K120420 · Shanghai Double Dove Industrial Co., Ltd. · FMI · Oct 18, 2012 · General Hospital
Device Facts
Record ID
K120420
Device Name
STERILE BLUNT FOR SINGLE USE
Applicant
Shanghai Double Dove Industrial Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Oct 18, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The Sterile Blunt for Single Use includes two models, which are blunt fill needle and blunt plastic cannula. The blunt fill needle is used in conjunction with a syringe to penetrate the vial/ampoules stopper intended for injection or aspiration of fluid. The blunt plastic cannula is used in conjunction with a syringe to access the pre-slit septum covering injection sites on I.V. System intended for injection or aspiration of fluid.
Device Story
Sterile Blunt for Single Use consists of two models: blunt fill needle (stainless steel tube, hub, sheath) and blunt plastic cannula (molded plastic blunt tip, hub, sheath). Devices attach to syringes to facilitate fluid aspiration or injection. Blunt fill needle penetrates vial/ampoule stoppers; blunt plastic cannula accesses pre-slit septa on I.V. systems. Designed to replace sharp hypodermic needles to prevent accidental needle-stick injuries. Used in clinical settings by healthcare providers. Sterilized via EO or radiation to SAL 10^-6. Benefits include reduced risk of sharps injury during fluid handling.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity), sterility validation (ISO 11135-1, ISO 11137-2), package integrity (ASTM F1140-07, ASTM F88-09), and physical performance (ISO 9626, ISO 7864, ISO 594-1).
Technological Characteristics
Materials: Stainless steel (needle), molded plastic (cannula). Connectivity: None. Energy: None. Sterilization: EO or Radiation (SAL 10^-6). Standards: ISO 9626, ISO 7864, ISO 594-1, ISO 10993 series, ISO 11135-1, ISO 11137-2, USP 34, ASTM F1140-07, ASTM F88-09.
Indications for Use
Indicated for use with a syringe for fluid injection or aspiration. Blunt fill needle penetrates vial/ampoule stoppers; blunt plastic cannula accesses pre-slit septa on I.V. systems. Intended for single use.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
OCT 1 8 2012
# Attachment III 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
KIZOYNO The assigned 510(k) Number:
- Date of Submission: 17 OCT 2012 1 .
- 2. Sponsor
Shanghai Double-Dove Industry Co, Ltd. No.1888 Huhang Road, Fengxian Economic Zone Shanghai, 201400, China
Establishment Registration No .: 3004416208
Contact Person Mr. Sanba Yang Vice General Manager Tel: +86-21-67104888 Fax: +86-21-67104666 Email: Yangsbmaster@hotmail.com
- 3. Submission Correspondent Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
### 4. Proposed Device Identification
Proposed Device Name: Sterile Blunt for Single Use Proposed Device Model: Blunt Fill Needle and Blunt Plastic Cannula
Classification: II Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital
l l l - 1
ﻴﻨﻴ
{1}------------------------------------------------
Intended Use Statement:
The Sterile Blunt for Single Use includes two models, which are blunt fill needle and blunt plastic cannula.
The blunt fill needle is used in conjunction with a syringe to penetrate the vial/ampoules stopper intended for injection or aspiration of fluid.
The blunt plastic cannula is used in conjunction with a syringe to access the pre-slit septum covering injection sites on I.V. System intended for injection or aspiration of fluid.
5. Predicate Device Identification
> 510(k) Number:K 102328 Product Name: Blunts Manufacturer: Shan Dong Wei Gao Group Medical Polymer Co., LTD
### 6. Device Description
The proposed device, Sterile Blunt for Single Use, includes two models, which are blunt fill needle and blunt plastic cannula.
The blunt fill needle consists of a stainless steel needle tube, a hub and a sheath. It is used to penetrate the vial and/or ampoules stopper intended for injection or drawl fluid. It is not intended to access the injection site on I.V. System.
The blunt plastic cannula consists of molded plastic blunt tipped cannula, a hub and a sheath. It is used to replace hypodermic needles to prevent from the unintended injury by metal needle.
Both two models are available in either EO sterilized to achieve a Sterilized to achieve a Sterility Assurance Level (SAL) of 106. The proposed device is sealed in a sterility maintenance package to maintain its sterility during its shelf life of three (3) years.
### 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ISO 9626:1991/Amendment 1:2001, Stainless steel needle tubing for the manufacture of medical devices.
{2}------------------------------------------------
ISO 7864:1993, Sterile hypodermic needles for single use.
ISO 594-1:1986, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements.
ISO 10993-1: 2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
ISO 10993- 4:2002/And 1:2006, Biological Evaluation of Medical Device. Part 4: Selection of rest for Interactions with Blood.
ISO 10993-5: 2009, Biological evaluation of medical devices - Part 5: Tests for In Viro cytotoxicity. 2-153
ISO 10993-7:2008, Biological evaluation of medical devices -Part 7: Ethylene oxide sterilization residuals
ISO 10993-10: 2002/Amd. 1:2006(E), Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity AMENDMENT I
ISO 10993-11: 2006. Biological evaluation of medical devices -- Part H : Tests for systemic toxicity. ISO 11135-1: 2007 Sterilization of health care products -- Ethylene oxide -- Part 1: Requirements Tor development, validation and routine control of a sterifization process for medical devices.
USP 34:2011, -: 85> Bacterial Endotoxins Test
ASTM F1140-07, Standard Test Methods for Internal Pressurization Failure Resisunce of Unrestrained Packages.
ASTM F88-09, Standard Test Method for Seal Strength of Flexible Barrier Materials.
ISO 11137-2: 2006. Sterifization of healtheare products-Radistion-Part2: Establishing ប្រើ sterifization dose.
#### Substantially Equivalent Conclusion 8.
The proposed device, Sterile Blunt for Single Use, has same intended use, configuration, operation mode with those of the predicate device, Blunts (K 102328). The materials are different, the biocompatibility test and leachable substance analysis demonstrated that the materails used for manufacturing of proposed dveice are safe and accpetable. The bevel of the proposed device is identical to that of the predicate device, while the gauges of proposed and predicate device is different. The physical performance test results submitted demonstrated the proposed device complied with the referenced standards with different gauge of the predicate device. The proposed device is available in EO sterilization or Radiation sterilization, the predicate device is only avaiable in radiation sterilization. Various tests, including biocompatibility, perfromance, package, endotoxin tests demonstrated that the both EO sterilization sterilization sterilization proposed device complied with acceptance criteria. In addition, both of the proposed and predicate device could achieve same Sterility Assurance Level (SAL)
Therefore, the proposed device, Sterile Blunt for Single Use, is determined to be Substantially Equivalent (SE) to the predicate device, Blunts (K102328), in respect of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shanghai Double-Dove Industry Company, Limited C/O Ms. Diana Hong General Manager Mid-Link Consulting Company, Limited P.O. Box 237-023 Shanghai, CHINA 200237
OCT 1 8 2012
Re: K120420
Trade/Device Name: Sterile Blunt for Single Use Model: Blunt Fill Needle, Blunt Plastic Cannula Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: August 30, 2012 Received: September 7, 2012
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
## Page 2 - Ms. Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthem um
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
arket Notification 510(k) Submission
K 12O H2O
# Section II Indications for Use
510(k) Number:
Device Name: Sterile Blunt for Single Use Model: Blunt Fill Needle, Blunt Plastic Cannula
Indications for Use:
Sterile Blunt for Single Use includes two models, which are blunt fill needle and blunt plastic cannula.
The blunt fill needle is used in conjunction with a syringe to penetrate the vial/ampoules stopper intended for injection or aspiration of fluid.
The blunt plastic cannula is used in conjunction with a syringe to access the pre-slit septum covering injection sites on I.V. System intended for injection or aspiration of fluid.
OPRESCRIPTION USE (Part 21 CFR 801 Subpart D) ■OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
RC Chap 10/12/12 Page 1 of 1
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesional Devices. Division of AneSthoolors.
Infection Control, Dental Devices.
510(k) Number: K120920
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.