← Product Code [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF) · K980580

# BECTON DICKINSON SYRINGES (K980580)

_Bd Becton Dickinson Vacutainer Systems Preanalytic · FMF · Jun 25, 1998 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K980580

## Device Facts

- **Applicant:** Bd Becton Dickinson Vacutainer Systems Preanalytic
- **Product Code:** [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF.md)
- **Decision Date:** Jun 25, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

These syringes are intended for general purpose fluid aspiration/injection and insulin injection.

## Device Story

Becton Dickinson single-use hypodermic and insulin syringes; manual devices for fluid aspiration/injection or insulin delivery. Device modification involves replacing natural rubber plunger tip with alternate latex-free rubber formulation. No changes to device design, manufacturing process, or site. Syringes available sterile/non-sterile, with/without needles. Used by clinicians or patients for medication administration. Output is physical delivery of fluid/insulin. Benefits include latex-free option for patients with sensitivities.

## Clinical Evidence

No clinical data. Equivalence demonstrated through bench testing: leakage resistance (ISO 7886/8537), autoclavability, infusion pump performance, chemical testing (ISO/USP), biocompatibility, and mechanical actuation force testing.

## Technological Characteristics

Manual hypodermic/insulin syringes. Components: barrel and molded rubber plunger tip. Material change: alternate latex-free rubber formulation. Performance standards: ISO 7886, ISO 8537. Sterilization: not specified. Connectivity: none. Software: none.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Becton Dickinson Single Use Hypodermic and Insulin Syringes

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUN 25 1998 SUMMARY OF SAFETY AND EFFECTIVENESS DATA

#### B-D Contact Person 1.0

Peter Zurlo Becton Dickinson & Company Manager, Regulatory Affairs 1 Becton Drive, Building 2 Franklin Lakes, NJ 07417-1884 (201) 847-6447 - Phone (201) 848-0457 - FAX

#### 2.0 Device Name

Becton Dickinson Single Use Hypodermic and Insulin Syringes

#### 3.0 . Predicate Device

Becton Dickinson Single Use Hypodermic and Insulin Syringes

#### Product Description/Function 4.0

- Types This change in syringe plunger tip formulation is for the Becton 4.1 Dickinson syringe. Products include:
	- General purpose and insulin syringes. 1.1.1
	- Sterile and non-sterile syringes. 1.1.2
	- 1.1.3 Syringes with and without needles attached.

### No design changes are being made.

- Intended Use These syringes are intended for general purpose fluid 4.2 aspiration/injection and insulin injection.
#### Comparison of Modified and Predicate Devices 5.0

- Design Changes No design changes are being made. 5,1
- 5.2 Changes
	- 5.2.1 Labeling - The syringe labeling will be revised to include :

## "This product does not contain natural rubber latex."

Or (optionally) :

- "Latex Free".

{1}------------------------------------------------

5.2.2 Material Change: Becton Dickinson added an alternate latex free rubber formulation used for its molded syringe plunger tip in 1988.

#### 6.0 Equivalence

.

.

The following data demonstrates functional equivalence to the Becton Dickinson predicate plunger tip and fitness for use.

Syringe barrels molded with the alternate plunger tip formulation have demonstrated equivalence to barrels molded with the current plunger tip formulation.

- 6.1 Leak Test - Syringes with the altemate plunger tip formulation passed ISO 7886 and ISO 8537 syringe leakage tests (as did control samples with current plunger tip formulation).
- 6.2 Autoclavibility (for syringes that require autoclavibility resistance) -Syringes assembled with the alternate plunger tip satisfactorily passed autoclavibility testing. (Syringes were autoclaved at 270°F, 15 minutes then tested for ISO leakage resistance testing. All samples tested passed ISO leak testing.)
- 6.3 Syringe Infusion Pump Performance (for syringes used in infusion pumps) - Syringes assembled with the altemate plunger tip demonstrated equivalent performance to natural rubber plunger tip samples in syringe pump application testing.
- 6.4 Chemical Testing - Testing has been performed per ISO and USP requirements. Results were acceptable.
- Biocompatibility Biocompatibility testing was satisfactorily completed on the 6.5 altemate plunger tip.
- 6.6 Plunger Actuation Forces - Forces for the alternate plunger tip are equivalent or superior to the current plunger tip.
- Manufacturing Process Changes No manufacturing process changes are 6.7 being made.
- ૯.૪ Manufacturing Site Changes - No manufacturing site changes are being made.
- 6.9 Packaging Component Changes - No packaging components are being changed.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, depicted in black, with its wings forming a shape that resembles human profiles.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 5 1998

Mr. Peter Zurlo ·Manager, Regulatory Affairs Becton Dickinson Consumer Products 1 Becton Drive Franklin Lakes, New Jersey 07417-1883

K980580 Re : Becton Dickinson Syringes Trade Name: Regulatory Class: II Product Code: EMF June 9, 1998 Dated: Received: June 10, 1998

Dear Mr. Zurlo:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Mr. Zurlo

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance a (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at it toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours

Timothy A. Ulatowski

Timoth U Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page ol

510(k) Number (if known):_

Becton Dickinson Syringes Device Name:

Indications For Use:

.

The general use syringes are intended for the aspiration and Insulin syringes are intended for the injection of fluids. injection of insulin.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Patricia Crescentz

(Division Sign-Off) Division of Dental, Infection Control. and General Hospital Devices

510(k) Number K9801580

Prescription Use_ (Per 21 CFR 801.109)

ಂದ

Over-The-Counter Use Insul

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K980580](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K980580)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).