K090111 · Novo Nordisk, Inc. · FMI · Feb 1, 2010 · General Hospital
Device Facts
Record ID
K090111
Device Name
FLEXPEN NEEDLE
Applicant
Novo Nordisk, Inc.
Product Code
FMI · General Hospital
Decision Date
Feb 1, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FlexPen® needles are intended for use with pen injector devices for the subcutaneous injection of insulin, liraglutide, and somatropin.
Device Story
FlexPen® needle is a sterile, single-use disposable hypodermic needle designed for attachment to Novo Nordisk pen injection devices. Operation involves removing a protective tab, screwing the needle onto the pen injector, removing protective caps, and performing the injection as directed by a healthcare professional or user manual. After use, the needle is detached and disposed of per local regulations. The device facilitates subcutaneous delivery of insulin, liraglutide, or somatropin. It is intended for use by patients or healthcare providers in clinical or home settings. The device does not incorporate electronic components or software.
Clinical Evidence
No clinical tests were performed or submitted. The device relies on design equivalency and non-clinical functional testing to demonstrate safety and effectiveness.
Technological Characteristics
Sterile, single-lumen, disposable hypodermic needle. Materials and performance comply with domestic and international standards. Mechanical design is consistent with existing pen needles for subcutaneous injection. No electronic, software, or energy-based components.
Indications for Use
Indicated for patients requiring subcutaneous injection of insulin, liraglutide, or somatropin using pen injector devices.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
NovoFine® 32G Tip (0.23/ 0.25 mm) x 6 mm disposable needle (K062500)
NovoFine® 32G Tip (0.23/ 0.25 mm) x 6 mm disposable needle (K053470)
Becton Dickinson BD Pen Needle (29G, 30G, 31G) x (5 mm, 8 mm, 12.7 mm) (K051899)
Submission Summary (Full Text)
{0}------------------------------------------------
CONFIDENTIAL
Date: Version: Status: Page:
17 September 2009 -Final
P. lot
Steve Nordick
5 of 7
# 807.87(h) 510(k) Summary
#### As required by Section 807.92(a)
## FEB - 1 2010
(1) DATE OF PREPARATION: September 16, 2009
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. FlexPen® needle meets all applicable product and quality standards.
#### SUBMITTER'S NAME AND ADDRESS:
Novo Nordisk Inc. 100 College Road West Princeton, New Jersey 08540
Contact Person: Rick Spring Tel: 609-987-5046 Fax: 609-987-3916
#### (2) NAME OF DEVICE:
Proprietary Name: Common or usual name: Classification: Class:
FlexPen® needle Sterile disposable hypodermic needle Hypodermic single lumen needle (21 CFR 880.5570) Class II
#### (3) SUBSTANTIAL EQUIVALENCE:
FlexPen® needle is a disposable which is substantially equivalent to K062500 cleared on November 21, 2006, NovoFine® 32G Tip (0.23/ 0.25 mm) x 6 mm disposable needle, cleared under 510/k) K053470. NovoFine Autocover® 30G x 8 mm, cleared under 510(k) K050106, and to the Becton Dickinson BD Pen Needle (29G, 30G, 31G) x (5 mm, 8 mm, 12.7 mm), cleared under 510(k) K051899.
The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food. Drug, and Cosmetic Act, as amended and as supplied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statement related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent laws or their application by the court.
{1}------------------------------------------------
FlexPen needle 510(k) - K090111 Novo Nordisk Inc.
CONFIDENTIAL
Date: Version: Status: Page:
LOROIII p.2 of 3
17 September 2009 | Novo Nordisk
Final
6 of 7
## (4) DEVICE DESCRIPTION:
FlexPen® needle is designed for single use in conjunction with Novo Nordisk injection delivery devices. Prior to giving an injection, the protective tab is removed from the outer needle cap of the single-use disposable needle. With the disposable needle remaining in the outer needle cap, it is then carefully screwed onto the injection delivery device until tight and then the needle outer and inner caps are removed. Prepare for injection by following the procedure described in the user manual provided with the pen injection device and instructions from your health care professional.
After the injection, the needle is removed from the skin. The needle is detached from the injection device and disposed of in accordance with national/local regulations. For each subsequent injection, another disposable needle must be used. Delivery device function checks can be performed with the FlexPen® needle by using the needle cap as described in the user manuals provided with the Novo Nordisk pen injection devices.
FlexPen® needle is used in exactly the same manner as the NovoFine® needles. The instructions for use are described in the user manuals for Novo Nordisk delivery devices for injection and are the same for all NovoFine® needles and FlexPen® needle.
## (5) INTENDED USE:
FlexPen® needle is intended for use with pen injector devices for the subcutaneous injection of insulin, liraglutide, and somatropin.
## (6) TECHNOLOGICAL CHARACTERISTICS:
FlexPen® needle is considered substantially equivalent to K062500, NovoFine® 32G Tip x 6 mm, NovoFine® Autocover® 30G x 8 mm, and to the Becton Dickinson BD Pen Needle (29G, 30G, 31G) in intended use (intended for use with pen injector devices for the subcutaneous injection of insulin, liraglutide and somatropin), technology/principle of operation, materials and performance. Differences between the devices do not raise any significant issues of safety and effectiveness.
{2}------------------------------------------------
090111 p. 3 of 3
FlexPen needlc 510(k) - K090111 Novo Nordisk Inc.
'CONFIDENTIAL
Date: Version: Status: Page:
#### As required by Section 807.92(b)
# (1) NON-CLINICAL TESTS PERFORMED: None
FlexPen® needle will be manufactured in accordance with current Good Manufacturing Practices for Medical Devices. Biocompatibility and performance tests have been performed and the results are in compliance with existing domestic and international standards.
(2) CLINICAL TESTS SUBMITTED:
No clinical tests are required.
# (3) CONCLUSIONS DRAWN FROM THE NON-CLINICAL AND CLINICAL TESTS:
Based on the design equivalency and the functional testing, Novo Nordisk had determined that FlexPen® needle is substantially equivalent to a device currently marketed in the United States. Differences between the devices do not raise any significant issues of safety and effectiveness.
(1) me Ellingto
Mary Ann McElligott, Ph D. Associate Vice President, Regulatory Affairs Novo Nordisk Inc.
18 Sept 04
Date
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The image is in black and white.
Food and Drug Administration 10903 New Hampshire Avenne Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Rick Spring Senior Manager, Regulatory Affairs Novo Nordisk, Incorporated 100 College Road West Princeton, New Jersey 08540
FEB - 1 2010
Re: K090111
Trade/Device Name: FlexPen Needle 32G Tip x 6mm (1/4") Disposable Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: January 26, 2010 Received: January 27, 2010
Dear Mr. Spring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Spring .
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
#### Enclosure
{5}------------------------------------------------
FlexPen needle 510(k) - K090111 Novo Nordisk Inc. Date: Version: Status: l'age:
Novo Nordisk
# 4 of 7
40
1
Final
4 of 7
1
Final
# Indications for Use Statement
510(k) Number (if known)
K090111
Device Name:
FlexPen needle 32G Tip x 6mm (1/4") Disposable Needle
Indications For Use:
FlexPen® needles are intended for use with pen injector devices for the subcutaneous injection of insulin, liraglutide, and somatropin.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Charles G. Koch
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K090111
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.