BD PEN NEEDLE

K131358 · Becton, Dickinson & CO · FMI · Jul 16, 2013 · General Hospital

Device Facts

Record IDK131358
Device NameBD PEN NEEDLE
ApplicantBecton, Dickinson & CO
Product CodeFMI · General Hospital
Decision DateJul 16, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

BD Pen Needle is intended for use with pen injector device for subcutaneous injection of drugs, including insulin and exenatide.

Device Story

BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with PentaPoint™ are single-use, sterile, hypodermic needles designed for use with pen injectors. The device consists of a double-ended cannula assembled into an injection-molded hub using adhesive; the hub features internal threads for attachment to pen injectors. The non-patient end penetrates the pen cartridge septum; the patient end is used for subcutaneous drug delivery. Both ends are lubricated with silicone-based lubricant to facilitate injection and septum penetration. An inner shield protects the patient end; an outer cover provides a sterile barrier and assists in hub removal. Used by patients, caregivers, and clinicians in various settings. The device facilitates drug administration, such as insulin or exenatide, by providing a sterile, consistent interface between the medication cartridge and the patient.

Clinical Evidence

Bench testing only. Performance evaluated per ISO 11608-2 and ISO 7864 standards, including tubing dimensions, lumen patency, needle point sharpness, cannula load testing (22N), lubrication, compatibility, freedom from defects, needle dislocation, penetration force, and thumb force testing.

Technological Characteristics

Double-ended stainless steel cannula; injection-molded hub; silicone-based lubricant. Sterilized via gamma irradiation (SAL 10-6). Dimensions vary by gauge (31G/32G) and length. Connectivity via internal threads for pen-injector compatibility. No software or electronic components.

Indications for Use

Indicated for subcutaneous injection of drugs, including insulin and exenatide, using a pen injector device. Suitable for use by consumers, caregivers, and healthcare professionals.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 131358 1 Becton Drive Franklin Lakes, New Jersey 07417 tel: 201.847.6800 www.bd.com JUL 1 6 2013 Image /page/0/Picture/3 description: The image shows the BD logo, which consists of a circular emblem with a stylized figure and sun rays above it. To the right of the emblem are the letters 'BD' in bold, sans-serif font. Below the letters is the tagline 'Helping all people' in a smaller font size. live healthy lives # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS Date of Summary Preparation: May 9, 2013 - 1 . Submitted By/Contact: Pasquale Amato Staff Regulatory Affairs Specialist Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417 201-847-4513 Phone: Fax: 201-848-8830 # Alternate Contact Person: Sheryl Robinson Bagalio Director, Regulatory Affairs Phone: 201-847- 5326 201-848- 8830 Fax: Email: Shery1 Bagalio@BD.com - 2. Device Name: Trade Name: BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with PentaPoint™ - Insulin Pen Needle Common Names: Needle, Hypodermic, Single Lumen Classification Name: Classification: Class II, 21 CFR 880.5570 FMI ### 3. Predicate Device: BD Pen Needles: - K123300, BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with . PentaPoint™ - K 100005 BD Pen Needle 32G x 4mm . {1}------------------------------------------------ K051899 31G x 8mm and the 31G x 5mm Pen Needles . Manufactured by: Becton, Dickinson and Company #### Device Description: 4. Like the predicate devices, BD pen needles are single use, sterile, medical devices designed to be used in conjunction with pen injectors and pen cartridges. Pen needles are used by consumers, caregivers and health care professionals. They are offered in various gauge sizes and lengths. BD Pen Needles are sterile (gamma irradiation sterilization), non-toxic and nonpyrogenic. The pen needle assembly consists of a doubled-ended cannula that is assembled into an injection-molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen-injector device. This allows the Non Patient (NP) end of the cannula to penetrate through the rubber septum of the cartridge. The Patient end and NP end of the cannula are lubricated using silicone based lubes for ease of injection and rubber septum penetration. An injection-molded inner shield is assembled over the Patient end of the Cannula to protect the point from damage and accidental needlesticks. This needle assembly is inserted into a protective injection-molded outer cover and sealed with a peel-away (tear drop) label to provide sterile barrier and tamper evidence. The outer cover is also used to remove the hub and cannula from the pen. The peel-away label is pre-printed with information, which includes the lot number and needle gauge / length. The individual needle assemblies are packaged in polybags and / or cartons, and placed into shippers with appropriate labeling. The shipper cases are palletized and sterilized to an SAL 10-6 by gamma irradiation. The above information is the same as the predicate device. The purpose of this Traditional 510(k) is to provide substantiation for the label claims. - Statement of Intended Use/Indications for Use: 5. BD Pen Needle is intended for use with pen injector device for subcutaneous injection of drugs, including insulin and exenatide. The intended use for the principal device remains the same as the predicate devices. #### Technological Characteristics: 6. The principal device, BD 31G and 32G XTW Pen Needles with PentaPoint™, of this Premarket Notification has the same design and all design control activities have been conducted in accordance with Quality System Regulations. The principal device is substantially equivalent to the predicate devices given: {2}------------------------------------------------ - Has the same intended use and indications for use as the predicate . devices - Uses the same operating principles . - Incorporates the same design . - Is manufactured from the same materials . - Is sterilized using the same mode . - Is sterilized to SAL of 10-6 . - Is packaged using same unit and case materials . - Same scientific technology . - Same manufacturing process . There are little to no differences between the principal device BD 31G and 32G XTW Pen Needles with PentaPoint™ and the predicate devices. #### 7. Performance: Bench tests relating to the performance of the BD 31G and 32G XTW Pen Needle with PentaPoint™ were conducted. The principal device demonstrated equivalent performance to the predicate devices during bench testing. Bench testing consisted of: | Tubing diameters | Per ISO 11608-2, section 4.3.1 (tubing dimensions meet OD and ID requirement). | |----------------------------------------------|---------------------------------------------------------------------------------------------------------------------| | Patency of lumen | Per ISO 11608-2, section 4.4 (stylet, having a diameter equivalent to 80% ±2% of lumen ID passes through freely). | | Needle points | Per ISO 11608-2, section 4.5 (visually sharp at 2.5X magnification, designed to minimize coring and fragmentation). | | Type A Needle (length) | Per ISO 11608-2, section 4.3.2 (patient end within indicated length ± 1.25 mm) | | Cannula load test<br>(No pre-conditioning) | Per ISO 11608-2, section 4.9 and 9.<br>ISO 7864 Section 13.1(cannula holds force of 22N for 5 seconds). | | Cannula load test<br>(with pre-conditioning) | Per ISO 11608-2, section 4.9 and 9.<br>ISO 7864 section 13(cannula holds force of 22N for 5 seconds). | | Lubrication | Per ISO 11608-2, section 4.7 (no visible droplets inside/outside surfaces of cannula). | | Compatibility Testing | Per ISO 11608-2, section 4.10 (connectivity (torque). | {3}------------------------------------------------ | Freedom form Defects | Per ISO 11608-2, Section 4.6 (needle to be<br>straight and of regular cross section and wall<br>thickness) | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Needle Dislocation | Per ISO 11608-2, section 4.8 and 8 (Patient<br>end max dislocation from center for 4mm<br>PN=0.58mm; 5mm PN=0.65mm; 8mm<br>PN=0.9mm.<br>The angular deviation on the patient end shall<br>not exceed ±3° | | Penetration Testing | Per BD Test Method TP700279 | | Thumb Force Testing | Per BD Test Method TP700381 | ### 8. Substantial Equivalence: The principal device, BD 31G and 32G XTW Pen Needles with PentaPoint™ is the same as the predicate device cleared in K123300, which was found substantially equivalent to the devices previously submitted and cleared in K100005 and K051899, and is therefore substantially equivalent in intended use, function and composition is not affected. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wing shapes, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 16, 2013 Becton Dickinson & Company Mr. Pasquale Amato Staff Regulatory Affairs Specialist 1 Becton Drive FRANKLIN LAKES NJ 07417-1885 Re: K131358 Trade/Device Name: BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with PentaPont™ Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: May 9, 2013 Received: May 10, 2013 Dear Mr. Amato: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commence prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ### Page 2 - Mr. Amato Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mary S. Runner -S Susan Run DDS, MA 2013.07.16 14:27:39-04'00' Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): 13/358 Device Name: BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with PentaPoint™ Indications For Use: BD Pen Needle is intended for use with pen injector device for subcutaneous injection of drugs, including insulin and exenatide. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ric 20 -0. Richard C. Chapman 2013.07.16 11:31:40 -04'00' Page 1 of ____________________________________________________________________________________________________________________________________________________________________ (Division Sign-Off) livision of Anesthestology, General Hospital ifection Control, Dental Devices 510(k) Number: _ K (313 56
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...