K033409 · Gettig Pharmaceutical Instrument Co. · FMI · Jan 21, 2004 · General Hospital
Device Facts
Record ID
K033409
Device Name
AMERSHAM HEALTH NEEDLE-GUARD
Applicant
Gettig Pharmaceutical Instrument Co.
Product Code
FMI · General Hospital
Decision Date
Jan 21, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Amersham Health Needle Guard is to withdraw medications from a vial or ampule and to inject these medications subcutaneously, intramuscularly, intravenously, or via I.V. access ports. This device may aid in the prevention of needlestick injury.
Device Story
Amersham Health Needle Guard is a sterile, single-use hypodermic needle for use with 3ml syringes. Device components include stainless steel needle, nylon hub, polypropylene cannula cover, polysulfone Luer hub, and PETG protective sleeve. Operation involves attaching needle to syringe, removing cannula cover for medication withdrawal or injection, and manually pushing the sleeve forward to a locked safety position post-injection for disposal. Used in clinical settings by healthcare providers. Active safety feature reduces risk of accidental needlestick injuries during disposal. Device provides a physical barrier between the user and the contaminated needle.
Clinical Evidence
No human clinical studies were conducted. Evidence consists of bench testing and simulated use testing. Bench testing verified compliance with ISO 7864, ISO 594-2, and ASTM A751 standards. Simulated use testing evaluated the effectiveness of the safety feature mechanism in preventing accidental sharp object injuries and the usability of the instructions for use. Risk analysis performed during simulated testing indicated no unexpected risks.
Technological Characteristics
Materials: Stainless steel 304 (ASTM A751), nylon hub, polypropylene cannula cover, polysulfone Luer hub, PETG protective sleeve. Design: Single-lumen hypodermic needle with active safety sleeve. Connectivity: None. Sterilization: Method 1 per ANSI/AAMI/ISO 11137-1. Standards: ISO 7864 (sterile hypodermic needles), ISO 594-2 (Luer fittings).
Indications for Use
Indicated for patients requiring subcutaneous, intramuscular, or intravenous medication administration via vial or ampule withdrawal. Intended for use by healthcare professionals to aid in the prevention of needlestick injury.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
## 510(k) SUMMARY
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is K033409.
### 1. Submitter Identification:
Gettig Pharmaceutical Instrument Company A Division of Gettig Technologies, Inc. One Streamside Place West P.O. Box 85 Spring Mills, PA 16875
Date Summary Prepared:
January 12, 2004
### 2. Name of the Device:
Amersham Health Needle Guard
#### 3 Predicate Device Information:
Gettig Guard, Gettig Pharmaceutical Instrument Company, Spring Mills, PA, K#000455.
### 4. Device Description:
The Amersham Health Needle Guard is a sterile, single use hypodermic needle designed to be used in conjunction with a disposable 3ml synnge. The device consists of six basic constituents, a Stainless Steel needle bonded to a Nylon hub, a Polypropylene cannula cover, a Nylon Luer attachment connected to a Polysulfone Luer hub used to connect to a syringe, and a PETG Protective sleeve. Once the Amersham Health Needle Guard is attached to a syringe, the cannula cover is removed, exposing the needle. In this condition the needle can be inserted into a vial or ampule to withdraw medication into the syringe or to administer the medication through injection. Once the needle is withdrawn from the patient, the sleeve is pushed forward to the locked safety position for disposal. The sharps safety feature is an active safety feature
### 5. Intended Use:
The intended use of the Amersham Health Needle Guard is to withdraw medications from a via| or ampule and to iniect these medications subcutaneously, intramuscularlay. intravenously, or via I.V. access ports. This device may aid in the prevention of needlestick injury.
Image /page/0/Picture/17 description: The image shows a logo with the word "GETTIG" on top of a square shape. The square is divided into sections, with the left side featuring a semi-circle shape. The right side of the square has a smaller square shape in the bottom right corner.
{1}------------------------------------------------
## 6. Comparison to Predicate Device:
## Similarities and Differences:
The Amersham Health Needle Guard incorporates attributes from several currently marketed devices. In reference to the Gettig Guard, the action and function of the mancedo devices: 'n rolar in both use and application. Both allow for the needle to be protective sibere to birninge with medicament, and permanent covering of the needle exposed for ming the oyeence is that the Gettig Guard allows for the protective sleeve to cover the needle prior to injection, for handling and transport, while the Amersham cover the needle Guard allows for a cannula cover to cover the needle during handling and transport.
## Substantial Equivalence:
Our testing results discussed in a later section, have shown that the Amersham Health Needle Guard meets our specifications for hypodermic needles. Based on this analysis we conclude the device to be safe and effective for its intended use. The Amersham we conclude Guard is similar in feel, function and intended use to currently marketed r Child Noodie Gallery, Therefore, we believe the device to be substantially equivalent to currently marketed devices.
## 7 Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
- a) ANSI/AAMI/ISO Sterilization Standards 11137-194 Section B 3.4.1, Method 1: Dose Setting Using Bioburden Information. Bioburden and Sterilization per ISO 11737-1 and ISO 11737-2.
- b) ISO 7864, "Sterile Hypodermic Needles for Single Use". Third Edition, 1993.
- c) ISO 594-2, "Conical Fittings with a 6% (Luer) Taper for Syringes, Needles and Certain Other Medical Equipment - Part 2: Lock Fittings, First Edition.
- d) ASTM A751 for Stainless Steel 304 Series
# 8. Discussion of Clinical Tests Performed:
Human clinical studies were not conducted on the Amersham Health Needle Guard device because the design and intended use of the devices are not significantly different to warrant these studies. However, simulated use testing was conducted on the Amersham Health Needle Guard safety feature designed to reduce the risk of needlestick accidents.
Simulated use testing was conducted to determine the effectiveness of the safety feature mechanism in preventing accidental sharp object injuries. The simulated use testing also determined the usability of the "Instructions for Use".
Attached as Exhibit #4 is the evaluation form that was utilized to record simulated use testing data, along with the results and conclusions of the study findings.
A risk analysis was assessed during the simulated testing. Based on the testing results and findings we have concluded that there were no unexpected risks in use of this device.
Image /page/1/Picture/16 description: The image shows a logo with the word "GETTIG" on top of a stylized graphic. The graphic appears to be a combination of geometric shapes, possibly representing a shield or emblem. The overall design is simple and clean, with a focus on the text and the geometric shapes.
{2}------------------------------------------------
## 9. Conclusions:
The Amersham Health Needle Guard has the same intended use and similar technological characteristics as the predicate device. Moreover, the bench testing and simulated use testing demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Amersham Health Needle Guard is substantially equivalent to the predicate device.
Image /page/2/Picture/3 description: The image shows a logo with the word "GETTIG" on top. Below the word is a design that includes a partial circle on the left and a square shape on the right. The square shape has some internal lines or patterns within it. The overall design appears to be a stylized emblem or symbol.
Gellig Pharmaceutica Instrumen
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized eagle with three wavy lines extending from its body, representing the department's mission to promote health and well-being.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 1 2004
Mr. James A. Benz Gettig Pharmaceutical Instrument Company 1 Stream Side Place West P.O. Box 85 Spring Mills, Pennsylvania 16875-0085
Re: K033409
Trade/Device Name: Amersham Health Needle Guard Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: October 20, 2003 Received: October 27, 2003
Dear Mr. Benz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in " interstate commerce prior to May 28, 1976, the enactment date of the Medical Device 4 mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Benz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Chiu Lin, Ph., D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K033409
# Exhibit B
Page of 1 ー
510(k) Number (if known): K033409
Device Name Amersham Health Needle Guard
## Indications For Use:
The intended use of the Amersham Health Needle Guard is to withdraw medications from a vial or ampule and to inject these medications subcutaneously, intramuscularly, intravenously, or via I.V. access ports. This device may aid in the prevention of needlestick injury.
Prescription Use V (Per 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Vick Hubbard, Interim Branch Chief
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control. Dental Devices
510(k) Number KO 3 -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.