. When the injection is needed the needle is inserted into the chosen site. While inserting the needle into the skin, the safety shield glides into the housing enabling the needle to penetrate the skin barrier and the subcutaneous tissue. While the safety shield glides into the housing the safety mechanism is activated. The pen-injector delivers the medicinal product through the needle.
After the injection, in order to remove the needle from the skin, the user withdraws the pen injector from the skin. As the pen and needle is withdrawn from the skin, the safety shield glides back in its initial position, completely covering the needle, where it will remain locked. The safety shield is designed to automatically cover the needle (providing passive protection) to minimize the risk of accidental needle-stick injury. Once the Clickfine AutoProtect Pen Needle is in the locked mode, it can no longer be used. The red safety lock indicator tells the user that the safety lock has been activated. The needle is detached from the injection device and disposed of in accordance with local regulations. For each subsequent injection, another disposable needle must be used.
### 8.8 Intended Use
The intended use of the modified device remains the same as the predicate device Clickfine AutoProtect Pen Needles K131123:
The Clickfine AutoProtect Pen Needle is intended for use with pen injector devices for the injection of fluids, including insulin and exenatide.
Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.
- 8.9 Technological Characteristics and Substantial Equivalence discussion For the 5 mm Clickfine AutoProtect, except for the material of the needle shield, all of the material, processes and operating principles are identical to the currently marketed pen needles.
{5}------------------------------------------------
| Characteristic | Current Ypsomed Clickfine AutoProtect<br>Injection Pen Needle 8 and 6 mm | Modified Ypsomed<br>Clickfine AutoProtect<br>Injection Pen Needle<br>5 mm |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| 510(k) | K131123<br>The Clickfine AutoProtect Pen Needle is<br>intended for use with pen injector devices for<br>the injection of fluids, including insulin and<br>exenatide. | N/A |
| Indications for<br>use | Additionally, the attached safety shield<br>automatically locks in place and reduces the<br>occurrence of accidental needle sticks from<br>the patient end of the needle. The shield also<br>serves to hide the needle before and after<br>injection. | Unchanged |
| OTC | OTC | Unchanged |
| Needle length | 6 and 8 mm | 5, 6 and 8 mm |
| Needle gauge | 31, 30 and 29 gauge | Unchanged |
| Needle outer<br>diameter | 0.25, 0.3 and 0.33 mm | Unchanged |
| | Normal walled needle:<br>$\ge$ 0.114, $\ge$ 0.133 and<br>$\ge$ 0.133 mm | Unchanged |
| Needle inner<br>diameter | Thin walled needle:<br>$\ge$ 0.125, 0.165 and<br>$\ge$ 0.190 mm | Unchanged |
| Needle grind | Front and back bevel | Unchanged |
| Needle tip: | 3-bevel | Unchanged |
| Method of<br>attachment to<br>pen | "Snap-on" & "Twist-on” attachment and<br>Twist-off removal | Unchanged |
| Needle<br>material | Stainless steel | Unchanged |
| Needle/hub<br>bond | Adhesive bonded | Unchanged |
| Hub material | PP | Unchanged |
| Housing | White colored PP | Unchanged |
| Needle shield | Blue colored MBS, using the Zylar 530<br>granulate | Blue colored MBS,<br>using the Zylar 550<br>granulate |
| Safety lock<br>indicator | Red colored PP | Unchanged |
| Spring | Stainless steel | Unchanged |
| Characteristic | Current Ypsomed Clickfine AutoProtect<br>Injection Pen Needle 8 and 6 mm | Modified Ypsomed<br>Clickfine AutoProtect<br>Injection Pen Needle<br>5 mm |
| Outer<br>protective<br>container<br>material | Blue colored PP | Unchanged |
| Peel tab | PET | Unchanged |
| Sterilization<br>method | Gamma irradiation | Unchanged |
| Sterilization<br>conditions | 25 kGy | 17.5 kGy |
| IFU | Leaflet | Unchanged |
## Comparison to Predicate Device
{6}------------------------------------------------
#### 8.10 Performance Data
As part of the Design Control Activities, Ypsomed evaluated the potential risks, hazards or hazardous situations, potential clinical impact on the patient, identified the appropriate risk control measure and acceptability criteria associated with each change in the modified devices. Then, each risk was mitigated through appropriate verification and/or validation activities which were determined to be the testing specified in the following international and internal standards and protocols. This testing confirmed that the subject devices are substantially equivalent to the predicate devices.
The Clickfine AutoProtect Pen Needles have met the requirements of the relevant sections of the following standards:
- ISO 11608-2:2012 Needle based injection systems for medical use -. Requirements and test methods - Part 2: Needles
- ISO 9626:1991/Amd.1:2001 Stainless steel needle tubing for the manufacture of medical devices
- ISO 7864:1993 Sterile hypodermic needles for single use
- . ISO 23908:2011 Sharps injury protection - Requirements and test methods
| Test | Requirement | Result |
|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Needle assembly<br>attachment shall<br>fit and function<br>with needle<br>based injection<br>systems<br>specified in<br>ISO 11608-1 | Clause 4.2.1 of ISO 11608-2<br>The pen needles were tested for compatibility with<br>needle based injection system as specified in ISO<br>11608-1 | Pass |
| Tubing<br>dimensions | Clause 4.2.2, Table 1 of ISO 11608-2<br>Length of patient end: +/- 1.25 mm<br>Length of cartridge end: 5.7 – 7.0 mm | Pass |
| Test | Requirement | Result |
| Flow rate | Clause 4.3 of ISO 11608-2:<br>Measurement of flow rate in accordance with<br>Annex A. | Pass |
| Bond between<br>hub and needle<br>tube | Clause 9 of ISO 11608-2 and clause 13.1 of ISO<br>7864<br>The union of the hub and needle tube shall not<br>break for at least 5 sec. while a force of at least 22<br>N is applied | Pass |
| Needle points | Clause 4.5 of ISO 11608-2<br>Visually sharp at a 2.5x magnification; designed to<br>minimize coring and fragmentation | Pass |
| Lubrication | Clause 4.7 of ISO 11608-2<br>No visible droplets on the outside surface of the<br>needle tube. | Pass |
| Dislocation of<br>measuring point<br>at the patient-<br>end | Clause 8 of ISO 11608-2:<br>maximum allowable dislocation acc. Clause 4.8,<br>Table 2 for 5mm: 0.65mm and for 8mm: 0.9mm | Pass |
| Functional<br>compatibility<br>with needle<br>based injection<br>systems | Clause 11 of ISO 11608-2:<br>Needle assembly torque: 0.07 +/- 0.01 Nm<br>Needle hub removal: less than 0.100 Nm<br>Dose accuracy:<br>for doses ≤ 20 ml the calculated values were within<br>± 0.01 ml of the targeted dose;<br>for doses > 20 ml the calculated values were within<br>± 5 % of the targeted dose. | Pass |
| Tubing<br>characteristics | Table 1 of ISO 9626/A1:<br>The tubing is made of austenitic stainless steel acc.<br>one of the designated types given in Table 1 of ISO<br>9626/A1. | Pass |
| Tubing<br>diameters | Table 2 of ISO 9626/A1<br>Tubing dimensions meet outer and inner diameter<br>requirements as defined in table 2 of ISO 9626/A1. | Pass |
| Stiffness | Clause 9 of ISO 9626:<br>When tested in accordance with annex C of ISO<br>9626 the tubing shall show a deflection not greater<br>than the relevant value given in table 3 of ISO<br>9626/A1. | Pass |
| Resistance to<br>breakage | Clause 10 of ISO 9626:<br>When tested in accordance with annex D of ISO<br>9626 and Table 4 of ISO 9626/A1 the tubing shall<br>not break | Pass |
| Test | Requirement | Result |
| Limits for<br>acidity and<br>alkalinity | Clause 6 of ISO 9626<br>Tested in accordance with annex A and extract<br>preparation of the tubing in accordance with Annex<br>B: correction for the volume of titrant required for<br>the control fluid shall not be more than 0.04 ml<br>NaOH or 0.12 ml HCl to reach the end-point<br>titration. | Pass |
| Resistance to<br>corrosion | Clause 11 of ISO 9626:<br>When tested in accordance with annex E, the<br>immersed half of the tubing shall show no evidence<br>of corrosion resulting from the test. | Pass |
| Freedom of<br>defects | Clause 11.3 in ISO 7864:<br>When examined by normal or corrected-to-normal<br>vision, the needle shall appear straight and of<br>regular cross-section and wall thickness | Pass |
| Limits for<br>extractable<br>metals | Clause 6 of ISO 7864:<br>When tested with a recognized method the content<br>of lead, tin, zinc and iron shall be less than 5mg/l.<br>The cadmium content shall be lower than 0.1 mg/l. | Pass |
| Freedom of<br>defects | Clause 11.3 of ISO 7864:<br>When examined by normal or corrected vision, the<br>needle lube shall appear straight and of regular<br>cross-section and wall thickness. | Pass |
| Activation of the<br>sharps injury<br>protection | In accordance with ISO 23908 clause 5.2 the sharps<br>injury protection feature shall be able to be<br>activated by an appropriate force that eases<br>actuation and avoids unintended actuation. An<br>appropriate force was determined using a risk-based<br>approach in accordance with ISO 14971. | Pass |
| Verification of<br>safety feature<br>activation | In accordance with ISO 23908, clause 5.3, needles<br>were tested to provide evidence that once the safety<br>feature was activated, they resist axial and lateral<br>forces so as to prevent unintended exposure to the<br>sharps. | Pass |
| Device design<br>verification for<br>needle tip<br>protection after<br>activation | In accordance to clause 5.4 of ISO 23908 it was<br>demonstrated that a sphere of 6mm radius did not<br>come into contact with the tip of the cannula, either<br>before or immediately after triggering the safety<br>mechanism | Pass |
{7}------------------------------------------------
{8}------------------------------------------------
The 5 mm Clickfine AutoProtect Pen Needles were tested, and met all acceptance criteria, for all the requirements as provided above. Based on the results Y psomed concluded that the device performance is acceptable for the product.
#### 8.11 Clinical Data
{9}------------------------------------------------
Clinical Data was not required to establish substantial equivalence for these devices.
- 8.12 Conclusion
Ypsomed AG concludes based on the information presented that the subject device is substantially equivalent to the predicate product legally marketed in the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.